BD Surgiphor™ Antimicrobial Irrigation System (910110)
K253996 · Becton, Dickinson and Company (BD) · FQH · Feb 25, 2026 · General Hospital
Device Facts
Record ID
K253996
Device Name
BD Surgiphor™ Antimicrobial Irrigation System (910110)
Applicant
Becton, Dickinson and Company (BD)
Product Code
FQH · General Hospital
Decision Date
Feb 25, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5475
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds.
Device Story
BD Surgiphor™ Antimicrobial Irrigation System is a manual wound irrigation device; consists of 450 mL aqueous solution (0.5% povidone-iodine in phosphate-buffered saline) in a high-density polyethylene (HDPE) bottle with a polycarbonate screw cap; used for wound debridement; operator squeezes bottle to dispense solution; mechanical fluid pressure loosens and removes debris and microorganisms from wound site; PVP-I acts as preservative; device is terminally sterilized; intended for prescription use.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-1), endotoxin/pyrogen testing (USP <85>, AAMI ST72), sterilization validation (ISO 13004, ISO 11737), and functional testing of mechanical debris removal.
Technological Characteristics
450 mL aqueous solution (0.5% PVP-I, phosphate-buffered saline, potassium iodide, vitamin E TPGS). Bottle material: HDPE. Cap: Polycarbonate. Sterilization: Gamma irradiation (SAL 10^-6). Manual squeeze-bottle delivery. No software or electronics.
Indications for Use
Indicated for wound cleansing and removal of wound debris in patients requiring mechanical irrigation of wounds.
Regulatory Classification
Identification
A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
Predicate Devices
BD Surgiphor™ Antimicrobial Irrigation System (K221504)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 25, 2026
Becton, Dickinson and Company (BD)
Daniel Ehrhardt
Staff Regulatory Affairs Specialist
75 N Fairway Dr.
Vernon Hills, Illinois 60061
Re: K253996
Trade/Device Name: BD Surgiphor™ Antimicrobial Irrigation System (910110)
Regulation Number: 21 CFR 880.5475
Regulation Name: Jet Lavage; Wound dressing, drug
Regulatory Class: Class II; Unclassified
Product Code: FQH; FRO
Dated: January 26, 2026
Received: January 26, 2026
Dear Daniel Ehrhardt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253996 - Daniel Ehrhardt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253996 - Daniel Ehrhardt
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MUSTAFA A. MAZHER
-S
For Yu-Chieh Chiu, Ph.D
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253996 | |
| Device Name BD Surgiphor™ Antimicrobial Irrigation System | |
| Indications for Use (Describe) BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K253996
# 510(k) Summary
Preparation Date: December 12, 2025
510(k) Number: K253996
Applicant: Becton, Dickinson and Company
75 N Fairway Dr
Vernon Hills, IL 60061
Contact Person: Daniel Ehrhardt
Staff Regulatory Affairs Specialist
Tel: (303) 217-1876
Device Trade Name: BD Surgiphor™ Antimicrobial Irrigation System
Classification Name: Jet Lavage
Device Classification: Class II (21 CFR 880.5475)
Unclassified (Pre-amendment)
Product Code: FQH; FRO
Predicate Device: BD Surgiphor™ Antimicrobial Irrigation System
Product Code: FQH (Jet Lavage), FRO (Dressing, Wound, Drug); Class II
(21 CFR 880.5475)
Applicant: Becton, Dickinson and Company (BD) K221504
Page 1 of 5
{5}
K253996
# Device Description
The subject BD Surgiphor™ Antimicrobial Irrigation System is a terminally sterilized 450 mL aqueous solution for irrigation and debridement of wounds. The device includes one screw cap and one bottle of Surgiphor solution (0.5% povidone-iodine (PVP-I)), which is used to loosen and remove wound debris via manual lavage. The mechanical action of fluid moving across the wound provides the mechanism of action and aids in the loosening and removal of debris, and foreign materials, including microorganisms, from wounds. The PVP-I in the Surgiphor solution serves as a preservative to ensure that no microbial growth occurs in the solution after the bottle is open.
# Indications for Use
BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds.
# Comparison of Technological Characteristics
The subject device is substantially equivalent to its predicate BD Surgiphor™ Antimicrobial Irrigation System (K221504). Table 1 compares the subject and predicate devices.
The subject device is unchanged from the legally marketed predicate BD Surgiphor™ Antimicrobial Irrigation System (K221504) in its intended use, packaging, and design, except for the material of the bottle itself. The mechanism of action to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds, is unchanged as well. The mechanism of action is defined by the fluid pressure of the solution dispensed upon a wound.
This 510(k) proposes a new material for the 450mL bottle containing the Surgiphor solution. The bottle material of the predicate is polypropylene and the proposed bottle material for the subject device is high-density polyethylene (HDPE). There are no changes to the screw cap, packaging, nor other design aspects, including the bottle shape and dimensions. The Instructions for Use of the subject device are identical to the predicate device. The Surgiphor solution remains unchanged from the predicate device, as a dilute PVP-I solution containing 0.5% PVP-I in phosphate-buffered saline with potassium iodide and vitamin E TPGS.
