PUREVUE™ Quick Connect Cannula System Tray; PUREVUE™ Quick Connect Cannula System Lid
K253791 · DePuy Mitek, Inc. · KCT · Feb 26, 2026 · General Hospital
Device Facts
Record ID
K253791
Device Name
PUREVUE™ Quick Connect Cannula System Tray; PUREVUE™ Quick Connect Cannula System Lid
Applicant
DePuy Mitek, Inc.
Product Code
KCT · General Hospital
Decision Date
Feb 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid are intended to hold PUREVUE™ Quick Connect Cannula System instruments during the sterilization process, storage, and transportation between uses.
Device Story
Modular, reusable tray and lid system; contains and protects PUREVUE™ Quick Connect Cannula System instruments during steam sterilization, storage, and transport. Components are perforated to allow steam ingress/egress. Used in healthcare facilities; operated by clinical staff. System integrates into a single unit; requires external FDA-cleared sterile barrier wrap to maintain sterility post-processing. Facilitates organized handling of surgical instruments; ensures effective sterilization via validated pre-vacuum steam cycles. Benefits include standardized instrument organization and protection during reprocessing cycles.
Modular tray/lid system; materials: anodized aluminum (base/lid), stainless steel (latches), silicone/aluminum/stainless steel (inserts). Perforated design for steam penetration. Dimensions: 21.23-21.50" L x 9.71-10.01" W x 0.63-2.86" H. Sterilization: Pre-vacuum steam (132°C/4 min or 134°C/3 min). No electrical or software components.
Indications for Use
Indicated for storing, organizing, and transporting PUREVUE™ Quick Connect Cannula System instruments during sterilization and surgical procedures. Not intended to maintain sterility independently; must be used with an FDA-cleared sterile barrier wrap. Validated for a maximum load of 22.1 lb.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 26, 2026
DePuy Mitek, Inc.
Jeffrey Dzialo
Senior Regulatory Affairs Specialist
325 Paramount Dr.
Raynham, Massachusetts 02767
Re: K253791
Trade/Device Name: PUREVUE™ Quick Connect Cannula System Tray; PUREVUE™ Quick Connect Cannula System Lid
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization wrap
Regulatory Class: Class II
Product Code: KCT
Dated: November 26, 2025
Received: November 28, 2025
Dear Jeffrey Dzialo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253791 - Jeffrey Dzialo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253791 - Jeffrey Dzialo
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN A. Digitally signed by STEPHEN A. ANISKO -S
ANISKO -S
Date: 2026.02.26
11:43:35 -05'00'
Stephen Anisko
Acting Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K253791
Device Name
PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid
Indications for Use (Describe)
The PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid are used to store, organize, and transport PUREVUE™ Quick Connect Cannula System instruments during sterilization and surgical procedures. The tray and lid are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, FDA-cleared sterile barrier wrap.
The tray and lid were validated for a maximum load of 22.1 lb (tray + contents + lid + wrap).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Table 1 includes validated stem sterilization (moist heat sterilization) cycle parameters.
Table 1: Validated Steam Sterilization Cycle Parameters.
| Method | Exposure Time (minutes) | Exposure Temperature | Minimum Drying Time (minutes) |
| --- | --- | --- | --- |
| Pre-vacuum Steam Sterilization Cycle | 4 | 132 °C (270°F) | 30 to 60 |
| Pre-vacuum Steam Sterilization Cycle | 3 | 134 °C (274°F) | 30 to 60 |
Table 2 includes information regarding the descriptions and dimensions of the tray and lid.
Table 2: Descriptions and Dimensions of PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid.
| Part Number | Part Description | Volume to Vent Ratio (in³/in²) (with Lid) | Weight (lb) | Dimensions (in) (L x W x H) Tray (L x W) Lid |
| --- | --- | --- | --- | --- |
| 242333 | PUREVUE™ Quick Connect Cannula System Tray | 10.08 | 4.73 | 21.23 x 9.71 x 2.86 |
| 242334 | PUREVUE™ Quick Connect Cannula System Lid | Not applicable | 1.81 | 21.50 x 10.01 x 0.63 |
FORM FDA 3881 (8/23)
{5}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
# 510(k) Summary K253791
| Submitter's Name and Address | DePuy Mitek, Inc.
