The Veritas Reverse Total Shoulder System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants, and a functional deltoid muscle is necessary. The Veritas Reverse Total Shoulder System standard components are indicated for primary, fracture and revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The system's patient specific components are only indicated for primary total shoulder replacement. Glenoid components with porous surface are indicated for uncemented application with the addition of screw fixation. Humeral components with porous surface are indicated for either cemented or uncemented applications. Note: A CT scan is used to create the Veritas™ patient specific components.
Device Story
Veritas Reverse Total Shoulder System is a reverse shoulder arthroplasty device featuring patient-specific glenoid baseplates and standard humeral components. Input: patient CT scans. Process: AI-based auto-segmentation software assists in designing patient-specific implants with custom backside geometry and fixation. Output: 3D-printed implants (glenoid baseplate, glenosphere, humeral stem, polymer bearing) and supporting instrumentation. Used in clinical settings by orthopedic surgeons for primary, fracture, or revision shoulder replacement. Benefits: provides patient-specific anatomical fit for patients with rotator cuff deficiency and severe arthropathy, aiming to relieve pain and restore function.
Clinical Evidence
Bench testing only. Performance testing included Glenoid Baseplate Loosening (ASTM F2028), Torsional Taper Resistance, Glenoid Baseplate Fatigue, Range of Motion Analysis (ASTM F1378), Impingement analysis, and Cadaver Validation. AI algorithm performance validated against manual segmentation ground truth; DICE scores ≥ 0.87 (Scapula 0.88, Humerus 0.90) and Mean Hausdorff Distance ≤ 3.0 (Scapula 0.98, Humerus 0.68).
Technological Characteristics
Additively manufactured metal implants with integrally built 3D-printed porous surfaces. Includes patient-specific glenoid baseplate and standard humeral components. Testing performed per ASTM F2028 and ASTM F1378. Software includes AI-based auto-segmentation for CT-based patient-specific design.
Indications for Use
Indicated for patients with grossly deficient rotator cuff and severe arthropathy or failed shoulder replacement. Requires functional deltoid muscle and anatomical/structural suitability for implants. Standard components for primary, fracture, and revision total shoulder replacement. Patient-specific components for primary total shoulder replacement only.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
restor3d Reverse Total Shoulder Arthroplasty System (K243643)
Reference Devices
Comprehensive Vault Reconstruction System (VRS) (K152754)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
May 7, 2026
Restor3d
Brianna Prindle
Director of Regulatory Affairs
4001 Nc-54 Hwy
Suite 2160
Durham, North Carolina 27709
Re: K253992
Trade/Device Name: Veritas Reverse Total Shoulder System
Regulation Number: 21 CFR 888.3660
Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Code: PHX, QHE
Dated: April 7, 2026
Received: April 7, 2026
Dear Brianna Prindle:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K253992 - Brianna Prindle
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253992 - Brianna Prindle
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FARZANA SHARMIN -S
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253992 | ? |
| Please provide the device trade name(s). | | ? |
| Veritas Reverse Total Shoulder System | | |
| Please provide your Indications for Use below. | | ? |
| The Veritas Reverse Total Shoulder System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants, and a functional deltoid muscle is necessary. The Veritas Reverse Total Shoulder System standard components are indicated for primary, fracture and revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The system's patient specific components are only indicated for primary total shoulder replacement. Glenoid components with porous surface are indicated for uncemented application with the addition of screw fixation. Humeral components with porous surface are indicated for either cemented or uncemented applications. Note: A CT scan is used to create the Veritas™ patient specific components. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
K253992
Page 1 of 3
restor3d
4001 NC-54 Hwy Suite 3160
Durham, NC 27709
# 510(k) Summary
Date Prepared: May 7, 2026
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
A. 510(k) Owner:
restor3d, inc.
