CMN Capillary Function with Virtual Expert for MRI, CT and CBCT
K253831 · Cercare Medical A/S · LLZ · May 18, 2026 · Radiology
Device Facts
Record ID
K253831
Device Name
CMN Capillary Function with Virtual Expert for MRI, CT and CBCT
Applicant
Cercare Medical A/S
Product Code
LLZ · Radiology
Decision Date
May 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K253831 · May 18, 2026
CMN Capillary Function with Virtual Expert for MRI, CT and CBCT
Cercare Medical A/S
Retrospective clinical patient imaging data (acute stroke patients)
Retrospective clinical patient data was used to conduct a blinded multi-reader multi-case (MRMC) study to confirm the non-inferiority of the CBCT-derived perfusion parameter maps compared to the predicate CT device.
Diagnostic acceptability of CBCT-derived perfusion parameter maps; non-inferiority on diagnostic-confidence endpoint; territory-agreement analysis.
Indications for Use
Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians. The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI, CT, and Cone-Beam CT (CBCT). The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI, CT, or CBCT, including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI, CT or CBCT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. The Virtual Expert module allows the calculation of regions of interest and the visual inspection of time attenuation curves. For MRI, one clinical application is to visualize the apparent blood perfusion and diffusion and to calculate ADC threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. For CT, one clinical application is to visualize the apparent blood perfusion and to calculate rCBF threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. Areas of decreased perfusion appear as areas of changed signal intensity: • Lower signal intensity for CBF and CBV • Higher signal intensity for TTP, MTT, and Tmax
Device Story
Software-only device; processes DICOM-compliant MRI, CT, and CBCT images; provides visualization, quantification, and analysis of hemodynamic status and tissue diffusion. Operates on standard off-the-shelf PCs or virtual platforms (e.g., VMware). Capillary Function module calculates perfusion parameters (e.g., rCBF, rCBV, Tmax) and DWI maps; Virtual Expert module performs automatic ROI delineation and volumetric calculations for stroke assessment. Used by neuroradiologists and technicians in clinical settings as decision support; clinician retains final treatment authority. Output maps and metrics assist in identifying brain tissue affected by acute stroke; helps clinicians assess perfusion/diffusion mismatch to guide clinical decision-making.
Clinical Evidence
Bench testing used 96 simulated CBCT perfusion phantoms; 10/13 parameter-algorithm combinations met MS-SSIM, patch-bias, and patch-correlation criteria. A blinded MRMC clinical study with 13 neuroradiologists evaluated 35 acute stroke patients. Results confirmed non-inferiority of CBCT-derived maps vs. CT predicate on diagnostic-confidence (mean difference -0.171, 95% CI: -0.249 to -0.093). Territory-agreement analysis showed inter-modality weighted κ (0.680) non-inferior to intra-rater MDCT baseline (κ=0.690, p=0.629).
Technological Characteristics
Software-only image processing system; DICOM-compliant; runs on standard PC/virtual hardware. Uses perfusion post-processing and DWI analysis algorithms. Supports MRI, CT, and CBCT modalities. Integrates into radiological networks via VPN/SSH. Conforms to ISO 13485, ISO 14971, IEC 62304, and NEMA PS 3.1-3.20 standards.
Indications for Use
Indicated for trained professionals, including physicians and medical technicians, to perform image viewing, processing, and analysis of MRI, CT, and CBCT datasets for patients, including those with acute stroke, to visualize tissue perfusion, diffusion, and hemodynamic status.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
CMN Capillary Function with Virtual Expert for MRI and CT (K241989)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 18, 2026
