K253922 · JJGC Indústria e Comércio de Materiais Dentários S.A. · NHA · Jun 15, 2026 · Dental
Device Facts
Record ID
K253922
Device Name
Neodent Scannable Healing Abutment
Applicant
JJGC Indústria e Comércio de Materiais Dentários S.A.
Product Code
NHA · Dental
Decision Date
Jun 15, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The Neodent Scannable Healing Abutments are used for soft tissue maintenance, during the osseointegration phase of Neodent Implants to be rehabilitated by the late loading technique, and for the transfer of implant position during scanning by means of the CAD/CAM technique. It can be used in implant installation surgery or reopening surgery (second surgical phase).
Device Story
Titanium alloy healing abutment; used during osseointegration phase of dental implants; maintains soft tissue; facilitates intraoral scanning via CAD/CAM technique. Features Encode coding scheme (machined markings) on transmucosal surface; allows transfer of implant position/orientation to digital workflow. Placed by dentists during implant surgery or reopening; used out of occlusion. Output is digital scan data used by dental labs/clinicians to design final restorations. Benefits include streamlined digital workflow for prosthetic fabrication and soft tissue management.
Clinical Evidence
Prospective pilot clinical investigation (n=7 patients, 8 implants) evaluated marginal bone remodeling, soft tissue parameters (PES), and plaque index (VPI) over 10 months. Results showed physiological bone adaptation, stable peri-implant mucosa, and no inflammation. VPI showed effective plaque control. Supported by published literature. Bench testing included SEM/EDS surface analysis, cytotoxicity (XTT method), and sterilization validation (SAL 1x10^-6).
Indicated for patients requiring endosseous dental implants in the upper or lower jaw to support prosthetic devices (artificial teeth) for chewing function restoration. Suitable for single or two-stage procedures, single or multiple unit restorations, and immediate loading when primary stability is achieved.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Neodent Implant System – GM Line (GM healings) (K163194)
MRI Compatibility for Existing Neodent Implant System (K182620)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 15, 2026
JJGC Indústria e Comércio de Materiais Dentários S.A.
% Jennifer Jackson
Sr. Director, Regulatory Affairs and Quality
Straumann USA, LLC
60 Minuteman Rd.
Andover, Massachusetts 01810
Re: K253922
Trade/Device Name: Neodent Scannable Healing Abutment
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: May 14, 2026
Received: May 14, 2026
Dear Jennifer Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253922 - Jennifer Jackson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253922 - Jennifer Jackson
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANDREW I. STEEN -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253922
Device Name
Neodent Scannable Healing Abutment
Indications for Use (Describe)
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# **Traditional 510(k) Submission**
**Neodent Scannable Healing Abutment**
510(k) Summary
K253922
# **510(k) Summary**
# **Submitter's Contact Information**
Submitter: Straumann USA, LLC
60 Minuteman Road
Andover, MA 01810, USA
Registration No.: 1222315 Owner/Operator No.: 9005052
On the behalf of:
JJGC Indústria e Comércio de Materiais Dentários S.A (dba Neodent)
Av. Juscelino Kubitschek de Oliveira, 3291
Curitiba, Paraná, Brazil 81270-200
Registration No.: 3008261720
Owner/Operator No.: 10031702
Contact Person: Jennifer M. Jackson, MS, RAC
Sr. Director, Regulatory Affairs and Quality
Phone Number: +1 978 747-2509
Fax Number: +1 978 747-0023
Email: jennifer.jackson@straumann.com
Prepared By & Alternate Contact: Leticia Milani
Regulatory Affairs Analyst
JJGC Indústria e Comércio de Materiais Dentários SA
Date of Submission: June 12, 2026
# **Name of the Device**
Trade Names: Neodent Scannable Healing Abutment
Common Name: Endosseous dental implant abutment
Classification Name: Endosseous dental implant abutment
Regulation Number: 21 CFR 872.3630
Device Classification: II
Product Code(s): NHA
Classification Panel: Dental
JJGC Indústria e Comércio de Materiais Dentários S.A.
1
{5}
# Traditional 510(k) Submission
## Neodent Scannable Healing Abutment
510(k) Summary
K253922
### Predicate Device(s)
#### Primary Predicate:
K163194 Neodent Implant System – GM Line (GM healings)
#### Reference Devices:
K212730 BellaTek Encode Emergence Healing Abutments
K182620 MRI Compatibility for Existing Neodent Implant System
### Device Description
This premarket notification includes new scannable abutments of Neodent Implant System, which are compatible with GM prosthetic interface. The Neodent Scannable Healing Abutments proposed on this submission are similar to devices already cleared in previous submissions, according to predicate devices described above. This submission intends to expand the portfolio of Neodent Implant System with new solutions to provide more treatment options to the customers.
