OsteoCool™ RF Ablation System; OsteoCool™ 2.0 RF Ablation System
K253656 · Medtronic Sofamor Danek USA, Inc. · GEI · Jun 5, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253656
Device Name
OsteoCool™ RF Ablation System; OsteoCool™ 2.0 RF Ablation System
Applicant
Medtronic Sofamor Danek USA, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 5, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K253656 · Jun 5, 2026
OsteoCool™ RF Ablation System; OsteoCool™ 2.0 RF Ablation System
Medtronic Sofamor Danek USA, Inc.
Systematic literature review; Retrospective chart review; Meta-analysis
The sponsor conducted a systematic literature review, retrospective chart review, and meta-analysis to assess the safety and performance of the OsteoCool system for treating bone lesions in patients with multiple myeloma, supporting a labeling change to remove a contraindication/warning.
Multiple myeloma; Bone lesion ablation; Retrospective chart review; Systematic literature review; Labeling change
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Systematic literature review, retrospective chart review, and meta-analysis; Retrospective chart review and meta-analysis of literature
Patients with multiple myeloma undergoing RFA for bone lesions
Not applicable for this study
Safety and performance of RFA for multiple myeloma bone lesions
Indications for Use
The OsteoCool™ 2.0 RF Ablation System is indicated for: • Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body. • Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy. • Ablation of benign bone tumors such as osteoid osteoma.
Device Story
OsteoCool™ 2.0 RF Ablation System is a radiofrequency (RF) ablation device used for bone tissue coagulation and necrosis. System components include an RF generator, peristaltic pump, connector hub, RF ablation probes, and independent thermocouples. During percutaneous procedures, physicians deliver bipolar RF energy to target bone lesions; internal water cooling circulates through probes to manage thermal ablation zones. Thermocouples monitor temperatures around the ablation site. The system provides controlled thermal coagulation necrosis to treat metastatic lesions or benign tumors. Output is used by physicians to guide palliative pain relief and tumor ablation. The device is intended for use in surgical settings.
Clinical Evidence
Evidence includes a systematic literature review, retrospective chart review, and meta-analysis evaluating the safety and performance of RFA for bone lesions, specifically focusing on multiple myeloma patients. No new bench testing was required as the hardware remains identical to the predicate.
Technological Characteristics
Energy: Radiofrequency (465.1 KHz). Power: 80W total (20W/channel). Materials: Stainless steel electrodes, polyimide dielectric. Cooling: Sterile water (30-40 mL/min). Dimensions: 17G probes, 16cm length, 7-20mm active tips. Connectivity: 4-channel generator with peristaltic pump and thermocouple monitoring. Sterilization: EtO, single-use disposables.
Indications for Use
Indicated for patients requiring palliative treatment of metastatic malignant lesions in a vertebral body, patients with metastatic bone lesions who failed or are not candidates for standard therapy, and patients with benign bone tumors (e.g., osteoid osteoma).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 5, 2026
Medtronic Sofamor Danek USA, Inc.
Jeff Sprague
Sr. Principal, Regulatory Affairs
1800 Pyramid Place
Memphis, Tennessee 38132
Re: K253656
Trade/Device Name: OsteoCool™ 2.0 RF Ablation System
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: May 7, 2026
Received: May 7, 2026
Dear Jeff Sprague:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253656 - Jeff Sprague
Page 2
(https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253656 - Jeff Sprague
Page 3
Sincerely,
**JESSE MUIR - S** Digitally signed by JESSE MUIR-S
Date: 2026.06.05 15:23:58
-04'00'
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253656 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| OsteoCool™ 2.0 RF Ablation System | | |
| Please provide your Indications for Use below. | | ? |
| • Palliative treatment in spinal procedures by ablation of metastatic, malignant lesions in a vertebral body. • Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy. • Ablation of benign bone tumors such as osteoid osteoma. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
Page 1 of 1
{4}
