The device is a digital monitor intended for use in measuring the systolic pressure and diastolic pressure, as well as the pulse rate of adults and adolescents via a non-invasive auscultatory method in medical facilities or at home, with an inflatable cuff wrapped around the upper arm for arm circumferences ranging from 22 to 42 cm. The measurement ranges of the device are systolic pressure 40 ~ 280 mmHg (supported by automatic mode for 40 to 210 mmHg and manual mode for 210 to 280 mmHg), diastolic pressure 20 ~ 220 mmHg, and pulse rate 40 ~ 240 beats/min. The device can detect Irregular Heart Beat (IHB), Atrial Fibrillation, Tachycardia, and evaluate Blood Pressure Variability (BPV) and Time in Target Range (TTR) according to the measured blood pressure values.
Device Story
Automatic non-invasive blood pressure monitor; uses Korotkoff sound method via inflatable upper arm cuff. Measures systolic/diastolic pressure and pulse rate. Device includes main unit with control circuit, LCD, and battery. Intended for use in medical facilities or home by patients or healthcare providers. Software analyzes pulse-to-pulse intervals to detect Irregular Heart Beat (IHB), Atrial Fibrillation (AFib), and Tachycardia; calculates Blood Pressure Variability (BPV) and Time in Target Range (TTR). Output displayed on screen for user review. Data storage function included. Provides informational alerts to assist in monitoring; does not replace clinical diagnosis or therapy.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2018 on models KSY 3050A and KSY 6600A. Study 1 (KSY 3050A): 88 subjects (20-90 years), 264 data sets. Study 2 (KSY 6600A): 87 subjects (20-84 years), 261 data sets. Mean error and SD within ISO limits. AFib detection validated in 74 subjects (31-93 years) with 97.3% sensitivity and 100% specificity. No adverse effects reported.
Technological Characteristics
Non-invasive sphygmomanometer; Korotkoff sound sensing principle. Components: main unit, cuff, optional AC adapter. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), and IEC 80601-2-30 (performance). Biocompatibility per ISO 10993-1 (cytotoxicity, sensitization, irritation). Software level of concern: moderate.
Indications for Use
Indicated for adults and adolescents requiring non-invasive blood pressure (systolic/diastolic) and pulse rate monitoring in medical facilities or home settings. Cuff fits arm circumferences 22-42 cm. Detects IHB, AFib, Tachycardia; evaluates BPV and TTR.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP Progress (BP3T01-1B) (K251120)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 21, 2026
Beijing Hanvon Health Technology Co., Ltd.
Qingjun Yi
Regulatory Responsible
Rm. 162, 1st Floor And Rm. 256, 2nd Floor,
Hanwang Bldg. 5, #8 Dongbeiwangxi Rd., Haidian District
Beijing, Beijing 100193
China
Re: K253982
Trade/Device Name: Korotkoff Sound Blood Pressure Monitor (KSY6600A, KSY6620B, KSY6600, KSY3500A, KSY3500B, KSY3500, KSY3200A, KSY3200B, KSY3200, KSY3160B, KSY3150B, KSY3150, KSY3100A, KSY3110, KSY3110B, KSY3118B, KSY3128B, KSY3120B, KSY3120, KSY3100B, KSY3100, KSY3050A, KSY3080B, KSY3080, KSY3060B, KSY3050B, KSY3050 and KSY3160)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: December 10, 2025
Received: December 12, 2025
Dear Qingjun Yi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253982 - Qingjun Yi
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253982 - Qingjun Yi
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**STEPHEN C. BROWNING -S**
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253982 | ☑ |
| Please provide the device trade name(s). | | ☑ |
| Korotkoff Sound Blood Pressure Monitor (KSY6600A, KSY6620B, KSY6600, KSY3500A, KSY3500B, KSY3500, KSY3200A, KSY3200B, KSY3200, KH3160B, KSY3150B, KSY3150, KSY3100A, KSY3110, KSY3110B, KSY3118B, KSY3128B, KSY3120B, KSY3120, KSY3100B, KSY3100, KSY3050A, KSY3080B, KSY3080, KSY3060B, KSY3050B, KSY3050 and KH3160) | | |
| Please provide your Indications for Use below. | | ☑ |
| The device is a digital monitor intended for use in measuring the systolic pressure and diastolic pressure, as well as the pulse rate of adults and adolescents via a non-invasive auscultatory method in medical facilities or at home, with an inflatable cuff wrapped around the upper arm for arm circumferences ranging from 22 to 42 cm. The measurement ranges of the device are systolic pressure 40 ~ 280 mmHg (supported by automatic mode for 40 to 210 mmHg and manual mode for 210 to 280 mmHg), diastolic pressure 20 ~ 220 mmHg, and pulse rate 40 ~ 240 beats/min.
