K253688 · Bayer Medical Care, Inc. · PQH · Jun 12, 2026 · General Hospital
Device Facts
Record ID
K253688
Device Name
MEDRAD® Imaging Bulk Package Transfer Spike
Applicant
Bayer Medical Care, Inc.
Product Code
PQH · General Hospital
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The MEDRAD® Imaging Bulk Package (IBP) Transfer Spike is indicated for the transfer of the following contrast media: - Ambelvist® (gadoquatrane) injection in IBP - Gadavist® (gadobutrol) injection in IBP The Transfer Spike is used to transfer contrast media to empty, sterile hand syringes for MR procedures. Discard the Transfer Spike after one of the following conditions has occurred first: (1) the Transfer Spike has been disconnected from the contrast media container; (2) the contrast media container has been depleted; or (3) after contrast media use time has elapsed since the container was penetrated. For a specific contrast media use time, refer to the contrast media prescribing information or the contrast media IBP container label.
Device Story
Sterile, single-use, pre-administration filling device; facilitates aseptic transfer of contrast media from Imaging Bulk Package (IBP) to empty, sterile hand syringes. Comprises vented spike and swabable valve. Used in MR suite by trained healthcare professionals. Device does not contact patient. Operation involves spiking IBP container; transferring contrast to syringe. Enables safe, aseptic contrast handling; reduces contamination risk. Discarded after container depletion, disconnection, or expiration of contrast use time.
Clinical Evidence
No clinical data. Evidence consists of bench testing (chemical integrity, particulate matter per USP <788>) and human factors knowledge-based assessment. Biocompatibility, sterilization, and MR compatibility data were leveraged from the predicate device (K200280).
Indicated for trained healthcare professionals to transfer Ambelvist® (gadoquatrane) or Gadavist® (gadobutrol) contrast media from Imaging Bulk Packages (IBP) to empty, sterile hand syringes for MR procedures.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
MEDRAD® Imaging Bulk Package Transfer Spike (K200280)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 12, 2026
Bayer Medical Care, Inc.
Nishanth Reddy Deva
Associate Director, Global Regulatory Strategy Devices
1 Bayer Dr.
Indianola, Pennsylvania 15051
Re: K253688
Trade/Device Name: MEDRAD® Imaging Bulk Package Transfer Spike
Regulation Number: 21 CFR 880.5440
Regulation Name: Intravascular Administration Set
Regulatory Class: Class II
Product Code: PQH
Dated: May 26, 2026
Received: May 26, 2026
Dear Nishanth Reddy Deva:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253688 - Nishanth Reddy Deva
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253688 - Nishanth Reddy Deva
Page 3
Sincerely,
DAVID WOLLOSCHECK
-S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253688
Device Name
MEDRAD® Imaging Bulk Package Transfer Spike
Indications for Use (Describe)
The MEDRAD® Imaging Bulk Package (IBP) Transfer Spike is indicated for the transfer of the following contrast media:
- Ambelvist® (gadoquatrane) injection in IBP
- Gadavist® (gadobutrol) injection in IBP
The Transfer Spike is used to transfer contrast media to empty, sterile hand syringes for MR procedures.
Discard the Transfer Spike after one of the following conditions has occurred first: (1) the Transfer Spike has been disconnected from the contrast media container; (2) the contrast media container has been depleted; or (3) after contrast media use time has elapsed since the container was penetrated.
For a specific contrast media use time, refer to the contrast media prescribing information or the contrast media IBP container label.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# MEDRAD® IMAGING BULK PACKAGE TRANSFER SPIKE 510(K) SUMMARY
K253688 – 510(k) Summary
Bayer Medical Care, Inc. MEDRAD® Imaging Bulk Package Transfer Spike
The Summary is prepared in conformance with 21 CFR 807.92
## I. SUBMITTER
Bayer Medical Care, Inc.
1 Bayer Drive
Indianola, Pa 15051-0780
US Correspondent Contact: Nishanth Reddy Deva
Bayer Medical Care, Inc.
1 Bayer Drive Indianola PA 15051 United States
Email: nishanth.reddy@bayer.com
Phone: +1 412-915-9561
US Alternate Contact: Hortense Allison
Bayer Medical Care, Inc.
