K253863 · Serim Research · FKJ · Mar 17, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K253863
Device Name
GUARDIAN™ Quick Dip Residual Chlorine (5212)
Applicant
Serim Research
Product Code
FKJ · Gastroenterology, Urology
Decision Date
Mar 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Indications for Use
GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment.
Device Story
Device is a semi-quantitative, single-use reagent test strip for detecting residual free chlorine in hemodialysis rinse water. Input is a liquid sample; user dips the strip for 2 seconds, shakes off excess, and compares the reagent pad color to a provided chart within 5 seconds. The reagent pad contains an indicator, non-ionic surfactant, and reducing agent; free chlorine oxidizes the redox indicator to produce a yellow-to-green color change proportional to concentration (0–5 ppm). Used by clinical staff or technicians in hemodialysis facilities to ensure adequate rinsing of dialysate lines post-disinfection. Output is a visual colorimetric reading; informs clinical decision-making regarding whether equipment is safe for patient use. Benefits include rapid, convenient verification of chemical removal, preventing patient exposure to residual bleach.
Clinical Evidence
Bench testing only. Performance evaluated using water samples spiked with sodium hypochlorite across 0, 0.5, and 1.0 ppm concentrations. Validation included instructions for use (dip/read time), shelf-life stability (12-month accelerated), use-life stability (30-month accelerated), and comparative sensitivity/specificity. Results confirmed performance equivalent to the predicate device.
Technological Characteristics
Single-use reagent test strip; 0.20-inch square reagent pad on 3.25 x 0.20-inch white opaque polystyrene handle. Sensing principle: chemical oxidation of a redox indicator by free chlorine. Detection range: 0–5 ppm. Colorimetric output: yellow/green complex. No electronic components or software.
Indications for Use
Indicated for use in hemodialysis settings to monitor residual chlorine bleach (sodium hypochlorite) concentrations in rinse water following equipment disinfection.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 17, 2026
Serim Research
Amy Sheets
Quality Operations and Regulatory Affairs Manager
3506 Reedy Dr.
Elkhart, Indiana 46514
Re: K253863
Trade/Device Name: GUARDIAN™ Quick Dip Residual Chlorine (5212)
Regulation Number: 21 CFR 876.5820
Regulation Name: Hemodialysis System And Accessories
Regulatory Class: Class II
Product Code: FJK
Dated: February 17, 2026
Received: February 17, 2026
Dear Amy Sheets:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K253863 - Jackie Nelson
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253863 - Jackie Nelson
Page 3
Sincerely,
# MAURA ROONEY -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253863 | |
| Device Name GUARDIAN™ Quick Dip Residual Chlorine (5212) | |
| Indications for Use (Describe) | |
| GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K253863
Page 1 of 5
CONFIDENTIAL
Serim Research Corporation
510(k) Premarket Notification
GUARDIAN™ Quick Dip Residual Chlorine (5212)
# 510 (k) Summary
Prepared: December, 2025
Submitter: Serim Research Corporation
Address: 3506 Reedy Drive
Elkhart IN 46514
Phone: 574-264-3440
Fax: 574-266-6222
Contact: Amy Sheets
Quality Operations and Regulatory Affairs Manager
Device Trade Name: GUARDIAN™ Quick Dip Residual Chlorine (5212)
Common or Usual Name: Free Chlorine Reagent Strips
Device Classification Name: Filter, Blood, Dialysis
Product Code: FKJ
Class: II
Regulation Number: 876.5820
Substantial Equivalence: The GUARDIAN™ Quick Dip Residual Chlorine (5212) is substantially equivalent to Serim® GUARDIAN™ RESIDUAL CHLORINE/ K901734
Subject Device Description: The device is a semiquantitative, single use, reagent test strip made up of a 0.20-inch square reagent pad that has been chemically treated to detect residual free chlorine in water. The pad is affixed to one end of a 3.25 inch by 0.20-inch white opaque polystyrene strip.
Subject Device Intended Use: GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment.
{5}
K253863
Page 2 of 5
Serim Research Corporation
510(k) Premarket Notification
GUARDIAN™ Quick Dip Residual Chlorine (5212)
CONFIDENTIAL
## Indications for use:
GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment.
## Technological Characteristics:
The GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips contain one indicator, a non-ionic surfactant, reducing agent, and other inactive ingredients. Free chlorine reacts with the indicator to form shades of yellow-green complex. The amount of green color formed is dependent on the concentration of free chlorine in the sample. The color blocks on the label are related to chlorine concentration in terms of parts per million (ppm). The device is used to detect free chlorine concentrations between 0 and 5 ppm. Refer to table 1 for more information.