With the packaging, manual screw cap, and Surgiphor solution remaining unchanged from the predicate device, the previously provided testing in K221504 and previous 510(k)s remains valid. For the subject device, there is no change to the intended use, and the change does not raise new safety and effectiveness concerns. Substantial equivalence has been demonstrated through conformity to standards and design verification and validation testing.
Page 2 of 5
{6}
K253996
Table 1 – Comparison of Subject and Predicate Devices
| Comparison Feature | Subject Device: BD Surgiphor™ Antimicrobial Irrigation System | Predicate Device: BD Surgiphor™ Antimicrobial Irrigation System | Subject to Predicate Comparison |
| --- | --- | --- | --- |
| 510(K) Number | K253996 | K221504 | N/A |
| Product Code | FQH, Jet Lavage
FRO, Dressing, Wound, Drug | FQH, Jet Lavage
FRO, Dressing, Wound, Drug | Identical to predicate |
| Product Classification | Class II (21 CFR 880.5475)
Unclassified (Pre-Amendment) | Class II (21 CFR 880.5475)
Unclassified (Pre-Amendment) | Identical to predicate |
| Device Description | BD Surgiphor™ Antimicrobial Irrigation System is an antimicrobial irrigation system containing 0.5% povidone-iodine (PVP-I) in phosphate-buffered saline, potassium iodide and vitamin E TPGS. PVP-I acts as a preservative to help inhibit microbial growth in the irrigation solution. | BD Surgiphor™ Antimicrobial Irrigation System is an antimicrobial irrigation system containing 0.5% povidone-iodine (PVP-I) in phosphate-buffered saline, potassium iodide and vitamin E TPGS. PVP-I acts as a preservative to help inhibit microbial growth in the irrigation solution. | Identical to predicate |
| Intended Use | Intended for wound cleansing and removal of wound debris | Intended for wound cleansing and removal of wound debris | Identical to predicate |
| Indications For Use | BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds. | BD Surgiphor™ Antimicrobial Irrigation System is intended to mechanically loosen and remove debris, and foreign materials, including microorganisms, from wounds. | Identical to predicate |
| Type of Use | Prescription use only | Prescription use only | Identical to predicate |
| Mechanism of Action | The mechanical action of fluid across the wound removes wound debris, including microorganisms. | The mechanical action of fluid across the wound removes wound debris, including microorganisms. | Identical to predicate |
| Solution | Surgiphor solution (0.5% PVP-I plus vitamin E TPGS in 0.9% saline) | Surgiphor solution (0.5% PVP-I plus vitamin E TPGS in 0.9% saline) | Identical to predicate |
| Solution Antimicrobial Preservative | 0.5% PVP-I | 0.5% PVP-I | Identical to predicate |
Page 3 of 5
{7}
K253996
| Comparison Feature | Subject Device: BD Surgiphor™ Antimicrobial Irrigation System | Predicate Device: BD Surgiphor™ Antimicrobial Irrigation System | Subject to Predicate Comparison |
| --- | --- | --- | --- |
| How Supplied | 1 – 450 mL Surgiphor solution (0.5% PVP-I plus vitamin E TPGS in 0.9% saline); pH 4.6 – 7.0
1 – Manual Screw Cap
Packed within a PETG heat-sealed tray with a Tyvek cover and sterilized by gamma irradiation to achieve a SAL of 10-6. Instructions for Use are included with the system. | 1 – 450 mL Surgiphor solution (0.5% PVP-I plus vitamin E TPGS in 0.9% saline); pH 4.6 – 7.0
1 – Manual Screw Cap
Packed within a PETG heat-sealed tray with a Tyvek cover and sterilized by gamma irradiation to achieve a SAL of 10-6. Instructions for Use are included with the system. | Identical to predicate |
| Storage Conditions | Store at room temperature. Avoid freezing or heating above 40°C (104°F). | Store at room temperature. Avoid freezing or heating above 40°C (104°F). | Identical to predicate |
| Applicator | Polycarbonate cap with spike threads onto the high-density polyethylene (HDPE) bottle. The user squeezes the bottle to dispense the solution onto the wound. | Polycarbonate cap with spike threads onto the polypropylene bottle. The user squeezes the bottle to dispense the solution onto the wound. | The material for the bottle containing Surgiphor solution is changed to HDPE from polypropylene. |
{8}
K253996
# Standards and Testing
The following tests were conducted to support the changes under this 510(k):
- Biocompatibility
- ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- Endotoxin and Pyrogens
- USP General Chapter <85>, Bacterial Endotoxins Test
- (AAMI) ST72: 2019/R2010, Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing, Includes Erratum
- Sterilization
- AANSI/AAMI/ISO 13004:2022, Sterilization of Health Care Products – Radiation – Substantiation of selected sterilization dose: Method VDmax$^{SD}$
- ANSI/AAMI/ISO 11737-1:2018, Sterilization of health care products — Microbiological methods — Part 1: Determination of a population of microorganisms on products
- ANSI/AAMI/ISO 11737-2:2019, Sterilization of health care products — Microbiological methods — Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- Functional Testing
- Customized tests to demonstrate mechanical action to loosen and remove wound debris/foreign materials
# Substantial Equivalence Conclusion
The subject BD Surgiphor™ Antimicrobial Irrigation System is substantially equivalent to the previously cleared BD Surgiphor™ Antimicrobial Irrigation System (K221504) with the change described within this submission. The change does not impact the safety or effectiveness of the subject device.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.