325 Paramount Drive
Raynham, MA 02767, USA
On behalf of:
Medos International SARL
Chemin-Blanc 38
CH-2400 Le Locle
Switzerland |
| --- | --- |
| Contact Person | Jeffrey Dzialo
Senior Regulatory Affairs Specialist
Phone: +1 800-382-4682
Email: jdzialo@its.jnj.com |
| Alternate Contact | Pol Fort Grébol
Project Manager Regulatory Affairs
Phone: +34 672-672-579
Email: pfortgre@its.jnj.com |
| Date Prepared | February 25, 2026 |
| Proprietary Name | PUREVUE™ Quick Connect Cannula System Tray
PUREVUE™ Quick Connect Cannula System Lid |
| Common Name | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories |
| Regulation Number | 21 CFR § 880.6850 (Sterilization wrap) |
| Regulatory Class | Class II |
| Product Code | KCT |
| Predicate Device | Synthes Graphic Case & Tray System (K241927)
No reference devices were used in this submission. |
| Device Description Summary | The PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid are a modular, reusable tray and lid system intended for use in health care facilities for the purpose of containing medical devices for sterilization. The system is composed of two pieces, designed to be integrated into a single unit that contains and protects PUREVUE™ Quick Connect Cannula System instruments during sterilization and transport. All components are perforated for steam penetration. |
K253791
Page 1 of 12
{6}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| Intended Use | The PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid are intended to hold PUREVUE™ Quick Connect Cannula System instruments during the sterilization process, storage, and transportation between uses. |
| --- | --- |
| Indications for Use | The PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid are used to store, organize, and transport PUREVUE™ Quick Connect Cannula System instruments during sterilization and surgical procedures. The tray and lid are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, FDA-cleared sterile barrier wrap.
The tray and lid were validated for a maximum load of 22.1 lb (tray + contents + lid + wrap).
Table 1 includes validated steam sterilization (moist heat sterilization) cycle parameters.
Table 1: Validated Steam Sterilization Cycle Parameters. |
| Method | Exposure Time (minutes) | Exposure Temperature | Minimum Drying Time (minutes) |
| Pre-vacuum Steam Sterilization Cycle | 4 | 132 °C (270°F) | 30 to 60 |
| Pre-vacuum Steam Sterilization Cycle | 3 | 134 °C (274°F) | 30 to 60 |
| Table 2 includes information regarding the descriptions and dimensions of the tray and lid.
Table 2: Descriptions and Dimensions of PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid. |
| Part Number | Part Description | Volume to Vent Ratio (in³/in²) (with Lid) | Weight (lb) | Dimensions (in) (L x W x H) Tray (L x W) Lid |
| 242333 | PUREVUE™ Quick Connect Cannula System Tray | 10.08 | 4.73 | 21.23 x 9.71 x 2.86 |
| 242334 | PUREVUE™ Quick Connect Cannula System Lid | Not applicable | 1.81 | 21.50 x 10.01 x 0.63 |
| Indications for Use Comparison | The subject and predicate devices share comparable intended use, indications for use, intended user, and user training requirements. The minor differences in the intended use statements primarily reflect the distinct formatting approaches adopted by DePuy Mitek, Inc. and |
K253791
Page 2 of 12
{7}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE OSTHOPAEDICS COMPANY OF Johnson & Johnson
| | Synthes. These variations are negligible and do not raise new questions of safety or effectiveness. | | | |
| --- | --- | --- | --- | --- |
| Comparison of Technological Characteristics with the Predicate Device | The subject devices, PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid, are compared to the previously cleared Synthes Graphic Case & Tray System (K241927) regarding intended use, indications for use, technological characteristics (design features), and operating principle. | | | |
| | Characteristic | Subject Devices (K253791) | Predicate Device (K241927) | Comparison |
| | General Information | | | |
| | Proprietary Name | (1) PUREVUE™ Quick Connect System Tray
(2) PUREVUE™ Quick Connect System Lid | (1) Synthes Graphic Case & Tray System | Not applicable |
| | Representative Image (Tray) | | | Similar |
| | Representative Image (Lid) | | | Similar |
| | Product Code | Same as the predicate device. | KCT | Same |
| | Intended Use | | | |
