K253769 · Instylla, Inc. · FMF · Dec 22, 2025 · General Hospital
Device Facts
Record ID
K253769
Device Name
Instylla Delivery Kit
Applicant
Instylla, Inc.
Product Code
FMF · General Hospital
Decision Date
Dec 22, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline, and other aqueous solutions.
Device Story
The Instylla Delivery Kit is a sterile, single-patient-use manual device designed for the simultaneous injection of two distinct fluids. It consists of a double-barreled syringe holder, two 1mL needleless syringes with calibrated polycarbonate barrels and movable plungers, and a plunger clip that links the two plungers. The device is operated manually by a clinician who depresses the plunger clip to advance both plungers simultaneously. The syringes feature fixed male luer connectors for secure attachment to female luer hubs. The device is used in clinical settings to deliver radiographic contrast media, saline, or other aqueous solutions into the arterial or venous system. By enabling the simultaneous delivery of two fluids, the device facilitates specific procedural workflows, potentially improving efficiency and consistency in fluid administration.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility evaluation per ISO 10993-1, packaging validation per ISO 11607-1, and sterilization validation per ISO 11135. Syringes are compliant with ISO 7886-1 and ISO 80369-7.
Technological Characteristics
Components: Polycarbonate (PC) syringes, ABS holder/clip, silicone. Principle: Manual dual-piston syringe injection. Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 10^-6 per ISO 11135. Standards: ISO 7886-1 (graduation), ISO 80369-7 (luer connector), ISO 10993-1 (biocompatibility), ISO 10993-7 (EO residuals), ISO 11607-1 (packaging).
Indications for Use
Indicated for intra-arterial or intra-venous administration of radiographic contrast media, saline, and other aqueous solutions. Contraindicated for use in pediatric patients.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
Instyla, Inc.
Jennifer Greer
Sr. Regulatory Manager
201 Burlington Rd.
North Bldg.
Bedford, Massachusetts 01730
Re: K253769
Trade/Device Name: Instyla Delivery Kit
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston syringe
Regulatory Class: Class II
Product Code: FMF
Dated: November 25, 2025
Received: November 25, 2025
Dear Jennifer Greer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
2K253769 - Jennifer Greer
Page
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
3K253769 - Jennifer Greer
Page
Sincerely,
KYRAN R. GIBSON -S
For
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253769 | |
| Device Name Instylla Delivery Kit | |
| Indications for Use (Describe) The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline, and other aqueous solutions. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
instylla
# K253769 510(k) Summary
## Submitter Information:
Instylla, Inc.
201 Burlington Rd
North Building
Bedford, MA 01730
## Contact Person:
Jennifer Greer
Sr. Regulatory Manager
Phone: 781-622-9293
E-mail: jennyg@instylla.com
## Date Prepared:
December 19, 2025
## Subject Device:
Proprietary Name: Instylla Delivery Kit
Common Name: Syringe, Piston
Classification Name: Piston syringe
Classification Regulation: 21 CFR 880.5860
Product Code: FMF
Device Classification: Class II
Classification Panel: General Hospital
## Predicate Device:
Proprietary Name: Instylla Delivery Kit
Manufacturer: Instylla. Inc.
510(k) Number: K213632
## Device Description:
The Instylla Delivery Kit is comprised of two delivery syringes, a syringe holder, and a plunger clip (link). The barrel contains a fixed male luer connector, which is compatible with female luer hubs.
The Instylla Delivery Kit is a sterile, single patient use device intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous
Page 1
{5}
solutions. The device consists of a double-barreled syringe holder which holds and couples two (2) 1mL commercially available needleless syringes with calibrated hollow barrels and movable plungers. A single plunger clip is attached to both syringe plungers to facilitate the simultaneous injection of two distinct fluids when depressed.
The syringes of the Instylla Delivery Kit are composed of polycarbonate. The plunger clip and holder are composed of Acrylonitrile Butadiene Styrene (ABS).
## Indications for Use:
The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions.
## Comparison of Technological Characteristics to the Predicate Device:
The subject Instylla Delivery Kit is substantially equivalent in intended use and fundamental technological characteristics to the legally marketed predicate devices. The table below summarizes the similarities in design and configuration of the subject Instylla Delivery Kit compared with the predicate devices.
