Electronic Medical Records (EMR); Clinical notes; Imaging results; Pathology results; Structured diagnosis codes; Transthoracic Echocardiogram (TTE) reports; 12-lead ECG databases
Retrospective clinical data from four health systems was used to validate the diagnostic performance (sensitivity and specificity) of the AI algorithm in a real-world patient population. Retrospective data was also used to train the neural network by identifying CA cases and propensity-matched controls from clinical health system databases.
Retrospective Clinical Validation Study; Retrospective cohort study; Follow-up/Duration: ECG and TTE within +/- 180 days
Adults aged 18 years and older with both an ECG and TTE obtained within +/- 180 days of each other, presenting with signs or symptoms consistent with cardiac amyloidosis.; Sample Size: 25,525 patient records; Number of Sites: 4
Not applicable for this study
Sensitivity and specificity for the detection of cardiac amyloidosis
Algorithm Training Cohort; Retrospective case-control study; Follow-up/Duration: ECG within +/- 180 days of TTE for controls
4,251 CA patients and 45,730 propensity-matched control patients; Sample Size: 49,981 patients; Number of Sites: Large health system with CA expertise
Propensity-matched non-CA patients
Distinguish between CA and non-CA patients
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cardiac Amyloidosis
Neural Network
sensitivity >= 75% and specificity >= 80%
sensitivity 78.9%, specificity 91.2%
Training: 4,251 CA patients and 45,730 control patients.
>1 (physicians)
Retrospective study of 25,525 patient records from four health systems in the United States.
—
Indications for Use
The ECG-AI Cardiac Amyloidosis 12-Lead Algorithm is software as a medical device intended to aid in earlier detection of Cardiac Amyloidosis in adults being evaluated for signs or symptoms consistent with cardiac amyloidosis. This population includes patients that have at least one condition commonly associated with cardiac amyloidosis, such as: unexplained heart failure, nephrotic syndrome, peripheral neuropathy, arrhythmias, or aortic stenosis. The ECG-AI Cardiac Amyloidosis 12-Lead Algorithm is not intended to be a stand-alone diagnostic device for cardiac amyloidosis nor to replace current clinical practice guidelines. A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Cardiac Amyloidosis. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out additional clinical evaluation. The ECG-AI Cardiac Amyloidosis 12-Lead algorithm should be applied jointly with clinician judgment.
Device Story
Software as a medical device (SaMD) analyzing 10+ seconds of 12-lead ECG voltage data; uses AI-based neural network to detect cardiac amyloidosis; integrated into clinical workflows via EMR or ECG management systems; accessed by clinicians via PC or mobile; outputs binary result (Detected/Not Detected/Error) to support clinical decision-making; aids in identifying patients requiring further workup or specialist referral; benefits patients through earlier detection of cardiac amyloidosis.
Clinical Evidence
Retrospective study of 25,525 patient records across four US health systems. Primary endpoint: sensitivity and specificity for cardiac amyloidosis detection. Results: sensitivity 78.9%, specificity 91.2%. Met prespecified acceptance criteria (sensitivity ≥75%, specificity ≥80%). Subgroup analyses performed across demographics, comorbidities, and ECG devices.
Technological Characteristics
Software module in Docker container; no GUI; integrates with EMR/EMS via API. Input: 12-lead ECG digital waveform. Processing: Neural Network (AI). Output: Binary classification. Compliant with IEC 62304 and FDA cybersecurity guidance.
Indications for Use
Indicated for adults (18+) being evaluated for signs/symptoms of cardiac amyloidosis, including those with unexplained heart failure, nephrotic syndrome, peripheral neuropathy, arrhythmias, or aortic stenosis. Not for stand-alone diagnosis; not to replace clinical guidelines.
Regulatory Classification
Identification
Viz HCM is a cardiovascular machine learning-based notification software intended to be used in parallel to the standard of care to analyze 12-lead ECG recordings from patients 18 years of age or older. It detects signs associated with hypertrophic cardiomyopathy (HCM) and allows the user to view the ECG and analysis results. It is not intended for use on patients with implanted pacemakers, does not replace standard diagnostic methods, and is not intended to rule out HCM or be used in lieu of a full patient evaluation.
