Tri-staple 2.0™ Reloads; Endo GIA™ Reloads with Tri-Staple™ Technology; Endo GIA™ Gray Articulating Reloads; Signia™ Small Diameter Reloads
K253657 · Covidien (Part of Medtronic) · GDW · Apr 10, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253657
Device Name
Tri-staple 2.0™ Reloads; Endo GIA™ Reloads with Tri-Staple™ Technology; Endo GIA™ Gray Articulating Reloads; Signia™ Small Diameter Reloads
Applicant
Covidien (Part of Medtronic)
Product Code
GDW · General, Plastic Surgery
Decision Date
Apr 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K253657 · Apr 10, 2026
Tri-staple 2.0™ Reloads; Endo GIA™ Reloads with Tri-Staple™ Technology; Endo GIA™ Gray Articulating Reloads; Signia™ Small Diameter Reloads
Covidien (Part of Medtronic)
Post-market clinical follow-up data; Published clinical literature
The sponsor utilized RWE, including post-market clinical follow-up and published literature, to demonstrate the safety and effectiveness of the devices in urologic procedures (e.g., nephrectomy, prostatectomy, cystectomies) to support a new indication for use.
The Tri-Staple™ 2.0 reloads are indicated for use in abdominal, urologic, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas. The Tri-Staple™ 2.0 curved tip reloads can be used to blunt dissect or separate target tissue from other tissue. The Endo GIA™ Ultra universal short, Endo GIA™ Ultra universal and Endo GIA™ Ultra universal XL staplers and Endo GIA™ reloads with Tri-Staple™ technology are indicated for use in abdominal, urologic, gynecologic, pediatric, and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas. The Endo GIA™ Ultra universal short, Endo GIA™ Ultra universal and Endo GIA™ Ultra universal XL staplers when used with the Endo GIA™ curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue. The Signia™ small diameter reloads are indicated for use in abdominal, urologic, gynecologic, pediatric and thoracic surgery for resection and transection of vasculature using gray reloads and thin tissue and vasculature using white reloads.
Device Story
Single-use surgical stapler reloads; titanium staples placed in staggered rows; simultaneously divides tissue; curved tip variants allow blunt dissection/tissue manipulation. Used in OR by surgeons; compatible with Endo GIA, iDrive, and Signia platforms. Provides mechanical tissue closure and transection; facilitates resection of organs/vasculature. Benefits include reliable hemostasis and tissue approximation during minimally invasive or open procedures.
Clinical Evidence
No new bench testing required due to lack of design changes. Evidence for urologic indication supported by combination of real-world evidence, post-market clinical follow-up, and published clinical literature demonstrating safety and effectiveness in procedures including nephrectomy, prostatectomy, and cystectomies.
Technological Characteristics
Titanium staples; staggered row configuration (2 or 3 rows); articulating and non-articulating configurations; curved tip options for dissection; compatible with powered and manual stapler platforms (Endo GIA, iDrive, Signia).
Indications for Use
Indicated for patients undergoing abdominal, urologic, gynecologic, pediatric, or thoracic surgery requiring resection, transection, or anastomosis. Includes specific use for liver, hepatic vasculature, biliary structures, pancreas, and general vasculature/thin tissue.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
ECHELON™ 3000 45mm Standard Stapler (ECH45S), ECHELON™ 3000 45mm Long Stapler (ECH45L), ECHELON™ 3000 60mm Compact (K241629)
Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip, and Endopath Echelon™ Vascular White Reloads for Advanced (K141952)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 10, 2026
Covidien (Part of Medtronic)
