← Product Code [LZA](/productcode/LZA) · K253824

# Powder Free Nitrile Examination Gloves Non-Sterile Black (K253824)

_Kanglongda Vietnam Protection Technology Company Limited · LZA · Jun 18, 2026 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K253824

## Device Facts

- **Applicant:** Kanglongda Vietnam Protection Technology Company Limited
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jun 18, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Powder Free Nitrile Examination Gloves Non-Sterile Black is a disposable device intended for medical purposes that is worn on an examiner's hand to prevent the contamination between patient and examiner

## Device Story

Disposable nitrile examination glove; provides single-use barrier protection for wearer. Used in clinical settings by healthcare examiners. Device acts as physical barrier to prevent cross-contamination between patient and examiner. Manufactured from nitrile compound; black color. Complies with ASTM D6319-19 standards for physical properties, dimensions, and barrier integrity. Device is non-sterile and powder-free.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM D6319-19 (dimensions, tensile strength, elongation), ASTM D5151-19 (freedom from holes), and ASTM D6124-06 (residual powder). Biocompatibility testing performed per ISO 10993-23 (skin irritation), ISO 10993-10 (skin sensitization), and ISO 10993-11 (acute systemic toxicity). All tests passed.

## Technological Characteristics

Nitrile compound; powder-free; non-sterile. Dimensions and physical properties conform to ASTM D6319-19. Biocompatibility per ISO 10993-23, ISO 10993-10, and ISO 10993-11. Available in sizes XS through XXL.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Medical Examination Nitrile Gloves ([K232008](/device/K232008.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** **U.S. FOOD & DRUG**  
ADMINISTRATION

June 18, 2026

Kanglongda Vietnam Protection Technology Company Limited  
Thamira Ranasinghe  
Vice President - Technical and Sales  
Lot Cn 05, Viglacera - Phong Dien Industrial Park  
Phong Dinh Ward  
Hue City, 530000  
Vietnam

Re: K253824

Trade/Device Name: Powder Free Nitrile Examination Gloves Non-Sterile Black  
Regulation Number: 21 CFR 880.6250  
Regulation Name: Non-Powdered Patient Examination Glove  
Regulatory Class: Class I, reserved  
Product Code: LZA  
Dated: May 19, 2026  
Received: May 19, 2026

Dear Thamira Ranasinghe:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration  
10903 New Hampshire Avenue  
Silver Spring, MD 20993  
www.fda.gov

{1}

K253824 - Thamira Ranasinghe

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K253824 - Thamira Ranasinghe

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

![img-0.jpeg](img-0.jpeg)

Bifeng Qian, M.D., Ph.D.  
Assistant Director  
DHT4C: Division of Infection  
Control Devices  
OHT4: Office of Surgical and  
Infection Control Devices  
Office of Product Evaluation and Quality  
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

# Indications for Use

510(k) Number (if known)
K253824

Device Name
Powder Free Nitrile Examination Gloves Non-Sterile Black

Indications for Use (Describe)

Powder Free Nitrile Examination Gloves Non-Sterile Black is a disposable device intended for medical purposes that is worn on an examiner's hand to prevent the contamination between patient and examiner

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)

☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}

康隆达防护
KANG LONG DA PROTECTION

# 510(k) SUMMARY (K253824)

**0.0 Summary Preparation Date:** 2026-06-18

**1.0 Applicant Information**

- Name: Kanglongda Vietnam Protection Technology Company Limited
- Address: Lot CN 05, Viglacera - Phong Dien Industrial Park, Phong Dinh Ward, Hue City, Vietnam
- Establishment Registration Number: 3024897455

**2.0 CORRESPONDENT**

- Contact Name: Thamira Ranasinghe
- Contact Title: Vice President -Technical
- Phone Number: +84788611640
- Email: thamira@kanglongda.com.cn

**3.0 Name of the Device**

- Proprietary/Trade name: Powder Free Nitrile Examination Gloves Non-Sterile Black
- Common Name: Nitrile Examination Gloves
- Classification Name: Non-powdered Patient Examination Glove
- Device Classification: Class I
- Regulation Number: 21 CFR 880.6250
- Product Code: LZA
- 510(K) Number: K253824

