K253693 · Riverpoint Medical, LLC · MBI · Mar 11, 2026 · Orthopedic
Device Facts
Record ID
K253693
Device Name
Strut Suture; No-Tie Button
Applicant
Riverpoint Medical, LLC
Product Code
MBI · Orthopedic
Decision Date
Mar 11, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Strut Suture is intended to be used for fixation of bone to bone, and is intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in ACL/PCL Repair. The No-Tie Button is intended to be used for fixation of bone to bone, intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in ACL/PCL Repair when used with Strut Suture (or equivalent) and soft tissue fixation to bone in the knee for meniscal root repair when used with RootMend MRR (or equivalent).
Device Story
Strut Suture and No-Tie Button are orthopedic fixation devices; used for bone-to-bone fixation, ligament/tendon repair (ACL/PCL), and meniscal root repair. Strut Suture consists of UHMWPE suture with a titanium (Ti-6Al-4V) button; No-Tie Button is a standalone titanium fixation anchor. Devices function as fixation posts or distribution bridges to manage suture tension. Used in surgical settings by orthopedic surgeons. Devices provide mechanical stability for soft tissue-to-bone or bone-to-bone attachment; benefit includes secure fixation during healing. Output is physical stabilization of tissue/bone constructs.
Clinical Evidence
Bench testing only. Includes cyclic and ultimate tensile strength (UTS) testing of anchor constructs, usability engineering per EN62366-1, sterilization validation (ISO 14937), biocompatibility (ISO 10993-1, ISO 10993-7), and packaging stability (ISO 11607-1). LAL and rabbit pyrogenicity testing confirmed non-pyrogenic status.
Technological Characteristics
Materials: UHMWPE (suture), Titanium Ti-6Al-4V (button). Principle: Mechanical bone fixation fastener. Sterilization: Ethylene oxide. Standards: ISO 10993-1/7 (biocompatibility), ISO 11607-1 (packaging), ISO 14937 (sterilization). Single-use, non-degradable.
Indications for Use
Indicated for patients requiring bone-to-bone fixation, ligament or tendon repair (ACL/PCL), or meniscal root repair. Used for fixation posts, distribution bridges, or suture tension distribution.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 11, 2026
Riverpoint Medical, LLC
Palmira Obeso
Regulatory Associate II
825 NE 25th Ave.
Portland, Oregon 97272
Re: K253693
Trade/Device Name: Strut Suture; No-Tie Button
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: MBI
Dated: February 9, 2026
Received: February 9, 2026
Dear Palmira Obeso:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253693 - Palmira Obeso
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253693 - Palmira Obeso
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253693 | ? |
| Please provide the device trade name(s). | | ? |
| Strut Suture;
No-Tie Button | | |
| Please provide your Indications for Use below. | | ? |
| Strut Suture:
The Strut Suture is intended to be used for fixation of bone to bone, and is intended as fixation posts, a
distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in ACL/PCL
Repair.
No-Tie Button
The No-Tie Button is intended to be used for fixation of bone to bone, intended as fixation posts, a
distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in ACL/PCL
Repair when used with Strut Suture (or equivalent) and soft tissue fixation to bone in the knee for meniscal
root repair when used with RootMend MRR (or equivalent). | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
RP RIVERPOINT
K253693
510(k) SUMMARY
Strut Suture, No-Tie Button
## Submitter Information
Submitter's Name: Riverpoint Medical
Address: 825 NE 25th Ave. Portland, OR 97232
Phone Number: (503) 517-8001
Fax Number: (503) 517-8002
Registration Number: 3006981798
Contact Person: Palmira Obeso (503) 517-8001
Date of Preparation: November 21st, 2025
## Device Name
Trade Name: Strut Suture; No-Tie Button
Common or Usual Names: Fastener, fixation, nondegradable, soft tissue
## Primary Device Classification
Product Code: MBI
FDA Class: 2
Regulation Number: 888.3040: Smooth or threaded metallic bone fixation fastener.
## Predicate Device
K202581 – Arthrex Tight Rope II
## Reference Devices
K230212 – OrthoButton AL
K243988 – RootMend MRR
K253693 - Page 1 of 3
{5}
RP RIVERPOINT
## Device Description
Strut Suture devices are a suture and button construct composed of uncoated UHMWPE (ultra-high molecular weight polyethylene) with UHMWPE tracer material and an attachable fixation anchor composed of a titanium (Ti-6Al-4V) button. Strut Suture devices will be provided sterile using ethylene oxide, for single use only, and are not to be re-sterilized.
No-Tie Button devices are an attachable fixation anchor composed of a titanium (Ti-6Al-4V) button. No-Tie Button devices will be provided sterile using ethylene oxide, for single use only, and are not to be re-sterilized.
## Intended Use / Indications for Use
The Strut Suture is intended to be used for fixation of bone to bone, and is intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in ACL/PCL Repair.
The No-Tie Button is intended to be used for fixation of bone to bone, intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in ACL/PCL Repair when used with Strut Suture (or equivalent) and soft tissue fixation to bone in the knee for meniscal root repair when used with RootMend MRR (or equivalent).
## Performance Data
Non-clinical mechanical testing was performed to verify the fixation strength of the Strut Suture and No-Tie Button devices and is compared to the predicate device. Testing conducted includes cyclic and UTS testing of anchor constructs and usability engineering testing with simulated use in cadaver lab performed per EN62366-1. In all instances of the testing referenced above, the acceptance criteria were met, and the proposed devices performed as intended.
Additional non-clinical testing for the Strut Suture and No-Tie Button materials and packaging included sterilization validation per ISO 14937:2009 Sterilization of health care products- General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices including biocompatibility ISO 10993-7:2008 for Biological evaluation of medical devices-Part 7 Ethylene Oxide Sterilization Residuals, biocompatibility testing per ISO 10993-1:2009 - Biological Evaluation of Medical Devices, stability testing on the product material and packaging per ISO 11607-1:2006 - Packaging for terminally sterilized medical devices -- Part 1: Requirements for materials, sterile barrier systems and packaging systems. LAL and rabbit pyrogenicity testing have demonstrated that the materials used in the Strut Suture devices are non-pyrogenic and do not raise any additional concerns.
K253693 - Page 2 of 3
{6}
RP RIVERPOINT
# Substantial Equivalence and Comparison of Technical Characteristics
The proposed Strut Suture and No-Tie Button devices are within the scope of the intended use and indications, has similar principles of operation, and has similar technical characteristics as the predicate Arthrex Tight Rope II (K151342) and references OrthoButton AL (K230212) and RootMend MRR (K243988) No notable technical differences were identified. Therefore, the subject devices do not raise any issues of safety or effectiveness.
# Conclusion
The information provided in this Traditional 510(k) demonstrates that the Strut Suture and No-Tie Button subject devices are substantially equivalent to the predicate device.
K253693 - Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.