The Radical 6F Access Catheter is indicated for the introduction of intravascular catheters into the peripheral, coronary, and neuro vasculature.
Device Story
The Radical 6F Access Catheter is a sterile, single-use intravascular catheter designed to facilitate access to target vascular regions. It is operated by physicians trained in interventional techniques. The device is advanced through the vasculature under fluoroscopy, typically using standard percutaneous techniques. A rotating hemostasis valve (RHV) is attached to the proximal luer to maintain a continuous lumen for saline infusion or device passage. The catheter features variable stiffness along its length, achieved through hybrid ribbon technologies, and a hydrophilic coating on the distal region to improve trackability. It serves as a conduit for other interventional devices, assisting in reaching target sites in the peripheral, coronary, or neuro vasculature. The device provides stability and navigation support, potentially improving procedural outcomes by allowing precise placement of therapeutic catheters.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including visual/dimensional inspection, simulated use in anatomical models, mechanical testing (tensile, torque, kink, burst, leak, corrosion), tip stiffness comparison, particulate analysis, and coating integrity assessment. Biocompatibility was evaluated per ISO 10993-1 for limited exposure, blood-contacting devices.
Indicated for the introduction of intravascular catheters into the peripheral, coronary, and neuro vasculature in patients requiring interventional procedures.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 9, 2026
Maduro Medical, Inc.
Janice Kemp
Director RA/QA
983 University Ave.
Bldg. A
Los Gatos, California 95032
Re: K253975
Trade/Device Name: Radical 6F Access Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: QJP, DQY
Dated: March 13, 2026
Received: March 13, 2026
Dear Janice Kemp:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253975 - Janice Kemp
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K253975 - Janice Kemp
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**JAIME RABEN -S**
for Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253975 | |
| Device Name Radical 6F Access Catheter | |
| Indications for Use (Describe) The Radical 6F Access Catheter is indicated for the introduction of intravascular catheters into the peripheral, coronary, and neuro vasculature. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
MADURO
# 510(k) SUMMARY K253975
Manufacturer/Sponsor: Maduro Medical, Inc.
983 University Avenue
Building A
Los Gatos, California 95032
Phone: (408) 600-2235
Contact: Janice Kemp
QA/RA Director
(408) 600-2235
janice@maduromed.com
Date Prepared: April 07, 2026
Device Trade Name: Radical 6F Access Catheter
Common/Usual Name: Catheter, Percutaneous, Neurovasculature
Classification: 21 CFR 870.1250, Percutaneous Catheter
Class: II
Product Code: QJP, DQY
Predicate Device: Radical the Dude® 7F Guide Catheter (K231393)
## Indications for Use
The Radical 6F Access Catheter is indicated for the introduction of intravascular catheters into the peripheral, coronary, and neuro vasculature.
## Device Description
The Radical 6F Access Catheters are sterile, single-use intravascular catheters intended to facilitate access to target vascular regions during interventional procedures. The Radical 6F Access Catheter is offered in various working lengths, and an outer diameter (OD) size designation of 6 French (Fr). The Radical 6F Access Catheter has variable stiffness along its length, incorporating hybrid ribbon technologies to maintain stability, and vary stiffness throughout the device length. The distal region of the Radical 6F Access Catheter has a hydrophilic coating.
{5}
MADURO
# Principles of Operation
The Radical 6F Access Catheter may be used with support catheters to assist in accessing target vasculature. During use, the male luer of the RHV is attached to the proximal luer of the Radical 6F Access Catheter to create a continuous lumen through the catheter and to the RHV ports. The female luer of the RHV is typically connected to a saline drip line while the Radical 6F Access Catheter is advanced through the vasculature. Use of the Radical 6F Access Catheter relies on standard percutaneous interventional techniques, including access site preparation, introducing the catheter portion of the device, advancing the catheter under fluoroscopy, withdrawing the catheter, and closing the access site. Intended users of the Radical 6F Access Catheter are physicians who have received appropriate training in interventional techniques. The devices are provided sterile, non-pyrogenic, and are intended for single use only.
# Comparison of Technological Characteristics with the Predicate Device
The predicate device is the Radical the Dude 7F Guide Catheter, K231393. The only differences being the smaller inner and outer diameters and shorter length of the subject device. The predicate and subject devices share the same intended use and basic technological characteristics as shown in Table 1. A comparison of the technological characteristics of the subject, predicate and reference devices is shown in Table 1.