The device can detect Irregular Heart Beat (IHB), Atrial Fibrillation, Tachycardia, and evaluate Blood Pressure Variability (BPV) and Time in Target Range (TTR) according to the measured blood pressure values. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
510(k) #: K253982
510(k) Summary
Prepared on: 2026-04-21
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Beijing Hanvon Health Technology Co., Ltd. |
| --- | --- |
| Applicant Address | Room 162, 1st Floor and Room 256, 2nd Floor, Hanwang Building, Building 5, No.8 Dongbeiwangxi Road, Haidian District Beijing 100193 China |
| Applicant Contact Telephone | (010) 82786525 |
| Applicant Contact | Mr. Qingjun Yi |
| Applicant Contact Email | yiqingjun@hanwang.com.cn |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Korotkoff Sound Blood Pressure Monitor (KSY6600A, KSY6620B, KSY6600, KSY3500A, KSY3500B, KSY3500, KSY3200A, KSY3200B, KSY3200, KH3160B, KSY3150B, KSY3150, KSY3100A, KSY3110, KSY3110B, KSY3118B, KSY3128B, KSY3120B, KSY3120, KSY3100B, KSY3100, KSY3050A, KSY3080B, KSY3080, KSY3060B, KSY3050B, KSY3050 and KH3160) |
| --- | --- |
| Common Name | Noninvasive blood pressure measurement system |
| Classification Name | System, Measurement, Blood-Pressure, Non-Invasive |
| Regulation Number | 870.1130 |
| Product Code(s) | DXN, DXN |
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K251120 | Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model BP Progress (BP3T01-1B) | DXN |
# Device Description Summary
21 CFR 807.92(a)(4)
The Korotkoff Sound Blood Pressure Monitor is designed as an automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the Korotkoff sound method. And the monitor will indicate possibility of AFib if an irregularity from pulse to pulse intervals is detected during a measurement. Also Blood Pressure Variability (BPV) and Time in Target Range (TTR) can be indicated on the screen after the measurement.
The device has the data storage function in order for data reviewing, low voltage indication, which will be triggered when the battery is low.
The proposed device is intended to be used in medical facilities or at home. And the effectiveness of this sphygmomanometer has not been established in pregnant (including pre-eclamptic) patients.
The product is provided non-sterile, and not to be sterilized by the user prior to use.
The monitor is composed of main unit (including control circuit, LCD display, battery and keys), cuff and AC Adapter (optional).
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
K253982
Page 1 of 4
{5}
The device is a digital monitor intended for use in measuring the systolic pressure and diastolic pressure, as well as the pulse rate of adults and adolescents via a non-invasive auscultatory method in medical facilities or at home, with an inflatable cuff wrapped around the upper arm for arm circumferences ranging from 22 to 42 cm. The measurement ranges of the device are systolic pressure 40 ~ 280 mmHg (supported by automatic mode for 40 to 210 mmHg and manual mode for 210 to 280 mmHg), diastolic pressure 20 ~ 220 mmHg, and pulse rate 40 ~ 240 beats/min.
The device can detect Irregular Heart Beat (IHB), Atrial Fibrillation, Tachycardia, and evaluate Blood Pressure Variability (BPV) and Time in Target Range (TTR) according to the measured blood pressure values.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device and predicate share the same primary intended use and target population, well-established, non-invasive, upper-arm BP measurement technologies with comparable accuracy and safety. All differences are secondary features, not changes to indication or patient group.
The subject device extends the range by 2 cm (upper limit only), covering a standard adult arm size subset already addressed by cleared upper arm cuffs. No new mechanical, biocompatibility, or measurement risks are introduced. Performance is validated over the full range to meet accuracy requirements.
Ranges of both devices are clinically appropriate for adult BP/PR monitoring and consistent with cleared electronic sphygmomanometers. Accuracy and reliability are verified per applicable standards (e.g., ISO 81060-1).
Both devices can detect the appearance of irregular heartbeat during measurement and gives a warning signal with the reading once the irregular heartbeat is detected. While the subject device can also evaluate the atrial fibrillation, tachycardia detection; BPV and TTR evaluation based on measured BP, which are provide as informational alerts/metrics only; they do not modify measurement, diagnosis, or therapy. No new safety hazards are introduced; performance is validated to minimize false alerts without compromising core measurement.
All the verification evidence (e.g., product specification comparison table, software function test report, clinical validation data) confirm that all differences are minor, well-characterized, and do not affect the safe and effective use of the device for its intended purpose.
## Technological Comparison
21 CFR 807.92(a)(6)
1) Different Application scenarios
The subject device is intended to be used in both home and clinical environment while only home healthcare environment is claimed for the predicate. The application of such product in the clinical environments decrease the application risks since it will be operated by healthcare person. Therefore, the difference does not raise any issues concerning safety and effective.
2) Displayed info.
The suggestive Atrial Fibrillation (AFib), Blood Pressure Variability (BPV) and Time in Target Range (TTR) is not available for the predicate device, these displayed information are results from software analysis base on the DIA, SYS and the pulse rate information, and these function has been well demonstration by the software verification and validation, so this difference will not the difference does not raise any issues concerning safety and effective.
