The Cubresa BrainPET is a Positron Emission Tomography (PET) scanner, containing MR transmit and Receive head coils, and is intended for use with the Siemens Biograph mMR with syngo software to provide high resolution physiologic and anatomic information via PET and MR modalities acquired simultaneously and isocentrically. The Cubresa BrainPET maintains independent functionality of the MR and PET devices, allowing for single-modality MR or simultaneous PET/MR imaging. The Cubresa BrainPET is intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders of the brain. The Cubresa BrainPET MR Coil is intended to produce transverse, sagittal, coronal, and oblique cross-sectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human brain. The Cubresa BrainPET images and measures the distribution of PET radiopharmaceuticals in human head to aid in the evaluation of various metabolic (molecular) and physiologic functions within the human brain for evaluation of diseases and disorders such as, but not limited to, neurological diseases and disorders. The Cubresa BrainPET utilizes the MR for radiation-free attenuation correction maps for PET studies. The system provides inherent anatomical reference for the PET and MR images due to precisely aligned MR and PET image coordinate systems.
Device Story
Removable PET scanner insert with integrated MR head coil; designed for insertion into Siemens Biograph mMR scanner; enables simultaneous or independent PET/MR brain imaging. PET component uses dual-layer offset LYSO scintillators and SiPM photodetectors; MR component uses 1Tx/32Rx head array. System includes EM-quiet cabinet for data acquisition, copper cabling, and workstation for reconstruction/calibration. Operated by trained healthcare professionals in MR suite. PET data reconstructed via OSEM; MR data processed via Siemens syngo software. Provides high-resolution anatomical/physiologic images; MR-based attenuation correction for PET. Benefits include simultaneous acquisition for precise anatomical reference and metabolic evaluation of neurological disorders.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing (NEMA NU 2-2018 for PET performance, NEMA MS series for MRI performance) and expert review of sample clinical images by board-certified radiologists/nuclear medicine physicians to confirm diagnostic quality.
Technological Characteristics
Hybrid PET/MR insert; PET: 4 rings, 64 LYSO scintillator modules, SiPM photodetectors. MR: 1Tx/32Rx head array, 3T field strength. Connectivity: Copper cabling to EM-quiet cabinet, fiber to workstation. Software: 64-bit Ubuntu Linux for PET, Siemens syngo for MR. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA NU 2-2018, ISO 10993-1.
Indications for Use
Indicated for use by trained healthcare professionals to aid in the detection, localization, and diagnosis of brain diseases and disorders in patients undergoing PET/MR imaging. Used for metabolic and physiologic evaluation of the brain via PET radiopharmaceutical distribution and MR imaging (transverse, sagittal, coronal, oblique cross-sections, and spectroscopy).
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
Biograph mMR with syngo MR E11P system software (K163234)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 16, 2026
Cubresa, Inc.
% Sanjay Shah
Regulatory Affairs Consultant
Device Pathways, LLC
5054 Merrimac Ln N
Plymouth, Minnesota 55446
Re: K253963
Trade/Device Name: Cubresa BrainPET™
Regulation Number: 21 CFR 892.1200
Regulation Name: Emission Computed Tomography System
Regulatory Class: Class II
Product Code: OUO, LNH, MOS, KPS
Dated: June 21, 2026
Received: June 22, 2026
Dear Sanjay Shah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253963 - Sanjay Shah
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253963 - Sanjay Shah
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.07.17 13:17:11 -04'00' For
Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253963 | ? |
| Please provide the device trade name(s). | | ? |
| Cubresa BrainPET™ | | |
| Please provide your Indications for Use below. | | ? |
| The Cubresa BrainPET is a Positron Emission Tomography (PET) scanner, containing MR transmit and Receive head coils, and is intended for use with the Siemens Biograph mMR with syngo software to provide high resolution physiologic and anatomic information via PET and MR modalities acquired simultaneously and isocentrically. The Cubresa BrainPET maintains independent functionality of the MR and PET devices, allowing for single-modality MR or simultaneous PET/MR imaging. | | |
| The Cubresa BrainPET is intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders of the brain. | | |
| The Cubresa BrainPET MR Coil is intended to produce transverse, sagittal, coronal, and oblique cross-sectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human brain. | | |
| The Cubresa BrainPET images and measures the distribution of PET radiopharmaceuticals in human head to aid in the evaluation of various metabolic (molecular) and physiologic functions within the human brain for evaluation of diseases and disorders such as, but not limited to, neurological diseases and disorders. | | |
| The Cubresa BrainPET utilizes the MR for radiation-free attenuation correction maps for PET studies. The system provides inherent anatomical reference for the PET and MR images due to precisely aligned MR and PET image coordinate systems. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
CUBRESA
Cubresa BrainPET 510(k) Summary
K253963
## 510(k) Summary
This 510(k) summary is submitted in accordance with the requirements of 21 CFR §807.92.