1 Bayer Drive Indianola PA 15051 United States
Email: Hortense.allison@bayer.com
Phone: +1 412-935-7777
Date Prepared: June 11, 2026
## II. DEVICE
Name of Device: MEDRAD® Imaging Bulk Package Transfer Spike
Common Name: Contrast Media Transfer Set
Classification Name: Intravascular Administration Set
Classification Regulation: 21 CFR 880.5440
Regulatory Class: Class II
Product Code: PQH
## III. PREDICATE DEVICE
Name of Device: MEDRAD® Imaging Bulk Package Transfer Spike, K200280
Common Name: Contrast Media Transfer Set
Classification Name: Intravascular Administration Set
Classification Regulation: 21 CFR 880.5440
Regulatory Class: Class II
Product Code: PQH
This predicate has not been subject to a design-related recall.
PAGE 1 OF 7
PROPRIETARY TO BAYER
{5}
# MEDRAD® IMAGING BULK PACKAGE TRANSFER SPIKE 510(K) SUMMARY
# IV. DEVICE DESCRIPTION
The MEDRAD® Imaging Bulk Package Transfer Spike ("Transfer Spike") is a sterile, single-use, pre-administration filling device designed for the aseptic transfer of contrast media from an Imaging Bulk Package (IBP) to sterile, empty hand syringes for use in magnetic resonance (MR) imaging procedures. The device is intended to be used by individuals with training and experience in diagnostic studies. It is intended to spike one bulk package of Gadavist® (gadobutrol) injection contrast media or Ambelvist® (gadoquatrane) injection contrast media. Each imaging bulk transfer spike consists of a spike and a swabable valve.
Transfer Spike is designed for single use with one IBP container; the Transfer Spike must not be reused or applied across multiple containers. It does not come in direct contact with patients, and each unit is provided sterile, individually packaged, and is not intended for reprocessing or resterilization.
# V. INDICATIONS FOR USE
The MEDRAD® Imaging Bulk Package (IBP) Transfer Spike is indicated for the transfer of the following contrast media:
- Ambelvist® (gadoquatrane) injection in IBP
- Gadavist® (gadobutrol) injection in IBP
The Transfer Spike is used to transfer contrast media to empty, sterile hand syringes for MR procedures.
Discard the Transfer Spike after one of the following conditions has occurred first: (1) the Transfer Spike has been disconnected from the contrast media container; (2) the contrast media container has been depleted; or (3) after contrast media use time has elapsed since the container was penetrated.
For a specific contrast media use time, refer to the contrast media prescribing information or the contrast media IBP container label.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A table comparing key features of the subject device (Transfer Spike) and predicate device (Transfer Spike) is provided below.
PAGE 2 OF 7
PROPRIETARY TO BAYER
{6}
## MEDRAD® IMAGING BULK PACKAGE TRANSFER SPIKE 510(K) SUMMARY
Table 1 – Key Feature Comparison
| Item | Predicate Device (K200280): MEDRAD® Imaging Bulk Package (IBP) Transfer Spike | Subject Device (K253688): MEDRAD® Imaging Bulk Package (IBP) Transfer Spike | Comparison |
| --- | --- | --- | --- |
| **Regulatory Classification** | | | |
| Class | II | II | Same |
| FDA Regulation Number | 880.5440 | 880.5440 | Same |
| Classification Product Code | PQH (Intravascular administration set) | PQH (Intravascular administration set) | Same |
| **Labeling** | | | |
| Indications for Use | The MEDRAD® Imaging Bulk Package Transfer Spike (Transfer Spike) is indicated for the transfer of Gadavist® (gadobutrol) Injection contrast media as supplied in an approved Imaging Bulk Package presentation (30 mL or 65 mL) to empty, sterile hand syringes and/or empty, sterile syringes on single-use only syringe-based contrast power injection systems indicated for the controlled, automatic venous administration of contrast agents for MR procedures. The Transfer Spike is to be discarded after one of the following conditions has occurred first: the contrast media container has been depleted, the Transfer Spike has been disconnected from the contrast vial, or after 24 hours has elapsed since the container was penetrated. | The MEDRAD® Imaging Bulk Package (IBP) Transfer Spike is indicated for the transfer of the following contrast media: - Ambelvist® (gadoquatrane) injection in IBP - Gadavist® (gadobutrol) injection in IBP The Transfer Spike is used to transfer contrast media to empty, sterile hand syringes for