Special Controls
In combination with the general controls of the FD&C Act, cardiovascular machine learningbased notification software is subject to the following special controls:
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 7, 2026
Anumana, Inc.
Taylor Gold West
Director, Regulatory Affairs
One Main St., Suite 400
East Arcade, 4th Floor
Cambridge, Massachusetts 02142
Re: K253801
Trade/Device Name: ECG-AI Cardiac Amyloidosis (CA) 12-Lead Algorithm
Regulation Number: 21 CFR 870.2380
Regulation Name: Cardiovascular Machine Learning-Based Notification Software.
Regulatory Class: Class II
Product Code: SHP
Dated: November 28, 2025
Received: November 28, 2025
Dear Taylor Gold West:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253801 - Taylor Gold West
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
{2}
K253801 - Taylor Gold West
Page 3
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**STEPHEN C. BROWNING -S**
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253801 | |
| Device Name ECG-AI Cardiac Amyloidosis (CA) 12-Lead Algorithm | |
| Indications for Use (Describe) The ECG-AI Cardiac Amyloidosis 12-Lead Algorithm is software as a medical device intended to aid in earlier detection of Cardiac Amyloidosis in adults being evaluated for signs or symptoms consistent with cardiac amyloidosis. This population includes patients that have at least one condition commonly associated with cardiac amyloidosis, such as: unexplained heart failure, nephrotic syndrome, peripheral neuropathy, arrhythmias, or aortic stenosis. The ECG-AI Cardiac Amyloidosis 12-Lead Algorithm is not intended to be a stand-alone diagnostic device for cardiac amyloidosis nor to replace current clinical practice guidelines. A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Cardiac Amyloidosis. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out additional clinical evaluation. The ECG-AI Cardiac Amyloidosis 12-Lead algorithm should be applied jointly with clinician judgment. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
eSTAR Traditional
510(k)
Submission – ECG-AI CA 12-Lead Algorithm
K253801 - 510(k) Summary
# 510(k) Summary
ECG-AI Cardiac Amyloidosis 12-Lead Algorithm
| Applicant Name: | Anumana, Inc.
One Main Street, Suite 400
East Arcade 4th Floor
Cambridge, MA 02142 |
| --- | --- |
| Submitter: | Vyaas Shankar
Senior Regulatory Specialist
Anumana, Inc.
vyaas.shankar@anumana.ai |
| Primary Contact: | Taylor Gold West, MBA, RAC
Director, Regulatory Affairs
Anumana, Inc.
tgoldwest@anumana.ai |
| Date Summary Prepared: | April 7, 2026 |
| Trade/Device Name: | ECG-AI Cardiac Amyloidosis (CA) 12-Lead Algorithm |
| Classification Panel: | Cardiology |
| Regulation Number: | 21 CFR § 870.2380 |
| Regulation Name: | Cardiovascular machine learning-based notification software |
| Regulatory Class: | Class II |
| Product Code: | SHP |
| Predicate Device(s): | |
| Device Name: | ECG-AI Low Ejection Fraction 12-Lead algorithm |
| Manufacturer: | Anumana, Inc. |
| Application Number: | K250652 |
# Device Description
The ECG-AI CA 12-Lead algorithm is software as a medical device (SaMD) which interprets 12-lead ECG voltage time series data using an artificial intelligence-based algorithm. The device analyzes 10 seconds or longer duration of a single 12-lead ECG acquisition, and within seconds detects the presence of cardiac amyloidosis in the intended patient population.
The ECG-AI CA 12-Lead algorithm is integrated into existing clinical workflows to support overall assessment and to enable physicians to determine whether further workup or referral to a cardiac amyloidosis specialist is clinically indicated. The ECG-AI CA 12-Lead algorithm is intended to be used by clinicians that diagnose and/or treat patients with signs, symptoms, or co-morbidities, associated with cardiac amyloidosis.