### I. SUBMITTER
**Establishment:** Cubresa Inc.
136 Market Ave. Suite 800 Winnipeg MB R3B 0P4
Canada
Registration Number: N/A
**Date Prepared:** July 16, 2026
**Applicant Contact** Lisa Bako
VP Sales & Strategic Partnerships
Telephone: 204-798-4579
Email: lbako@cubresa.com
**Correspondent Contact** Device Pathways LLC
Mr. Sanjay Shah
Regulatory Affairs Consultant
Telephone: 612-812-6462
Email: sshah@devicepathways.com
### II. DEVICE
**Device Trade Name:** Cubresa BrainPET™
**Common Name:** Emission computed tomography system
**Classification Name:** Tomographic Imager Combining Emission Computed Tomography with Nuclear Magnetic Resonance
**Regulatory Class:** II
**Regulation Number:** 21 CFR § 892.1200
21 CFR § 892.1000
**Product Codes:** Primary: OUO
Secondary: LNH, MOS, KPS
### III. PREDICATE DEVICE
**Device Name:** Biograph mMR with syngo MR E11P system software
**510(k) Number:** K163234
**Common Name:** Emission computed tomography system
**Classification Name:** Tomographic Imager Combining Emission Computed Tomography with Nuclear Magnetic Resonance
**Regulatory Class** II
**Regulation Number:** 21 CFR § 892.1200
21 CFR § 892.1000
Page 1 of 15
{5}
[LOGO]
CUBRESA
Cubresa BrainPET 510(k) Summary
Product Code:
Primary: OUO
Secondary: LNH, LNI, KPS
### IV. DEVICE DESCRIPTION
The Cubresa BrainPET system is a removable positron emission tomography (PET) scanner with an integrated magnetic resonance imaging (MRI) coil. It is designed for insertion into a magnetic resonance imager (MRI) scanner, enabling hybrid PET/MR imaging of the brain, and is configured for use with the Siemens Biograph mMR with syngo software. The Cubresa BrainPET system maintains independent functionality of the MRI for single modality MR imaging or quickly enables simultaneous PET/MRI imaging of the brain.
The Cubresa BrainPET insert assembly is MR-conditional and has a cylindrical shape, with an axial length of 200 mm, ensuring ample coverage of the entire human head, and produces images of the distribution of PET radiopharmaceuticals within the brain of the subject being imaged. The PET camera has a dual layer offset scintillator design with Lutetium Yttrium Orthosilicate (LYSO), with 64 detector modules arranged in a cassette fashion across 4 identical PET rings. The custom-designed actively decoupled Transmit (Tx) and Receive (Rx) head coil slides into the inner diameter of the PET camera and is plugged directly into the MRI scanner to enable acquisition of MR images simultaneously with PET data. The Tx portion of the coil consists of 8 elements arranged in a circular birdcage orientation around the cylinder, while the 32 circular Rx elements are distributed over the entire helmet area to ensure adequate brain coverage. The insert is stored on an MR-conditional cart which facilitates movement of the insert around the MR suite as well as mounting and demounting of the insert on to the bed of the MRI scanner.