{6}
MADURO
Table 1: Comparison of Subject and Predicate Devices
| Device Attribute | Subject Device | Predicate Device | Reference Device |
| --- | --- | --- | --- |
| Product Name | Radical 6F Access Catheter | Radical the Dude 7F Guide Catheter | Benchmark Delivery Catheter |
| 510(k) Number | K253975 | K231393 | K212838 |
| Indications for Use | The Radical 6F Access Catheter is indicated for the introduction of intravascular catheters into the peripheral, coronary, and neuro vasculature. | The Radical the Dude 7F Guide Catheter is indicated for the introduction of intravascular catheters into the peripheral, coronary, and neuro vasculature. | The Benchmark Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. |
| Product Code | QJP, DQY | QJP, DQY | QJP, DQY |
| Regulation No. | 21 CFR 870.1250 | 21 CFR 870.1250 | 21 CFR 870.1250 |
| Classification | Class II | Class II | Class II |
| Components Supplied | Catheter, Peel-away Sheath, Rotating Hemostasis Valve (RHV) | Same | n/a |
| Materials | | | |
| Catheter Shaft Material | Urethane (Tecoflex), Pebax, Nylon | Same | Commonly used medical grade plastics & stainless steel |
| Inner Liner | PTFE | Same | n/a |
| Hub Material | Polypropylene Copolymer (Repsol HPR35CMD) | Nylon (Grilamid) | n/a |
| Strain Relief | Polyolefin | Same | n/a |
| Catheter Shaft Reinforcement | Braid: 304V Stainless Steel Coil: 304V Stainless Steel | Same | n/a |
| Lubricious Coating | Hydrophilic Coating | Same | n/a |
| Radiopaque Marker Band | Platinum/Iridium | Same | n/a |
| Peel-away Sheath | PTFE | Same | n/a |
{7}
MADURO
| Device Attribute | Subject Device | Predicate Device | Reference Device |
| --- | --- | --- | --- |
| RHV | Polycarbonate, Silicone | Same | n/a |
| Dimensions | | | |
| Working Length | 80, 90, 95, 100, 105, 110, 115 cm | 95, 105, 115 cm | 95, 105, 115 cm |
| Inner Diameter | 0.072 inches | 0.082 inches | 0.070 inches minimum |
| Outer Diameter | Proximal: 0.085 inches
Distal: 0.083 inches | Proximal: 0.098 inches
Distal: 0.094 inches | 0.081 – 0.083 inches |
| Packaging | Tyvek/ Nylon/polyethylene (PE) Pouch, PE tube, Packaging card, SBS carton | Same | Commonly used medical device packaging materials |
| Sterilization | Ethylene Oxide (EO) SAL 10^{-6} | Same | Same |
| Pyrogenicity | Nonpyrogenic | Same | Same |
| Number of Uses | Single Use | Same | Same |
4
{8}
MADURO
# Nonclinical Performance Testing
The following nonclinical performance testing was conducted to demonstrate substantial equivalence.
## Bench Testing
Table 2 lists the bench testing performed to demonstrate substantial equivalence.
Table 2: Bench Testing Summary
| Test | Test Method/Applicable Standard | Result |
| --- | --- | --- |
| Visual Inspection | Visual inspection completed for surface defects. | Pass |
| Dimensional Inspection | Critical dimensions were verified. | Pass |
| Simulated Use Test | Simulated use in a bench anatomical model with femoral artery and radial artery access. | Pass |
| PTFE Delamination | Assessed for PTFE delamination at distal tip following simulated use testing. | Pass |
| Tensile Testing | Tensile strength measured along entire catheter length. | Pass |
| Torque Strength | The distal end of the catheter was constrained from movement while the proximal end was turned until failure in a simulated anatomy model. | Pass |
| Kink Resistance | Resistance to kink tested at various locations along the catheter shaft using successively smaller radii to challenge the catheter. | Pass |
| Catheter Burst | Catheter burst tested per ISO 10555-1. | Pass |
| Liquid Leak Test | Liquid leak tested per ISO 10555-1. | Pass |
| Air Leak Test | Air leak tested per ISO 10555-1. | Pass |
| Corrosion | Corrosion tested per ISO 10555-1. | Pass |
| Tip Stiffness | Compared the tip stiffness of the Radical 6F Catheter with the reference device. | Pass |
| Particulate Testing | During simulated use testing in an in vitro model the particle size and count were analyzed using light obscuration method and compared to the predicate. | Pass |
| Coating Integrity | The integrity of the hydrophilic coating was inspected before and after simulated use testing in an in vitro model. | Pass |
## Biocompatibility
The subject Radical 6F Access Catheter is categorized as a limited exposure ( $\leq 24$ hours), externally communicating device with circulating blood contact in accordance with ISO 10993-1 and FDA guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". The subject device is constructed using materials that are commonly used in the medical device industry. The subject device uses the same materials or substantially equivalent materials and same manufacturing processes as the predicate device. The following biocompatibility endpoints were assessed to confirm biocompatibility of the subject device:
{9}
MADURO
- Cytotoxicity
- Sensitization
- Intracutaneous Reactivity
- Acute Systemic Toxicity
- Material-mediated Pyrogenicity
- Direct and Indirect Hemolysis
- Complement Activation
- Thrombogenicity
## Animal Testing
An animal study was not deemed necessary to support the substantial equivalence of the subject device to the predicate device. Bench testing was determined sufficient to support substantial equivalence.
## Sterilization and Shelf-Life
The Radical 6F Access Catheter is sterilized using a validated ethylene oxide (EO) process with a sterility assurance level of $1 \times 10^{-6}$. The sterilization method is identical for the subject and predicate devices.
The subject device’s shelf-life and packaging configuration remain identical to that of the predicate device. Therefore, no shelf life testing or packaging validation was required to demonstrate substantial equivalence between the subject and predicate devices.
## Clinical
The non-clinical performance testing was determined to be sufficient to support the substantial equivalence of the subject device.
## Conclusion
The subject Radical 6F Access Catheter has similar intended use, indications for use, principles of operation, and technological characteristics as the predicate device. The technological differences identified do not raise new questions of safety or effectiveness between the subject and predicate devices. Performance testing demonstrates that the Radical 6F Access Catheter is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.