| | Intended Use | The PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid are intended to hold PUREVUE™ Quick Connect Cannula System instruments during the sterilization process, storage, and transportation between uses. | The Synthes Graphic Case and Tray System is designed to hold surgical implants and instruments in order to organize, steam sterilize and transport the devices between uses. The cases and trays are intended to be wrapped with an FDA-cleared sterilization wrap or placed within a reusable sterilization container during the pre-vacuum autoclave sterilization process. | Similar. The predicate device can be used with both instruments and surgical implants, and is qualified for use with a rigid sterilization container. |
| | Indications for Use | The PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect | The Synthes Graphic Case & Tray System is used in healthcare facilities to store, organize, and | Similar. The predicate device is validated for a greater maximum load and the subject |
K253791
Page 3 of 12
{8}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | | Cannula System Lid are used to store, organize, and transport PUREVUE™ Quick Connect Cannula System instruments during sterilization and surgical procedures. The tray and lid are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, FDA-cleared sterile barrier wrap.
The tray and lid were validated for a maximum load of 22.1 lb (tray + contents + lid + wrap).
Table 1 includes validated steam sterilization (moist heat sterilization) cycle parameters.
Table 2: Descriptions and Dimensions of PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid.
Table 1 includes the following descriptions and dimensions of PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid:
PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid
PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid
PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ |
| --- | --- | --- |
transport DePuy Synthes orthopedic instruments and implants during sterilization and surgical procedures. The Synthes Graphic Case & Tray System are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, FDA-cleared sterile barrier (e.g., wraps or reusable rigid sterilization containers).
The Synthes Graphic Case & Tray System is validated for use with orthopedic medical devices including lumen (cannulated) devices and mated surface devices (devices or device configurations with conjoined surfaces which meet, touch or unite). Compatible instrument and implant materials include intrinsically stable metals, composites, thermoplastics and thermosetting polymers with constant use temperatures above 135°C.
The Synthes Graphic Case & Tray System were validated for a maximum load of 25 lb (case + contents+ lid+ weight of sterile barrier wrap).
K253791
Page 4 of 12
{9}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | | | | |
| --- | --- | --- | --- | --- |
| | Contraindications | Same as the predicate device. | There are no contraindications specific to these devices. | Same |
| | Technological Characteristics | | | |
| | Principles of Operation | The PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect System Lid consist of general, non-sterile, reusable sterilization cases intended to contain and organize instruments for sterile processing. All components are perforated to allow ingress and egress of steam during sterilization. The tray and lid facilitate storage, handling, and transport of instruments between uses. | The Synthes Graphic Case and Tray System is a modular, reusable case and tray system intended for use in health care facilities for the purpose of containing medical devices for sterilization. It is composed of multiple pieces, designed to be integrated into a single unit that contains and protects instruments and implants during sterilization and transport. All components are perforated for steam penetration. | Same |
| Device Features | • Basic lid/base design
• General use trays
• Perforated for steam sterilization
• Handles
• Latches | • Basic lid/base design
• General use trays
• Perforated for steam sterilization
• Handles
• Latches | Same | |
K253791
Page 5 of 12
{10}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | Dimensions | Range of Dimensions • Length: 21.23 inches to 21.50 inches • Width: 9.71 inches to 10.01 inches • Height: 0.63 inches to 2.86 inches Total Validated System • Weight: 22.1 lb (container + contents + lid + wrap) | Range of Dimensions • Length: 6.34 inches to 20 inches • Width: 9.25 inches to 9.7 inches • Height: 1.77 inches to 6.6 inches Total Validated System • Weight: 25 lb (container + contents + lid + wrap) | Similar |