| Attribute | Subject Device: Instylla Delivery Kit | Predicate Device: Instylla Delivery Kit (K213632) | Substantial Equivalence |
| --- | --- | --- | --- |
| Indications for Use | The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions. | The Instylla Delivery Kit is intended to be used for the intra-arterial or intra-venous administration of radiographic contrast media, saline and other aqueous solutions. | Same |
| Contraindications | The safety and effectiveness of the Instylla Delivery Kit in pediatric patients has not been established. | None | The subject device includes a contraindication. |
| Basic Design | Two standard delivery piston syringes | Two standard delivery piston syringes | Same |
| Components | Two syringes, syringe holder, plunger clip | Two syringes, syringe holder, plunger clip | Same |
{6}
| Attribute | Subject Device: Instyla Delivery Kit | Predicate Device: Instylia Delivery Kit (K213632) | Substantial Equivalence |
| --- | --- | --- | --- |
| Materials | Generic Polycarbonate (PC), silicone, ABS | Polycarbonate Calibre Megarad 2081-15, silicone, ABS | The PC used in the barrel has been changed to generic PC based on the qualification of an additional source for the PC. |
| Principle of Operation: | Manually operated by advancing the dual syringe plungers simultaneously with the aid of a plunger clip. | Manually operated by advancing the dual syringe plungers simultaneously with the aid of a plunger clip. | Same |
| Operational Volume | 1 mL | 1mL or 3mL | The subject device only includes a 1mL operational volume. |
| Tip Type | ISO 80369-7 compliant fixed male luer connector | ISO 594-2 compliant fixed male luer connector | The syringe of the subject device is compliant with ISO 80369-7. |
| Gradation | Printed with accurate graduation lines that are compliant with ISO 7886-1. | Printed with accurate graduation lines that are compliant with ISO 7886-1. | Same |
| Packaging | Device is packed with a die card and placed into a tray with Tyvek lid | Device is packed with a die card and placed into a tray with Tyvek lid | Same |
| Sterilization Method | Ethylene Oxide (EO) to a Sterility Assurance Level (SAL) of 10^{-6} | EO to a SAL of 10^{-6} | Same |
| Shelf Life | 6 months | 6 months | Same |
{7}
The subject device includes a contraindication for pediatric use. This does not impact the safety and effectiveness of subject device as the intended use is still within the scope of the predicate device. Similarly, the subject device is only offered with a syringe volume of 1mL which differs from the predicate device that is offered with a syringe volume of 1mL and 3mL. This does not impact the safety or effectiveness since the operational volume is still within the cleared volume of the predicate device.
Lastly, the supplier of the syringes has changed the material of the barrel from polycarbonate (PC), Calibre Megarad 2081-15 to generic polycarbonate. A biocompatibility evaluation was conducted by the supplier to demonstrate this change does not impact the safety or effectiveness of the Instylla Delivery Kit. The supplier of the syringes also confirmed that the syringes now conform with the requirements of ISO 80369-7. This differs from the predicate device where the barrel of each syringe contained an ISO 594-2 compliant fixed male luer connector. This does not impact the safety and effectiveness of subject device as there are no changes being made to the design or materials of the syringes.
The subject and predicate device differ from one another in packaging and labeling. The intended use, indications, principle of operation, and technological characteristics remain identical between the subject and predicate device.
## Performance Data
Verification testing was conducted for product specifications related to packaging changes to the Instylla Delivery Kit. Modifications include updating the label stock suppliers for the shelf box and 2x2 label, adding patient chart peelable labels to the shelf box, and adding a heat-resistant aqueous coating to the shelf box. These changes do not impact the functional design specifications of the Instylla Delivery Kit. Therefore, only packaging validation was repeated for the Instylla Delivery Kit and no additional performance testing was performed.
The syringes of the subject device are compliant against FDA recognized standards ISO 7886-1 and ISO 80369-7.
## Biocompatibility Testing
The supplier of the syringes has changed the material of the barrel from polycarbonate (PC), Calibre Megarad 2081-15 to generic polycarbonate. A biocompatibility evaluation was conducted in accordance with the FDA Guidance for Industry, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." The Instylla Delivery Kit with modified packaging and labeling is classified as an external communicating device, indirectly contacting blood for a limited duration (≤24 hours). Biocompatibility testing on the updated syringes was performed per the FDA Guidance. The results of the testing concluded that the updated barrel material supplier does not raise any additional biocompatibility concerns.
Page 4
{8}
Page 5
# Sterility
The Instylla Delivery Kit with modified packaging and labeling is sterilized via a validated ethylene oxide (EO) process to a Sterility Assurance Level (SAL) of 10⁻⁶. The sterilization process was validated per ISO 11135 Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices. The Instylla Delivery Kit with modified packaging and labeling was evaluated and adopted into the validated process per AAMI TIR28 Product adoption and process equivalence for ethylene oxide sterilization. The EO and ECH levels were determined to be acceptable in accordance with ISO 10993-7 Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals.
Since the manufacturing process and materials of the Instylla Delivery Kit with modified packaging and labeling did not change, bacterial endotoxin test (BET) validation, also known as the Limulus amebocyte lysate (LAL) test, was not repeated for the subject device.
# Packaging
Packaging testing was conducted on sterilized Instylla Delivery Kit with modified packaging and labeling to evaluate the effectiveness of the packaging configuration to maintain the sterile barrier in accordance with ISO 11607-1 Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems. The subject device and updated packaging were evaluated after accelerated aging to demonstrate that the packaging maintains the sterile barrier and package integrity over the shelf life.
# Shelf Life
The Instylla Delivery Kit with modified packaging and labeling has a shelf life of 6 months. Shelf life studies have been conducted to demonstrate that the device maintains its performance and the packaging will maintain its sterile barrier over the entirety of the intended shelf life.
# Conclusion
Instylla has demonstrated that the Instylla Delivery Kit is substantially equivalent in fundamental design, function, device materials, packaging, sterilization, operating principle, intended use/ indication for use and fundamental technology to the legally marketed predicate device, Instylla Delivery Kit, which was cleared under 510(k) Premarket Notification K213632 on February 10, 2022.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.