K253638 · Spinex, Inc. · SDO · Apr 20, 2026 · Physical Medicine
Device Facts
Record ID
K253638
Device Name
xStep (xStep)
Applicant
Spinex, Inc.
Product Code
SDO · Physical Medicine
Decision Date
Apr 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5851
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The xStep™ is intended to deliver programmed transcutaneous electrical spinal cord stimulation in conjunction with functional task practice in the clinic to improve hand sensation and strength in individuals between 18 and 75 years old that present with a chronic, non-progressive neurological deficit resulting from an incomplete spinal cord injury (C2-C8 inclusive). The xStep™ device is for prescription use in patients ages 18-75 years old in a medical center.
Device Story
xStep delivers transcutaneous electrical spinal cord stimulation to improve hand function in patients with incomplete cervical spinal cord injury. Device consists of stimulator, charger, electrode cables, and hydrogel electrodes. Operation involves placing electrodes on skin near spine; device delivers custom waveform combining burst (therapeutic) frequency and high-frequency carrier wave. Burst frequency activates spinal neurons; carrier wave provides analgesic effect on superficial skin, allowing deeper penetration without pain. Used in clinical settings under professional supervision during functional task practice. Therapy aims to tune spinal networks based on proprioceptive feedback and descending brain signals, inducing neuroplasticity to improve hand strength and sensation. Output is controlled via manual interface; no remote connectivity or data storage. Benefits include improved coordination of muscle activity and potential for functional recovery in patients with chronic paralysis.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical performance testing, including electrical safety (IEC 60601-1, IEC 60601-2-10), EMC (IEC 60601-1-2), software verification/validation (IEC 62304), battery safety (IEC 62133-2), and mechanical/packaging testing.
Indicated for individuals 18-75 years old with chronic, non-progressive neurological deficits resulting from incomplete spinal cord injury (C2-C8 inclusive, ASIA Impairment Scale Grade B, C, or D) to improve hand sensation and strength.
Regulatory Classification
Identification
A transcutaneous electrical spine stimulator to improve skeletal muscle strength and sensation is a device that can be programmed to apply an electrical current via electrodes on a patient's skin over the spine to improve muscle strength and sensation after neurological deficit. The ARCEX System is intended to deliver programmed, transcutaneous electrical spinal cord stimulation in conjunction with functional task practice in the clinic to improve hand sensation and strength in individuals between 18 and 75 years old that present with a chronic, non-progressive neurological deficit resulting from an incomplete spinal cord injury (C2-C8 inclusive).
Special Controls
In combination with the general controls of the FD&C Act, the transcutaneous electrical spine stimulator to improve skeletal muscle strength and sensation is subject to the following special controls:
- (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
- (i) Characterization of the electrical stimulation parameters, including the following: waveforms; output modes; maximum output voltage and maximum output current; pulse duration; frequency; net charge per pulse; maximum phase charge, maximum current density, maximum average current, and maximum average power density:
- Characterization of the impedance monitoring system; and (ii)
- Characterization of electrode performance, including the electrical performance, (iii) adhesive integrity, shelf life, reusability, and current distribution of the electrode surface area.
- (2) Performance data must demonstrate the electromagnetic compatibility, electrical safety and performance, battery safety, and wireless compatibility of the device.
- Software verification, validation and hazard analysis must be performed. (3)
- (4) The patient-contacting components of the device must be demonstrated to be biocompatible.
- (ર) Labeling must include:
- Summaries of electrical stimulation parameters; (i)
- Instructions for user management of the device in the event of adverse effects; (ii)
- A contraindication for patients with active implantable devices or wearable (iii) defibrillators:
- (iv) Information on the typical sensations experienced during treatment;
- Instructions for accurate placement of the device on the patient; and (v)
- Cleaning instructions. (vi)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 20, 2026
Spinex, Inc.