An electromagnetic (EM)-quiet cabinet containing a data acquisition chassis is located within the MR suite and connected to the insert via copper cabling. The cabinet is connected via fiber to a workstation in the MR equipment room and then routed to the control room. The workstation consists of a computer pre-loaded with Cubresa acquisition and reconstruction software which enables system calibration, attenuation map creation, data reconstruction, and export of reconstructed DICOM-tagged data to the user's picture archiving system and subsequent visualization on third party software.
### V. INDICATIONS FOR USE
The Cubresa BrainPET is a Positron Emission Tomography (PET) scanner, containing MR transmit and Receive head coils, and is intended for use with the Siemens Biograph mMR with syngo software to provide high resolution physiologic and anatomic information via PET and MR modalities acquired simultaneously and isocentrically. The Cubresa BrainPET maintains independent functionality of the MR and PET devices, allowing for single-modality MR or simultaneous PET/MR imaging.
The Cubresa BrainPET system is intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders of the brain.
The Cubresa BrainPET MR Coil is intended to produce transverse, sagittal, coronal, and oblique cross-sectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human brain.
Page 2 of 15
{6}
[Non-Text]
CUBRESA
Cubresa BrainPET 510(k) Summary
The Cubresa BrainPET images and measures the distribution of PET radiopharmaceuticals in the human head to aid in the evaluation of various metabolic (molecular) and physiologic functions within the human brain for evaluation of diseases and disorders such as, but not limited to, neurological diseases and disorders.
The Cubresa BrainPET utilizes the MR System for attenuation correction maps for PET studies. The system provides inherent anatomical reference for the PET and MR images due to precisely aligned MR and PET image coordinate systems.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Cubresa BrainPET is shown to be substantially equivalent to the predicate device Siemens Biograph mMR with syngo software by comparing the indications for use, principles of operation, technological characteristics, and performance testing similarities and differences.
The technological characteristic similarities and differences between the proposed Cubresa BrainPET and the Siemens Biograph mMR with syngo software predicate device are listed in Table 1.
The components used in the Cubresa BrainPET, including the insert assembly, electromagnetic-quiet cabinet, MR-conditional lift and storage device and workstation are substantially equivalent to the predicate Siemens Biograph mMR. While there are differences in technological characteristics for the Cubresa BrainPET compared to the predicate device, the differences have been tested and the conclusions from the verification and validation data support that the technological characteristics of the Cubresa BrainPET have an equivalent safety and performance profile to that of the Biograph mMR with syngo software predicate device (K163234). Successful completion of the standard QA tests and expert review of sample clinical images demonstrate that the Cubresa BrainPET maintains clinically acceptable MR and PET imaging performance. Based on the substantial equivalence analysis, the proposed Cubresa BrainPET is substantially equivalent to the predicate device.
