K253814 · Beijing Honghugaoxiang Science and Technology Development CO · HBO · Jul 21, 2026 · Neurology
Device Facts
Record ID
K253814
Device Name
Raney Clips Kit
Applicant
Beijing Honghugaoxiang Science and Technology Development CO
Product Code
HBO · Neurology
Decision Date
Jul 21, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Raney Clips Kit is indicated for use in temporary hemostasis of the scalp edge.
Device Story
Raney Clips Kit provides temporary hemostasis during scalp surgery. System comprises pistol-grip clip applier and pre-loaded cartridges containing 12, 13, or 16 scalp clips. Operator triggers applier to fire clips into scalp tissue; clips exert squeezing force on tissue surrounding bleeding vessels to restrict blood flow. Device is single-use and provided sterile. Used in clinical settings by surgeons. Benefits include rapid, effective control of scalp bleeding during procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included appearance, dimension, clamping force, tooth spacing, simulated use, package integrity (ASTM F88, F1929, F1886), simulated transportation (ASTM D4169), EO/ECH residuals (ISO 10993-7), bacterial endotoxins (USP <85>), shelf life, and biocompatibility (ISO 10993-5, 10, 11, 23).
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
July 21, 2026
Beijing Honghugaoxiang Science and Technology Development CO
Zichen Jia
RA Specialist
5 Bldg. W. Side, #3, Tiangui St., Zhongguancun Science
Park Daxing Biomedical Industry Base, Daxing District
Beijing, Beijing 102600
China
Re: K253814
Trade/Device Name: Raney Clips Kit
Regulation Number: 21 CFR 882.4150
Regulation Name: Scalp Clip
Regulatory Class: Class II
Product Code: HBO
Dated: July 1, 2026
Received: July 2, 2026
Dear Zichen Jia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253814 - Zichen Jia
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253814 - Zichen Jia
Page 3
Sincerely,
JULIA E.
Digitally signed by
JULIA E. SLOCOMB -S
SLOCOMB -S
Date: 2026.07.21
15:55:15 -04'00'
for Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253814 | ? |
| Please provide the device trade name(s). | | ? |
| Raney Clips Kit | | |
| Please provide your Indications for Use below. | | ? |
| The Raney Clips Kit is indicated for use in temporary hemostasis of the scalp edge. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253814 Page 1 of 4
### 510(k) Summary
1. Date of Preparation: 07/20/2026
2. Sponsor Identification
Beijing Honghugaoxiang Science and Technology Development Co., Ltd.
5 Building On The West Side, No. 3, Tiangui Street, Zhongguancun Science Park
Daxing Biomedical Industry Base, Daxing District, 102600 Beijing, P.R. China
Contact Person: Zichen Jia
Position: RA Specialist
Tel: +86 (0)10 6125 2128
Email: hhgxzc@hhgx.cn
3. Designated Submission Correspondent
Beijing Honghugaoxiang Science and Technology Development Co., Ltd.
5 Building On The West Side, No. 3, Tiangui Street, Zhongguancun Science Park
Daxing Biomedical Industry Base, Daxing District, 102600 Beijing, P.R. China
Contact Person: Zichen Jia
Position: RA Specialist
Tel: +86 (0)10 6125 2128
Email: hhgxzc@hhgx.cn
4. Identification of Subject Device
Trade Name: Raney Clips Kit
Common Name: Scalp Clip and Applier
Classification Name: Clip, Scalp; Classification: II
Product Code: HBO
Regulation Number: 21 CFR 882.4150
Review Panel: Neurology
5. Identification of Predicate Device
510(k) Number: K012219
Trade Name: ScalpFix Clip System
6. Indications for Use
The Raney Clips Kit is indicated for use in temporary hemostasis of the scalp edge.
7. Device Description
The subject device, Raney Clips Kit, is indicated for use in temporary hemostasis of the scalp edge. The device is provided sterile and single use.
The subject device consists of a clip applier and cartridge(s) that is pre-loaded with scalp clips. With the pull of a trigger, the scalp clips are fired into the scalp tissue and create a squeezing force on the scalp tissue directly surrounding the bleeding vessels, causing them to constrict or close. As a result, blood flow through these vessels is greatly restricted, leading
{5}
K253814 Page 2 of 4
to temporary hemostasis (stopping bleeding).
The shelf life of the subject device is three years. The subject device is sterilized by Ethylene Oxide to achieve a SAL of 10-6 and supplied in sterility maintenance package which could maintain the sterility of the device during the shelf life of 3 years.
