The Nasal Alar SpO2 Sensor is intended for single patient use, for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (at least 4 years and older and weighing ≥15 kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments, where compatible pulse oximetry monitors are indicated for use, under professional supervision.
Device Story
Disposable, single-patient-use pulse oximetry sensor; attaches to nasal ala via soft silicone rubber cushions; integrates LEDs and photodiode. Connects to compatible Philips FAST SpO2 technology monitors via DB-9 cable. Operates via transmissive spectrophotometric measurement of functional arterial oxygen saturation. Used in various healthcare environments under professional supervision. Provides continuous SpO2 and pulse rate data to monitor; assists clinicians in assessing patient oxygenation status; benefits patients by providing reliable perfusion monitoring at nasal site, avoiding potential flow interruptions associated with distal sites like digits.
Clinical Evidence
Clinical SpO2 testing conducted per ISO 80601-2-61. Results confirmed an ARMS (Accuracy Root Mean Square) of ≤3% across the 70-100% range. Bench testing included biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and mechanical reliability.
Technological Characteristics
Materials: silicone rubber cushions, plastic body, new glue/coverlay formulations. Sensing: transmissive spectrophotometric pulse oximetry (LEDs/photodiode). Connectivity: wired DB-9 connector to compatible monitors. Ingress Protection: IP34. Standards: ISO 10993-1, ISO 80601-2-61, IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for continuous noninvasive monitoring of functional SpO2 and pulse rate in adult and pediatric patients (≥4 years, ≥15 kg) who are well or poorly perfused. For use in healthcare environments with compatible pulse oximetry monitors under professional supervision.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 1, 2026
Philips Medizin Systeme
Jamie Palmer
Principal, Regulatory Affairs Specialist
Boblingen Gmbh
Hewlett-Packard St. 2
Boblingen, 71034
Germany
Re: K253887
Trade/Device Name: Nasal Alar SpO2 Sensor (989803205381); Nasal Alar SpO2 Sensor (989803205391); Nasal Alar SpO2 Sensor (989803205401)
Regulation Number: 21 CFR 870.2700
Regulation Name: Oximeter
Regulatory Class: Class II
Product Code: DQA
Dated: March 31, 2026
Received: April 1, 2026
Dear Jamie Palmer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253887 - Jamie Palmer
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253887 - Jamie Palmer
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253887 | ? |
| Please provide the device trade name(s). | | ? |
| Nasal Alar SpO2 Sensor (989803205381); Nasal Alar SpO2 Sensor (989803205391); Nasal Alar SpO2 Sensor (989803205401) | | |
| Please provide your Indications for Use below. | | ? |
| The Nasal Alar SpO2 Sensor is indicated for single patient use, for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (at least 4 years and older and weighing ≥15 kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments, where compatible pulse oximetry monitors are indicated for use, under professional supervision. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
# 510K SUMMARY
| 1. Submitter | |
| --- | --- |
| Date Prepared: | 29 Apr 2026 |
| Submitter/Owner: | Philips Medizin Systeme Böblingen, GmbH FDA Establishment Number: 9610816 Hewlett-Packard Str. 2 Böblingen, 71034 Germany Phone: +49 7031 4630 Fax: 07031-463 -2202 |
| Key Contact: | Jamie L. Palmer Principal, Regulatory Affairs Specialist Jamie.Palmer@philips.com |
| 510(k) Submission Type: | Traditional 510(k) |
| 2. Device | |
| --- | --- |
| Trade Name(s): | Nasal Alar SpO2 Sensor (989803205381); Nasal Alar SpO2 Sensor (989803205391); Nasal Alar SpO2 Sensor (989803205401) |
| Common Name: | Oximeter |
| Classification Name: | Device Class: II (Performance Standards) Product Code: DQA (Oximeter) Regulation Number: 21 CFR 870.2700 Review Panel: Cardiology |
| 3. Predicate Device | | | |
| --- | --- | --- | --- |
| Predicate Device: | 510(k) No. | Company Name & Device Name | Product Code |
| | K171423 | Company: Xhale Assurance, Inc. Device Name: Nasal Alar SpO₂ Sensor | DQA |
PHILIPS
Philips CAD (MCS/OBS)
Page 1 of 10
{5}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
The Philips branded Nasal Alar SpO₂ Sensor is substantially equivalent to the legally marketed predicate, Xhale Assurance, Inc.'s, Nasal Alar SpO₂ Sensor cleared via K171423.
## 4. Device Description [21CFR 807.92 (a) (4)]
The Nasal Alar SpO₂ Sensor is a disposable, single patient use pulse oximetry sensor. It has been designed to attach to the patient's nasal alar region (the fleshy region at the side of the nose) as the nasal ala provides a source of perfusion without the potential physiological interruption of sufficient blood flow that can occur when monitoring distal sites, such as the digits. The Nasal Alar SpO₂ Sensor attaches to the patient's ala via skin contacting, adhesive free, soft, silicone rubber cushions which encapsulate the optical components of the sensor. The Nasal Alar SpO₂ Sensor has been designed for connection to monitors that include or are compatible with Philips FAST SpO₂ technology, utilizing a DB-9 connector cable.
