K253895 · Arthrex, Inc. · MBI · Dec 30, 2025 · Orthopedic
Device Facts
Record ID
K253895
Device Name
Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm
Applicant
Arthrex, Inc.
Product Code
MBI · Orthopedic
Decision Date
Dec 30, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm is intended to be used for suture or tissue fixation in the foot/ankle and hand/wrist. Specific indications for use are listed below: • Hand/Wrist: Repair of Collateral Ligaments, Repair of Flexor and Extensor Tendous at the PIP, DIP and MCP joints for all digits and Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty) • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus Reconstruction
Device Story
Knotless, two-component suture anchor; consists of fully threaded, vented PEEK anchor body and closed eyelet; preassembled on disposable inserter with suture threader. Used by surgeons in orthopedic procedures (foot/ankle, hand/wrist) for soft tissue fixation. Device provides mechanical fixation of sutures to bone; surgeon inserts anchor into pre-drilled pilot hole to secure tissue. Benefits include stable fixation for ligament/tendon repairs. Sterile, single-use device.
Clinical Evidence
Bench testing only. Conducted cyclic pull-out testing and Failure Torque/Insertion Torque Testing per ASTM F3690-24. Packaging validation and 5-year real-time aging shelf-life testing performed per ISO 11607. No clinical data.
Technological Characteristics
Material: PEEK Optima (ASTM F2026). Design: Knotless, two-component, fully threaded, vented anchor with closed eyelet. Dimensions: 3.5 x 10 mm. Sterilization: Ethylene Oxide (EtO). Connectivity: None. Energy source: None (mechanical).
Indications for Use
Indicated for suture or tissue fixation in foot/ankle and hand/wrist, including collateral ligament repair, flexor/extensor tendon repair at PIP, DIP, and MCP joints, carpometacarpal joint arthroplasty, lateral/medial stabilization, metatarsal ligament repair, and hallux valgus reconstruction.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Arthrex DX SwiveLock with Forked Eyelet 3.5 x 8.5 mm (K150648)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
December 30, 2025
Arthrex, Inc.
Stacy Valdez
Principal Regulatory Affairs Specialist
1370 Creekside Blvd.
Naples, Florida 34108
Re: K253895
Trade/Device Name: Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth or threaded metallic bone fixation fastener
Regulatory Class: Class II
Product Code: MBI
Dated: December 4, 2025
Received: December 4, 2025
Dear Stacy Valdez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K253895 - Stacy Valdez
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253895 - Stacy Valdez
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253895 | ? |
| Please provide the device trade name(s). | | ? |
| Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm | | |
| Please provide your Indications for Use below. | | ? |
| The Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm is intended to be used for suture or tissue fixation in the foot/ankle and hand/wrist. Specific indications for use are listed below:
• Hand/Wrist: Repair of Collateral Ligaments, Repair of Flexor and Extensor Tendous at the PIP, DIP and MCP joints for all digits and Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty)
• Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus Reconstruction | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253895
Page 1 of 4
# 510(k) Summary
| Date Prepared | 12/04/2025 |
| --- | --- |
| Submitter | Arthrex Inc.
1370 Creekside Boulevard
Naples, FL 34108-1945 |
| Contact Person | Name: Stacy Valdez
Title: Principal Regulatory Affairs Specialist
Phone: 1-239-643-5553, ext. 72010
Email: stacy.valdez@arthrex.com |
| Trade Name | Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm |
| Common Name | Fastener, Fixation, Nondegradable, Soft Tissue |
| Product Code | MBI |
| Classification Name | 21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener |
| Regulatory Class | II |
| Primary Predicate Device | K150648: Arthrex DX SwiveLock with Forked Eyelet 3.5 x 8.5 mm |
| Reference Device | K230435: Arthrex 3.9 mm SwiveLock Anchor |
| Purpose of Submission | This Special 510(k) premarket notification is submitted to obtain clearance for the Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm as a line extension of the Arthrex SwiveLock Suture Anchors cleared via K150648. |
| Device Description | The proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm is a knotless, two-component suture anchor consisting of a fully threaded and vented anchor body and a closed eyelet. The anchor and eyelet are manufactured from polyetheretherketone (PEEK Optima per ASTM F2026). The anchor and eyelet are preassembled on a disposable inserter with a suture threader. Arthrex recommends the use of the Arthrex 1.3 mm SutureTape (sold separately) previously cleared within Arthrex SutureTape (K193575). The proposed device is offered sterile, single-use, and is packaged in a single-pack (one per box). |
| Indications for Use | The Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm is intended to be used for suture or tissue fixation in the foot/ankle and hand/wrist. Specific indications for use are listed below:
• Hand/Wrist: Repair of Collateral Ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits and Carpometacarpal Joint Arthroplasty (basal thumb joint arthroplasty)
• Foot/Ankle: Lateral Stabilization, Medial Stabilization, Metatarsal Ligament Repair, Hallux Valgus Reconstruction |