3) Different Blood pressure measurement&Heart rate range and accuracy
The Blood pressure measurement&Heart rate ranges of the subject device is minor different from that of the predicate device, but the measurement range of the subject device has been verified by IEC 60601-1 and IEC 80601-2-30, so the different range will be acceptable for the subject device.
Regarding the accuracy, both are also well evaluated per the international standards IEC 80601-2-30 and ISO 81060-2, and the results show compliance with the requirements defined, no safety or effectiveness will be affected by this difference.
4) Different Cuff circumference
The range of cuff circumference is minor differently from that of the predicate device with different blood pressure measurement accuracy. The differences are very slightly and they both contain the blood pressure range of most people, and the measurement range of proposed device is fully verified according to IEC 80601-2-30 and the system clinical accuracy validation per ISO 81060-2, so the different range and accuracy will be acceptable for the subject device.
5) Operating and storage environment
The operation and storage conditions of subject device is different from that of predicate devices, but they all complied with the IEC 60601-1-11 and IEC 60601-1, so the difference will not raise any new safety and effectiveness issues.
6) Different Power supply and IP rating
The power supply and IP rating between the subject device and the predicate device are different, the battery of the subject device has been tested to the IEC 62133-2. In addition, the subject device has been tested to IEC 60601-1, IEC 60601-1-11 and IEC 60601-1-2 to evaluate the electrical safety including the IP rating. Therefore, the difference does not raise any new safety and effectiveness issues.
The subject and predicate devices have same design principle, similar design features and performance specifications. The different technological characteristics between the subject and predicate devices will not raise different questions of safety or effectiveness as demonstrated in the non-clinical and clinical evidence.
K253982
Page 2 of 4
{6}
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The following performance data were provided in support of the substantial equivalence determination.
1) Non-Clinical Testing:
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility testing
The biocompatibility evaluation for the Korotkoff Sound Bood Pressure Monitor was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests, results of which demonstrate the biocompatibility of the subject device:
- Cytotoxicity
- Skin Sensitization
- Skin Irritation
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the IEC 60601-1: 2005 +AMD1 (2012) +AMD2 (2020) Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2014/AMD1:2020 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC.
Bench Testing
Bench testing was conducted and the results show that the subject device complies with the IEC 80601-2-30: Medical electrical equipment – Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers. And blood pressure Accuracy meets the requirements defined in IEC 80601-2-30.
Home-used medical equipment requirements and environmental test:
Environmental testing was conducted and the test results show that the subject device complies with the IEC 60601-1-11:2015 +A1:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
Usability Evaluation
The monitor is evaluated per IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
Software Verification and Validation Testing
Software documentation including verification & validation was provided in accordance with FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern.
Software Verification and Validation Testing
Software verification and validation testing were conducted and Basic Documentation Level was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." since a failure or latent flaw of the device software function(s) (e.g., software fails to properly measure the blood pressure or fail to display any prompt) would not present a hazardous situation with a probable risk of death or serious injury to either a patient, user of the device, or others in the environment of use prior to the implementation of risk control measures..
Clinical data:
Substantial equivalence was based in part on the clinical study.
Blood pressure measurement accuracy has been performed on KSY 3050A and KSY 6600A respectively to cover all the models involved in this submission in accordance with ISO 81060-2:2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2: 2018.
For the clinical study on KSY 3050A, a total of 90 people aged 20 to 90 years old volunteered to participate in the clinical verification trial of the Korotkoff Sound Blood Pressure Monitor. All completed the trial and successfully withdrew from the group. The data of 88 people met the clinical evaluation requirements, generating 264 sets of evaluation data. There are 47 males, accounting for 53.4%, and 41 females, accounting for 46.6%. The systolic blood pressure range is 89.0-200.0mmHg, and the diastolic blood pressure range is 50.0-117.0 mmHg.
For the clinical study on KSY 6600A, a total of 90 people aged 20 to 84 years old volunteered to participate in the clinical verification trial of the Korotkoff Sound Blood Pressure Monitor. All completed the trial and successfully withdrew from the group. The data of 87 people met the clinical evaluation requirements, generating 261 sets of evaluation data. There are 36 males, accounting for 41.4%, and 51 females, accounting for 58.6%. The systolic blood pressure range is 86.0-179.0mmHg, and the diastolic blood pressure range is 46.0-116.0 mmHg.
And clinical verification regarding the Atrial Fibrillation notification is also conducted with a 97.3% sensitivity and 100% specificity. A
K253982
Page 3 of 4
{7}
total of 74 volunteers participated in the clinical verification test of the AFib function of the Korotkoff Sound Blood Pressure Monitor. They all completed the trial and successfully withdrew from the group, generating 74 sets of evaluation data. Aged 31 to 93, 27 males accounting for $36.5\%$ , and 47 females, accounting for $63.5\%$ . Systolic blood pressure range is 88-180mmHg, diastolic blood pressure range is 53-112mmHg, pulse range is 43-140 beats/min.
No adverse effect and/or complication is found in the clinical study.
Summary
Based on the clinical performance as documented in the clinical study, the monitor was found to have a safety and effectiveness profile that is similar to the predicate device.
K253982
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.