These proposed components are intended for temporary use, during the maintenance of soft tissues in the osseointegration phase of implants to be rehabilitated by the late loading technique. Additionally, they have the function of transferring the position and orientation of the implants position during intraoral scanning by means of the CAD/CAM technique. They are abutments to be placed out of occlusion.
### Intended Use
The Neodent Scannable Healing Abutments are used for soft tissue maintenance, during the osseointegration phase of Neodent Implants to be rehabilitated by the late loading technique, and for the transfer of implant position during scanning by means of the CAD/CAM technique. It can be used in implant installation surgery or reopening surgery (second surgical phase).
JJGC Indústria e Comércio de Materiais Dentários S.A.
2
{6}
# **Traditional 510(k) Submission**
**Neodent Scannable Healing Abutment**
510(k) Summary
K253922---
## Indications for Use
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
## Technological Characteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following tables:
JJGC Indústria e Comércio de Materiais Dentários S.A.
3
{7}
## Neodent Scannable Healing Abutment
510(k) Summary
K253922
**Table 1 – Table of Substantial Equivalence**
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE |
| --- | --- | --- | --- |
| **K Number** | **K253922** *Neodent Scannable Healing Abutment* | **K163194** *Neodent Implant System – GM Line (GM healings)* | **K212730** *BellaTek Encode Emergence Healing Abutments* |
| **Indications for Use** | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. | The BellaTek Encode Emergence Healing Abutments are intended for use as an accessory to endosseous dental implants during endosseous and gingival healing to prepare gingival tissue for acceptance of a final abutment and restoration. |
| **Material** | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. | Titanium alloy, according to ASTM F136. |
| **Interface Connection** | Grand Morse (GM) | Grand Morse (GM) | N/A |
| **Diameter** | 3.5, 4.5 and 5.5 mm | 4.5 mm | 3.5 to 5.7 mm |
| **Gingival Height** | 0.8, 1.5, 2.5, 3.5, 4.5, 5.5 and 6.5 mm | 0.8, 1.5, 2.5, 3.5, 4.5 and 5.5 mm | 3.8 to 7.5 mm |
| **Surface Treatment** | Sandblasting + Electrolysis (yellow anodized) | Electrolysis (yellow anodized) | Electrolysis (pink anodized) |
| **Scanning Identification** | Encode coding scheme (machined markings) can be visualized in STL files. | N/A | Encode coding scheme (machined markings) can be visualized in STL files. |
| **Single use** | Yes | Yes | Yes |
| **Sterilization Method** | Provided sterile via Ethylene Oxide. | Provided sterile via Ethylene Oxide. | Provided sterile via Gamma Radiation. |
JJGC Indústria e Comércio de Materiais Dentários S.A.
4
{8}
## Neodent Scannable Healing Abutment
510(k) Summary
K253922
### Performance Testing
#### Bench Testing
A surface analysis was performed in line with FDA's guidance document 'Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments', to characterize the surface topography using SEM (Scanning Electron Microscopy) and EDS (Energy Dispersive Spectroscopy). The detected chemical composition is consistent with titanium alloy (Ti6Al4V-ELI), with no significant variation among the samples. Samples also have a uniform morphology, and do not contain stains, debris, contamination, or manufacturing residues.
The anodization process applied to the subject device is exactly the same to that used for the primary predicate device (K163194), with no variation of parameters, formulation and equipments.
#### MRI Compatibility Testing
An assessment was made to demonstrate that the MR conditional labeling from K182620 is applicable to the subject devices, and a patient treated with them can be safely scanned observing the parameters previously established per reference devices.
#### Biocompatibility Testing
Assessments regarding biological compatibility were performed according to ISO 10993-1 '*Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process*' and to the FDA Guidance document '*Use of International Standard ISO 10993- 1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016*'.
To evidence the biological safety of subject devices, a cytotoxicity evaluation was conducted in accordance with ISO 10993-5 and ISO 10993-12, using the XTT method (inhibition of the cellular proliferation), after 72 hours of extraction time. Results ensure the biological safety of subject
JJGC Indústria e Comércio de Materiais Dentários S.A.
