K253633 · Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. · GEX · Jan 24, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253633
Device Name
Holmium Laser Therapeutic Apparatus (HZ-40)
Applicant
Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and stone fragmentation in use in medical specialties including: Urology, Gastroenterology, ENT, Gynecology and general surgery.
Device Story
HZ-40 is a pulsed solid-state Holmium:YAG laser system (2100nm wavelength) for soft tissue surgery and stone fragmentation. System uses flashlamp-pumped laser medium; energy delivered via SMA905-compatible optical fibers. Operated by clinicians in surgical settings (open, laparoscopic, endoscopic). Includes touchscreen interface, single-pedal footswitch, and integrated liquid cooling. Safety features include emergency stop, door interlock, and real-time laser output monitoring. Device enables precise tissue interaction and lithotripsy, facilitating minimally invasive surgical interventions.
Clinical Evidence
Bench testing only. Device demonstrated compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (laser safety), and IEC 60825-1 (laser product safety). Software verified per IEC 62304.
Technological Characteristics
Pulsed Ho:YAG laser; 2100nm wavelength; 40W max average power; 90-2000μs pulse duration; 0.2-6J energy per pulse; 30Hz frequency. Delivery via SMA905-compatible optical fibers. Includes touchscreen, footswitch, liquid cooling. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1. Software lifecycle per IEC 62304.
Indications for Use
Indicated for patients requiring surgical incision, excision, resection, ablation, vaporization, coagulation, haemostasis of soft tissue, or stone fragmentation in Urology, Gastroenterology, ENT, Gynecology, and general surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 24, 2026
Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd.
Zhengzhou Li
General Manager
Rm. 208, Bldg. C, # 3, Juquan Rd.
Huangpu District
Guangzhou, Guangdong 510000
China
Re: K253633
Trade/Device Name: Holmium Laser Therapeutic Apparatus (HZ-40)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: November 14, 2025
Received: November 19, 2025
Dear Zhengzhou Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253633 - Zhengzhou Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253633 - Zhengzhou Li
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU-S
Digitally signed by YAN FU-S
Date: 2026.01.24 09:12:14
-05'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253633 | ? |
| Please provide the device trade name(s). | | ? |
| Holmium Laser Therapeutic Apparatus (HZ-40) | | |
| Please provide your Indications for Use below. | | ? |
| Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and stone fragmentation in use in medical specialties including: Urology, Gastroenterology, ENT, Gynecology and general surgery. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary - K253633
1 Submitter / Contact Information
| Item | Information |
| --- | --- |
| Submitter Name | Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. |
| Address | Room 208, Building C, No. 3 Juquan Road, Huangpu District, Guangzhou, Guangdong, 510000, China. |
| Phone | +86 -20-37396970 |
| Fax | +86 -20-37376979 |
| Contact Person | Zhengzhou Li |
| Email | potent_medical_public@potent-medical.com |
| Date | January 23, 2026 |
2 Subject Device Information
| Item | Information |
| --- | --- |
| Device Trade Name | Holmium Laser Therapeutic Apparatus |
| Models | HZ-40 |
| Common Name | Powered Laser Surgical Instrument |
| Regulation Name | Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology |
| Regulation Number | 21 CFR 878.4810 |
| Regulatory Class | Class II |
| Product Code | GEX |
| Review Panel | General and Plastic Surgery |
3 Predicate Devices
| Type | Device | Manufacturer | 510(k) Number |
| --- | --- | --- | --- |
| Primary Predicate Device | Litho DK30 | Quanta System SPA | K141403 |
| Secondary Predicate Device | MultiPulse HoPlus(1911) | Asclepion Laser Technologies Gmbh | K161257 |
4 Device Description
The HZ-40 Holmium Laser Therapeutic Apparatus is a family of pulsed solid-state Holmium:YAG laser systems that deliver energy at a wavelength of $2100\mathrm{nm}$ through optical fibers for soft tissue surgery and stone fragmentation.
1 / 6
{5}
The system utilizes a flashlamp-pumped Ho:YAG laser medium, and the laser energy is transmitted to the target site via SMA905-compatible optical fibers.
The system includes a touchscreen interface, single-pedal footswitch, integrated liquid cooling, and multiple built-in safety features such as an emergency stop, door interlock, and real-time monitoring of laser output.
# 5 Intended Use
Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and stone fragmentation in use in medical specialties including: Urology, Gastroenterology, ENT, Gynecology and general surgery.
# 6 Indications for use
Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and stone fragmentation in use in medical specialties including: Urology, Gastroenterology, ENT, Gynecology and general surgery.