| --- | --- | --- | --- | --- |
| | Compatible Devices | PUREVUE™ Quick Connect Cannula System instruments | DePuy Synthes Implants and Instruments | Similar |
| | Materials | • Lid/base – Aluminum • Inserts – Silicone, Aluminum, Stainless Steel • Latch – Stainless Steel | • Lid/base – Aluminum • Inserts – Silicone, Aluminum, Stainless Steel • Latch – Stainless Steel | Same |
| | Sterilization Method | Same as the predicate device. | Pre-vacuum Steam Sterilization Cycle | Same |
| | Cycle Temperature | The subject devices are validated for the same sterilization cycle as the predicate device. In addition, the subject devices are validated for a pre-vacuum cycle with a 3-minute exposure time at 134 °C (274°F). | 270°F (132°C) | Similar. The subject devices include an additional cycle at 134 °C (274°F). |
| | Cycle Time | The subject devices are validated for the same sterilization cycle as the predicate device. In addition, the subject devices are validated for a pre-vacuum cycle with a 3-minute exposure time at 134 °C (274°F). | 4 minutes | Similar. The subject devices include an additional cycle with 3-minute exposure time. |
| | Dry Time | 30 minutes to 60 minutes | 20 minutes | Similar |
K253791
Page 6 of 12
{11}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | Sterile Barrier | The subject devices are intended to be used in conjunction with a legally marketed, FDA-cleared sterile barrier wrap. | FDA cleared sterilization wrap or rigid container. | Same. The predicate device is also qualified for use with a rigid container. |
| --- | --- | --- | --- | --- |
| | Patient Contact | Same as the predicate device. | No direct or indirect patient contact. | Same |
| | Performance Data | | | |
| | Non-clinical Performance Testing | • Steam Sterilization Efficacy
• Dry Time Validation
• Device Processing Cleaning Efficacy – Manual Cleaning
• Device Processing Cleaning Efficacy – Automated (Mechanical) Cleaning
• Durability Assessment (Limits of Reuse)
• Biocompatibility (Cytotoxicity of Ink Test Coupon)
• Design Verification for Tray/Lid Functional Testing | • Steam Sterilization Efficacy
• Dry Time Validation
• Device Processing Cleaning Efficacy – Manual Cleaning
• Device Processing Cleaning Efficacy – Automated (Mechanical) Cleaning
• Durability Assessment (Limits of Reuse)
• Biocompatibility (Cytotoxicity for Ink Test Coupon) | Same. The subject devices include design verification functional testing. |
The subject and predicate devices utilize the same design characteristics and fundamental technology, as presented in the previous table. The subject and predicate devices are perforated trays and lids composed of anodized aluminum bases with aluminum, stainless steel, and silicon inserts, as well as stainless-steel latches. The subject devices are intended for the storage, organization, transport, and sterile processing of medical devices. The subject and predicate devices are not intended to maintain sterility on their own and are reliant on an FDA-cleared
K253791
Page 7 of 12
{12}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | sterile barrier. The subject and predicate devices share the same principles of operation, device features, materials, and sterilization method.
The primary differences in technological characteristics (dimensions and weight) stem from the compatible devices which are intended to be sterilized within the container systems. The predicate case and tray system can be used with a broad set of instruments and implants, whereas the subject devices are utilized with the PUREVUE™ Quick Connect Cannula System instruments specifically. Because of the differences in compatible devices there are some differences in device dimensions and intended maximum weight. These differences are negligible as they did not result in unexpected design verification or validation results and are consistent with the respective intended use. In summary, all differences in technological characteristics have been identified and do not raise new questions of safety or effectiveness. | | | |
| --- | --- | --- | --- | --- |
| Non-clinical and Clinical Performance Data | Design verification and validation testing were performed on the subject PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid as a result of the risk analysis and product requirements.