{5}
eSTAR Traditional 510(k) Submission – ECG-AI CA 12-Lead Algorithm
K253801 - 510(k) Summary
The software module can be integrated into a client application to be accessed by clinicians and results viewed through an Electronic Medical Record (EMR) system or an ECG Management System (EMS) accessed via a PC, mobile device, or another medical device. In each case, the physician imports 12-lead ECG data in digital format. The tool analyzes the 10 seconds or longer duration of voltage data collected during a standard 12-lead ECG and outputs a binary result of the likelihood of cardiac amyloidosis as an Application Programming Interface (API) result.
## Indications for Use
The ECG-AI Cardiac Amyloidosis 12-Lead algorithm is software as a medical device intended to aid in earlier detection of Cardiac Amyloidosis in adults being evaluated for signs or symptoms consistent with cardiac amyloidosis. This population includes patients that have at least one condition commonly associated with cardiac amyloidosis, such as: unexplained heart failure, nephrotic syndrome, peripheral neuropathy, arrhythmias, or aortic stenosis.
The ECG-AI Cardiac Amyloidosis 12-Lead algorithm is not intended to be a stand-alone diagnostic device for cardiac amyloidosis nor to replace current clinical practice guidelines. A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Cardiac Amyloidosis. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out additional clinical evaluation. The ECG-AI Cardiac Amyloidosis 12-Lead algorithm should be applied jointly with clinician judgment.
## Technological Characteristics
The ECG-AI CA 12-Lead algorithm is provided as a software module packaged in a Docker container to facilitate installation. Technical installation details including access to docker hub, docker hub path, software upgrades, and associated access rights can be found in the Installation Manual - ECG-AI Cardiac Amyloidosis 12-Lead algorithm. The ECG-AI CA 12-Lead algorithm does not provide a graphical user interface (GUI) of its own. It is integrated with other medical systems such as Electronic Medical Record (EMR) systems or ECG Management Systems (EMS). The third-party integrating software will furnish a 12-lead ECG digital waveform as input to ECG-AI CA 12-Lead algorithm and record the output from the device for display or for printing in an offline report.
## Predetermined Change Control Plan
This device has been cleared by the U.S. Food and Drug Administration (FDA) with a Predetermined Change Control Plan (PCCP). As part of this authorization, the device software may be updated periodically to enhance performance, including for higher sensitivity and/or specificity.
The first modification, related to device performance improvement, will be validated through a multi-center retrospective clinical study that uses a combination of new and existing data. This approach involves supplementing the validation dataset with at least 20% new data for each iteration to ensure it remains current and representative. To be implemented, a modified version must demonstrate improved performance by meeting pre-specified acceptance criteria. These criteria require the new version's sensitivity and specificity point estimates and lower bound of the 95% confidence interval to be greater than those of the previous version.
The second modification, which pertains to the expansion of ECG device compatibility, will be validated through a retrospective clinical study utilizing newly acquired clinical data from ECG devices targeted for inclusion under the expanded compatibility claim. To be implemented, the modified version must
{6}
eSTAR Traditional 510(k) Submission – ECG-AI CA 12-Lead Algorithm
K253801 - 510(k) Summary
demonstrate that performance on ECGs obtained from the target cardiographs exceeds the minimum performance observed on accepted cardiographs from the original clinical validation study, plus any performance gain resulting from the implementation of the first modification. These criteria require the new version's sensitivity point estimate and lower bound of the 95% confidence interval to be greater than the minimum threshold, plus any performance gain resulting from implementation of the first modification, and specificity lower bound of the 95% confidence interval to be greater than the minimum threshold, plus any performance gain resulting from implementation of the first modification.
Users will be informed of each applicable update through revised labeling, release notes, or other appropriate communication channels. It is important to review updated instructions and performance summaries accompanying each software version to ensure continued proper use of the device.
## Summary of Non-Clinical Performance Data
The performance characteristics for the ECG-AI CA 12-Lead Algorithm have been evaluated with the following non-clinical testing: software verification and validation (per IEC 62304), cybersecurity, labeling verification and validation, and human factors.
