K253871 · Radcliffe Watts, LLC · FTC · Mar 4, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253871
Device Name
Klär Lite (RCW-KL1000)
Applicant
Radcliffe Watts, LLC
Product Code
FTC · General, Plastic Surgery
Decision Date
Mar 4, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Klär Lite is an Ultraviolet Light Emitting Medical Device. It is intended for use in localized phototherapeutic treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leukoderma on all skin types (I-VI).
Device Story
Klär Lite is a handheld, prescription-use, narrowband UVB LED phototherapy device for home or clinical use. It emits 311 nm light to treat dermatologic conditions. The device features an integrated diffuser and thermal sensor for temperature-compensated dose control. A microcontroller manages exposure via preset or manual modes, utilizing real-time temperature feedback to trigger auto-shutdown if necessary. The healthcare provider or patient uses the device to deliver localized light therapy; the output affects clinical decision-making by providing a controlled, targeted therapeutic dose. Benefits include localized treatment of skin conditions with integrated safety features like optical filters and thermal interlocks.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including electrical/mechanical safety (IEC 60601-1, IEC 60601-1-11), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), biocompatibility (ISO 10993-1, 5, 10, 11), software verification (IEC 62304), and usability engineering (IEC 62366-1).
Technological Characteristics
Narrowband UVB LED array (311 ± 2 nm); irradiance 3.0-5.0 mW/cm²; handheld form factor with integrated diffuser. Powered by medical-grade DC supply. Software-controlled timer and temperature-compensated dose control. Biocompatible materials tested per ISO 10993. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 62471 (photobiological safety).
Indications for Use
Indicated for localized phototherapeutic treatment of psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis, and leukoderma in patients of all skin types (I-VI).
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 4, 2026
Radcliffe Watts, LLC
Dan Smith
Chief Executive Officer
201 E 5th St.
Mansfield, Ohio 44902
Re: K253871
Trade/Device Name: Klär Lite (RCW-KL1000)
Regulation Number: 21 CFR 878.4630
Regulation Name: Ultraviolet Lamp For Dermatologic Disorders
Regulatory Class: Class II
Product Code: FTC
Dated: November 23, 2025
Received: December 4, 2025
Dear Dan Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K253871 - Dan Smith
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253871 - Dan Smith
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.03.04
15:46:19 -05'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K här Lite
Page 9 of 44
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253871 | ? |
| Please provide the device trade name(s). | | ? |
| Klär Lite (RCW-KL1000) | | |
| Please provide your Indications for Use below. | | ? |
| The Klär Lite is an Ultraviolet Light Emitting Medical Device. It is intended for use in localized phototherapeutic treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leukoderma on all skin types (I-VI). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) #: K253871
510(k) Summary
Prepared on: 2025-12-04
| Contact Details | 21 CFR 807.92(a)(1) |
| --- | --- |
| Applicant Name | Radcliffe Watts, LLC |
| Applicant Address | 201 E 5th St Mansfield OH 44902 United States |
| Applicant Contact Telephone | 714-292-5532 |
| Applicant Contact | Mr. Daniel Smith |
| Applicant Contact Email | admin@radcliffewatts.com |
| Device Name | 21 CFR 807.92(a)(2) |
| Device Trade Name | Klär Lite (R CWKL10®) |
| Common Name | Ultraviolet lamp for dermatologic disorders |
| Classification Name | Light, Ultraviolet, Dermatological |
| Regulation Number | 878.4630 |
| Product Code(s) | FTC |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) |
| Product Code | Flowering |
| K170489 | Clarify Medical Phototherapy System |
| | FTC |
| Device Description Summary | 21 CFR 807.92(a)(4) |
| The Klär Lite is a narrowband UVB light emitting phototherapy medical device intended for treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leukoderma on all skin types (I-VI). | |
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) |
| The Klär Lite is an Ultraviolet Light Emitting Medical Device. It is intended for use in localized phototherapeutic treatment of dermatologic conditions such as psoriasis, vitiligo, atopic dermatitis (eczema), seborrheic dermatitis and leukoderma on all skin types (I-VI). | |
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| The indications of use are the same between this device and predicate/reference devices. | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| Please reference the attached document. | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
{5}
The Klär Lite phototherapeutic device was evaluated through a comprehensive series of non-clinical tests to support a determination of substantial equivalence with the predicate device.
## Electrical and Mechanical Safety:
Testing was performed in accordance with IEC 60601-1:2012 (Edition 3.1) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, and IEC 60601-1-11:2015 Medical electrical equipment — Part 1-11: General requirements for basic safety and essential performance — Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment for home-use medical electrical equipment. All results demonstrated compliance with requirements for protection against electrical shock, mechanical hazards, fire, and thermal safety.
