K253760 · Sight Sciences, Inc. · MRH · Aug 4, 2026 · General Hospital
Device Facts
Record ID
K253760
Device Name
OMNI Ultra Surgical System (1-110)
Applicant
Sight Sciences, Inc.
Product Code
MRH · General Hospital
Decision Date
Aug 4, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OMNI Ultra Surgical System is indicated for canaloplasty (microcatheterization and transluminal viscodilation of Schlemm's canal) followed by trabeculotomy (cutting of trabecular meshwork) to reduce intraocular pressure in adult patients with primary open-angle glaucoma.
Device Story
Handheld, manually operated ophthalmic surgical device; used by ophthalmologists in clinical settings. Device accesses, microcatheterizes, and viscodilates Schlemm's canal; re-accesses canal to cut trabecular meshwork. Input: manual rotation of advancement wheels by surgeon. Transformation: internal reservoir and plunger mechanism dispenses cohesive viscoelastic fluid (OVD) into target anatomy. Output: controlled delivery of OVD (6 μl on advance, 15 μl on retraction) and mechanical cutting of trabecular meshwork. Benefits: reduction of intraocular pressure in glaucoma patients. Device is sterile, single-use, and compatible with standard OVD cartridges.
Clinical Evidence
Bench testing only. Includes dimensional/visual inspections, mechanical joint strength and actuation force testing, simulated use testing, sterilization validation, packaging/shelf-life testing, biocompatibility, bacterial endotoxin testing, and comparative human cadaver eye performance testing. No clinical data presented.
Indicated for adult patients with primary open-angle glaucoma requiring canaloplasty and trabeculotomy to reduce intraocular pressure.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}

August 4, 2026
Sight Sciences, Inc.
Rachel Franco
Lead Regulatory Affairs Specialist
4040 Campbell Ave., Suite 100
Menlo Park, California 94025
Re: K253760
Trade/Device Name: OMNI Ultra Surgical System (1-110)
Regulation Number: 21 CFR 880.5725
Regulation Name: Infusion Pump
Regulatory Class: Class II
Product Code: MRH, HMZ
Dated: June 29, 2026
Received: June 30, 2026
Dear Rachel Franco:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253760 - Rachel Franco
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253760 - Rachel Franco
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CLAUDINE H. KRAWCZYK -S
Claudine Krawczyk
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253760 | ? |
| Please provide the device trade name(s). | | ? |
| OMNI Ultra Surgical System (1-110) | | |
| Please provide your Indications for Use below. | | ? |
| The OMNI Ultra Surgical System is indicated for canaloplasty (microcatheterization and transluminal viscodilation of Schlemm's canal) followed by trabeculotomy (cutting of trabecular meshwork) to reduce intraocular pressure in adult patients with primary open-angle glaucoma. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
S
Sight
Sciences®
OMNI Ultra Surgical System
# 510(k) SUMMARY
## Submitter Information
510(k)Number: K253760
510(k) Owner: Sight Sciences, Inc.
4040 Campbell Ave., Suite 100
Menlo Park, CA 94025
Tel: (877) 266-1144
Contact Person: Rachel M. Franco
Lead Regulatory Affairs Specialist
4040 Campbell Ave., Suite 100
Menlo Park, CA 94025
Tel: 408-221-7376
Date Prepared: July 28, 2026
## Device Name and Classification
| TRADE NAME: | OMNI Ultra Surgical System |
| --- | --- |
| COMMON NAME: | Ophthalmic Infusion Pump |
| CLASSIFICATION NAME: | Infusion pump |
| REGULATION NUMBER: | 21 CFR 880.5725 |
| DEVICE CLASSIFICATION: | Class II |
| PRODUCT CODE: | MRH |
| DEVICE | Pump, infusion, ophthalmic |
| ASSOCIATED PRODUCT CODE | HMZ |
| DEVICE | trabeculotome |
## Predicate Device
Device Name: OMNI Surgical System
510(k) Holder: Sight Sciences, Inc.
510(k) Number: K232214
Clearance Date: August 25, 2023
## Indications for Use
The OMNI® Ultra Surgical System is indicated for canaloplasty (microcatheterization and transluminal viscodilation of Schlemm's canal) followed by trabeculotomy (cutting of trabecular meshwork) to reduce intraocular pressure in adult patients with primary open-angle glaucoma.