5
{9}
## Neodent Scannable Healing Abutment
510(k) Summary
K253922
devices per ISO 10993-1:2020 and FDA guidance, proving that the benefits provided by the device outweigh the residual risk of biological safety.
### Sterilization validation
The subject devices are supplied sterile via Ethylene Oxide (EO), the method was validated to a sterility assurance level (SAL) of 1x10-6 in accordance with ISO 11135:2014, “*Ethylene Oxide – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices*”. EO sterilization residuals have been verified to be less than the maximum allowable limits as defined in ISO 10993-7 “*Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals*”. The sterilization of the subject devices is identical to the sterilization already cleared for the primary predicate device (K163194).
### Shelf Life
The expiration date of the devices was determined considering the integrity of the product and the packaging tests after shelf life testing. The packaging of the proposed abutments is identical to the packaging of the primary predicate device (K163194). The shelf life for Neodent Scannable Healing Abutments is 5 years.
### Clinical Testing
A prospective pilot clinical investigation evaluating the clinical performance and safety of the subject devices was conducted. The research was submitted and approved by local ethics committee, with the objectives of assessing marginal bone remodeling over time, as well as evaluating the peri-implant soft tissue parameters, clinical and esthetic outcomes as measured by the Pink Esthetic Score (PES) and the Visible Plaque Index (VPI). The study included 7 patients and 8 implants placed in posterior regions (4 maxilla and 4 mandible), in combination with 8 Neodent Scannable Healing Abutment. Teeth treated included premolars and molars.
Marginal bone remodeling was radiographically assessed at baseline, 90 days, 180 days, and after prosthetic loading. Results demonstrate a remodeling pattern consistent with the expected
JJGC Indústria e Comércio de Materiais Dentários S.A.
6
{10}
## Neodent Scannable Healing Abutment
510(k) Summary
K253922
physiological bone adaptation during early osseointegration and the establishment of the peri-implant biological width.
The Pink Esthetic Score (PES) assessment was conducted to assess the peri-implant soft tissue response and demonstrated the maintenance of peri-implant mucosal stability throughout follow-up. The parameters of soft tissue color, texture, and gingival contour were scored within ranges consistent with gingival health and showed no clinical evidence of inflammation. No clinical signs of mucosal breakdown, suppuration, or peri-implantitis were observed. No device-related or procedure-related adverse events were reported during the study period.
The plaque biofilm was evaluated by the Visible Plaque Index (VPI), assessed at baseline (before surgery), 90 days, 180 days, and post-loading (10 months after surgery). Results demonstrated a statistically significant reduction in plaque presence from baseline to 90 days. A statistically significant reduction was not demonstrated between all other timepoints. These findings indicate no increase in biofilm retention associated with the transmucosal roughened surface and no evidence of increased inflammatory burden. To support effective plaque control during the healing phase, patients enrolled in the clinical study were provided with standardized post-surgical oral hygiene instructions, which were effective and demonstrate adequate plaque removal from the abutment surface.
In addition, published clinical literature evidences were analyzed to support and demonstrate the consistence of results found in the pilot clinical study. Collectively, these characteristics evidence the safety and performance of the subject device under the proposed conditions of use, and the available evidence does not raise concerns regarding the risk associated with the device.
## Conclusion
The subject, primary predicate and reference devices have similar indications for use, intended use, design, raw material and overall dimensions. The main technological difference of the subject devices, when compared to the primary predicate device (K163194) and reference device (K212730), is the sandblasted roughened transmucosal surface, which allows the abutment to be scanned. Surface analysis showed a composition consistent with the raw material (titanium alloy), and does not raise concerns on residual elements. These results are consistent with those found in biocompatibility tests, which demonstrated no cytotoxic response under the tested conditions.
JJGC Indústria e Comércio de Materiais Dentários S.A.
7
{11}
# **Traditional 510(k) Submission**
**Neodent Scannable Healing Abutment**
510(k) Summary
K253922
Clinical evidence was also submitted to support the new surface technology safety for the intended indications. The prospective pilot clinical investigation evidenced results consistent with the published literature data. Together these data demonstrate the sandblasted roughened transmucosal surface does not adversely affect safety or effectiveness. Therefore, it is concluded that the proposed devices are substantially equivalent to the primary predicate and reference devices.
JJGC Indústria e Comércio de Materiais Dentários S.A.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.