## Electromagnetic Compatibility (EMC):
Conducted per IEC 60601-1-2:2014 (Edition 4.0) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. The device met all emissions and immunity limits including ESD, radiated/conducted RF, EFT/Burst, Surge, Power Frequency Magnetic Field, and Voltage Dips/ Interruptions.
## Photobiological Safety and Performance:
Evaluated per IEC 62471:2006 Photobiological safety of lamps and lamp systems and ISO 15004-2, confirming the emitted optical radiation is within safe exposure limits for intended therapeutic use.
## Biocompatibility:
Evaluated per the ISO 10993 series (specifically ISO 10993-1, Biological evaluation of medical devices—Part 1: Evaluation and testing within a risk management process. Part 5: Tests for in vitro cytotoxicity, Part-10, Tests for skin sensitization, and Part-11: Tests for systemic toxicity. Tests confirmed the safety of all patient-contacting materials for intact skin contact, including cytotoxicity, sensitization, and irritation testing performed under GLP conditions.
## Software Verification and Validation:
Conducted in accordance with IEC 62304:2006 + AMD 1:2015, Medical device software — Software life cycle processes, supporting software reliability and safety at the determined risk classification with basic level of documentation.
## Risk Management:
Completed per ISO 14971:2019 Medical devices — Application of risk management to medical devices, to ensure identified hazards are effectively controlled and residual risk is acceptable.
## Labeling and Usability:
Evaluated per IEC 62366-1:2015 EC 62366-1:2015 and validated through formative and summative usability studies.
Collectively, these tests demonstrate that Klär Lite meets applicable FDA-recognized consensus standards.
Clinical testing was not required to establish substantial equivalence for the Klär Lite phototherapeutic device.
The results of all nonclinical verification and validation testing demonstrate that the Klär Lite phototherapeutic device is as safe and effective.
Conclusion: The Klär Lite device is substantially equivalent to the predicate device in terms of safety, effectiveness, and performance.
{6}
| Feature | Predicate Device – Skylit Phototherapy System (K170489) | Subject Device – Klär Lite K253871 | Comparison Result |
| --- | --- | --- | --- |
| Technology Type | Narrowband UV-B LED phototherapy | Narrowband UV-B LED phototherapy | Identical |
| Wavelength | 300-320 nm | 311 ± 2 nm (UV-B) (nominal peak within UV-B emission range 295-320 nm) | Comparable |
| Light Source | LED array | LED array | Identical |
| Output Irradiance | 3.0-15.0 mW/cm² | 3.0-5.0 mW/cm² | Within Predicate Range |
| Treatment Area | 4.5 cm² | 5.08 cm x 5.08 cm | Comparable |
| Exposure Control | Software-controlled timer and preset dose by skin type | Microcontroller algorithm with skin-type presets and real-time temperature feedback | Equivalent / Enhanced |
| Safety Features | Optical filter, thermal shutdown, firmware interlocks | Optical filter, integrated thermal sensor with auto-shutdown | Comparable |
| Applicator Type | Handheld home-use applicator with comb guide | Handheld home-use applicator with integrated diffuser | Equivalent |
| Operating Modes | Preset and manual | Preset and manual with temperature-compensated dose control | Equivalent / Enhanced |
| Cleaning Method | 70 % IPA wipe (per IFU) | Validated 70 % IPA wipe per (RFV-1010A, CEV-1010A, SLJ-1010A) – no material damage | Equivalent / Enhanced |
{7}
| Power Input | Rechargeable battery or DC supply | Medical-grade DC supply | Comparable |
| --- | --- | --- | --- |
| Software Classification | IEC 62304 Class B | IEC 62304 Class B | Identical |
| Biocompatibility | ISO 10993 tested for skin contact | ISO 10993 tested for skin contact | Identical |
| Electrical Safety | IEC 60601-1 Ed 3.1 | IEC 60601-1 Ed 3.1 | Identical |
| EMC | IEC 60601-1-2 Ed 4.0 | IEC 60601-1-2 Ed 4.0 | Identical |
| Photobiological Safety | IEC 62471 | IEC 62471 | Identical |
| Usability Engineering | IEC 62366-1 | IEC 62366-1 | Identical |
| Prescription Status | Rx-only | Rx-only | Identical |

Figure 1. Skylit Phototherapy System (K170489)

Figure 2. Subject Device - Klär Lite Phototherapy System
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.