Performance testing results were compared to those of the predicate device and demonstrated that the proposed devices are suitable for their intended use.
**Performance Testing**
Performance testing was conducted to ensure the device design meets user needs in the environment of use as shown in the table below. | | | |
| | Test Performed | Objective of the Test | Pre-defined Acceptance Criteria | Results Summary |
| | Steam Sterilization Efficacy | The purpose of the testing was to validate the sterilization efficacy of the PUREVUE™ Quick Connect Cannula System Tray and Lid when wrapped and processed in a steam pre-vacuum sterilization cycle at 132°C (270°F) with four (4) minutes exposure time. | All biological indicator test samples shall be negative for growth following the minimum incubation period.
The positive controls shall be positive for growth.
The negative and environmental controls shall be negative for growth.
The Chemical Integrators shall demonstrate steam | The subject devices passed the acceptance criteria and met the requirements.
The study results demonstrate the efficacy of the steam sterilization process in attaining a minimum microbiological lethality (F₀) of 12.0 minutes and therefore, achieving a Sterility Assurance Level (SAL) of 10⁻⁶ in the worst case set |
K253791
Page 8 of 12
{13}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | | | penetration.
The sterilizer cycle tapes shall verify that the parameters per were achieved. were achieved.
NOTE: In the event that a positive sample can clearly be identified as a source other than that of a sterilization efficacy failure, the sample shall be rendered invalid. Specifics to the cause of such a positive sample can be identified with a justification in the final report. | configuration container within the PUREVUE™ Quick Connect Cannula System Tray and Lid. |
| --- | --- | --- | --- | --- |
| | Dry Time Validation | The purpose of the testing was to determine the proper drying time of the DePuy Synthes PUREVUE™ Quick Connect System (including the tray and lid) when processed in a hospital steam pre-vacuum sterilization cycle at 132°C (270°F) with four (4) minutes exposure time and 134°C (273°F) for 3 minutes exposure time. | The wrapped instrument case shall produce a pre- and post-sterilization weight difference of less than 0.2% within five (5) minutes of cycle completion.
Each instrument tray of the case and sterilization wrap shall exhibit no visible moisture following thirty (30) minutes of cooling.
The integrators shall demonstrate steam penetration.
The cycle tapes shall confirm that the required cycle parameters were achieved. | The subject devices passed the acceptance criteria and met the requirements. |
| | Device Processing Cleaning Efficacy – Manual Cleaning | The purpose of the testing was to validate the manual cleaning procedure of the PUREVUE™ Quick Connect Cannula System (including the tray and lid). | Protein level shall be ≤ 6.4 μg/cm².
Hemoglobin level shall be ≤ 2.2 μg/cm².
Each test sample shall have no visible soil after cleaning.
The extraction efficiency shall be | The subject devices passed the acceptance criteria and met the requirements. |
K253791
Page 9 of 12
{14}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | | | greater than or equal to 70%.
The standard deviation of the sample set added to the concentration of the highest data point shall not exceed the acceptance criterion of 6.4 μg/cm². | |
| --- | --- | --- | --- | --- |
| | Device Processing Cleaning Efficacy – Automated (Mechanical) Cleaning | The purpose of the testing was to validate the automated cleaning procedure of the PUREVUE™ Quick Connect Cannula System (including the tray and lid). | The study results shall validate that the manufacturer’s automated cleaning instructions are efficacious in removing gross amounts of soil from the DePuy Synthes PUREVUE™ Quick Connect Cannula System (including the tray and lid).
Protein levels shall be less than 6.4 μg/cm².