Table 1 – Substantial Equivalence
| Characteristic | Predicate Device Siemens Biograph mMR with syngo MR E11P system software (K163234) | Proposed Device Cubresa BrainPET | Equivalence Comparison |
| --- | --- | --- | --- |
| Trade/Device Name | Biograph mMR with syngo MR E11P system software | Cubresa BrainPETTM | N/A |
| Regulation Number: | 21 CFR § 892.1200 21 CFR § 892.1000 | 21 CFR § 892.1200 21 CFR § 892.1000 | Same |
| Device Class | Class II | Class II | Same |
| Product Code | Primary: OUO Secondary: LNH, LNI, KPS | Primary: OUO Secondary: LNH, MOS, KPS | Same |
| Classification | Tomographic Imager | Tomographic Imager | Same |
Page 3 of 15
{7}
CUBRESA
Cubresa BrainPET 510(k) Summary
| Characteristic | Predicate Device Siemens Biograph mMR with syngo MR E11P system software (K163234) | Proposed Device Cubresa BrainPET | Equivalence Comparison |
| --- | --- | --- | --- |
| Name: | Combining Emission Computed Tomography with Nuclear Magnetic Resonance | Combining Emission Computed Tomography with Nuclear Magnetic Resonance | |
| Panel | Radiology | Radiology | Same |
| Intended Use/Indications for Use | The Siemens MR-PET system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders. The MR is intended to produce transverse, sagittal, coronal and oblique cross-sectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human body. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, approved contrast agents may be used, as described in their labeling. This | The Cubresa BrainPET is a Positron Emission Tomography (PET) scanner, containing MR transmit and Receive head coils, and is intended for use with the Siemens Biograph mMR with syngo software to provide high resolution physiologic and anatomic information via PET and MR modalities acquired simultaneously and isocentrically. The Cubresa BrainPET maintains independent functionality of the MR and PET devices, allowing for single-modality MR or simultaneous PET/MR imaging. The Cubresa BrainPET is intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders of the brain. The Cubresa BrainPET MR Coil is intended to produce transverse, sagittal, coronal, and oblique cross-sectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human brain. | Same |
Page 4 of 15
{8}
[Non-Text]
CUBRESA
Cubresa BrainPET 510(k) Summary
| Characteristic | Predicate Device Siemens Biograph mMR with syngo MR E11P system software (K163234) | Proposed Device Cubresa BrainPET | Equivalence Comparison |
| --- | --- | --- | --- |
| | system may also be used for imaging during interventional procedures when performed with MR compatible devices, such as MR safe biopsy needles.The PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.The combined system utilizes the MR for radiation-free attenuation correction maps for PET studies. The system provides inherent anatomical reference for the fused PET and MR images due to precisely aligned MR and PET image coordinate systems. | The Cubresa BrainPET images and measures the distribution of PET radiopharmaceuticals in human head to aid in the evaluation of various metabolic (molecular) and physiologic functions within the human brain for evaluation of diseases and disorders such as, but not limited to, neurological diseases and disorders.The Cubresa BrainPET utilizes the MR for radiation-free attenuation correction maps for PET studies. The system provides inherent anatomical reference for the PET and MR images due to precisely aligned MR and PET image coordinate systems. | |
| PET system technology | | | |
| Axial field of view | 258 mm | 200 mm | Equivalent |
| Transaxial field of view | 594 mm | 320 mm | Equivalent |
| Scintillators | Lutetium Orthosilicate (LSO) | Lutetium Yttrium Orthosilicate (LYSO) | Equivalent |
| Crystal cut | Crystal cut | Dual-layer offset design | Equivalent |
Page 5 of 15
{9}
CUBRESA
Cubresa BrainPET 510(k) Summary
| Characteristic | Predicate Device Siemens Biograph mMR with syngo MR E11P system software (K163234) | Proposed Device Cubresa BrainPET | Equivalence Comparison |
| --- | --- | --- | --- |
| Photodetectors | Avalanche Photodiodes (APDs) (water-cooled) | Silicon Photomultipliers (SiPMs) | Equivalent |
