Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 6, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, ENT, Gynecology and general surgery.
Device Story
Pulsed solid-state Holmium:YAG laser system; delivers 2100nm energy via optical fibers; used for soft tissue surgery and stone fragmentation. System includes touchscreen interface, single-pedal footswitch, liquid cooling, and safety features (key switch, emergency stop, door interlock). Operated by clinicians in OR or clinic settings. Laser energy transmitted to target site; enables precise tissue interaction. Output monitored in real-time; allows surgeon to perform incision, ablation, or coagulation. Benefits include minimally invasive access and effective stone lithotripsy.
Clinical Evidence
Bench testing only. Conformance to IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and IEC 60825-1 (laser product safety). Software verified per IEC 62304.
Indicated for patients requiring soft tissue surgery (incision, excision, resection, ablation, vaporization, coagulation, haemostasis) or stone fragmentation in urology, gastroenterology, ENT, gynecology, and general surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Litho 150 / Cyber Ho 150 Holmium Laser System (K201455)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 6, 2026
Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd.
Zhengzhou Li
Rm. 208, Bldg. C, # 3, Juquan Rd.
Huangpu District,
Guangzhou, Guangdong 510000
China
Re: K253951
Trade/Device Name: Holmium Laser Therapeutic Apparatus (HZ-A); Holmium Laser Therapeutic Apparatus (HZ-B); Holmium Laser Therapeutic Apparatus (HZ-E)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: November 18, 2025
Received: December 10, 2025
Dear Zhengzhou Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253951 - Zhengzhou Li
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K253951 - Zhengzhou Li
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2026.03.06
18:26:50 -05'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253951 | |
| Device Name Holmium Laser Therapeutic Apparatus (HZ-A); Holmium Laser Therapeutic Apparatus (HZ-B); Holmium Laser Therapeutic Apparatus (HZ-E) | |
| Indications for Use (Describe) Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, ENT, Gynecology and general surgery. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
1 Submitter / Contact Information
| Field | Information |
| --- | --- |
| 510 (K) # | K253951 |
| Date Preparation | February 27, 2026 |
| Submitter Name | Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. |
| Address | Room 208, Building C, No. 3 Juquan Road, Huangpu District, Guangzhou, Guangdong, 510000, China. |
| Phone | +86 -20-37396970 |
| Fax | +86 -20-37376979 |
| Contact Person | Zhengzhou Li |
| Email | potent_medical_public@potent-medical.com |
2 Device Information
| Field | Information |
| --- | --- |
| Device Trade Name | Holmium Laser Therapeutic Apparatus |
| Models | HZ-A, HZ-B, HZ-E |
| Common Name | Powered Laser Surgical Instrument |
| Regulation Name | Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology |
| Regulation Number | 21 CFR 878.4810 |
| Regulatory Class | Class II |
| Product Code | GEX |
1 / 7
{5}
2 / 7
| Panel | General and Plastic Surgery |
| --- | --- |
# 3 Predicate and Reference Devices
| Type | Device | Manufacturer | 510(k) Number |
| --- | --- | --- | --- |
| Primary Predicate | Litho 150 / Cyber Ho 150 Holmium Laser System | Quanta System S.p.A. | K201455 |
# 4 Device Description
The HZ Series is a family of pulsed solid-state Holmium:YAG laser systems that deliver energy at a wavelength of $2100\mathrm{nm}$ through optical fibers for soft tissue surgery and stone fragmentation. The system utilizes a diode-pumped Ho:YAG laser medium, and the laser energy is transmitted to the target site via SMA-905 compatible optical fibers. All models in the series share the same fundamental design and operational principles, differing in maximum average power output, Energy per pulse and frequency.
The system includes a touchscreen interface, single-pedal footswitch, integrated liquid cooling, and multiple built-in safety features such as a key switch, emergency stop and door interlock, and real-time monitoring of laser output.
# 5 Intended Use/Indications for use
- Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, ENT, Gynecology and general surgery.
# 6 Comparison of Technological Characteristics
| | Subject device | Predicate device | Basis for SE |
| --- | --- | --- | --- |
| 510 (k) | / | K201455 | / |
| Model name | Holmium Laser Therapeutic Apparatus | Litho 150, Cyber Ho 150 | / |
| Manufacturer | Potent Medical | Quanta System Spa | / |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Regulation Name: | Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology | Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology | Same |
| Regulation Class | Class II | Class II | Same |
{6}
| | Subject device | Predicate device | Basis for SE |
| --- | --- | --- | --- |
| Product Code | GEX | GEX | Same |
| Device Panel | General & Plastic Surgery | General & Plastic Surgery | Same |
| Indications for use | Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology,ENT, Gynecology and general surgery. | Full indications for use: See Appendix A. | Substantially Equivalent |
| Laser Source | Pulsed Holmium laser (CHT:YAG) | Pulsed Holmium laser (CHT:YAG) | Same |
| Wavelength (nm) | 2,100 nm | 2,100 nm | Same |
| Emission | Pulsed | Pulsed | Same |
| Pulse duration | 90-1000 μs | Up to 1100 μs | Substantially Equivalent |
| Energy per pulse | HZ-A (up to 4.5 J)HZ-E (up to 4.5 J)HZ-B (up to 5.0 J) | up to 5.0 Joule | Substantially Equivalent |
| Frequency | HZ-A (up to 40 Hz)HZ-E (up to 40Hz)HZ-B (up to 60 Hz) | up to 100 Hz | Substantially Equivalent |
| Max average power | HZ-A (up to 80W)HZ-E (up to 70W)HZ-B (up to 90W) | 152W | Substantially Equivalent |
| Delivery system | optical fibers | optical fibers | Substantially Equivalent |
| Aiming beam | Green diode laser < 3 mW | Green diode laser < 5 mW | Substantially Equivalent |
# Conclusion:
The HZ Series Holmium Laser Therapeutic Apparatus is a similar device to the identified predicate device, sharing the same intended use and fundamental technological principles. Differences in parameters such as pulse duration, energy, frequency, and aiming beam wavelength do not raise new questions of safety or effectiveness. The HZ Series is considered to be substantially equivalent to the identified predicate device.