Hemoglobin levels shall be less than 2.2 μg/cm². | The subject devices passed the acceptance criteria and met the requirements. |
| | Durability Assessment (Limits of Reuse) | The purpose of the testing was to demonstrate that the subject devices can withstand repeated clinical reprocessing cycles without unacceptable effects occurring to the material, joints, surfaces, color, labeling, function, etc. which could compromise patient / user safety or product’s function during its intended lifetime. | The PUREVUE™ Quick Connect Cannula System Tray shall have direct part markings identifying instrument profiles with instrument description
The PUREVUE™ Quick Connect Cannula System Tray shall allow for insertion and removal of instruments without damage.
The PUREVUE™ Quick Connect Cannula System Tray shall group 10 cannulas, 2 bridges, 6 obturators, and 1 switching stick, accommodating either length cannula | The subject devices passed the acceptance criteria and met the requirements. |
K253791
Page 10 of 12
{15}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | | | or obturator.
The PUREVUE™ Quick Connect Cannula System Tray and Lid shall lock and unlock via side latch, forming a sterilization container.
All products in scope shall meet the criteria of 500 reprocessing cycles.
There shall be no observations of unacceptable damage, including but not limited to, corrosion (rust, pitting), discoloration, excessive scratches, flaking, cracks and wear, or illegible part markings post testing.
The devices shall demonstrate proper function, including but not limited to, connection and functional fit. | |
| --- | --- | --- | --- | --- |
| | Biocompatibility (Cytotoxicity for Ink Test Coupon) | The purpose of this study was to determine the potential of the Aluminum Coupon (Fujifilm Ink) test article to cause cytotoxicity. | The results of the testing shall show Grade ≤ 2 cytotoxicity (mildly reactive). | The subject devices passed the acceptance criteria and met the requirements.
The test article extract showed no evidence of causing cell lysis or toxicity.
The test article extract met the requirements of the test since the grade was less than a grade 2 (mild reactivity). |
| | Design Verification for Tray/Lid Functional Testing | The objective of this testing was to establish evidence that the PUREVUE™ Quick Connect Cannula System Tray and Lid meet the form, fit, and functional | The PUREVUE™ Quick Connect Cannula System Tray and Lid shall meet the following pre-determined acceptance criteria:
• Dimensional | The subject devices passed the acceptance criteria and met the requirements. |
K253791
Page 11 of 12
{16}
510(k) Premarket Notification (Traditional)
PUREVUE™ Quick Connect Cannula System Tray and Lid
^{}[]
DePuy Synthes
THE ORTHOPAEDICS COMPANY OF Johnson & Johnson
| | | requirements outlined in the Requirements Traceability Matrix and mitigate associated risk. | • Functional fit
• Visual
• Loaded weight | |
| --- | --- | --- | --- | --- |
| **Software Verification and Validation**
No software verification or validation testing was required or performed for the subject devices, as they do not incorporate software functions. | | | | |
| **Electrical Safety and Electromagnetic Compatibility (EMC)**
No electrical safety or EMC testing was required or performed for the subject devices, as they do not contain electrical components. | | | | |
| **Animal Study**
No animal testing was required or performed for the subject devices, as their design and materials do not necessitate such evaluation to establish safety and effectiveness. | | | | |
| **Clinical Studies**
No human clinical testing was required or performed for the subject devices, as their design and intended use do not necessitate such evaluation to establish safety and effectiveness. | | | | |
| **Comparison of Subject and Predicate Devices** | The assessment of the comparison between the subject devices and the predicate device was based on a detailed device description, performance data, and conformance to consensus and voluntary standards.
The testing included in this submission demonstrates that:
• Any differences in technological characteristics of the predicate device do not raise any new questions of safety or effectiveness.
• The proposed devices are at least as safe and effective as the predicate device. | | | |
| **Conclusion** | The conclusions drawn from the non-clinical tests demonstrate that the subject devices in 510(k) submission K253791, the PUREVUE™ Quick Connect Cannula System Tray and PUREVUE™ Quick Connect Cannula System Lid, are as safe and effective as, and perform as well as or better than, the legally marketed predicate device cleared under K241927. | | | |
K253791
Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.