Raney Clips Kit is available in two series with six models, HZ Sa and HZ Sb. They differ only in the number of clips and accompanying cartridges. Cartridges are available in three numbers of clips: 12,13, and 16. The clip applier, scalp clip and cartridge in each series have the same material and manufacturing process and are different only in number of cartridges. All models of the Raney Clips Kits share identical material composition, device design, sterile barrier and manufacturing process.
# 8. Comparison of Technological Characteristics
Table 1 General Comparison
| ITEM | Subject device | Predicate Device, K012219 | Remark |
| --- | --- | --- | --- |
| Device Name | Raney Clips Kit | SCALPFIX CLIP SYSTEM | / |
| Product Code | HBO | HBO | Same |
| Regulation No. | 882.4150 | 882.4150 | Same |
| Class | Class II | Class II | Same |
| Indication for Use | The Raney Clips Kit is indicated for use in temporary hemostasis of the scalp edge. | Aesculap's Scalpfix is indicated for use in temporary hemostasis of the scalp edge. | Same |
| Principle of Operation | The clip applier has a pistol-like design and holds an exchangeable, pre-loaded cartridge of scalp clips. With the pull of a trigger, the scalp clips are inserted into the scalp tissue, and grip the margin of the scalp wound tightly to provide temporary hemostasis. | Aesculap's scalp clip gun (FF012R) is used in conjunction with Aesculap's scalp clip (FF013P) to insert sterile plastic scalp clips at the margin of scalp wounds, providing temporary hemostasis of the opened scalp. | Similar |
| Configuration | Clip Applier Cartridge | Clip Applier Cartridge | Same |
| Single-use | Single-use only | Scalp Clip - Singe-use only Clip Applier - Reusable | Different |
| Clip Material | Polyoxymethylene (POM) | Polyoxymethylene (POM) | Same |
| Clip Pattern | Serrated jaw | Straight jaw | Different |
| Number of clips in each cartridge | 12, 13, 16 | 10 | Different |
| Sterile | Supplied STERILE | Scalp Clip - Supplied STERILE Clip Applier - Non-sterile | Different |
| Shelf life | 3 Years | Scalp Clip - 5 Years Clip Applier - N/A | Different |
{6}
K253814 Page 3 of 4
# 9. Summary of Nonclinical Testing
Nonclinical tests were conducted to verify that the subject device met all design specifications and to support the Substantially Equivalent (SE) determination.
# Bench testing
The bench test performed on the subject device and predicate device includes appearance, dimension, clamping force, tooth spacing and simulated use tests. The test results demonstrated that the subject device and predicate devices have similar product performance.
# Package integrity
The package integrity of the subject sterility maintenance package includes visual inspection, seal strength and dye penetration resistance. The test results demonstrated that the subject device complies with the following standards:
- ASTM F88/F88M-23, Standard Test Method for Seal Strength of Flexible Barrier Materials.
- ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration.
- ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
# Simulated transportation
A simulated transportation test was conducted on the subject device following the standard below. The test results demonstrate that packing can protect the device from damage during distribution environments.
- ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems
# EO and ECH Residue
Examination of the Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) residuals have been conducted in accordance with ISO 10993-7:2008. The test results demonstrated that the subject device complies with the allowable limits.
# Bacterial Endotoxins
Medical device endotoxin contamination is associated with a risk of pyrogenic reactions in patients. Endotoxin limit is established as 20 USP Endotoxin Unit (EU) for the subject device. Test has been conducted by Limulus Amebocyte Lysate (LAL) method according to USP <85> Bacterial Endotoxin Limit. The test results demonstrated that the subject device complies with the allowable limits.
# Shelf life
A maximum shelf life of 3 years has been assigned to the subject device. In order to ensure the safety and effectiveness of subject device in the shelf life, following testing was conducted to provide valid data for the device shelf life:
- Package integrity after simulated distribution
- Package integrity of the accelerated aging device and unaged device, Performance testing of accelerated aging device
# Biocompatibility
The subject device has limited contact duration and tissue contact level, Therefore, it was tested for Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Intracutaneous Reactivity (ISO 10993-23), Acute Systemic Toxicity (ISO 10993-11) and Material-mediated Pyrogenicity (ISO 10993-11 and USP <151>).
{7}
K253814 Page 4 of 4
# 10. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject devices are as safe and effective when compared to the legally marketed predicate device K012219. The Indications for Use, technological characteristics, and performance of the proposed Raney Clips Kit are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.