OsteoCool™ RF Ablation System
Page 1 of 8
# **K253656**
**510(k) Summary**
**Medtronic Sofamor Danek USA, Inc.**
| **Submitter** | Medtronic Sofamor Danek USA, Inc. 1800 Pyramid Place Memphis, Tennessee 38132 |
| --- | --- |
| **Contact** | Jeff Sprague Sr. Principal, Regulatory Affairs Direct Telephone: 901-498-2697 Email: jeffrey.a.sprague@medtronic.com |
| **Date Prepared** | November 3, 2025 |
| **Common Name** | OsteoCool™ 2.0 RF Ablation System |
| **Regulatory Class Regulation Number Regulation Name and Device Product Classification Code** | Class II 21 CFR 878.4400 Electrosurgical cutting and coagulation device and accessories GEI |
| **Predicate Devices** | OsteoCool™ 2.0 RF Ablation System K233830 (S.E. 2/28/2024) – primary predicate The predicate devices have not been subject to a design related recall. |
Medtronic
K253656 - Page 1 of 8
{5}
OsteoCool™ RF Ablation System
Page 2 of 8
| **Description of Device** | This 510(k) submission is for labeling changes to the OsteoCool™ 2.0 RF Ablation System cleared in K233830 (S.E. 2/28/2024). The primary change was removal of a warning for ablation of lesions due to multiple myeloma. The subject OsteoCool™ 2.0 RF Ablation Systems include the following hardware components: 1. OsteoCool™ 2.0 Radiofrequency (RF) Generator 2. OsteoCool™ 2.0 Peristaltic Pump & Pump Cable 3. OsteoCool™ 2.0 Connector Hub 4. OsteoCool™ Cart The OsteoCool™ 2.0 RF Ablation System includes the following commercially available sterile disposables that remain unchanged from the those cleared in K182497 (S.E. 12/17/2018): 5. OsteoCool™ RF Ablation Probe with Tube Kit 6. OsteoCool™ Independent Thermocouple with Introducer Kit: The OsteoCool™ RF Ablation System delivers controlled radiofrequency (RF) energy in a bipolar manner with a cooling mechanism to facilitate RF lesions in target tissue. The RF Generator operates together with the RF Ablation Probe to deliver the RF energy to the target ablation site(s). The Tube Kit is used with the Peristaltic Pump to circulate water internally through the RF Ablation Probe(s) during RF energy delivery. The Pump Cable connects the Peristaltic Pump to the RF Generator, which controls the pump's flow. The Connector Hub connects the RF Ablation Probe(s) and Thermocouple(s) to the RF Generator. The Thermocouple Monitor is used with the Thermocouple Introducer and enables temperature monitoring around the thermal ablation zones during procedures. |
| --- | --- |
Medtronic
K253656 - Page 2 of 8
{6}
OsteoCool™ RF Ablation System
Page 3 of 8
| **Indications for Use:** | The OsteoCool™ 2.0 RF Ablation System is indicated for: • Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body. • Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy. • Ablation of benign bone tumors such as osteoid osteoma. |
| --- | --- |
| **Comparison of Technological Characteristics with the Predicate Device** | The intended use of the subject devices is substantially equivalent to the predicate devices. |
| | The subject and predicate devices are identical in fundamental scientific technology, principles of operation and mechanism of action, and design and technological aspects. |
Medtronic
K253656 - Page 3 of 8
{7}
OsteoCool™ RF Ablation System
Page 4 of 8
| | *OsteoCool™ 2.0 RF Ablation System* **SUBJECT DEVICE** | *OsteoCool™ 2.0 RF Ablation System* **PREDICATE DEVICE** | |
| --- | --- | --- | --- |
| 510(k) Clearance | TBD | K233830 (S.E. 2/28/2024) | |
| Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 | Identical |
| Product Code | GEI | GEI | Identical |
| Product Class | Class II | Class II | Identical |
| Indication for Use | The OsteoCool™ 2.0 RF Ablation System is intended for: • Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body. • Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy. • Ablation of benign bone tumors such as osteoid osteoma. | The OsteoCool™ 2.0 RF Ablation System is intended for: • Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body. • Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy. • Ablation of benign bone tumors such as osteoid osteoma. | Identical |
| User | Physicians trained in the use of OsteoCool RF ablation system | Physicians trained in the use of OsteoCool RF ablation system | Identical |
| Anatomical Site of Use | Bone | Bone | Identical |
| Method of Access | Percutaneous | Percutaneous | Identical |
Medtronic
K253656 - Page 4 of 8
{8}
OsteoCool™ RF Ablation System
Page 5 of 8