{7}
4 / 7
# 7 Non-Clinical Performance Data
The HZ Series Holmium Laser System has undergone comprehensive performance testing to ensure compliance with applicable international standards and FDA guidance. The following evaluations were conducted:
## 7.1 Electrical Safety and Electromagnetic Compatibility
The device was tested and found to conform to the following standards:
- IEC 60601-1:2005 + A1:2012 + A2:2020
Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 + A1:2020
Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances – Requirements and tests.
- IEC 60601-2-22:2019
Medical electrical equipment – Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic, and diagnostic laser equipment.
- IEC 60825-1:2014
Safety of laser products – Part 1: Equipment classification and requirements.
## 7.2 Software Verification and Validation
Software development and validation were performed in accordance with:
- IEC 62304:2006 + A1:2015
Medical device software – Software life cycle processes.
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
# 8 Substantial Equivalence Conclusion
The HZ Series Holmium Laser Therapeutic Apparatus has the same intended use and similar technological characteristics as the predicate device. Minor differences between the subject and predicate device do not raise new risks regarding safety or effectiveness. The subject device is substantially equivalent to the identified predicate device.
## Appendix A: Indications for Use (Predicate device)
The Multicavity Holmium laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Gynaecology, ENT and General Surgery.
{8}
5 / 7
# Urology
Open and endoscopic surgery (incision, excision, resection, ablation,
vaporization, coagulation and haemostasis) including:
- Urethral Strictures
- Bladder Neck Incisions (BNI)
- Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors,
- Ablation of Benign Prostatic Hypertrophy (BPH),
- Transurethral incision of the prostate (TUIP)
- Holmium Laser Resection of the Prostrate (HoLRP)
- Holmium Laser Enucleation of the Prostate (HoLEP)
- Holmium laser Ablation of the Prostate (HoLAP)
- Condylomas
- Lesions of external genitalia
# Lithotripsy and Percutaneous Urinary Lithotripsy
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate
- dehydrate stones.
- Endoscopic fragmentation of kidney calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed.
# Gastroenterology
Open and endoscopic Gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:
- Appendectomy
- Polyps
- Biopsy
- Gall Bladder calculi
- Biliary/Bile duct calculi
- Ulcers
- Gastric ulcers
- Duodenal ulcers
- Non Bleeding Ulcers
- Pancreatitas
- Haemorrhoids
- Cholecy stectomy
- Benign and Malignant Neoplasm
{9}
- Angiody splasia
- Colorectal cancer
- Telangiectasias
- Telangiectasias of the Osler-Weber-Renu disease
- Vascular Malformation
- Gastritis
- Esophagitis
- Esophageal ulcers
- Varices
- Colitis
- Mallory-Weiss tear
- Gastric Erosions
## Arthroscopy
Arthroscopy/Orthopaedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue) in small and large joints of the body, excluding the spine but including:
- Ligament and tendon Release
- Contouring and sculpting of articular surfaces
- Capsulectomy in the Knee
- Chondroplasty in the Knee
- Debridement of inflamed synovial tissue
- Chondromalacia Ablation
- Chondromalacia and tears
- Plica Removal
- Meniscectomy
- Loose Body Debridement
- Lateral retinecular release
Ablation of soft, cartilaginous and bony tissue in Minimal Invasive Spinal Surgery including
- Percutaneous Laser Disc Decompression/Discectomy of the L4-5 and L5-SI lumbar discs, including
- Foraminoplasty
- Percutaneous Cervical Disc Decompression/Discectomy
- Percutaneous Thoracic Disc Decompression/Discectomy
## Gynaecology
Open and laparoscopic gynaecological surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) of soft tissue.
## ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and
{10}
haemostasis of soft tissue and cartilage) including:
- Endonasal/sinus Surgery
- Partial turbinectomy
- Polypectomy
- Dacryocystorhinostomy
- Frontal Sinusotomy
- Ethmoidectomy
- Maxillary antrostomy
- Functional endoscopic sinus surgery
General Surgery
Open, laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:
- Appendectomy
- Skin incision
- Excision of external and internal lesions
- Complete of partial resection of internal organs, tumors and lesions
- Biopsy
7 / 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.