# 7 Comparison of Technological Characteristics
| Device | Subject device | Primarily Predicate device | Secondary Predicate device | Remark |
| --- | --- | --- | --- | --- |
| 510(k) number | Pending | K141403 | K161257 | / |
| Trade name | Holmium Laser Therapeutic Apparatus | Litho DK30 | MultiPulse HoPlus | / |
| Model | HZ-40 | / | 1911 | / |
| Manufacturer | Guangzhou Potent Medical Equipment Joint-Stock Co., Ltd. | Quanta System SPA | Asclepion Laser Technologies Gmbh | / |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Regulation Name | Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology | Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology | Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology | Same |
| Regulation Class | Class II | Class II | Class II | Same |
| Product Code | GEX | GEX | GEX | Same |
{6}
| Device | Subject device | Primarily Predicate device | Secondary Predicate device | Remark |
| --- | --- | --- | --- | --- |
| 510(k) number | Pending | K141403 | K161257 | / |
| Device Panel | General & Plastic Surgery | General & Plastic Surgery | General & Plastic Surgery | Same |
| Intended use | Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and stone fragmentation in use in medical specialties including: Urology, Gastroenterology, ENT, Gynecology and general surgery. | The device Litho including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue and in Lithotripsy of stones. | The MultiPulse HoPLUS Laser system and its fiber optic delivery system are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery. | Similar
**Note 1** |
| Indications for use | Holmium Laser Therapeutic Apparatus are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and haemostasis of soft tissue and stone fragmentation in use in medical specialties including: Urology, | The device Litho including a fiber optic delivery system is indicated for use in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation.vaporization, coaaulation and hemostasis of soft tissue and in Lithotripsy of | Incision, excision, resection, ablation, vaporisation, coagulation and hemostasis in medical specialties. | Similar
**Note 1** |
3 / 6
{7}
| Device | Subject device | Primarily Predicate device | Secondary Predicate device | Remark |
| --- | --- | --- | --- | --- |
| 510(k) number | Pending | K141403 | K161257 | / |
| | Gastroenterology, ENT, Gynecology and general surgery. | stones in medicalspecialties including, but not limited to: Urology, Gastroenterology, Arthroscopy, Neurosurgery, Pulmonary, Gynecology, ENT, Dermatology, Plastic Surgery and General Surgery. | | |
| Laser Source | Pulsed Holmium laser (CHT:YAG) | Pulsed Holmium laser (CHT:YAG) | Pulsed Holmium laser (CHT:YAG) | Same |
| Wavelength (nm) | 2, 100 nm | 2, 100 nm | 2, 100 nm | Same |
| Emission | Pulsed | Pulsed | Pulsed | Same |
| Pulse duration | 90-2000μs | Not publicly available | 150-1700 μs | Similar
**Note 2** |
| Energy per pulse | 0.2-6 J | Not publicly available | 0.25-6 Joule | Similar |
| Frequency | 30Hz | Not publicly available | 5-100 Hz | Similar
**Note 3** |
| Max average power | 40W | 30W | 140W | Similar
**Note4** |
| Power supply | AC 100-240V, 50/60 Hz, 3.5 kVA | Not publicly available | 400 V 50 Hz/16 A 3-phase | Different
**Note 5** |
| Delivery system | Optical fibers | Optical fibers | Optical fibers | Same |
| Aiming beam | Power < 3mW, Wavelength | Not publicly available | Green diode laser < 5 mW | Similar |
4 / 6
{8}
| Device | Subject device | Primarily Predicate device | Secondary Predicate device | Remark |
| --- | --- | --- | --- | --- |
| 510(k) number | Pending | K141403 | K161257 | / |
| | 520nm±10nm, Class 3R | | | **Note 6** |
**Note 1:**
The intended use and indications for use of the subject device is the same as predicate device(K141403), with only a difference in wording.
**Note 2:**
The pulse duration of the subject device is similar with the secondary predicate device(K161257), and the energy per pulse of subject device is quite similar with the predicate device, also the subject device has passed the IEC 60601-1 testing and IEC 60601-2-22 testing. So this minor difference of pulse duration will not raise any safety and effectiveness issues.
**Note 3:**
The frequency of the subject device is within the range of the secondary predicate device(K161257), so this minor difference will not raise any safety and effectiveness issues.
**Note 4:**
The max average power of the subject device is similar with the primarily predicate device(K141403) and within the range of the secondary predicate device(K161257), also the subject device has passed the IEC 60601-1 testing and the IEC 60601-2-22 testing, so the difference will not raise any safety and effectiveness issues.
**Note 5:**
Though the power supply of the subject device is different from the predicate devices, the subject device has passed the IEC 60601-1 testing, so the difference will not raise any safety and effectiveness issues.
**Note 6:**
Though the aiming beam of the subject device is a little different from the predicate device and reference device, the subject device adopted the safer laser and has passed the IEC 60825-1 testing and the IEC 60601-2-22 testing, so this difference will not raise any safety and effectiveness issues.
## 8 Non-Clinical Performance Data
The HZ-40 Holmium Laser Therapeutic Apparatus has undergone comprehensive performance testing to ensure compliance with applicable international standards and FDA guidance. The following evaluations were conducted:
### 8.1 Electrical Safety and Electromagnetic Compatibility
The device was tested and found to conform to the following standards:
- IEC 60601-1:2005 + A1:2012 + A2:2020
Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 + A1:2020
{9}
Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances – Requirements and tests.
- IEC 60601-2-22:2019
Medical electrical equipment – Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic, and diagnostic laser equipment.
- IEC 60825-1:2014
Safety of laser products – Part 1: Equipment classification and requirements.
## 8.2 Software Verification and Validation
Software development and validation were performed in accordance with:
- IEC 62304:2006 + A1:2015
Medical device software – Software life cycle processes.
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005).
## 9 Conclusion
Based on the above analysis and non-clinical tests performed, the Holmium Laser Therapeutic Apparatus(Model: HZ-40) has the same intended use/indications for use and similar technological characteristics as the predicate devices. Minor differences between the subject and predicate devices do not raise new risks regarding safety or effectiveness. The subject device is substantially equivalent to the identified predicate devices.
6 / 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.