| Performance Data | To demonstrate product performance, Arthrex has conducted cyclic pull-out testing and Failure Torque/Insertion Torque Testing on the proposed Arthrex SwiveLock Anchor, 3.5 x 10 mm |
{5}
K253895
Page 2 of 4
| | comparing the results to the Arthrex SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648) in accordance with ASTM F3690-24 Standard Test Method for Evaluating Suture Anchor Insertion and Pull Displacement Resistance and well-established methods reviewed and accepted by FDA in predicate device Arthrex DX SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648). Furthermore, Arthrex has completed packaging validation and 5-year real-time aging shelf-life testing to demonstrate that the packaging configuration is capable of maintaining and protecting the product and sterility (EtO sterilized) of the device throughout the shipping and handling environment. Testing was performed using the same test methods submitted and cleared under Arthrex 3.9 mm SwiveLock Anchor (K230435) and/or in accordance with FDA recognized standards. The packaging configuration met all the packaging testing acceptance criteria in conformance to ISO 11607 and applicable standards. Assessment of the physical product attributes including product, design, size, and materials has determined that the Arthrex SwiveLock Anchor, 3.5 x 10 mm does not introduce additional risks or concerns regarding sterilization and shelf-life. |
| --- | --- |
| Technological Comparison | Compared to the predicate device Arthrex DX SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648), the proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm has the same intended use, basic design features, fundamental scientific technology, material, anchor configuration, principle of operation, anchor diameter, sterility, and shelf-life. The differences between the proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm and the predicate device Arthrex DX SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648) are as follows:
• Principle of Operation:
• The proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm will have a different pilot hole (drill bit) diameter compared to the predicate device Arthrex DX SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648).
• Anchor and Eyelet:
• The total length (anchor and eyelet) of the proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm is longer than the predicate device Arthrex DX SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648). However, the proposed anchor length falls within the length range |
{6}
K253895
Page 3 of 4
cleared under the predicate devices within K150648.
- The proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm will be offered as vented; whereas the predicate device Arthrex DX SwiveLock with Forked Eyelet 3.5 x 8.5 mm (K150648) is non-vented. However, the vented design was previously cleared under the predicate device K150648 (Arthrex DX SwiveLock Suture Anchor, 3.5 x 13.5 mm).
- The proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm will have a closed eyelet; whereas the predicate device Arthrex DX SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648) has an open (forked) eyelet. However, the closed eyelet design was previously cleared under the predicate device K150648 (Arthrex DX SwiveLock Suture Anchor, 3.5 x 13.5 mm).
- **Arthrex Suture:**
- The proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm will be used with existing Arthrex 1.3 mm SutureTape (sold separately) previously cleared within Arthrex SutureTape (K193575), whereas the predicate device Arthrex DX SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648) was cleared using existing Arthrex Suture (Size #4-0 to 4 mm) previously cleared under K1010673 (Arthrex FiberWire), K012923 (Arthrex FiberWire USP Size 5 Suture), K041589 (Arthrex FiberWire), K041553 (Arthrex FiberWire and FiberTape) and K122374 (Arthrex Suture).
- **Packaging Configuration:**
- The proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm will be packaged in a double PETG blister tray with Tyvek lidding inside a carton; whereas the predicate device Arthrex SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648; AR-8978P) is packaged within a double poly/Tyvek pouch inside a carton. The double PETG blister tray with Tyvek lidding is an existing configured used on Arthrex anchors (PEEK) cleared within Arthrex PushLock Tenodesis Anchor (K181513).
- **MRI Safety:**
- The proposed device will be labeled as MR Safe; whereas the predicate device Arthrex DX
{7}
K253895
Page 4 of 4
| | SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648) was not evaluated for MRI.
Based on the intended use, fundamental scientific technology, and the data provided in this Special 510(k), Arthrex has determined that the proposed Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm is substantially equivalent to the predicate device Arthrex DX SwiveLock SL with Forked Eyelet 3.5 x 8.5 mm (K150648). Any differences between the proposed and predicate devices are considered minor and do not raise different questions concerning safety and effectiveness. |
| --- | --- |
| **Conclusion** | The Arthrex SwiveLock Suture Anchor, 3.5 x 10 mm is substantially equivalent to the predicate devices cleared under K150648 in which the basic design features and intended use are the same. Based on the indications for use, technological characteristics, and the summary of data submitted, Arthrex Inc. has determined that the proposed device is substantially equivalent to the currently marketed predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.