Edi Catheter ENFit 6Fr/48 cm (6893166); Edi Catheter ENFit 6Fr/49cm (6893273); Edi Catheter ENFit 6Fr/50cm (6893274); Edi Catheter ENFit 8Fr/50cm (6893275); Edi Catheter ENFit 8Fr/100cm (6893276); Edi Catheter ENFit 8Fr/125cm (6893277); Edi Catheter ENFit 12Fr/125cm (6893278); Edi Catheter ENFit 16Fr/125cm (6893279)
K253715 · Maquet Critical Care AB · PIF · Aug 26, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K253715
Device Name
Edi Catheter ENFit 6Fr/48 cm (6893166); Edi Catheter ENFit 6Fr/49cm (6893273); Edi Catheter ENFit 6Fr/50cm (6893274); Edi Catheter ENFit 8Fr/50cm (6893275); Edi Catheter ENFit 8Fr/100cm (6893276); Edi Catheter ENFit 8Fr/125cm (6893277); Edi Catheter ENFit 12Fr/125cm (6893278); Edi Catheter ENFit 16Fr/125cm (6893279)
Applicant
Maquet Critical Care AB
Product Code
PIF · Gastroenterology, Urology
Decision Date
Aug 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Edi Catheter ENFit is intended for: • administrating nutrition, fluids and medications via the naso-gastroenteric or oro-gastroenteric route • aspiration via the naso-gastroenteric or oro-gastroenteric route • transfer of electrical activity of the diaphragm (Edi signal) to compatible Servo ventilator systems on which neurally controlled ventilation modes are available.
Device Story
Nasogastric feeding tube with integrated electrode array; records diaphragm electrical activity (Edi). Accessory to Servo ventilator systems; provides Edi signal to improve patient-ventilator synchrony; enables neurally controlled ventilation modes. Monitors respiratory drive in standby mode. Used in clinical settings; operated by healthcare professionals. Single-use; 5-day duration. Gamma sterilized. Provides enteral access for nutrition/medication/aspiration while simultaneously facilitating neural respiratory monitoring.
Clinical Evidence
No clinical tests were conducted. Evidence consists of bench-top design verification and validation, including tensile strength, leakage, kinking, flow rate, suction, electrical functionality, and animal testing. Biocompatibility, packaging, and sterilization validated per ISO 10993-1, ISO 11607, and ISO 11137 standards.
Technological Characteristics
Nasogastric tube with electrode array for Edi signal acquisition. Materials: biocompatible polymers. Connectors: ENFit (ISO 80369-3). Sterilization: Gamma radiation. Shelf-life: 2 years. Connectivity: Wired interface to Servo ventilator systems. Dimensions: 6Fr to 16Fr diameters; 48cm to 125cm lengths.
Indications for Use
Indicated for neonates, infants, children, adolescents, and adults requiring naso-gastroenteric or oro-gastroenteric nutrition, fluid/medication administration, gastric aspiration, or neurally controlled ventilation via compatible Servo ventilator systems.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 26, 2026
Maquet Critical Care Ab
Jerker Åberg
Director Regulatory Affairs Maquet Critical Care AB
Röntgenvägen 2 Se-171 54
Solna, Sweden
Re: K253715
Trade/Device Name: Edi Catheter ENFit 6Fr/48 cm (6893166);
Edi Catheter ENFit 6Fr/49cm (6893273);
Edi Catheter ENFit 6Fr/50cm (6893274);
Edi Catheter ENFit 8Fr/50cm (6893275);
Edi Catheter ENFit 8Fr/100cm (6893276);
Edi Catheter ENFit 8Fr/125cm (6893277);
Edi Catheter ENFit 12Fr/125cm (6893278);
Edi Catheter ENFit 16Fr/125cm (6893279)
Regulation Number: 21 CFR 876.5980
Regulation Name: Gastrointestinal Tube And Accessories
Regulatory Class: Class II
Product Code: PIF, CBK, MOD
Dated: July 24, 2026
Received: July 27, 2026
Dear Jerker Åberg:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253715 - Jerker Åberg
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253715 - Jerker Åberg
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# ANTHONY LEE -S
Anthony Lee, Ph.D., MBA
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253715 | ? |
| Please provide the device trade name(s). | | ? |
| Edi Catheter ENFit 6Fr/48 cm (6893166); Edi Catheter ENFit 6Fr/49cm (6893273); Edi Catheter ENFit 6Fr/50cm (6893274); Edi Catheter ENFit 8Fr/50cm (6893275); Edi Catheter ENFit 8Fr/100cm (6893276); Edi Catheter ENFit 8Fr/125cm (6893277); Edi Catheter ENFit 12Fr/125cm (6893278); Edi Catheter ENFit 16Fr/125cm (6893279) | | |
| Please provide your Indications for Use below. | | ? |
| The Edi Catheter ENFit is intended for: • administrating nutrition, fluids and medications via the naso-gastroenteric or oro-gastroenteric route • aspiration via the naso-gastroenteric or oro-gastroenteric route • transfer of electrical activity of the diaphragm (Edi signal) to compatible Servo ventilator systems on which neurally controlled ventilation modes are available. | | |
| Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
Page 1 of 3
| 510(k) #: K253715 | 510(k) Summary | Prepared on: 2026-08-24 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Maquet Critical Care AB | |
| Applicant Address | Röntgenvägen 2 SE-171 54, Solna, Sweden | |
| Applicant Contact Telephone | +46 70 7827985 | |
| Applicant Contact | Mr. Jerker Åberg | |
| Applicant Contact Email | jerker.aberg@getinge.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Edi Catheter ENFit 6Fr/48 cm (6893166); Edi Catheter ENFit 6Fr/49cm (6893273); Edi Catheter ENFit 6Fr/50cm (6893274); Edi Catheter ENFit 8Fr/50cm (6893275); Edi Catheter ENFit 8Fr/100cm (6893276); Edi Catheter ENFit 8Fr/125cm (6893277); Edi Catheter ENFit 12Fr/125cm (6893278); Edi Catheter ENFit 16Fr/125cm (6893279) | |