| TECHNICAL DETAILS - System | | | |
| --- | --- | --- | --- |
| FEATURE | OsteoCool™ 2.0 RF Ablation System | OsteoCool 2.0 RF Ablation System | |
| | **SUBJECT DEVICE** | **PREDICATE DEVICE** | |
| Energy Type | Radiofrequency | Radiofrequency | Identical |
| Principle of Operation | Radiofrequency energy delivered by operator from generator | Radiofrequency energy delivered by operator from generator | Identical |
| Mechanism of Action | Thermal coagulation necrosis | Thermal coagulation necrosis | Identical |
| Energy Delivery Control Mechanism | Temperature controlled | Temperature controlled | Identical |
| Probe Active Tip Size | 7mm, 10mm, 15mm, 20mm | 7mm, 10mm, 15mm, 20mm | Identical |
| Probe Outer Diameter | 17Gauge | 17Gauge | Identical |
| Probe length | 16 cm | 16 cm | Identical |
| Probe Electrodes Material | Stainless Steel | Stainless Steel | Identical |
| Probe Dielectric Materials | Polyimide | Polyimide | Identical |
| Electrode Sterilization | EtO, single use | EtO, single use | Identical |
| Compatible RF Generator | OsteoCool 2.0 | OsteoCool 2.0 | Identical |
| Number of independent thermocouples | 4 | 4 | Identical |
| Number of probes | 4 | 4 | Identical |
Medtronic
K253656 - Page 5 of 8
{9}
OsteoCool™ RF Ablation System
Page 6 of 8
| TECHNICAL DETAILS - System | | | |
| --- | --- | --- | --- |
| FEATURE | OsteoCool™ 2.0 RF Ablation System SUBJECT DEVICE | OsteoCool 2.0 RF Ablation System PREDICATE DEVICE | |
| Generator Power Output Channels | 4 | 4 | Identical |
| Generator Maximum Output Power | • 80 Watts total • 20 W/channel | • 80 Watts total • 20 W/channel | Identical |
| Generator Maximum Voltage | 224 V_{RMS} | 224 V_{RMS} | Identical |
| Generator Maximum Current | 1.2 A_{RMS} | 1.2 A_{RMS} | Identical |
| Generator Output Frequency | 465.1 KHz | 465.1 KHz | Identical |
Medtronic
K253656 - Page 6 of 8
{10}
OsteoCool™ RF Ablation System
Page 7 of 8
| TECHNICAL DETAILS - System | | | |
| --- | --- | --- | --- |
| FEATURE | OsteoCool™ 2.0 RF Ablation System SUBJECT DEVICE | OsteoCool 2.0 RF Ablation System PREDICATE DEVICE | |
| Default Ablation Temperature | 70C (Gradual) | 70C (Gradual) | Identical |
| | 75C (Accelerated) | 75c (Accelerated) | Identical |
| Electrode cooling flowrate | 40 mL/min (Gradual) | 40 mL/min (Gradual) | Identical |
| | 30 mL/min (Accelerated) | 30 mL/min (Accelerated) | Identical |
| Cooling medium | Sterile water | Sterile water | Identical |
| TECHNICAL DETAILS - System | | | |
| --- | --- | --- | --- |
| FEATURE | OsteoCool™ 2.0 RF Ablation System SUBJECT DEVICE | OsteoCool 2.0 RF Ablation System PREDICATE DEVICE | |
| System components | Pump OC02-100 Hub OC04-100 | Pump OC02-100 Hub OC04-100 | Identical |
| | Single Probe Kit 7 mm -OCP107 Single Probe Kit 10 mm -OCP110 Single Probe Kit 15 mm -OCP115 Single Probe Kit 20 mm -OCP120 Dual Probe Kit 7 mm x2 -OCP207 Dual Probe Kit 10 mm x2 -OCP210 Dual Probe Kit 15 mm x2 -OCP215 Dual Probe Kit 20 mm x2 -OCP220 RF Pump - OC02 ITC 28G - OCN001 | Single Probe Kit 7 mm -OCP107 Single Probe Kit 10 mm -OCP110 Single Probe Kit 15 mm - OCP115 Single Probe Kit 20 mm -OCP120 Dual Probe Kit 7 mm x2 -OCP207 Dual Probe Kit 10 mm x2 -OCP210 Dual Probe Kit 15 mm x2 -OCP215 Dual Probe Kit 20 mm x2 -OCP220 RF Pump - OC02 ITC 28G - OCN001 | Identical |
Medtronic
K253656 - Page 7 of 8
{11}
OsteoCool™ RF Ablation System
Page 8 of 8
| **Performance Data** | To support the labeling changes, a systematic literature review was conducted to evaluate the safety of RFA for patients with multiple myeloma to ensure labeling is aligned with current practice. In addition to the literature review, a retrospective chart review and meta-analysis were conducted to assess all available clinical evidence for the performance and safety of RFA, and specifically the OsteoCool bone tumor ablation system, for the treatment of bone lesions associated with multiple myeloma. |
| --- | --- |
| **Conclusion** | The subject and predicate device are identical in fundamental scientific technology, including principles of operation and mechanism of action. Differences in the labeling between the subject and predicate devices do not raise any new concerns of safety and effectiveness. Therefore, Medtronic believes the subject devices are substantially equivalent to the predicate OsteoCool™ 2.0 RF Ablation System cleared in K233830 (S.E. 2/28/2024). |
Medtronic
K253656 - Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.