The Nasal Alar SpO₂ Sensor is used as an accessory with the Oximeter (Monitors) as part of a System. The Philips Nasal Alar SpO₂ Sensor comprises three major components; a wire assembly attached to a flexible electrical circuit integrating the LEDs and photodiode; a molded plastic body; and molded silicone rubber cushions that consolidate the assembly and provide the point of skin contact to the patient. An applicator clip is provided with the sensor to allow for application of the sensor on the patient's nasal ala.
## 5. Intended Use as required per 21 CFR 807.92(a)(5)
The Nasal Alar SpO₂ Sensor is intended for single patient use, for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate from the nasal ala of adult and pediatric patients (at least 4 years and older and weighing ≥15 kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments, where compatible pulse oximetry monitors are indicated for use, under professional supervision.
PHILIPS
Philips CAD (MCS/OBS)
Page 2 of 10
{6}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
| 6. Comparison of Technological Characteristics with Predicate Device Acquired by Philips from Xhale Assurance, Inc. | |
| --- | --- |
| SIMILARITIES | |
| Item of Comparison | Description/Rationale |
| Oximeter (Monitor) Type | Those that include or are compatible with Philips FAST (Fourier Artifact Suppression Technology) SpO₂ technology. |
| Measurement | - SpO₂ - Pulse Rate |
| Measurement Method | Spectrophotometric measurement of functional arterial oxygen saturation by transmissive mode pulse oximetry. |
PHILIPS
Philips CAD (MCS/OBS)
Page 3 of 10
{7}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
| Biocompatibility Assessment (ISO 10993-1) | Contact Classification Direct and indirect patient contact |
| --- | --- |
| Sensor Specifications | - Single-Patient Use - Worn up to 7 days each, 29 days total cumulative use - ARMS range of ≤3% across the range of 70-100% - 30-240 bpm ± 3 bpm |
| Sensor Storage | -40° to 70°C (-40° to 158°F) |
| Sensor Shelf-Life | Max 5 years Shelf Life |
| DIFFERENCES | |
| Item of Comparison | Description/Rationale |
| Materials | The Nasal Alar SpO₂ Sensor utilizes a new glue formulation, updated colorants for the heat shrink tubing, and a new coverlay material that the Xhale Assurance, Inc, Nasal Alar SpO₂ Sensor predicate device did not. These materials also met the acceptance criteria of testing to ISO 10993-1 and are considered substantially equivalent. |
| LEDS | A minor adjustment was made to the OPTO SMD Dual LED chip of the Nasal Alar SpO₂ Sensor. This change was designed to improve process yield at the component level, while maintaining the previously established specifications shared with the Xhale Assurance, Inc Nasal Alar SpO₂ Sensor predicate device. |
| Labeling | Several labeling changes have been implemented. These labeling changes are all updates from the labeling of the Xhale Assurance, Inc. Nasal Alar SpO₂ Sensor predicate device. |
PHILIPS
Philips CAD (MCS/OBS)
Page 4 of 10
{8}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
| Ingress Protection Rating | The Nasal Alar SpO₂ Sensor has an ingress protection rating of IP34 which is more stringent than the predicate, Xhale Assurance, Inc Nasal Alar SpO₂ Sensor’s rating of IPX1. |
| --- | --- |
| **SUBSTANTIAL EQUIVALENCE SUMMARY** | |
| Operational and technological characteristics form the basis for the determination of substantial equivalence of the Nasal Alar SpO₂ Sensor with the legally marketed predicate device, the Xhale Assurance, Inc Nasal Alar SpO₂ Sensor (K171423). | |
| • The differences do not raise new safety and effectiveness questions as compared to the legally marketed predicate device. • The minor differences in materials of the subject device do not raise different questions of safety and effectiveness. • Performance testing was conducted to support the subject device’s substantial equivalence to the predicate device. | |
| The Nasal Alar SpO₂ Sensor is a modification of the Xhale Assurance, Inc Nasal Alar SpO₂ Sensor (K171423) and therefore is substantially equivalent to the predicate device. | |
## 7. Performance Data
### Recognized Consensus Standards
The Nasal Alar SpO₂ Sensor has passed all safety tests for demonstrated compliance with the consensus standards listed below:
| Standard | FDA Recognition # | Title # |
| --- | --- | --- |
| ISO 10993-1 | 2-258 | Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process |
| ISO 10993-5 | 2-245 | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity |
PHILIPS
Philips CAD (MCS/OBS)
Page 5 of 10
{9}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
| ISO 10993-6 | 2-247 | Biological evaluation of medical devices- Part 6: Tests for local effects after implantation |
| --- | --- | --- |
| ISO 10993-10 | 2-296 | Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization |
| ISO 10993-11 | 2-255 | Biological evaluation of medical devices-Part 11: Tests for systemic toxicity |
| ISO 10993-23 | 2-291 | Biological evaluation of medical devices- Part 23: Tests for irritation |
| ISO 14971 | 5-125 | Medical Devices – Application of risk management to medical devices |
| IEC 60601-1 | 19-49 | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance |
| IEC 60601-1-2 | 19-36 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| IEC 60601-1-6 | 5-132 | Medical electrical equipment. Part 1-6: General requirements for basic safety and essential performance. Collateral standard: Usability |
| ISO 14155 2^{nd} Ed | 2-282 | Clinical investigation of medical devices for human subjects — Good clinical practice |
| ISO 80601-2-61 | 1-139 | Medical electrical equipment Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment |
PHILIPS
Philips CAD (MCS/OBS)
Page 6 of 10
{10}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
# Non-Clinical Tests
1. Sterilization and Shelf Life
Sterilization is not applicable to the Nasal Alar SpO₂ Sensor. The subject device is not designed to be sterilized.