Parag Gad, PhD
Chief Executive Officer
37917 Lavender Commons
Fremont, California 94536
Re: K253638
Trade/Device Name: xStep (xStep)
Regulation Number: 21 CFR 890.5851
Regulation Name: Transcutaneous Electrical Spine Stimulator To Improve Skeletal Muscle Strength And Sensation
Regulatory Class: Class II
Product Code: SDO
Dated: November 12, 2025
Received: November 19, 2025
Dear Parag Gad:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253638 - Parag Gad, PhD
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253638 - Parag Gad, PhD
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
xStep
Page 7 of 52
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253638 | ? |
| Please provide the device trade name(s). | | ? |
| xStep (xStep) | | |
| Please provide your Indications for Use below. | | ? |
| The xStep™ is intended to deliver programmed transcutaneous electrical spinal cord stimulation in
conjunction with functional task practice in the clinic to improve hand sensation and strength in individuals
between 18 and 75 years old that present with a chronic, non-progressive neurological deficit resulting from
an incomplete spinal cord injury (C2-C8 inclusive). The xStep™ device is for prescription use in patients
ages 18-75 years old in a medical center. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
K253638
(per 21 CFR §807.92)
1. Submitter Information
Submission type: Traditional 510(k)
Company Name: SpineX Inc.
Address: 37917 Lavender Commons, Fremont, CA 94536, USA
Local Address: 37917 Lavender Commons, Fremont, CA 94536, USA
Contact Person: Parag Gad, PhD, CEO
Telephone: +1 (408) 203-5061
Email: parag@spinex.co
Date Prepared: April 17, 2026
2. Device Identification
Trade Name: xStep
Common Name: Transcutaneous electrical spine stimulator
Product Code: SDO
Regulatory Class: Class II
Regulation Number: 21 CFR 890.5851
3. Predicate Device
Predicate Manufacturer: Onward
Medical Predicate Device: ARCEX
510(k) Number: DEN240014
4. Device Description
The xStep™ is a medical device intended to deliver programmed transcutaneous electrical spinal cord stimulation in conjunction with functional task practice in the clinic to improve hand sensation and strength in individuals between 18 and 75 years old that present with a chronic, non-progressive neurological deficit resulting from an incomplete spinal cord injury (C2-C8 inclusive). The xStep™ device is for prescription use in patients ages 18-75 years old in a medical center.
The xStep device consists of four components:
- xStep™ Stimulator (Part Number 030-0200)
- Charger (Part Number 030-0251)
- Electrode Cable Assembly (Part Number 030-0257)
- 1 Set Active electrodes (Axelgaard, Model Number CF3200, 1.25" Round)
- 1 Set Return electrodes (Axelgaard, Model Number UF2040, 2" x 4" Rectangle)
Both electrodes are manufactured and have been previously cleared by FDA (K132422) for the same nature of body contact.
- 1 Storage Case (Part Number 030-0366)
{5}
# Mechanism of Action:
Due to paralysis caused by an incomplete cervical spinal cord injury, the brain may either send abnormal motor signals to the spinal networks, or the signals from the brain may not reach the spinal networks; thus resulting in abnormal signals being projected to muscles. Motor dysfunction in individuals with incomplete spinal cord injury results in the inability to control hand muscles due to the aberrant neural connectivity between the brain and spinal cord. Thus, the descending commands from the brain generates abnormal signals among the spinal networks that increases the co-contraction of flexors and extensor muscles. In addition, abnormal proprioceptive signals ascend to activate multiple supraspinal nuclei, thus completing a continuous loop of maladaptive sensory-motor signals resulting in disordered networks that reflect the initial supraspinal pathology.
The electrical stimulation delivered by the xStep device is intended to be delivered transcutaneously, and the active electrodes (K132422) are designed to be placed in direct contact with intact skin, in appropriate locations along or near the spine that activate the nerves in the spinal cord responsible for controlling motor and sensory function of that hand that is impaired from incomplete paralysis due to spinal cord injury.
xStep is built around a custom waveform that consists of 2 unique frequencies, the burst frequency and the carrier frequency.