## Summary of Clinical Performance Data
The performance characteristics for the ECG-AI CA 12-Lead Algorithm have been clinically validated for detection of cardiac amyloidosis in the intended patient population, with a sensitivity and specificity greater than the study's predetermined acceptance criteria.
Table 1: Performance Characteristics
| Performance Characteristic | Value |
| --- | --- |
| Sensitivity | 78.9 % |
| Specificity | 91.2 % |
## Algorithm Training
The ECG-AI CA 12 Lead Algorithm was developed by training a Neural Network (NN), a type of artificial intelligence, to learn patterns specific to CA directly from standard 12-lead ECGs. The algorithm was trained by providing the NN with ECGs from two distinct groups of patients, CA patients (cases) and matched non-CA patients (controls).
To identify cases, a database from a large health system with CA expertise was used. The database was rigorously examined to identify patients with evidence of CA reported in natural language in clinical notes, in imaging results, in pathology results, or in structured diagnoses codes. Physicians reviewed all possible CA cases and identified a final case cohort of 4,251 patients with a confirmed diagnosis of CA and supporting evidence of the diagnosis in their clinical records. The 4,251 patients contributed a total of 4,251 ECGs (one per patient) for model training.
Controls were selected by identifying patients with a complete Transthoracic Echocardiogram (TTE) and an ECG within ± 180 days of the TTE. Patients with any evidence of cardiac amyloidosis or amyloidosis in other organs were then excluded. Control candidates were then propensity matched to cases based on
{7}
eSTAR Traditional 510(k) Submission – ECG-AI CA 12-Lead Algorithm
K253801 - 510(k) Summary
age at the time of ECG acquisition, biological sex, race, and the year of the ECG acquisition. A final group of 45,730 control patients were identified, which contributed a total of 45,730 ECGs (one per patient) for model training. The ECGs from these 4,251 CA patients and 45,730 control patients were then used to train the NN to distinguish between the two groups.
# Summary of Clinical Validation
# Study Design
The performance of the Anumana ECG-AI CA 12-Lead algorithm was validated in a retrospective study of 25,525 patient records from four health systems in the United States. The study population included adults aged 18 years and older with both an ECG and TTE obtained within $+/-$ 180 days of each other. The study was designed to evaluate the device's ability to identify cardiac amyloidosis using AI-based analysis of a standard 12-lead ECG.
CA cases were defined as patients with a confirmed diagnosis of cardiac amyloidosis. For these individuals, the ECG-TTE pair closest to the date of diagnosis was selected. Controls were selected from patients with medical conditions associated with CA but without a CA diagnosis. Records with a diagnosis of non-cardiac forms of amyloid were excluded.
# Summary of Study Results
In total, 25,525 patient records were used for the primary analysis. The study sample was representative of the US adult population and included $52.0\%$ male and $48.0\%$ female participants. The racial and ethnic distribution was $72.7\%$ White, $7.5\%$ Hispanic, $11.3\%$ Black/African American, $0.7\%$ American Indian or Alaska Native, $1.9\%$ Asian, and $13.4\%$ Other/Multiracial. Approximately $63.0\%$ of participants were aged 60 years or older.
The study was designed with a joint primary performance endpoint assessing sensitivity and specificity, with prespecified acceptance criteria of sensitivity $\geq 75\%$ and specificity $\geq 80\%$ . The device met these acceptance criteria.
# Subgroup Analyses
Subgroup assessments of diagnostic performance were conducted to determine if there was heterogeneity in device performance across clinical sites, demographics, clinical characteristics, conduction disorders, ECG manufacturers, and ECG devices. To assess for heterogeneity, Breslow-Day tests for the strata were conducted on the odds ratio for condition presence given a positive ECG-AI result. The results are summarized in Table 2 below.