| PET data acquisition electronics | Housed within the body of the predicate device's PET system | Electronics housed within an EM-quiet cabinet located in the MR suite | Equivalent |
| PET camera storage mechanism | Predicate device, which has an integrated PET camera, is permanently fixed in place within a dedicated MR suite and does not required storage | Insert portion of the proposed device stored on a rolling cart that permits movement of the device around the MR suite to facilitate loading/unloading onto the bed of the MRI (when in use) and storage at the back of the MR suite (when not in use). | Equivalent |
| MRI System Technology Characteristics | | | |
| Field strength | 3T | 3T | Same |
| Bore strength | 60 cm | 60 cm | Same |
| System length | 199 cm | 199 cm | Same |
| System weight (in operation) | 6.3 tons | 6.3 tons | Same |
| Gradient strength | MQ gradients (45 mT/m @ 200 T/m/s) | MQ gradients (45 mT/m @ 200 T/m/s) | Same |
| Coil | Siemens mMR Head/Neck Matrix coil | 1H 1Tx32Rx Head Array | Different |
| MR atomic nuclei | 1H | 1H | Same |
| RF polarization | Linear | Quadrature | Equivalent |
| Field strength B0 | 3T | 3T | Equivalent |
Page 6 of 15
{10}
CUBRESA
Cubresa BrainPET 510(k) Summary
| Characteristic | Predicate Device Siemens Biograph mMR with syngo MR E11P system software (K163234) | Proposed Device Cubresa BrainPET | Equivalence Comparison |
| --- | --- | --- | --- |
| Operating frequency | 123.2 MHz | 123.2 MHz | Same |
| Signal to noise ratio (SNR) | Compliance with recognized standard test method | Compliance with NEMA MS1 test method | Equivalent |
| RF induced heating (SAR) | Compliance with IEC 60601-2-33 | Compliance with IEC 60601-2-33 | Equivalent |
| Software & Algorithms | | | |
| PET-related | - Operating system: Microsoft Windows - Listmode data acquisition in 3D mode (static, multibed) - PET Reconstruction: - Analytical: filtered back-projection (FBP) - Iterative: 3D ordinary Poisson ordered subset expectation maximization (OSEM) - Attenuation correction approaches: - MRAC (4 DIXON sequences) - MRextendedFOV (HUGE) - UTE-based - Scatter correction: Relative and Absolute Scaling - Dynamic imaging - Gating approaches: | - Operating system: 64-bit Ubuntu Linux - Static listmode data acquisition in 3D mode - PET Reconstruction: - Iterative: Ordered Subset Expectation Maximization (OSEM) - Attenuation correction hybrid approach: - DIXON+bone sequence (MR-based subject attenuation correction) plus CT-based hardware attenuation correction - Scatter correction: Single Scatter Simulation and Relative Scaling - Calibration includes generation of crystal, energy, activity, and timing calibration as well as normalization factors | Equivalent |
Page 7 of 15
{11}
[LOGO]
CUBRESA
Cubresa BrainPET 510(k) Summary
| Characteristic | Predicate Device Siemens Biograph mMR with syngo MR E11P system software (K163234) | Proposed Device Cubresa BrainPET | Equivalence Comparison |
| --- | --- | --- | --- |
| | optimal respiratory gating ○ multi-phase respiratory gating • Calibration includes generation of crystal, energy and activity calibration and normalization factors • Export of PET raw data (sinogram, list mode, normalization and physio files) Output PET images are DICOM-compliant | Output PET images are DICOM-compliant | |
| MR-related | Siemens syngo software is used for all MR data acquisition and processing | Siemens syngo software is used for all MR data acquisition and processing | Same |
| Safety | | | |
| Electrical safety | IEC 60601-1 compliance | IEC 60601-1 compliance | Same |
| EMC | IEC 60601-1-2 compliance | IEC 60601-1-2 compliance | Same |
| MRI Coil Safety and Performance | IEC 60601-2-33 compliance | ISO 10993-1 compliance | Same |
| Biocompatibility | ISO 10993-1 compliance | IEC 60601-1 compliance | Same |
| Sterility & Reprocessing | Instructions for cleaning/reprocessing are in the operator manual | Instructions for cleaning/reprocessing are in the operator manual | Same |
Page 8 of 15
{12}
CUBRESA
Cubresa BrainPET 510(k) Summary
The Cubresa BrainPET conforms to the FDA recognized consensus standards listed in Table 2.