K253760
Page 1 of 5
{5}
S
Sight Sciences®
OMNI Ultra Surgical System
## Device Description
The Sight Sciences OMNI Ultra Surgical System is a handheld, manually operated device used by ophthalmologists to access, microcatheterize, and viscodilate Schlemm's canal ("canaloplasty") and to re-access Schlemm's canal and cut trabecular meshwork tissue ("trabeculotomy"). The OMNI Ultra Surgical System is provided sterile and disposed after single-patient use. The device is fabricated from biocompatible materials standard to the medical device industry. Each OMNI Ultra device dispenses fluid on the principle of exchanging volumes much like a syringe and is designed to function with commercially available cohesive viscoelastic fluids (also known as ophthalmic viscosurgical device, or "OVD").
The OMNI Ultra device includes a stainless-steel cannula, polymeric microcatheter, removable priming lock, internal reservoir, a Luer fitting for direct connection with an OVD cartridge to prime the internal reservoir, and two advancement wheels. The stainless-steel cannula has a curved shape with a beveled tip for entry through the trabecular meshwork into Schlemm's canal. A single advancement wheel is located on each side of the handle. This allows the OMNI Ultra device to be used in either eye (OD or OS) and in either hand of the surgeon (left or right), by turning the device 180 degrees along its vertical axis. These wheels are used to advance and retract the microcatheter. To perform the combined and sequential canaloplasty/trabeculotomy procedures, the canaloplasty is performed first, followed by trabeculotomy. The subject OMNI Ultra Surgical System device dispense characteristics are described in Table 1 below.
Table 1 OMNI Ultra Surgical System device dispense characteristics
| Catalog Number | Description | Volume OVD Dispense on Advance | Volume OVD Dispense on Retraction |
| --- | --- | --- | --- |
| 1-110 | OMNI Ultra Surgical System | 6 μl | 15 μl |
## Comparison of Technological Characteristics with the Predicate
The primary predicate in this 510(k) submission is the OMNI Surgical System described in K232214. The OMNI Ultra Surgical System has the same intended use, indications for use, and similar technological characteristics as the predicate OMNI device. The proposed OMNI Ultra Surgical System updates aim to enhance the user experience while maintaining the same overall device performance with modifications limited to handle length, microcatheter travel length during canaloplasty, and dual OVD dispensing (on advancement and retraction) with the total volume dispensed equivalent to that of the predicate device (21 μl).
OMNI Ultra Surgical System Device updates do not significantly affect how the device is used, no new significant risks were identified with the changes, no unintended consequences of the changes were identified, and the changes did not significantly impact the safety or effectiveness of the device. A comparison of the technological characteristics between the subject OMNI Ultra Surgical System compared with the predicate OMNI Surgical System device is shown in Table 2 below.
K253760
Page 2 of 5
{6}
S
Sight Sciences®
OMNI Ultra Surgical System
Table 2 Comparison of Technological Characteristics with the Predicate Device
| Characteristic | OMNI Ultra Surgical System | OMNI Surgical System |
| --- | --- | --- |
| | SUBJECT DEVICE | PREDICATE DEVICE (K232214) |
| Intended Use | Ophthalmic surgical tool for delivery of controlled amounts of viscoelastic fluid into the anterior segment and used to cut trabecular meshwork when a trabeculotomy is indicated | Ophthalmic surgical tool for delivery of controlled amounts of viscoelastic fluid into the anterior segment and used to cut trabecular meshwork when a trabeculotomy is indicated |
| Indications for Use | The OMNI® Ultra Surgical System is indicated for canaloplasty (microcatheterization and transluminal viscodilation of Schlemm's canal) followed by trabeculotomy (cutting of trabecular meshwork) to reduce intraocular pressure in adult patients with primary open-angle glaucoma. | The OMNI® Ultra Surgical System is indicated for canaloplasty (microcatheterization and transluminal viscodilation of Schlemm's canal) followed by trabeculotomy (cutting of trabecular meshwork) to reduce intraocular pressure in adult patients with primary open-angle glaucoma. |
| Target Anatomy | Schlemm's Canal and Trabecular Meshwork | Schlemm's Canal and Trabecular Meshwork |
| Operating Principle | Manual | Manual |
| Dispensing Control | After priming, viscoelastic dispensing control occurs through manual rotation of either advancement wheel at the distal end of the device | After priming, viscoelastic dispensing control occurs through manual rotation of either advancement wheel at the distal end of the device |
| Dispensing Mechanism | Internal reservoir with plunger tube and AD rod (syringe-like volume exchange). | Internal reservoir with plunger tube (syringe-like volume exchange). |