**4.0 Predicate**

- Device Name: Disposable Medical Examination Nitrile Gloves
- Company name: Raxwell Industries LLC
- Device Classification: Class I
- Regulation Number: 21 CFR 880.6250
- Product Code: LZA
- 510(K) Number: K232008

**5.0 Device Description:**

"Powder Free Nitrile Examination Gloves Non-Sterile Black" is a patient examination glove made from nitrile compound, nonsterile (as per 21 CFR 880.6250, Class I). The principal operation of this medical device is to provide single use barrier protection for the wearer and the device meets the specifications for barrier protection and physical

1

{5}

**康隆达防护**^{}[] KANG LONG DA PROTECTION

properties as defined in ASTM D6319-19, Standard specification for Nitrile Examination Gloves.

### 6.0 Device Indications for Use:

“Powder Free Nitrile Examination Gloves Non-Sterile Black” is a nonsterile disposable device intended for medical purpose that is worn on the examiner’s hand to prevent contamination between patient and examiner.

### 7.0 Comparison of Device Technological Characteristics:

|  Device Characteristic | Predicate Device | Subject Device | Comparison  |
| --- | --- | --- | --- |
|  Product Name | Disposable Medical Examination Nitrile Gloves | Powder Free Nitrile Examination Gloves Non-Sterile Black | similar  |
|  510(K) No. | K232008 | K253824 | n/a  |
|  Applicant | Raxwell Industries LLC | Kanglongda Vietnam Protection Technology Company Limited | Different  |
|  Product Code | LZA | LZA | Same  |
|  Regulation | 21 CFR 880.6250 | 21 CFR 880.6250 | Same  |
|  Class | 1 | 1 | Same  |

2

{6}

**康隆达防护**^{}[] KANG LONG DA PROTECTION

|  Indications for Use | The Disposable Medical Examination Nitrile Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | 'Powder Free Nitrile Examination Gloves Non-Sterile Black' is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same  |
| --- | --- | --- | --- |
|  Powder free | Yes | Yes | Same  |
|  Material | Nitrile | Nitrile | Same  |
|  Colorant | Black | Black | Same  |
|  Size | XS/S/M/L/XL/XXL | XS/S/M/L/XL/XXL | Same  |
|  Single Use | Yes | Yes | Same  |
|  Non-Sterile | Yes | Yes | Same  |
|  Dimensions - Length | Complies with ASTM D6319-19 XS/S: ≥220; M/L/XL/XXL: ≥230; | Complies with ASTM D6319-19 XS/S: ≥220; M/L/XL/XXL: ≥230; | Same  |

3

{7}

**康隆达防护**^{}[] KANG LONG DA PROTECTION

|  Dimensions - Palm Width | Complies with ASTM D6319-19 XS: 70 ± 10 mm; S: 80 ± 10 mm; M: 95 ± 10 mm; L: 110 ± 10 mm; XL: 120 ± 10 mm; XXL:130± 10 mm; | Complies with ASTM D6319-19 XS: 70 ± 10 mm; S: 80 ± 10 mm; M: 95 ± 10 mm; L: 110 ± 10 mm; XL: 120 ± 10 mm; XXL:130± 10 mm; | Same  |
| --- | --- | --- | --- |
|  Dimensions - Thickness | Complies with ASTM D6319-19 Palm ≥0.05 mm; Finger ≥0.05 mm | Complies with ASTM D6319-19 Palm ≥0.05 mm; Finger ≥0.05 mm | Same  |
|  Physical Properties - Tensile Strength | Complies with ASTM D6319-19 Before aging ≥14MPa After aging ≥14MPa | Complies with ASTM D6319-19 Before aging ≥14MPa After aging ≥14MPa | Same  |
|  Physical Properties - Elongation | Complies with ASTM D6319-19 Elongation: Before Aging: ≥ 500% After Aging: ≥400% | Complies with ASTM D6319-19 Elongation: Before Aging: ≥ 500% After Aging: ≥400% | Same  |
|  Residual powder | Complies with ASTM D6319-19 < 2.0mg per glove | Complies with ASTM D6319-19 < 2.0mg per glove | Same  |
|  Freedom from Holes | In accordance with ASTM D6319-19 (G-1 with AQL 2.5) | In accordance with ASTM D6319-19 (G-1 with AQL 2.5) | Same  |