Cercare Medical A/S
% Jennifer Willner
Official Correspondent
JW Regulatory Consulting, LLC
406 Wacouta St.
Unit 417
Saint Paul, MN 55101
Re: K253831
Trade/Device Name: CMN Capillary Function with Virtual Expert for MRI, CT and CBCT
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: LLZ
Dated: April 16, 2026
Received: April 16, 2026
Dear Jennifer Willner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253831 - Jennifer Willner
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K253831 - Jennifer Willner
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253831 | ? |
| Please provide the device trade name(s). | | ? |
| CMN Capillary Function with Virtual Expert for MRI, CT and CBCT | | |
| Please provide your Indications for Use below. | | ? |
| Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians. The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI, CT, and Cone-Beam CT (CBCT). The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI, CT, or CBCT, including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI, CT or CBCT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. The Virtual Expert module allows the calculation of regions of interest and the visual inspection of time attenuation curves. For MRI, one clinical application is to visualize the apparent blood perfusion and diffusion and to calculate ADC threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. For CT, one clinical application is to visualize the apparent blood perfusion and to calculate rCBF threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. Areas of decreased perfusion appear as areas of changed signal intensity: • Lower signal intensity for CBF and CBV • Higher signal intensity for TTP, MTT, and Tmax | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
Cercare Medical
# 510(K) SUMMARY
510(k) Number: K253831
Date Prepared: April 15, 2026
Table 1: Submitter Information
| Manufacturer: Cercare Medical A/S Inge Lehmanns Gade 10 DK-8000 Aarhus C Denmark US FDA ERN: 3019844085 | Manufacturer's Contact Person: Mikkel Bo Hansen Chief Scientific Officer Phone: +45 22890125 Email: mbh@cercare-medical.com |
| --- | --- |
Table 2: Device Information
| Trade Name | Cercare Medical Neurosuite (CMN) Capillary Function with Virtual Expert for MRI, CT and CBCT |
| --- | --- |
| Common Name | CMN Capillary Function with Virtual Expert |
| Classification Name | System, Image Processing, Radiological |
| Regulation | 21 CFR 892.2050 |
| Product Code | LLZ |
| Regulatory Classification: | Class II |
| Device Panel: | Radiology (OHT8) |
The CMN Capillary Function with Virtual Expert for MRI, CT and CBCT device is substantially equivalent to the CMN Capillary Function with Virtual Expert for MR and CT Predicate Device (Table 3). The Predicate Device has not been subject to a design-related recall.
Table 3: Predicate Devices
| Predicate Device | Manufacturer | FDA 510(k) |
| --- | --- | --- |
| CMN Capillary Function with Virtual Expert for MRI and CT | Cercare Medical A/S | K241989 |
# Device Description
Cercare Medical Neurosuite is a software-only device designed to streamline medical image processing by providing for the visualization and study of medical images. CMN can be installed on a customer PC, or it can be accessed remotely using remote desktop technologies. CMN provides viewing, quantification, analysis and reporting capabilities. CMN is not intended as a dedicated PACS system for long-term persistent storage of patient data.
CMN is software that is intended for use by trained professionals, including physicians and medical technicians. The software provides cerebral image processing capabilities. CMN is intended to be used as decision support software only and the clinician continues to provide all treatment decisions.
Cercare Medical A/S
April 2026
{5}
Cercare Medical
The software is intended to visualize and study neuroimaging by image viewing and registration of medical images. CMN works with MRI (magnetic resonance image), CT (computed tomography), and CBCT (Cone-Beam CT) technologies.
CMN accepts and produces data sets in the DICOM format. DICOM is a standard format for storing and transmitting medical image data in vendor neutral format and is managed by the DICOM Standards Committee.
CMN is a platform that allows for the addition of certain modules for further analysis. Two of these modules are used in this device: Capillary Function and Virtual Expert.
## CMN Capillary Function
Capillary Function, when activated in the installed Cercare Medical Neurosuite, provides further functionalities for reading, writing, visualizing and studying medical images.
Capillary Function provides perfusion post-processing technologies, where dynamically acquired perfusion MRI or perfusion CT/CBCT series can be processed to yield information relevant for assessment of the hemodynamic status of a patient.
Capillary Function generates hemodynamic markers, which can be used for management of diseases with possibly compromised hemodynamic function, such as ischemic stroke and tumors.
The generated output maps can be viewed by standard DICOM image viewers. In addition, Capillary Function includes the possibility for post-processing diffusion-weighted imaging (DWI) MRI data. Post-processing of DWI data results in maps reflective of local water diffusion properties. The post-processed DWI-derived maps can be viewed in standard DICOM image viewers. Capillary Function thus works with MRI, CT and CBCT technologies.
## CMN Virtual Expert
Virtual Expert, when activated in the installed CMN Capillary Function, provides further functionalities for reading, writing, visualizing, and studying medical images.