Table 2: Subgroup Analyses
| Subgroup Analysis | Result of Test for Heterogeneity |
| --- | --- |
| Clinical Site | Diagnostic performance varied across sites (p<0.01), with the lowest diagnostic performance observed for UTSW, which had a high proportion of hereditary ATTR, and the highest observed for VUMC. |
{8}
eSTAR Traditional 510(k) Submission – ECG-AI CA 12-Lead Algorithm
K253801 - 510(k) Summary
| Subgroup Analysis | Result of Test for Heterogeneity |
| --- | --- |
| Cardiac Amyloidosis Subtypes | Diagnostic performance differed by cardiac amyloidosis subtype (p<0.01), with sensitivity highest for AL amyloidosis (87.6%) and lowest for unknown/unclear subtype (33.3%; small n), and intermediate for hereditary ATTR (69.4%), wild-type ATTR (73.2%), and ATTR not otherwise specified (84.6%). |
| Biological Sex | No statistically significant heterogeneity observed (p>0.01) |
| Race/Ethnicity | Diagnostic performance varied across racial groups (p<0.01), with the lowest diagnostic performance observed for Black participants (sensitivity 76%, specificity 88.0%) and the highest observed for White participants (sensitivity 80.6%, specificity 91.6%). |
| Age Group at Index Visit | Diagnostic performance varied across age groups (p<0.01), with the lowest diagnostic performance observed for the 70+ group (sensitivity 76%, specificity 91.3%) and the highest observed for the 50–59 group (sensitivity 85.6%, specificity 88.0%). |
| Body Mass Index | Diagnostic performance varied across BMI categories (p<0.01), with the lowest diagnostic performance observed for the 35+ group (sensitivity 66.2%, specificity 92.5%) and the highest observed for the 30–34 group (sensitivity 84.4%, specificity 92.4%). |
| ECG Manufacturer | Diagnostic performance varied across ECG manufacturer (p<0.01), with the lowest diagnostic performance observed for GE Healthcare (sensitivity 78.1%, specificity 80.3%) and the highest observed for Philips Medical Products (sensitivity 80.6%, specificity 92.9%), though this may be influenced by study site. |
| ECG Device Model | No statistically significant heterogeneity observed (p>0.01). |
| Comorbidities | Diagnostic performance was generally consistent across comorbidity subgroups. There was no statistically significant heterogeneity for aortic stenosis, glomerular disorders, or non-specific symptoms/signs (p>0.01), with similar performance in patients with and without these conditions. In contrast, diagnostic performance varied by cardiac arrhythmias, heart failure, and neuropathies, with lower performance observed in patients with these comorbidities and higher performance in those without; statistically significant heterogeneity was observed for each of these subgroups (p≤0.01). |
| Conduction Disorders | Diagnostic performance was broadly similar when stratified by conduction disorders identified from ECG XML diagnoses, and no statistically significant heterogeneity was observed between patients with and without a specified conduction disorder on ECG (p>0.01). In contrast, when conduction disorders were defined using ICD codes, diagnostic performance was lower in patients with an ICD code for a conduction disorder and higher in those without, and statistically significant heterogeneity was observed across these groups (p<0.01). |
Page 5 of 8
{9}
eSTAR Traditional 510(k) Submission – ECG-AI CA 12-Lead Algorithm
K253801 - 510(k) Summary
# Substantial Equivalence
The following table captures the substantial equivalence comparison between the subject and predicate device, including administrative information, patient population, use environment, and key technological features.