Fei Li
Principal Regulatory Affairs Specialist
60 Middletown Ave.
North Haven, Connecticut 06473
Re: K253657
Trade/Device Name: Tri-staple 2.0™ Reloads; Endo GIA™ Reloads with Tri-Staple™ Technology; Endo GIA™ Gray Articulating Reloads; Signia™ Small Diameter Reloads
Regulation Number: 21 CFR 878.4750
Regulation Name: Implantable Staple
Regulatory Class: Class II
Product Codes: GDW, GAG
Dated: March 10, 2026
Received: March 10, 2026
Dear Fei Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253657 - Fei Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253657 - Fei Li
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**TEK N. LAMICHHANE -S**
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253657 | ? |
| Please provide the device trade name(s). | | ? |
| Tri-staple 2.0™ Reloads;
Endo GIA™ Reloads with Tri-Staple™ Technology;
Endo GIA™ Gray Articulating Reloads;
Signia™ Small Diameter Reloads | | |
| Please provide your Indications for Use below. | | ? |
| The Tri-Staple™ 2.0 reloads are indicated for use in abdominal, urologic, gynecologic, pediatric and
thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and
resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of
pancreas.
The Tri-Staple™ 2.0 curved tip reloads can be used to blunt dissect or separate target tissue from other
tissue.
The Endo GIA™ Ultra universal short, Endo GIA™ Ultra universal and Endo GIA™ Ultra universal XL
staplers and Endo GIA™ reloads with Tri-Staple™ technology are indicated for use in abdominal, urologic,
gynecologic, pediatric, and thoracic surgery for resection, transection and creation of anastomosis. It may
be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for
transection and resection of pancreas.
The Endo GIA™ Ultra universal short, Endo GIA™ Ultra universal and Endo GIA™ Ultra universal XL
staplers when used with the Endo GIA™ curved tip single use reloads can be used to blunt dissect or
separate target tissue from other tissue.
The Signia™ small diameter reloads are indicated for use in abdominal, urologic, gynecologic, pediatric and
thoracic surgery for resection and transection of vasculature using gray reloads and thin tissue and
vasculature using white reloads. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253657
# 510(k) Summary
DATE PREPARED:
April 10, 2026
SUBMITTER:
Covidien LLC
60 Middletown Avenue
North Haven, CT 06473 USA
CONTACT PERSON:
Fei Li
Principal Regulatory Affairs Specialist
Telephone: (213) 618-6681
fei.li4@medtronic.com
ALTERNATE CONTACT PERSON:
Sanja Jahr
Regulatory Affairs Manager
sanja.jahr@medtronic.com
IDENTIFICATION OF DEVICE:
Proprietary/Trade Name:
Tri-staple 2.0™ Reloads
Endo GIA™ Reloads with Tri-Staple™ Technology
Endo GIA™ Gray Articulating Reloads
Signia™ Small Diameter Reloads
Classification Name:
Staples, Implantable
Regulation Number:
21 CFR 878.4750, 21 CFR 878.4740
Device Class:
Class II
Product Code:
GDW, GAG
Review Panel:
General and Plastic Surgery
Common Name:
Implantable staple
PREDICATE DEVICES:
| | Primary Predicate 1 | Primary Predicate 2 | Primary Predicate 3 | Secondary Predicate 1 | Secondary Predicate 2 |
| --- | --- | --- | --- | --- | --- |
| 510(k) Number | K202864 | K111825 | K222641 | K241629 | K141952 |
| Proprietary/Trade Name | Tri-Staple™ 2.0 Reloads | Endo GIA™ Staplers, DST Series™ GIA Staplers, DST Series™ TA Staplers | Signia™ Small Diameter Reloads | ECHELON™ 3000 45mm Standard Stapler (ECH45S)
ECHELON™ 3000 45mm Long Stapler (ECH45L)
ECHELON™ 3000 60mm Compact | Echelon Flex™ Powered Vascular Stapler with Advanced Placement Tip, and Endopath Echelon™ Vascular White Reloads for Advanced |
Traditional 510(k)
Covidien
April 2026
{5}
K253657
| Classification Name | Staples, Implantable | Staple, Implantable | Staple, Implantable | Stapler, Surgical | Staple, Implantable |
| --- | --- | --- | --- | --- | --- |
| Regulation Number | 21 CFR 878.4750 | 21 CFR 878.4750 | 21 CFR 878.4750 / 21 CFR 878.4740 | 21 CFR 878.4740 | 21 CFR 878.4750 |
| Device Class | Class II | Class II | Class II | Class II | Class II |
| Product Code | GDW | GDW | GDW, GAG | GAG | GDW |
| Review Panel | General and Plastic Surgery | General and Plastic Surgery | General and Plastic Surgery | General and Plastic Surgery | General and Plastic Surgery |
| Common Name | Implantable staples | Implantable staples | Implantable Staples | Surgical Stapler | Surgical Stapler with Implantable Staples |
# DEVICE DESCRIPTION:
## Tri-Staple™ 2.0 Reloads
The Tri-Staple™ 2.0 reloads and Tri-Staple™ 2.0 curved tip reloads place staggered rows of titanium staples and simultaneously divides the tissue so that three staggered rows of staples are placed on either side of the cut line. The size of the staples is determined by the selection of the single use reload:
The Tri-Staple™ 2.0 reload is available in articulating 30 mm, 45 mm and 60 mm length, the Tri-Staple™ 2.0 curved tip reload is available in articulating 30 mm, 45 mm and 60 mm lengths and the Tri-Staple™ 2.0 black reload is available in articulating 45 mm and 60 mm lengths.