Virtual Expert provides automatic delineation of regions of interest (ROI) relevant for stroke patient assessment based on perfusion and diffusion image output generated by the Capillary Function module. Specifically, diffusion MRI images are used to generate threshold masks of perceived core lesions, whereas MRI, CT or CBCT perfusion images are used to generate threshold masks of perceived perfusion restriction. For CT/CBCT perfusion, derived perfusion images are used to generate threshold masks of perceived core lesions. Virtual Expert thus works with MRI, CT and CBCT technologies.
The generated masks can be combined into a mismatch ROI. Volumetric calculations and ratios can be calculated from the computed ROI.
Cercare Medical A/S
April 2026
{6}
Cercare Medical
# Indications for Use
Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians.
The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI, CT, and Cone-Beam CT (CBCT).
The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI, CT or CBCT including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI, CT or CBCT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data.
The Virtual Expert module allows the calculation of regions of interest and the visual inspection of time attenuation curves. For MRI, one clinical application is to visualize the apparent blood perfusion and diffusion and to calculate ADC threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke.
For CT, one clinical application is to visualize the apparent blood perfusion and to calculate rCBF threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke.
Areas of decreased perfusion appear as areas of changed signal intensity:
- Lower signal intensity for CBF and CBV
- Higher signal intensity for TTP, MTT, and Tmax
# Technological Characteristics
CMN Capillary Function with Virtual Expert for MRI, CT and CBCT provides the following functions:
- processes DICOM images from multiple sources to provide visualization of changes of tissue perfusion, diffusion and change
- receives DICOM images from external DICOM image providers (modalities (MRI Scanners), PACS and Workstations) and sends DICOM images to external image consumers
- storage of status and results, and references therein, in a searchable database
Cercare Medical A/S
April 2026
{7}
Cercare Medical
CMN Capillary Function with Virtual Expert is a DICOM-compliant PACS software that provides comprehensive functionality to transfer, process and display modality specific imaging data. CMN runs on standard off-the-shelf computer and networking hardware and is entirely independent from CT, MRI or PACS platforms. It supports secure VPN networking or encapsulated Secure Shell (SSH) and seamlessly integrates into an existing radiological data network.
The primary users of CMN software are medical imaging professionals who analyze tissue using MRI, CT or CBCT images. The images generated by CMN provide additional diagnostic information, which is derived from the temporal/diffusion/density features of the native MRI image.
## Differences in Technical Characteristics/Performance with Respect to the Predicate Device
CMN Capillary Function with Virtual Expert for MR, CT and CBCT is identical to the Predicate CMN Capillary Function with Virtual Expert for MR and CT (K241989) with the exception of the following features:
- CBCT perfusion post-processing
- CBCT-based detection of stroke lesions
The perfusion processing algorithms cleared under K241989 for CT are reused in whole for the processing of CBCT data, since the fundamental physiological principles are the same, only the modality differs. Both modalities are based on tissue and contrast agent absorption of X-rays (attenuation) but the so-called Hounsfield unit scale is not directly applicable in the case of CBCT. This necessitates a few additional pre-processing steps to technically align the data in terms of image intensity and temporal sampling rate to allow processing with the algorithms cleared under K241989.
The Subject and Predicate devices are based on the same technological elements of viewing, processing and analyzing DICOM image data to assist the clinician during diagnostic procedures. These differences do not impact the intended use or raise new questions with the safety and performance of the device.
## Performance Standards
CMN Capillary Function with Virtual Expert has been developed in conformance with the following standards and FDA guidance, as applicable:
- ISO 13485:2016, Quality management systems – Requirements for regulatory purposes
- ISO 14971:2019, Medical devices – Application of risk management to medical devices
- IEC 62304:2015, Medical device software – Software lifecycle processes
- IEC 82304-1:2016, Health Software Part 1: General Requirements for Product Safety
- ISO 15223-1:2021, Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements
Cercare Medical A/S
April 2026
{8}
Cercare Medical
- ISO 20417:2021, Medical devices – Information to be supplied by the manufacturer
- NEMA PS 3.1-3.20:2024, Digital Imaging and Communications in Medicine (DICOM)
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, October 2018
- Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data – Premarket Notification [510(k)] Submissions, September 2022
- Content of Premarket Submissions for Device Software Functions, June 2023
## Performance Data
CMN Capillary Function with Virtual Expert for MR, CT and CBCT complies with DICOM (Digital Imaging and Communications in Medicine), developed by the American College of Radiology and the National Electrical Manufacturers Association – NEMA PS 3.1-3.20.