**Table 2 – FDA Recognized Consensus Standards**
| Recognition Number | Product Area | Standard Designation Number and Date | Title of Standard | Standards Development Organization |
| --- | --- | --- | --- | --- |
| 5-125 | General I (QS/ RM) | ANSI AAMI ISO 14971:2019 | Medical devices - Application of risk management to medical devices | ANSI AAMI ISO |
| 19-46 | General II (ES/ EMC) | ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010 /(R)2012 (Cons. Text) [Incl. AMD2:2021] | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) | ANSI AAMI |
| 19-36 | General II (ES/ EMC) | 60601-1-2:2014 [Including AMD 1:2021] | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | ANSI AAMI IEC |
| 13-79 | Software/ Informatics | 62304:2006/A1:2016 | Medical device software - Software life cycle processes | ANSI AAMI IEC |
| 12-295 | Radiology | 60601-2-33 Ed. 3.2 b:2015 | Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis | IEC |
| 12-382 | Radiology | NEMA NU 2-2024 | Performance Measurements of Positron Emission Tomographs | NEMA |
| 2-258 | Biocompatibility | AAMI ANSI ISO 10993-1:2018 | Biological evaluation of medical devices - Part 1: evaluation and testing within a risk management process | AAMI ANSI ISO |
Page 9 of 15
{13}
CUBRESA
Cubresa BrainPET 510(k) Summary
### VII. PERFORMANCE TESTING
The Cubresa BrainPET has been designed to provide MR and PET imaging in the same manner as the predicate Siemens Biograph mMR with syngo software (K163234). CUBRESA performed testing according to the applicable guidance and regulatory standards to demonstrate that the Cubresa BrainPET diagnostic imaging features are substantially equivalent to features of the predicate Siemens Biograph mMR with syngo software. The Cubresa BrainPET does not raise any new safety or effectiveness issues related to the use with the MRI system. No clinical testing was required to demonstrate substantial equivalence. Sample clinical images were assessed and provided as required by the guidance document Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices. The Cubresa BrainPET has passed performance testing and met product specifications. CUBRESA performed non-clinical testing which is summarized in Table 3.
Table 3 – Non-Clinical Testing
| Testing Performed | Tested Hardware | Source/ Rationale for Test |
| --- | --- | --- |
| Non-clinical Performance Testing | Cubresa BrainPET (finished device) | Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices |
| Sample Clinical Images | Cubresa BrainPET (finished device) | Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices |
| Electrical, mechanical, structural, and related system safety | Cubresa BrainPET (finished device) | ANSI /AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021], ASTM F2213-25, ASTM F2052-21 |
| Electromagnetic compatibility | Cubresa BrainPET (finished device) | IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION |
| Software | Cubresa BrainPET (finished device) | Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices |
| Cybersecurity | Cubresa BrainPET (finished device) | Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions |
Cubresa has conducted and submitted the results of a comprehensive series of non-clinical tests. These evaluations were designed to verify and validate the device's safety performance, and MRI environment compatibility in accordance with applicable standards and FDA guidance documents. The following non-clinical tests were conducted or referenced:
### 1. Electrical Safety and Electromagnetic Compatibility (EMC):
Electrical Safety and Electromagnetic Compatibility testing of the Cubresa BrainPET System was conducted by an independent 3rd party test lab, and confirmed compliance with the following standards:
Page 10 of 15
{14}
CUBRESA
Cubresa BrainPET 510(k) Summary
- IEC 60601-1:2005 / A1:2012 / A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-2:2014 / A1:2020 (Ed 4.1) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
## 2. Software Verification and Validation:
Software verification and validation was performed in accordance with internal company processes which follow IEC 62304:2006 Medical device software — Software life cycle processes. Documentation is provided as recommended by the FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions” (CDRH, June 2023).
## 3. Cybersecurity
Cybersecurity testing was performed by an independent 3rd party test lab, and the documentation as recommended by the FDA’s Guidance for Industry and FDA Staff, “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” (CDRH, June 27, 2025) is provided.