- The burst or therapeutic frequency activates the neurons in the spinal cord and in turn the targeted muscles.
- The high-frequency carrier wave that creates an analgesic effect on the superficial skin layers allowing the burst frequency to penetrate and reach the deeper neurons in the spinal cord.
The degree of penetration into the spinal cord and the number of neurons activated is based on the intensity of stimulation. However, despite the higher currents needed to activate nerves present at deeper locations, no superficial pain or discomfort is experienced because of the coordinated function of the burst and carrier waves. xStep therapy activates the nerves in the spinal cord resulting in improved coordination of muscle activity and neuroplasticity being induced in the brain and spinal cord leading to improved hand sensation and strength in individuals with incomplete spinal cord injury. It is intended to be used in conjunction with functional task practice in the clinic to improve hand sensation and strength by inducing neuroplasticity in the nervous system in individuals with chronic, non-progressive neurological deficits resulting from incomplete paralysis due to SCI
# Description of how the device functions
xStep is built around a custom waveform that consists of 2 unique frequencies, the burst frequency and the carrier frequency. The burst or therapeutic frequency 'tunes' the spinal neurons and in turn the targeted muscles towards an activation or inhibition pattern based on the proprioceptive (sensory) information received during the functional task being performed and/or the descending brain signals (voluntary commands). In the case of a patient with incomplete cervical spinal cord injury, the tuning of the spinal cord could result in assisting the patient to increase strength in their hands, while the sensory information from the hand informs the spinal cord and brain that the hands are performing the task. This results in the spinal cord (which is now excited due to xStep therapy) triggering activation of the hands muscles allowing the patient to complete the task.
{6}
The second component is a high-frequency carrier wave that creates an analgesic effect on the superficial skin layers allowing the burst frequency to penetrate and reach the deeper neurons in the spinal cord. The degree of penetration into the spinal cord and the number of neurons activated is based on the intensity of stimulation. However, despite the higher currents needed to activate nerves present at deeper locations, no superficial pain or discomfort is experienced because of the coordinated function of the burst and carrier waves. In comparison, off-the-shelf stimulators (TENS, muscle stimulators, etc.) use only a single burst frequency and are unable to activate deeper nerve structures, stimulating only superficial muscles and nerves and NOT the targeted spinal neurons.
## Any novel technology or features
The xStep device does not introduce any new or novel technologies, features, or mechanisms of action when compared to the predicate device, ARC-EX by ONWARD. Both devices utilize non-invasive, targeted transcutaneous electrical stimulation delivered through external hydrogel electrodes to activate spinal neural circuits below the level of injury. The core technological characteristics—such as stimulation waveform, delivery method, electrode configuration, and intended use—are functionally equivalent. The xStep device operates using the same fundamental principles of spinal cord stimulation and shares the same therapeutic objective: to improve hand muscle strength and sensation by inducing neuroplasticity in the nervous system in individuals with chronic, non-progressive neurological deficits resulting from incomplete paralysis due to SCI. As such, the xStep device does not raise new questions of safety or effectiveness relative to the predicate device.
## 5. Intended Use / Indications for Use
### Indications for Use
The xStep™ is intended to deliver programmed transcutaneous electrical spinal cord stimulation in conjunction with functional task practice in the clinic to improve hand sensation and strength in individuals between 18 and 75 years old that present with a chronic, non-progressive neurological deficit resulting from an incomplete spinal cord injury (C2-C8 inclusive). The xStep™ device is for prescription use in patients ages 18-75 years old in a medical center.
### Intended Use
The xStep™ is intended for individuals aged 18-75 old with chronic, non-progressive, incomplete (Grade B, C or D on the American Spinal Injury Association (ASIA) Impairment Scale (AIS)) cervical spinal cord injury (C2-C8 inclusive).