Cercare conducted extensive performance validation testing and software verification and validation testing of the CMN Capillary Function with Virtual Expert. This performance validation testing demonstrated that the CMN system provides accurate representation of key processing parameters under a range of clinically relevant parameters associated with the intended use of the software. Software performance, validation and verification testing demonstrated that the CMN system met all design requirements and specifications.
Comparative performance testing was performed for the Subject Device with respect to the Predicate Device through bench testing including retrospective clinical data.
For the Capillary Function module, a direct paired device comparison was conducted using 96 digitally simulated CBCT perfusion phantoms spanning clinically relevant ranges of motion, noise, and signal strength. Each phantom was processed by both the CBCT Subject Device and the CT Predicate Device, and the resulting parameter maps were compared using three acceptance metrics: multi-scale structural similarity index (MS-SSIM), patch-level bias, and patch-level correlation. The acceptance criteria were met for 10 of 13 parameter-algorithm combinations. Three timing-sensitive parameters (Delay, Tmax, TTP) showed reduced MS-SSIM acceptance rates attributable to the lower temporal sampling rate of CBCT; patch-bias and patch-correlation acceptance remained above acceptance threshold for these parameters.
For the Virtual Expert module, comparison testing of retrospective patient CT perfusion imaging consisted of primary analyses of volumetric and spatial agreement per-patient and per-lesion. Secondary clinical application assessments (high-level context of the devices not associated with clinical claims) were then conducted in terms of image-driven decision to treat analysis through the so-called DEFUSE3 criteria. The Virtual Expert comparison testing was modeled by the comparison testing conducted for the Predicate Device (K241989) and recognizing that direct comparison across different modalities acquired with varying time delay between acquisitions can be governed by several unknown sources of variation – both technical and disease development-wise.
Cercare Medical A/S
April 2026
{9}
Cercare Medical
A blinded, multi-reader multi-case (MRMC) clinical study was additionally conducted in which 13 expert neuroradiologists (8 board-certified and 5 with at least one year of neuroradiology experience) independently scored the diagnostic acceptability of CBCT-derived perfusion parameter maps from 35 acute stroke patients. The MRMC analysis confirmed non-inferiority of the CBCT device relative to the CT Predicate Device on the primary diagnostic-confidence endpoint (observed mean difference -0.171, 95% CI: -0.249 to -0.093; non-inferiority margin -0.5). The co-primary territory-agreement analysis showed that inter-modality weighted $\kappa$ (0.680) is non-inferior to the matched intra-rater delayed MDCT re-read baseline ( $\kappa = 0.690$ ; difference +0.019, 95% CI: -0.065, 0.128; $p = 0.629$ ).
The primary known limitation of CBCT perfusion is the lower temporal sampling rate inherent to the modality, which reduces MS-SSIM for timing-sensitive parameters. The spatial structure and clinical diagnostic utility of the maps are preserved, as confirmed by the $\geq 90\%$ patch-bias and patch-correlation acceptance rates and the non-inferior reader scores in the clinical study.
Together with software verification and validation, the performance validation demonstrated that CMN Capillary Function with Virtual Expert (MRI, CT & CBCT) satisfies all design requirements and device specifications and is substantially equivalent to the Predicate Device.
# Substantial Equivalence
In comparison with the Predicate CMN Capillary Function with Virtual Expert device (K241989), CMN Capillary Function with Virtual Expert (MR, CT & CBCT) has the same intended use and similar indications, technological characteristics and principles of operation as described in the comparison table below.