MR procedures. Discard the Transfer Spike after one of the following conditions has occurred first: (1) the Transfer Spike has been disconnected from the contrast media container; (2) the contrast media container has been depleted; or (3) after contrast media use time has elapsed since the container was penetrated. For a specific contrast media use time, refer to the contrast media prescribing information | Same intended use, as the device continues to serve for aseptic transfer of contrast media from an IBP to sterile, empty syringes for use in Magnetic resonance (MR) imaging procedures. The expansion of the transfer spike to include Ambelvist® (gadoquatrane) and the removal of contrast use time from the indications for use statement do not introduce new intended use. |
PAGE 3 OF 7
PROPRIETARY TO BAYER
{7}
## MEDRAD® IMAGING BULK PACKAGE TRANSFER SPIKE 510(K) SUMMARY
| | | or the contrast media IBP container label. | The subject device has been modified to remove reference to use with syringe-based contrast power injection systems, limiting use to hand syringes only from the indications for use do not introduce a new intended use. |
| --- | --- | --- | --- |
| Intended User | Trained healthcare professionals | Trained healthcare professionals | Same |
| **Construction** | | | |
| Spike | Vented Spike | Vented Spike | Same |
| Valve | Swabable | Swabable | Same |
| **Materials** | | | |
| Vented Spike | ABS | ABS | Same |
| Check Valve Air Filter | Spike Filter Housing: Polystyrene Spike Filter: Glass Fiber | Spike Filter Housing: Polystyrene Spike Filter: Glass Fiber | Same |
| Swabbable Valve | Medical Grade Polycarbonate, Silicone | Medical Grade Polycarbonate, Silicone | Same |
| Spike Guard | Low Density Polyethylene | Low Density Polyethylene | Same |
| Adhesive | Cyclohexane or 50/50 Cyclohexane/ MEK solvent and Loctite 3341 UV Core | Cyclohexane or 50/50 Cyclohexane/ MEK solvent and Loctite 3341 UV Core | Same |
| **Packaging** | | | |
| Type | Individually packaged in a Tyvek pouch | Individually packaged in a Tyvek pouch | Same |
| Shelf Life | 3 years at release | 3 years at release | Same |
| **Biological** | | | |
| Biocompatibility | Compliant to applicable sections of ISO/AAMI 10993-1:2018 | Compliant to applicable sections of ISO/AAMI 10993-1:2018 | Same |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Same |
| Sterilization | Radiation (Gamma) | Radiation (Gamma) | Same |
| Sterilization Assurance Level | 10^{-6} | 10^{-6} | Same |
| **Performance** | | | |
| Spike Valve Separation Torque | 6 in/lbs | 6 in/lbs | Same |
| Fill (Load) Rate | 10mL/sec for manual fill 4mL/sec for autoload | 10mL/sec for manual fill 4mL/sec for autoload | Same |
| Use Environment | Ambient (MR Suite) | Ambient (MR Suite) | Same |
PAGE 4 OF 7
PROPRIETARY TO BAYER
{8}
## MEDRAD® IMAGING BULK PACKAGE TRANSFER SPIKE 510(K) SUMMARY
| Transfer Spike Use Life | 24 hours | 24 hours | Same |
| --- | --- | --- | --- |
| Transfer Spike Assembly Use Time | 24 hours for Gadavist® (Gadobutrol) injection in IBP | 6 hours for Ambelvist® (gadoquatrane) injection in IBP 24 hours for Gadavist® (Gadobutrol) injection in IBP | Different and substantially equivalent. This difference does not change the intended use of the device. The safety and effectiveness of the MEDRAD® Imaging Bulk Package Transfer Spike has been confirmed through labeling verification, Human Factors Knowledge assessment, and chemical integrity testing to show the transfer spike can be used with Ambelvist® (gadoquatrane) injection in IBP with no additional risk. |
## VII. NON-CLINICAL TESTING
The following performance data has been provided in support of the substantial equivalence determination.
### A. Performance Testing – Bench
Performance bench testing was conducted, where applicable, to support the use of the Imaging Bulk Package Transfer Spike with Ambelvist® (gadoquatrane) injection in Imaging Bulk Package (IBP) presentations. Chemical Integrity was performed to demonstrate there are no increased concerns of safety when using the Transfer Spike with Ambelvist contrast media in an IBP presentation. Particulate testing was completed according to methodology prescribed in USP <788> Particulate Matter in Injections. The Transfer Spike disposables met all acceptance criteria and demonstrated that the device performs adequately with Ambelvist® IBP Injection.