### Patient Population
The xStep™ device is a medical device and is intended to treat patients with chronic, non-progressive neurological deficit resulting from an incomplete spinal cord injury (C2-C8 inclusive). The xStep™ device is for prescription use in patients ages 18-75 years old in a medical center.
{7}
# 6. Substantial Equivalence Discussion
The following comparison demonstrates that the xStep device is substantially equivalent to the predicate device in terms of intended use, indications for use, and technological characteristics. Where technological differences exist, the differences represent parameter values within clinically used ranges and within the range of the predicate. Non-clinical performance testing confirms that these differences do not raise new questions of safety or effectiveness
| Parameter | ARCex | xStep | Comparison |
| --- | --- | --- | --- |
| Waveform | Monophasic or Biphasic | Monophasic and Delayed Biphasic | Identical |
| Biphasic amplitude | 0-250 mA | 0-200 mA | Substantially Equivalent |
| Intra-burst pulse repetition frequency | 10000-20000 Hz | 10000 Hz | Substantially Equivalent |
| Intra-burst pulse width | 50 or 100 μs | 100 μs | Substantially Equivalent |
| Carrier frequency | 5000 or 10000 Hz | 10000 Hz | Substantially Equivalent |
| Frequency | 0.2-100 Hz | 30 Hz | Substantially Equivalent |
| Pulse (burst) width | 0.1-5 ms | 1 ms | Substantially Equivalent |
| Ramp-up duration | 2-60 s | Manual | Substantially Equivalent |
| Program duration | 1-180 min | 60 min | Substantially Equivalent |
| Channels | 4 | 2 | Substantially Equivalent |
| Output current | Rectangular pulses with carrier frequency | Rectangular pulses with carrier frequency | Identical |
| Display | Digital | Digital | Identical |
| Enclosure | ABS plastic | ABS plastic | Identical |
| Control interface | Manual button | Manual button | Identical |
| Electrodes | Axelgaard PALS (K132422) | Axelgaard PALS (K132422) | Identical |
| Power source | Rechargeable battery | Rechargeable battery | Identical |
| EMC standards | IEC 60601-1-2, IEC 60601-4-2 | IEC 60601-1-2, IEC 60601-4-2 | Identical |
| Electrical/mechanical safety | IEC 60601-1, IEC 60601-2-10 | IEC 60601-1, IEC 60601-2-10 | Identical |
| Battery safety | IEC 62133-2:2017, UN38.3, EN IEC 55035 | IEC 62133-2:2017, UN38.3, EN IEC 55035 | Identical |
| Software documentation level | Enhanced | Enhanced | Identical |
The minor differences in design specifications and stimulation parameters do not significantly affect the safety and effectiveness of the xStep System relative to the predicate, and the non-clinical performance testing (Section 7) demonstrates acceptable safety and performance to support substantial equivalence
{8}
7. Performance Data
Non-clinical testing was conducted to support substantial equivalence:
- Electrical Safety and EMC: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ANSI C63.4
- Electrical Stimulation Parameter Characterization: Verification of waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse; maximum phase charge, maximum current density, maximum average current, and maximum average power density, polarity, and rise/fall time
- Software Verification and Validation: IEC 62304, per FDA guidance
- Battery Safety: IEC 62133-2
- Mechanical/Packaging Testing (per IEC 60601-1): Push, impact, and drop testing
- Cybersecurity: Not applicable; the xStep device has no communication capabilities, no remote access, and stores no sensitive health information, per FDA guidance (October 2023)
The xStep device is provided non-sterile with an expected lifetime of 3 years.
Electrodes (K132422) are provided sterile by the manufacturer with a labeled shelf life of 2 years.
8. Conclusion
The xStep™ device has: - The same intended use, - Similar technological characteristics, and - No new questions of safety or effectiveness.
Therefore, xStep is substantially equivalent to the predicate device ARC ex™ (DEN240014).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.