4001 NC-54 Hwy, Suite 3160
Durham, NC 27709
B. Primary Correspondent:
Brianna Prindle
Director of Regulatory Affairs
brianna@restor3d.com
C. Premarket Notification:
Submission Type: Traditional 510(k)
Trade Name: Veritas Reverse Total Shoulder System
Classification Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulation Number: 888.3660
Product Code: PHX, QHE
Classification: II
Review Panel: Orthopedic
D. Predicate Devices:
The Veritas Reverse Total Shoulder Arthroplasty System is substantially equivalent to the following devices:
| 510(k) | Trade Name |
| --- | --- |
| Primary Predicate Device | |
| K243643 | restor3d Reverse Total Shoulder Arthroplasty System |
| Reference Device | |
| K152754 | Comprehensive Vault Reconstruction System (VRS) |
E. Indications for Use:
The Veritas Reverse Total Shoulder System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The patient must be anatomically and structurally suited to receive the implants, and a functional deltoid muscle is necessary.
Veritas Reverse Total Shoulder Arthroplasty System
{5}
K253992
restor3d
4001 NC-54 Hwy Suite 3160
Durham, NC 27709
The Veritas Reverse Total Shoulder System standard components are indicated for primary, fracture and revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency. The Veritas system's patient specific components are only indicated for primary total shoulder replacement.
Glenoid components with porous surface are indicated for uncemented application with the addition of screw fixation.
Humeral components with porous surface are indicated for either cemented or uncemented applications.
Note: A CT scan is used to create the Veritas™ patient specific components.
F. Device Description:
The Veritas Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The Veritas Reverse Total Shoulder Arthroplasty System consists of the patient-specific glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting instrumentation, both patient-specific and standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The proposed device is a robust patient-specific offering for a reverse total shoulder including baseplate offerings with patient specific backside geometry and central fixation location.
The Veritas Reverse Total Shoulder Arthroplasty System includes an AI-based algorithm for auto segmentation of the CT scans within the patient-specific software workflow. The auto segmentation is intended to be used with human intervention for manual cleanup and review. Similar to the indications for use of the Veritas Reverse Total Shoulder Arthroplasty, auto segmentation has only been validated for use in primary shoulder arthroplasty. Ground truth was established through manual segmentation of the CT Scans. The testing data was independent from the training and tuning data, and testing data was from independent clinical sites. CT Scans were collected following the proposed CT Scan protocol for the Veritas Reverse Total Shoulder System. The data set information is provided in the table below.
The algorithm met predefined DICE score-based acceptance criteria of $\geq 0.87$ compared to ground truth manual segmentation and was consistent across Scapula (95% CI: 0.88), and Humerus (95% CI: 0.90). The algorithm also met predefined Mean Hausdorff Distance (MHD) based acceptance criteria of $\leq 3.0$ and the result was consistent across Scapula (95% CI: 0.98), and Humerus (95% CI: 0.68)
G. Substantial Equivalence Comparison:
Substantial equivalence of the subject Veritas Reverse Shoulder System to the primary predicate restor3d rTSA (K243643) is based on the following:
- The subject Veritas Reverse Shoulder System has indications for use that are a subset of the indications for use of the predicate restor3d rTSA, with the same intended use and meeting the same acceptance criteria for performance testing.
- The subject Veritas Reverse Shoulder System and reference device VRS (cleared via K152754) share similar technological characteristics, with patient-specific design from CT imaging.
Veritas Reverse Total Shoulder Arthroplasty System
{6}
K253992
Page 3 of 3
# restor3d
4001 NC-54 Hwy Suite 3160
Durham, NC 27709
- The subject Veritas Reverse Shoulder System has the same technological characteristics as the primary predicate restor3d rTSA (K243643) as both devices are additively manufactured from the same material and have an integrally built 3D-printed porous surface.
## H. Comparison of Performance:
The subject restor3d Reverse Total Shoulder Arthroplasty System was subject to the following non-clinical performance tests to support the assertion of substantial equivalence:
- Glenoid Baseplate Loosening per ASTM F2028
- Torsional Taper Resistance
- Glenoid Baseplate Fatigue
- Range of Motion Analysis per ASTM F1378
- Impingement analysis
- Cadaver Validation
## I. Conclusions:
Based on the comparison of indications for use, intended use, and performance characteristics, as well as the performance testing conducted, the subject Veritas Reverse Total Shoulder System is shown to be substantially equivalent to the primary predicate restor3d rTSA, cleared via K243643.
Veritas Reverse Total Shoulder Arthroplasty System
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.