PAGE 5 OF 7
PROPRIETARY TO BAYER
{9}
# MEDRAD® IMAGING BULK PACKAGE TRANSFER SPIKE 510(K) SUMMARY
All other performance bench testing, including functional performance, was leveraged from the predicate device (K200280), as no changes were made to the device design, materials, or mechanism of action.
# B. Biocompatibility and Sterilization.
Biocompatibility testing and sterilization validation were leveraged from the predicate device (K200280). The subject device utilizes the same materials, patient-contacting components, and sterilization methods as the predicate. As no design, material, or processing changes were introduced, previously completed biocompatibility evaluations and sterilization validations remain applicable and sufficient to support the changes in this submission.
- ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
- ISO 11737-1:2018+AMD1:2021 Sterilization of health care products – Microbiological methods – Part 1: Determination of a population of microorganisms on product
- ISO 11737-2:2019 Sterilization of medical devices – Microbiological methods – Part 2: Tests of sterility performed in the definition validation and maintenance of a sterilization process
- ISO 11137-1: 2006+AMD2:2018 Sterilization of health care products - Radiation - Part 1: Requirements for development validation and routine control of a sterilization process for a medical device
- ISO 11137-2:2013+AMD1:2022 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
# C. MR Compatibility
MR compatibility information was leveraged from the predicate device (K200280). Transfer Spike is composed of materials that are non-magnetic, non-metallic, and non-conductive. Based on its material composition and design, the device does not introduce MR-related hazards such as magnetic field interactions, torque, or RF-induced heating, and does not alter the interaction of the system with the MR environment. Hence, there is no impact of Transfer Spike MR safety in MR environment.
The subject device does not include any changes in materials, design, or configuration that would impact MR safety compared to the previously cleared device. Therefore, no additional MR-specific testing was conducted, and the previously established MR compatibility conclusions remain applicable.
# D. Human Factors
A Human Factors knowledge-based assessment was conducted to evaluate the proposed labeling modifications associated with this submission. Human Factors usability validation previously completed for the predicate device (K200280) was leveraged, as the device user interface, use environment, and primary operating tasks remain unchanged. The assessment was performed in accordance with the following standards and guidance:
- IEC 62366-1 Edition 1.1 2020-06 Consolidated version. Medical Devices – Part 1: Application of Usability engineering to medical devices.
- Applying Human Factors and Usability Engineering to Medical Devices (Guidance for Industry and FDA Staff), February 2016.
PAGE 6 OF 7
PROPRIETARY TO BAYER
{10}
# MEDRAD® IMAGING BULK PACKAGE TRANSFER SPIKE 510(K) SUMMARY
Based on this assessment, the proposed labeling modifications for the Imaging Bulk Package Transfer Spike do not introduce new use-related risks for the intended users, uses, or use environment.
## Summary
In summary, the non-clinical testing and verification and validation activities demonstrate that the changes proposed in this submission do not affect the safety or effectiveness of the Imaging Bulk Package Transfer Spike. Testing leveraged from the predicate device (K200280), in combination with targeted performance testing and Human Factors knowledge-based assessment, supports the conclusion that the subject device is as safe and effective as the currently cleared Imaging Bulk Package Transfer Spike (K200280). No new non-clinical testing was conducted, as the changes in this submission are limited to labeling updates and the expansion of the indications to include new contrast media.
## VIII. CONCLUSION
Bayer considers the subject device, Imaging Bulk Package Transfer Spike to be substantially equivalent to the predicate device, Imaging Bulk Package Transfer Spike, K200280. This conclusion is based upon the devices having the same intended use and same technological characteristics. While there are changes to the Transfer Spike labeling for the expanded indications to allow use of Ambelvist, these differences do not raise new questions of safety or effectiveness. Imaging Bulk Package Transfer Spike has demonstrated the ability to perform within the specified parameters and operate as intended by the users of the device. As a result, its performance is deemed acceptable and substantially equivalent to the predicate device.
---PAGE 7 OF 7
PROPRIETARY TO BAYER
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.