## Predicate Device:
Serim® GUARDIAN™ RESIDUAL CHLORINE/K901734
## Predicate Device Description:
The device is a semi-quantitative, single use, reagent test strip made up of a 0.20-inch square reagent pad that has been chemically treated to detect residual free chlorine in water. The pad is affixed to one end of a 3.25 inch by 0.20-inch white opaque polystyrene strip.
## Predicate Device Intended use:
Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment
## Performance:
The performance of the GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips was evaluated using water samples, sodium hypochlorite was added to give a range of free chlorine levels. The performance of the GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips is equivalent to the predicate device, Serim® GUARDIAN™ RESIDUAL CHLORINE Reagent Strips.
Table 1 Summarize the technological characteristic for the subjected device against predicate device
{6}
Serim Research Corporation
510(k) Premarket Notification
GUARDIAN™ Quick Dip Residual Chlorine (5212)
CONFIDENTIAL
Table 1 Summary of Technological Characteristics Table
| Parameters | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Identification | GUARDIAN™ Quick Dip Residual Chlorine (5212) | Serim® GUARDIAN™ RESIDUAL CHLORINE | Different |
| Intended use | Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment. | Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment | Similar |
| Indications for use | Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment. | Test Strips provide a convenient means for indicating the concentration of chlorine bleach (sodium hypochlorite) remaining in the solution used to rinse dialysate lines following disinfection of hemodialysis equipment | Similar |
| Analyte | Free Chlorine | Free Chlorine | Similar |
| Indicator strip | Paper indicator adhered to a plastic handle | Paper indicator adhered to a plastic handle | Similar |
| Technological characteristics | Free chlorine oxidizes the redox indicator, to form a color that is directly proportional to the amount of chlorine in the sample | Free chlorine oxidizes the redox indicator to form a color that is directly proportional to the amount of chlorine in the sample | Similar |
| Indicator agents | Redox indicator | Redox indicator | Similar |
| Semi-quantitative Test Method | Dip the test strip for 2 seconds | Immerse indicator pad of test strip in sample solution | Different |
| | Gently shake strip to remove excess sample | Move strips back and forth vigorously for 30 seconds | |
| | Compare the pad to color chart within 5 seconds | Remove strip and compare to color chart within 10 seconds | |
| Color Blocks | 0, 0.5, 1.0, 2.0, and 5.0 ppm | 0, 0.5, 1.0, 2.0, and 5.0 ppm | Similar |
{7}
Serim Research Corporation
510(k) Premarket Notification
GUARDIAN™ Quick Dip Residual Chlorine (5212)
CONFIDENTIAL
K253863
Page 4 of 5
| Color Development | Yellow / shades of green | White / shades of purple | Different |
| --- | --- | --- | --- |
## Summary of Non-Clinical Performance Date:
The performance of the GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips was evaluated in non-clinical tests in a range of free chlorine concentration at 0, 0.5 and 1.0 ppm free chlorine. The results show that the proposed device is an effective monitor of residual concentration of free chlorine.
Table 2 summarizes the non-clinical testing performed by Serim Research to demonstrate safety and effectiveness of the GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips in monitoring the concentration of free chlorine in rinse water.
Table 2
| Study | Results |
| --- | --- |
| Performance | Met reference value 0, 0.5, and 1.0 ppm free chlorine. |
| Instructions for Use Validation | Met reference value with +/- 1 second dip time validation, and + 8 seconds read time validation. |
| Close Bottle Shelf-Life Stability | Met reference value for 12 months (Accelerated) |
| Open Bottle Use-Life Stability | Met reference value for 30 months (Accelerated) |
| Comparative Sensitivity and Specificity | Met reference value at <0.5ppm and ≥ 0.5ppm free chlorine concentration. |
| Light study | Met reference value under cool white-fluorescent light and day light. |
The performance of the GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips was evaluated using water samples, sodium hypochlorite was added to give a range of free chlorine levels. The performance of the GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips is equivalent to the predicate device, Serim® GUARDIAN™ RESIDUAL CHLORINE Reagent Strips.
{8}
Kerim Research Corporation
510(k) Premarket Notification
GUARDIAN™ Quick Dip Residual Chlorine (5212)
CONFIDENTIAL
K253863
Page 5 of 5
# Conclusion:
The GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips have the same intended use as the predicate device. Both test strips measure the free chlorine levels in water. The GUARDIAN™ Quick Dip Residual Chlorine (5212) Test Strips have no characteristics that raise new types of safety or effectiveness questions. The GUARDIAN™ Quick Dip Residual Chlorine (5212) test strips can be used to detect residual chlorine levels in rinse water from dialysis equipment which had been disinfected with chlorine bleach
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.