## 4. Performance Bench Testing:
Performance characteristics recommended in the FDA’s Guidance for Industry “Guidance for the Submission of Premarket Notifications for Emission Computed Tomography Devices and Accessories (SPECT and PET) and Nuclear Tomography Systems” IX.C.2 (CDRH,1998) were evaluated on the Cubresa BrainPET System by Cubresa Inc., and supporting documentation has been provided.
The following performance characteristics, as specified in NEMA NU 2-2018: Performance Measurements of Positron Emission Tomography, were measured in accordance with the standard under two configurations: (1) PET-only mode (with no MRI sequences running) and (2) simultaneous PET/MR mode (with an MR protocol running). The measured characteristics include:
| Purpose | Methodology | Outcome |
| --- | --- | --- |
| Spatial Resolution | NEMA NU2-2018 (Section 3). Measures intrinsic spatial resolution of the scanner using a point or line source of ^{18}F at varying distances from the center of the field-of-view (FOV). Results are reported as Full Width at Half Maximum (FWHM) and Full Width at Tenth Maximum (FWTM). | Pass- The test results meet the Cubresa BrainPET requirements. |
Page 11 of 15
{15}
CUBRESA
Cubresa BrainPET 510(k) Summary
| Scatter Fraction, Losses, and Randoms (Count Rate Performance) | NEMA NU2-2018 (Section 4, Method 1). Uses a specialized cylindrical polyethylene phantom. Cubresa designed a modified phantom with an outer diameter of 177.8mm and is curved at the back of the phantom to fit with the curved helmet shape of the Cubresa BrainPET Tx/Rx Array. | Pass- The test results meet the Cubresa BrainPET requirements. |
| --- | --- | --- |
| Sensitivity | NEMA NU2-2018 (Section 5). Evaluates the scanner's efficiency by measuring count rates from a line source surrounded by varying thicknesses of metal sleeves (simulating different levels of attenuation). Sensitivity phantom modified for Cubresa BrainPET FOV. | Pass- The test results meet the Cubresa BrainPET requirements. |
| Image Quality, Accuracy of Corrections | NEMA NU2-2018 (Section 7). Employs a phantom containing background and hot/cold spheres to simulate clinical conditions like whole-body lesion detection, testing the accuracy of scatter and attenuation corrections. IQ Phantom spherical array per NEMA specification, modified tank body (no lung insert), scaled to fit within Cubresa BrainPET FOV. | Pass- The test results meet the Cubresa BrainPET requirements. |
### 5. MRI Compatibility Testing:
For use in conjunction with the Biograph mMR, the Cubresa BrainPET System was evaluated in accordance with the FDA's Guidance for Industry and FDA Staff, "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" (CDRH, 2023). Supporting documentation is provided. This evaluation included:
| Purpose | Methodology | Outcome |
| --- | --- | --- |
| Specific Absorption Rate (SAR) characterization | Surface temperature measurements and specific absorption rate (SAR) evaluations were conducted in accordance with NEMA MS 8-2016 Characterization of the Specific Absorption Rate (SAR) for Magnetic Resonance Imaging Systems, Section 3 Method. | Pass- The test results meet the Cubresa BrainPET requirements. |
Page 12 of 15
{16}
CUBRESA
Cubresa BrainPET 510(k) Summary
| Device Function and PET Performance in MRI Environment | The Cubresa BrainPET System's performance in the MRI environment was assessed to ensure it maintained functionality and met key performance characteristics, as defined in NEMA NU 2-2018. | Pass- The test results meet the Cubresa BrainPET requirements. |
| --- | --- | --- |
| MRI Performance Assessment | MRI-related performance was evaluated through measurements of RF noise, main magnetic field homogeneity, and RF field homogeneity. Additionally, signal-to-noise ratio (SNR) and uniformity were measured and characterized using methods NEMA MS1-2008 (R2014) Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging and NEMA MS3-2008 (R2014) Determination of Image Uniformity in Diagnostic Magnetic Resonance Images, respectively, in accordance with NEMA MS 9-2008 (R2014) Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images. | Pass- The test results meet the Cubresa BrainPET requirements. |