Table 4: Substantial Equivalence Comparison Table
| Description | Subject Device | Predicate Device (K241989) | Conclusion |
| --- | --- | --- | --- |
| Product Name | CMN Capillary Function with Virtual Expert MRI, CT & CBCT | CMN Capillary Function with Virtual Expert CT & MRI | |
| Manufacturer | Cercare Medical A/S | Cercare Medical A/S | |
| Product Code / Regulation | LLZ / 21CFR 892.2050 | LLZ / 21CFR 892.2050 | Same |
| Indications for Use | Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians. The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. | Cercare Medical Neurosuite (CMN) and associated modules is an image processing software package to be used by trained professionals, including physicians and medical technicians. The CMN software package runs on standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, | Similar; Subject Device MR & CT capability identical to Predicate Device and CBCT capability is similar to CT. No new questions of safety or effectiveness. |
Cercare Medical A/S
April 2026
{10}
Cercare Medical
| Description | Subject Device | Predicate Device (K241989) | Conclusion |
| --- | --- | --- | --- |
| | Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with MRI, CT, and Cone-Beam CT (CBCT). The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI, CT, or CBCT, including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI, CT or CBCT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. The Virtual Expert module allows the calculation of regions of interest and the visual inspection of time attenuation curves. For MRI, one clinical application is to visualize the apparent blood perfusion and diffusion and to calculate ADC threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. For CT, one clinical application is to visualize the apparent blood perfusion and to calculate rCBF threshold volume, Tmax | processing, and analysis of images. Data and images are acquired through DICOM (Digital Imaging and Communications in Medicine) compliant imaging devices. CMN provides viewing capabilities of datasets acquired with CT and MRI. The Capillary Function module provides analysis capabilities for functional and dynamic imaging datasets acquired with MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast-enhanced imaging data for MRI and CT). The Capillary Function module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. In addition, the Capillary Function module's DWI technology is used to visualize local water diffusion properties from the analysis of diffusion-weighted MRI data. The Virtual Expert module allows the calculation of regions of interest and the visual inspection of time attenuation curves. For MRI, one clinical application is to visualize the apparent blood perfusion and diffusion and to calculate ADC threshold volume, Tmax threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. For CT, one clinical application is to visualize the apparent blood perfusion and to calculate rCBF threshold volume, Tmax | |
Cercare Medical A/S
April 2026
{11}
Cercare Medical
| Description | Subject Device | Predicate Device (K241989) | Conclusion |
| --- | --- | --- | --- |
| | brain tissue affected by acute stroke. Areas of decreased perfusion appear as areas of changed signal intensity: • Lower signal intensity for CBF and CBV • Higher signal intensity for TTP, MTT, and Tmax | threshold volume, and Mismatch Ratio in the brain tissue affected by acute stroke. Areas of decreased perfusion appear as areas of changed signal intensity: • Lower signal intensity for CBF and CBV • Higher signal intensity for TTP, MTT, and Tmax | |
| PACS Functionality | | | |
| Basic PACS Functions | View, process and analyze medical images. Communication of results through service class user protocols. | Yes | Same |
| Computer Platform | Standard off-the-shelf PC workstation / server | Yes | Same |
| | Virtual platform such as VMware | Yes | Same |
| DICOM Compliance | Yes | Yes | Same |
| Functional Overview | CMN is software package that provides for the visualization and study of changes of tissue in digital images captured by CT, CBCT and MRI. CMN provides viewing and quantification. | CMN is software package that provides for the visualization and study of changes of tissue in digital images captured by CT and MRI. CMN provides viewing and quantification. | Same; Subject Device MR & CT capability identical to Predicate Device and CBCT capability similar to CT. No new questions of safety or effectiveness. |
| Data Acquisition and Data Types | Acquires medical image data from DICOM compliant imaging devices and modalities; including CT, CBCT and MRI | Acquires medical image data from DICOM compliant imaging devices and modalities; including CT and MRI | Similar; Subject Device MR & CT capability identical to Predicate Device and CBCT capability similar to CT. No new questions of safety or effectiveness. |