## Tri-Staple™ 2.0 Extra Thin/Vascular Reloads:
Gray 30 mm reload, gray 30 mm curved tip reload, gray 45 mm reload and gray 45 mm curved tip reload – three rows of 2.0 mm titanium staples on either side of the cut line.
## Tri-Staple™ 2.0 Vascular/ Medium Reloads:
Tan 30 mm reload, tan 30 mm curved tip reload, tan 45 mm reload, tan 45 mm curved tip reload, tan 60 mm reload and tan 60 mm curved tip reload - three height progressive rows of 2.0 mm, 2.5 mm, 3.0 mm titanium staples on either side of the cut line.
## Tri-Staple™ 2.0 Medium/ Thick Reloads:
Purple 30 mm reload, purple 45 mm reload, purple 45 mm curved tip reload, purple 60 mm reload and purple 60 mm curved tip reload - three height progressive rows of 3.0 mm, 3.5 mm and 4.0 mm titanium staples on either side of the cut line.
## Tri-Staple™ 2.0 Extra Thick Reload:
Traditional 510(k)
Covidien
April 2026
{6}
# K253657
Black 45 mm reload and black 60 mm reload - three height progressive rows of 4.0 mm, 4.5 mm and 5.0 mm titanium staples on either side of the cut line.
The curved tip on the distal-end of the curved tip reload can be used to dissect and manipulate tissue/vessels when locating target tissue for subsequent firing and placement of staples.
The Tri-Staple™ 2.0 reloads and Tri-Staple™ 2.0 curved tip reloads can be used with the Endo GIA™ Universal, Endo GIA™ Ultra, iDrive™, and Signia™ platforms.
## Endo GIA™ Reloads with Tri-Staple™ Technology, Endo GIA™ Gray Articulating Reloads
The Endo GIA™ Reloads with Tri-Staple™ Technology and Endo GIA™ Gray Articulating Reloads place staggered rows of titanium staples and simultaneously divide the tissue so that three staggered rows of staples are placed on either side of the cut line. The size of the staples is determined by the selection of the single use reload:
Endo GIA™ single use articulating Extra Thin/Vascular reload:
Gray reload and gray curved tip reload - three rows of 2.0 mm titanium staples on either side of the cut line.
Endo GIA™ single use articulating Vascular/Medium reload with Tri-Staple™ technology:
Tan reload and tan curved tip reload - three height progressive rows of 2.0 mm, 2.5 mm, 3.0 mm titanium staples on either side of the cut line.
Endo GIA™ single use articulating Medium/Thick reload with Tri-Staple™ technology:
Purple reload and purple curved tip reload - three height progressive rows of 3.0 mm, 3.5 mm and 4.0 mm titanium staples on either side of the cut line.
Endo GIA™ single use articulating Extra Thick reloads with Tri-Staple™ technology:
Black reload - three height progressive rows of 4.0 mm, 4.5 mm and 5.0 mm titanium staples on either side of the cut line.
The curved tip on the distal-end of the curved tip reload can be used to dissect and manipulate tissue/vessels when locating target tissue for subsequent firing and placement of staples.