| RF Noise Measurement | NSD, Siemens RF Noise Spectrum Analysis. Performed during simultaneous PET/MR acquisition. | Pass- The test results meet the Cubresa BrainPET requirements. |
| B0 Homogeneity evaluation | B0 homogeneity analysis performed via Siemens rf_field scan, interference lines analysis. Performed during simultaneous PET/MR acquisition. | Pass- The test results meet the Cubresa BrainPET requirements. |
| B1 Homogeneity evaluation | B1 homogeneity measurements performed via Siemens rf_field scan, RF field symmetry analysis. Performed during simultaneous PET/MR acquisition. | Pass- The test results meet the Cubresa BrainPET requirements. |
| Signal-to-Noise (SNR) evaluation | NEMA MS1-2008 (R2014). Determination of Signal-to-Noise (SNR) in Diagnostic Magnetic resonance images. | Pass- The test results meet the Cubresa |
Page 13 of 15
{17}
[LOGO]
CUBRESA
Cubresa BrainPET 510(k) Summary
| | Performed during simultaneous PET/MR acquisition. | BrainPET requirements. |
| --- | --- | --- |
| Image Uniformity | NEMA MS3-2008 (R2014) Determination of Image Uniformity in Diagnostic Magnetic Resonance Images. PIU analysis. Performed during simultaneous PET/MR acquisition. | Pass- The test results meet the Cubresa BrainPET requirements. |
| Magnetically Induced Torque on Cubresa BrainPET in the Magnetic Resonance Environment | ASTM F2213-25 - Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment | Pass- The test results meet the Cubresa BrainPET requirements. |
| Measurement of Magnetically Induced Displacement Force on Cubresa BrainPET in the Magnetic Resonance Environment | ASTM F2052-21 - Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment | Pass- The test results meet the Cubresa BrainPET requirements. |
### 6. Biocompatibility Testing:
Biocompatibility testing of patient-contacting materials was conducted in accordance with ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and included evaluations for cytotoxicity, sensitization, and irritation.
### 7. Clinical Images:
Documentation is provided regarding the clinical effectiveness of Cubresa BrainPET PET/MR images as recommended by the FDA's Guidance for Industry "Guidance for the Submission of Premarket Notifications for Emission Computed Tomography Devices and Accessories (SPECT and PET) and Nuclear Tomography Systems" Section IX.F (CDRH,1998) and Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices (CDRH, Oct 2023).
Cubresa BrainPET PET and MR images acquired simultaneously in patient volunteers were evaluated by dual-trained American Board of Radiologists/Nuclear Medicine certificated physicians who have attested to the diagnostic image quality achieved by the Cubresa BrainPET PET/MR system.
### 8. Summary of Performance Testing:
The Cubresa BrainPET has been tested and the conclusions from the verification and validation data support that the technological characteristics of the proposed device have an equivalent safety and performance profile to the predicate device. Successful completion of the standard QA
Page 14 of 15
{18}
CUBRESA
Cubresa BrainPET 510(k) Summary
tests and expert review of sample clinical images demonstrate that the Cubresa BrainPET maintains clinically acceptable MR imaging performance. The Cubresa BrainPET technological characteristics do not raise different questions of safety and effectiveness. The verification and validation testing of the Cubresa BrainPET supports a determination of substantial equivalence.
### VIII. CONCLUSIONS
The Cubresa BrainPET has the same intended use and similar basic technological characteristics as the Siemens Biograph mMR with syngo software (K163234, predicate device). While there are some differences in technological characteristics between the proposed and predicate device, the differences were tested, and verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device.
CUBRESA has concluded that the Cubresa BrainPET is substantially equivalent to the currently marketed predicate device Siemens Biograph mMR with syngo software.
Page 15 of 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.