4

{8}

**康隆达防护**^{}[] KANG LONG DA PROTECTION

|  Biocompatibility | ISO 10993-10, under the conditions of the study, the device is not a sensitizer | ISO 10993-10, under the conditions of the study, the device is not a sensitizer | Same  |
| --- | --- | --- | --- |
|   |  ISO 10993-10, under the conditions of the study, the device is not an irritant | ISO 10993-23, under the conditions of the study, the device is not an irritant | Similar*  |
|   |  ISO 10993-5, under conditions of the study, device extract is cytotoxic | Not test, addressed by alternative testing | Different – Evaluation results demonstrate conformity with the requirements for acute systemic toxicity.  |
|   |  ISO 10993-11, under the conditions of the study, the device showed no evidence of acute systemic toxicity | ISO 10993-11, under the conditions of the study, the device showed no evidence of acute systemic toxicity | Same  |

*Note 1: The skin irritation test was conducted in accordance with the latest applicable standard, ISO 10993-23, which supersedes and refines the methodology previously found in ISO 10993-10. The test is equivalent in its objective and acceptance criteria, and the passing result confirms the device meets the required safety profile for skin irritation.

5

{9}

**康隆达防护**^{}[] KANG LONG DA PROTECTION

## 8.0 Assessment of Non-Clinical Performance Data:

Non-clinical testing was performed to verify that the subject devices meet the acceptance criteria of the performance test and all design specifications. The test results demonstrated that the subject device complies with the following standards as shown

|  Test | Purpose | Criteria | Result  |
| --- | --- | --- | --- |
|  Standard Test Method for Detection of Holes in Medical Gloves ASTM D5151-19 (2023) | To demonstrate glove integrity. | Freedom from holes. AQL 2.5. | Pass  |
|  Standard Test Method for Residual Powder on Medical Gloves ASTM D6124-06 (2022) | To demonstrate the gloves are 'powder free'. | Average less than 2 mg per glove. | Pass  |
|  Dimensional Conformance ASTM D6319-19(2023) | To demonstrate appropriate dimensions for labeled sizes. | Conforms to ASTM D6319 width, thickness, and length requirements for sizes XS, S, M, L, XL and XXL AQL 4 | Pass  |
|  Tensile Performance ASTM D6319-19(2023) | To demonstrate adequate tensile and elongation properties. | Conforms to ASTM D6319 requirements: - Tensile Strength (Before & After Aging): ≥ 14 MPa - Ultimate Elongation (Before Aging): ≥ 500% - Ultimate Elongation (After Aging): ≥ 400% AQL 4 | Pass  |

6

{10}

**康隆达防护**^{}[] KANG LONG DA PROTECTION

|  Biocompatibility: Skin Irritation ISO 10993-23 | To demonstrate low potential for skin irritation. | Under the conditions of the study, the device is not an irritant. | Pass  |
| --- | --- | --- | --- |
|  Biocompatibility: Skin Sensitization ISO 10993-10 | To demonstrate low potential for skin sensitization. | Under the conditions of the study, the device is not a sensitizer. | Pass  |
|  Biocompatibility: Acute Systemic Toxicity ISO 10993-11 | To demonstrate low acute systemic toxicity. | Under conditions of the study, the device showed no evidence of acute systemic toxicity | Pass  |

### **9.0 Assessment of Clinical Performance Data:**

Clinical data is not needed for this type of device.

### **10.0 Conclusion:**

The conclusions drawn from the nonclinical tests demonstrate that the subject device “Powder Free Nitrile Examination Gloves Non-Sterile Black” is as safe, as effective, and performs as well as or better than the legally marketed device, K232008.

7

---

**Source:** [https://fda-staging.innolitics.com/device/K253824](https://fda-staging.innolitics.com/device/K253824)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).