| Common Name | Gastrointestinal Tube with Enteral Specific Connectors | |
| Classification Name | Gastrointestinal Tube and Accessories | |
| Regulation Number | 21 CFR 876.5980 | |
| Product Code(s) | PIF, CBK, MOD | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| --- | --- | --- |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K153688 | Edi Catheter ENFit | PIF |
| Device Description Summary | 21 CFR 807.92(a)(4) |
| --- | --- |
| The Edi Catheter ENFit is a nasogastric feeding tube with an array of electrodes that record diaphragm electrical activity (Edi). The Edi Catheter ENFit is an accessory to be used with Servo ventilator systems that have neurally controlled ventilation modes available. The Edi signal is used as an additional detector to improve the synchrony between the patient and the ventilator and to give the patient corresponding ventilatory support. The Edi Catheter ENFit may also be used to record and monitor diaphragm electrical activity while the Servo ventilator system is in standby. The Edi Catheter ENFit has three functionalities: 1. To detect diaphragmatic electrical activity (Edi signals) by which the neurally ventilatory modes can be controlled. 2. To detect diaphragm electrical activity (Edi signals) by which the respiratory drive from the brain can be monitored even without ventilation, in stand-by mode. 3. To function as a naso-gastric tube for feeding and medications and to aspirate from the gastric ventricle. The Edi Catheter ENFit is available in different sizes depending on the patient population. The Edi Catheter ENFit is for single-use only and may be used for up to five days. The catheter is sterilized by the use of gamma radiation and has a validated shelf-life of two years. | |
{5}
Page 2 of 3
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Edi Catheter ENFit is intended for:
- administrating nutrition, fluids and medications via the naso-gastroenteric or oro-gastroenteric route
- aspiration via the naso-gastroenteric or oro-gastroenteric route
- transfer of electrical activity of the diaphragm (Edi signal) to compatible Servo ventilator systems on which neurally controlled ventilation modes are available.
# Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device Edi Catheter ENFit has the same indications for use as the predicate device Edi Catheter ENFit K152688. There is a slight clarification of the indications for use for the subject device, however not affecting directions for use or safety and effectiveness.
# Technological Comparison
21 CFR 807.92(a)(6)
The subject device Edi Catheter ENFit and the predicate device Edi Catheter ENFit (K153688) have the following similarities:
- The same intended use
- The same operating principles
- The same design
- The same sterilization method
- The same usage time of 5 days
- The same shelf life of 2 years
The subject device incorporates the following modifications:
- An additional shorter catheter version for premature patients, 6Fr/48cm with 2-lumen design
- Material change of the Male body of the ENFit connector
- Change of the sterile barrier of the packaging, the material in contact with the device remains unchanged
The design verification/validation testing demonstrates that the modification to the Edi Catheter ENFit has equivalent technological characteristics compared to the predicate devices, or in the case where differences are noted, such differences do not raise new questions of safety or effectiveness. Such conclusions are also supported by the successful design verification and validation conducted and included in the submission.
# Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Testing was performed to demonstrate that the incorporated modifications do not affect the device's ability to perform as intended. Specifically, the objective of the testing was to confirm that the modifications do not compromise safety and performance.
The design verification activities consisted of:
- Product verification
- Tensile Strength
- Liquid Leakage
- Forces Tip to Tissue and Insertion/Withdrawal Force
- Horizontal Kinking
- Flow Rate
- Suction
- Resistance/Capacitance and Electrical Functionality
- Animal testing
ENFit Connector verification in accordance with ISO 80369-3, Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications, using the test methods provided in ISO 80369-20, Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods.
Biocompatibility testing was conducted in accordance with ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" and applicable sub-standards.
Packaging validation was conducted in accordance with ISO 11607-1, "Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems" and ISO 11607-2, "Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes".
{6}
Page 3 of 3
Sterilization validation was conducted in accordance with ISO 11137-1, "Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices" and applicable sub-standards.
Transport and storage testing was conducted in accordance with ASTM D4169, "Standard Practice for Performance Testing of Shipping Containers and System".
No clinical tests were conducted and submitted in this 510(k) submission.
The results of the verification/validation testing demonstrate that the Edi Catheter ENFit meets the established performance requirements for the device and that the proposed device is as safe and effective as the predicate device (K153688).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.