The Nasal Alar SpO₂ Sensor is a single-use, single-patient device and is disposed of upon use life or completion of treatment.
The Nasal Alar SpO₂ Sensor has a defined maximum shelf-life of 5 years. This is supported by V&V testing to ensure product integrity, safety, and device performance over the defined shelf-life period has not been compromised or affected.
2. Biocompatibility
The subject device met the acceptance criteria as defined in the test requirements. Results were deemed as 'PASS'. There were no unexpected results or significant deviations that would have affected the tests.
Tests selected were based on contact and duration, as outlined in ISO 10993-1.
- Cytotoxicity
- Sensitization
- Irritation
- Implantation
- Acute Systemic Toxicity
- Subacute/Subchronic Toxicity
'Irritation' testing was conducted following ISO 10993-10. Based on testing, it was deemed that there was enough data to support the safety of the device without performing additional animal testing as per ISO 10993-2.
3. EMC, Wireless, Electrical, Mechanical, Thermal Safety and Ingress Protection
Prior Electrical Safety and EMC testing of the predicate device, Xhale Assurance, Inc Nasal Alar SpO2 Sensor (K171423) was determined to be representative of the subject device, Nasal Alar SpO2 Sensor after assessment of the modifications.
The Nasal Alar SpO2 Sensor is connected to the applicable patient monitor via cable and is not capable of wireless communication.
PHILIPS
Philips CAD (MCS/OBS)
Page 7 of 10
{11}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
The Nasal Alar SpO2 Sensor was subjected to mechanical and thermal safety testing and was found to perform substantially equivalent to the predicate sensor.
The Nasal Alar SpO2 Sensor and extender cable were evaluated for solid and liquid ingress protection. It was determined that the sensor, when mated with the corresponding SpO2 cable, met the requirements for an ingress protection rating of IP34.
## 4. Performance - Bench
The following Performance Bench Tests were performed:
| Test | Test Method | Acceptance Criteria | Result |
| --- | --- | --- | --- |
| Pull Test | Comparison of electrical continuity before pulling versus after pulling | 0.5V to 0.8V, verified by measuring the forward voltage drop of the diode in the connected probe. 1.5V to 1.8V verified by measuring the forward voltage drop of the LED in the connected probe. | Pass |
| Shelf-life | Accelerated and real time aging per ASTM F1980 | 5 years | Pass |
| Environmental-Reliability | Comparison of force values measured before and after testing | Force Measurement: Force after cycling is equal to the Force before cycling +/- 10% | Pass |
| DB-9 Mating Force | Analysis of insertion and withdrawal force | Not Applicable – this test was to gather average insertion and withdrawal force of the Alar Sensor DB-9 connector | Not Applicable – this test was to gather average insertion and withdrawal force of the Alar Sensor DB-9 connector |
| LED Transmission | Measured | LED RED | Pass |
PHILIPS
Philips CAD (MCS/OBS)
Page 8 of 10
{12}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
| | transmission/current of the LED with a 20mA square wave at 1.2kHz | Transmission: Tred < 300 nA/mA LED IR Transmission: TIR < 700 nA/mA | |
| --- | --- | --- | --- |
| Pulse Rate | Comparison of displayed readings from the connected monitor versus a simulated pulse rate. | 30-240 bpm ± 3bpm | Pass |
PHILIPS
Philips CAD (MCS/OBS)
Page 9 of 10
{13}
PHILIPS
Nasal Alar SpO₂ Sensor
Traditional 510(k)
510(k) Summary
## Clinical Tests:
Clinical SpO₂ testing was conducted per ISO 80601-2-61 to verify the accuracy of the Nasal Alar SpO₂ Sensor. The testing confirmed and overall ARMS of ≤3%.
The data from the non-clinical and clinical testing supports the substantial equivalence of the subject device.
## 8. CONCLUSIONS
The Nasal Alar SpO₂ Sensor is substantially equivalent to the predicate device in terms of:
- The Indications and Intended Use,
- Operating Principles,
- Conditions of Use, and
- Technological Characteristics.
Verification and Validation (V&V) test results support that the technological characteristics of the Nasal Alar SpO₂ Sensor are substantially equivalent to the predicate device, Xhale Assurance, Inc. Nasal Alar SpO₂ Sensor (K171423). Risk Management activities, V&V testing, and Performance testing were conducted to support the substantial equivalence of the subject device per FDA recognized consensus standards, recognized industry standards, and internal protocol/procedures.
PHILIPS
Philips CAD (MCS/OBS)
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.