Table 5: Substantial Equivalence Comparison of Subject Device to the Predicate
| | Subject Device | Predicate | Discussion |
| --- | --- | --- | --- |
| Manufacturer | Anumana, Inc. | Anumana, Inc. | |
| Product Name | ECG-AI CA 12-Lead algorithm | ECG-AI LEF 12-Lead algorithm | |
| Application No. | K253801 | K250652 | |
| Product Code | SHP | QYE | Equivalent. Both product codes are assigned to the Regulation number 21 CFR 870.2380. |
| Regulation No. | 21 CFR 870.2380 | 21 CFR 870.2380 | Identical. |
| Regulation Name | Cardiovascular machine learning-based notification software | Cardiovascular machine learning-based notification software | Identical. |
| Intended Use/Indications for Use | The ECG-AI Cardiac Amyloidosis 12-Lead Algorithm is software as a medical device intended to aid in earlier detection of Cardiac Amyloidosis in adults being evaluated for signs or symptoms consistent with cardiac amyloidosis. This population includes patients that have at least one condition commonly associated with cardiac amyloidosis, such as: unexplained heart failure, nephrotic syndrome, peripheral neuropathy, arrhythmias, or aortic stenosis. | The ECG-AI LEF 12-Lead algorithm is software intended to aid in earlier detection of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for heart failure. This population includes, but is not limited to: • patients with cardiomyopathies • patients who are post-myocardial infarction • patients with aortic stenosis • patients with chronic atrial fibrillation • patients receiving pharmaceutical therapies that are cardiotoxic, and | Equivalent. The subject device's intended use is equivalent to the predicate in that both are intended to aid in earlier detection of a cardiovascular disease or condition. The only difference between the indications for use is the disease condition, which was clinically validated. |
{10}
eSTAR Traditional 510(k) Submission – ECG-AI CA 12-Lead Algorithm
K253801 - 510(k) Summary
| | Subject Device | Predicate | Discussion |
| --- | --- | --- | --- |
| Manufacturer | Anumana, Inc. | Anumana, Inc. | |
| Product Name | ECG-AI CA 12-Lead algorithm | ECG-AI LEF 12-Lead algorithm | |
| | The ECG-AI Cardiac Amyloidosis 12-Lead Algorithm is not intended to be a stand-alone diagnostic device for cardiac amyloidosis nor to replace current clinical practice guidelines.
A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Cardiac Amyloidosis. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out additional clinical evaluation.
The ECG-AI Cardiac Amyloidosis 12-Lead algorithm should be applied jointly with clinician judgment. | • postpartum women.
The ECG-AI LEF 12-Lead algorithm is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm.
A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.
The ECG-AI LEF 12-Lead algorithm should be applied jointly with clinical judgement. | |
| Diagnostic Application | Cardiac Amyloidosis | Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% | Equivalent. The predicate device is designed for a different disease but the subject device similarly helps to detect cardiac amyloidosis. Additional clinical validation was conducted to support equivalence. |
Page 7 of 8
{11}
eSTAR Traditional 510(k) Submission – ECG-AI CA 12-Lead Algorithm
K253801 - 510(k) Summary
| | Subject Device | Predicate | Discussion |
| --- | --- | --- | --- |
| Manufacturer | Anumana, Inc. | Anumana, Inc. | |
| Product Name | ECG-AI CA 12-Lead algorithm | ECG-AI LEF 12-Lead algorithm | |
| Target Population | Adults presenting with signs and symptoms consistent with cardiac amyloidosis | Adults at risk for heart failure | Equivalent. Both devices are intended to target adults with cardiovascular symptoms and/or risk factors. |
| Intended User | Clinicians | Clinicians | Identical. |
| Principal of Operation | Machine learning-based algorithm | Machine learning-based algorithm | Identical. |
| Data Acquisition | Acquires data from 12-Lead ECGs | Acquires data from 12-Lead ECGs | Identical. |
| Device Output Format | Algorithm output is provided to third party software that displays a binary result to clinicians. Output provided for each ECG is “Detected,” “Not Detected” or “Error”. | Algorithm output is provided to third party software that displays a binary result to clinicians. Output provided for each ECG is “Detected,” “Not Detected” or “Error”. | Identical. |
| Cybersecurity | Compliant with FDA Guidance on Cybersecurity | Compliant with FDA Guidance on Cybersecurity | Identical. |
## Substantial Equivalence Conclusion
The subject device, Anumana ECG-AI CA 12-Lead algorithm, is substantially equivalent to the predicate device, Anumana ECG-AI LEF 12-Lead algorithm (K250652). The devices have equivalent intended uses, principles of operation and technical characteristics. Where differences occur between the subject device and the predicate, results of clinical performance and non-clinical verification and validation demonstrate that the subject device is as safe and as effective as the predicate.
Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.