| Acquisition and Modalities Features | | | |
Cercare Medical A/S
April 2026
{12}
Cercare Medical
| Description | Subject Device | Predicate Device (K241989) | Conclusion |
| --- | --- | --- | --- |
| **MRI** | Diffusion Weighted Image (DWI) | Yes | Same as Predicate Device |
| | Dynamic Analysis tissue flow (perfusion) and tissue blood volume | Yes | |
| **CT** | Dynamic Analysis tissue flow (perfusion) and tissue blood volume | Yes | |
| **Computed Parameter Maps** | | | |
| **Diffusion MRI** | Isotropic DWI (isoDWI) | Yes | Same as Predicate Device |
| | ADC | Yes | |
| | Zero=gradient image (B=0 image) | Yes | |
| **Perfusion MRI** | Relative cerebral blood flow (rCBF) | Yes | Same as Predicate Device |
| | Relative cerebral blood volume (rCBV) | Yes | |
| | Mean transit time (MTT) | Yes | |
| | Delay | Yes | |
| | Capillary transit time heterogeneity (CTH) | Yes | |
| | Coefficient of variation (COV) | Yes | |
| | Model-based oxygen extraction fraction (OEF (model-based)) | Yes | |
| | Model-based relative cerebral metabolic rate of oxygen (rCMRO2 (model-based)) | Yes | |
| | Relative extravasation correction (rLeakage) | Yes | |
| | rCBF-basic | Yes | |
| | rCBV-basic | Yes | |
| | MTT-basic | Yes | |
| | Tmax-basic | Yes | |
| **Perfusion CT** | Relative cerebral blood flow (rCBF) | Yes | Same as Predicate Device |
Cercare Medical A/S
April 2026
{13}
Cercare Medical
| Description | Subject Device | Predicate Device (K241989) | Conclusion |
| --- | --- | --- | --- |
| | Relative cerebral blood volume (rCBV) | Yes | |
| | Mean transit time (MTT) | Yes | |
| | Delay | Yes | |
| | Capillary transit time heterogeneity (CTH) | Yes | |
| | Coefficient of variation (COV) | Yes | |
| | Model-based oxygen extraction fraction (OEF (model-based)) | Yes | |
| | Model-based relative cerebral metabolic rate of oxygen (rCMRO2 (model-based)) | Yes | |
| | Relative extravasation correction (rLeakage) | Yes | |
| | rCBF-basic | Yes | |
| | rCBV-basic | Yes | |
| | MTT-basic | Yes | |
| Perfusion CBCT | Relative cerebral blood flow (rCBF) | No | Similar to Predicate Device CT; Subject Device comparative performance testing demonstrates SE to Predicate Device |
| | Relative cerebral blood volume (rCBV) | No | |
| | Mean transit time (MTT) | No | |
| | Delay | No | |
| | Capillary transit time heterogeneity (CTH) | No | |
| | Coefficient of variation (COV) | No | |
| | Model-based oxygen extraction fraction (OEF (model-based)) | No | |
| | Model-based relative cerebral metabolic rate of oxygen (rCMRO2 (model-based)) | No | |
| | Relative extravasation correction (rLeakage) | No | |
| | rCBF-basic | No | |
| | rCBV-basic | No | |
Cercare Medical A/S
April 2026
{14}
Cercare Medical
| Description | Subject Device | Predicate Device (K241989) | Conclusion |
| --- | --- | --- | --- |
| | MTT-basic | No | |
| | Tmax-basic | No | |
| Measurement Tools | | | |
| MR/CT/CBCT Tools | Arterial input function (AIF) | Yes – MR and CT only | Same; Subject Device MR & CT capability identical to Predicate Device and CBCT capability similar to CT. Minor design differences do not raise new questions of safety or effectiveness. |
| | Time-course | Yes – MR and CT only | |
| | Brain mask | Yes – MR and CT only | |
| | Region of interest (ROI) and Volumetry | Yes – MR and CT only | |
| | Volumetric comparison between 2 ROIs | Yes – MR and CT only | |
| | Motion correction | Yes – MR and CT only | |
| | Export parameter maps PACS and DICOM file systems | Yes – MR and CT only | |
| | Acquire, transmit, process, and store medical images | Yes – MR and CT only | |
Cercare Medical A/S
April 2026
{15}
Cercare Medical
# Conclusions
The CMN Capillary Function with Virtual Expert (MRI, CT & CBCT) device performs as intended and presents no unacceptable risks to the intended patient population. Non-clinical bench testing demonstrated that 10 of 13 perfusion parameter-algorithm combinations meet the acceptance criterion when compared to the Predicate Device; the 3 timing-sensitive parameters showed reduced structural similarity attributable to the lower temporal sampling rate of CBCT, while maintaining acceptable spatial agreement. A blinded MRMC clinical study confirmed non-inferiority of the CBCT device relative to the Predicate Device. Together with software verification and validation, these results demonstrate that CMN Capillary Function with Virtual Expert (MRI, CT & CBCT) is substantially equivalent in intended use, technological characteristics, safety, and performance characteristics to the legally marketed Predicate Device (K241989).
Cercare Medical A/S
April 2026
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.