| Viscoelastic Fluid (OVD) and Priming Method | Cohesive viscoelastic fluid (OVD or ophthalmic viscosurgical device) is supplied separately. Viscoelastic loaded into device (primed) prior to use by attaching OVD cartridge directly to Luer fitting on proximal end of OMNI device handle | Cohesive viscoelastic fluid (OVD or ophthalmic viscosurgical device) is supplied separately. Viscoelastic loaded into device (primed) prior to use by attaching OVD cartridge directly to Luer fitting on proximal end of OMNI device handle |
| OVD Volume Dispensed | Total 21 μL: 6 ± 3 μL during advancement and 15 ± 3 μL during retraction | 21 ± 3μL – 10.5 μL/each during the first two retractions |
| Materials | Medical grade materials, including ABS, polycarbonate, stainless steel, silicone, parylene coating, cyanoacrylate, acrylated urethane, polyimideMicrocatheter material: polyamide (Nylon) with laser markings and additives | Medical grade materials, including ABS, polycarbonate, stainless steel, silicone, parylene coating, cyanoacrylate, acrylated urethane, polyimideMicrocatheter material: polyimide (Nylon) |
| User Interface | Handheld | Handheld |
| Microcatheter length | 35 mm | 20 mm |
| Microcatheter Shaft Outer Diameter | 200 microns | 200 microns |
K253760
Page 3 of 5
{7}
Sight Sciences®
OMNI Ultra Surgical System
| Characteristic | OMNI Ultra Surgical System SUBJECT DEVICE | OMNI Surgical System PREDICATE DEVICE (K232214) |
| --- | --- | --- |
| Microcatheter Tip Outer Diameter Range | 0.0095 to 0.0115 inches | 0.0090 to 0.0110 inches |
| Sterile and Single Use | Provided sterile. Single patient use | Provided sterile. Single patient use |
| Sterilization Method | Gamma radiation | Gamma radiation |
| Sterility Assurance Level | 10^{-6} | 10^{-6} |
| Packaging | Thermoformed plastic tray with heat-sealed Tyvek lid | Thermoformed plastic tray with heat-sealed Tyvek lid |
### Risk Analysis
The risk management process at Sight Sciences complies with ISO 14971:2019 “Medical devices - - Application of risk management to medical devices.” As required by this standard, risk analyses are conducted according to defined procedures, using experienced, qualified personnel from multiple functions throughout the organization with prior experience in risk assessment. All the identified hazards were mitigated to an acceptable level of risk. The potential benefits to patients outweigh the low residual risk, taking into consideration the indications for use of the OMNI Ultra Surgical System.
### Performance Data – Bench Testing
The descriptive characteristics of the OMNI Ultra Surgical System are well-defined and adequate to ensure equivalence to the predicate device. Additionally, the following performance testing and inspections were conducted on the OMNI Ultra device: dimensional and visual inspections, visual inspection of labeling and component inspections, mechanical testing of joint strength and actuation force, simulated use testing, sterilization validation, packaging and shelf-life testing, biocompatibility testing, bacterial endotoxin testing, and comparative human cadaver eye performance testing. Acceptance criteria were based on design inputs including predicate OMNI performance, relevant standards, and where applicable, comparison to the predicate device performance. Testing demonstrated that the OMNI Ultra device performs as intended and meets defined specifications and functions as intended and demonstrates that the OMNI Ultra Surgical System is substantially equivalent to the legally marketed OMNI Surgical System predicate device cleared under K232214.
K253760
Page 4 of 5
{8}
S
Sight Sciences®
OMNI Ultra Surgical System
## Conclusions Drawn from Testing
Based on the testing conducted on the physically identical OMNI Ultra Surgical System, the Sight Sciences modified OMNI Ultra Surgical System has complied with applicable standards and met all product design requirements. The subject device shares the same principle of operation, intended use, indications for use, and many of the same key technological characteristics as the predicate device. Performance testing and inspections conducted on the OMNI Ultra device included: dimensional and visual inspections, visual inspection of labeling and component inspections, mechanical testing of joint strength and actuation force, simulated use testing, sterilization validation, packaging and shelf-life testing, biocompatibility testing, bacterial endotoxin testing, and comparative human cadaver eye performance testing. This testing demonstrated that the OMNI Ultra device performs as intended, meets defined specifications, and functions as intended and demonstrates that the OMNI Ultra Surgical System is substantially equivalent to the legally marketed OMNI Surgical System predicate device cleared under K232214. Acceptance criteria were based on design inputs including predicate OMNI performance, relevant standards, and where applicable, comparison to the predicate device performance. Therefore, the OMNI Ultra Surgical System is substantially equivalent to the legally marketed OMNI Surgical System predicate device cleared under K232214.
K253760
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.