Endo GIA™ reloads are compatible with the following Covidien staplers: Endo GIA™ universal, Endo GIA™ Ultra, iDrive™, and Signia™ platforms.
## Signia™ Small Diameter Reloads
The Signia™ small diameter reloads place staggered rows of titanium staples and simultaneously divide the tissue so that two staggered rows of staples are placed on either side of the cut line. The size of the staples is determined by the selection of the single use reload.
The Signia™ small diameter reloads place two rows of titanium staples on either side of the cut line and are available in multiple articulating configurations with following features:
- Open Staple Height: 2.0 mm (gray cartridge) and 2.5 mm (white cartridge)
- Cartridge Length: 30 mm and 45 mm
- Anvil Tip: Curved tip
- Shaft Length: Short (15 cm) and Long (24 cm)
- Shaft Diameter: 8 mm
The curved tip on the distal-end of the curved tip reload can be used to aid in positioning the reload around target tissue/vessels for subsequent firing and placement of staples.
The Signia™ small diameter reloads can be used with the Endo GIA™ Universal, Endo GIA™ Ultra, iDrive™, and Signia™ platforms.
Traditional 510(k)
Covidien
April 2026
{7}
K253657
# INDICATIONS FOR USE:
## Tri-Staple 2.0™ Reloads
The Tri-Staple™ 2.0 reloads are indicated for use in abdominal, urologic, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas.
The Tri-Staple™ 2.0 curved tip reloads can be used to blunt dissect or separate target tissue from other tissue.
## Endo GIA™ Reloads with Tri-Staple™ Technology, Endo GIA™ Gray Articulating Reloads
The Endo GIA™ Ultra universal short, Endo GIA™ Ultra universal and Endo GIA™ Ultra universal XL staplers and Endo GIA™ reloads with Tri-Staple™ technology are indicated for use in abdominal, urologic, gynecologic, pediatric, and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas.
The Endo GIA™ Ultra universal short, Endo GIA™ Ultra universal and Endo GIA™ Ultra universal XL staplers when used with the Endo GIA™ curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue.
## Signia™ Small Diameter Reloads
The Signia™ small diameter reloads are indicated for use in abdominal, urologic, gynecologic, pediatric and thoracic surgery for resection and transection of vasculature using gray reloads and thin tissue and vasculature using white reloads.
# TECHNOLOGICAL CHARACTERISTICS:
With the exception of the newly added urologic indication, the subject devices (Tri-Staple™ 2.0 Reloads, Endo GIA™ Reloads with Tri-Staple™ Technology, Endo GIA™ Gray Articulating Reloads, Signia™ Small Diameter Reloads) are identical to the respective primary predicate devices (K202864, K111825, K222641) in regard to intended use, material, design, and operational principles.
# SUBSTANTIAL EQUIVALENCE COMPARISON:
The subject devices, primary predicate devices and secondary predicate devices share the same intended use. Regarding the indications for use, the subject devices are IDENTICAL to primary predicates (K202864, K111825, K222641), except for the urology indication. Both the subject devices and the secondary predicates (K241629, K141952) have urology indications.
# PERFORMANCE DATA:
Except for the newly added urologic indication, there are no changes to the subject devices regarding the intended use, material, design, and operational principles. Therefore, the subject devices remain substantially equivalent to the predicate devices and no testing is required.
A combination of real-world evidence, post market clinical follow-up, and published clinical literature is provided to support the addition of the urology indication. This data demonstrates the safe and effective use of the subject devices in urologic procedures such as nephrectomy, prostatectomy, and cystectomies, and captures performance metrics across diverse patient populations and surgical settings.
# CONCLUSION:
Based upon the supporting data summarized above, the subject devices (Tri-staple 2.0™ Reloads, Endo GIA™ Reloads with Tri-Staple™ Technology, Endo GIA™ Gray Articulating Reloads, Signia™ Small Diameter Reloads) are substantially equivalent to the predicate devices (K202864, K111825, K222641, K241629, K141952).
Traditional 510(k)
Covidien
April 2026
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.