The Q Guidance System, when used with the Spine Guidance 5 Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients. The system is indicated for orthopedic and neurological spine procedures in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative stereotaxic guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified. The system assists in the positioning of instruments for surgical procedures on the spine and pelvis, including: - Screw and Needle Placement in the spine or pelvis - Bone resection in the spine - Interbody device placement in the lumbar spine (adults only)
Device Story
System provides image-guided stereotaxic planning and intraoperative navigation for spine/pelvis surgery; inputs include DICOM images (CT, MR, PET, 2D SCOUT); software processes images to display planned trajectories and anatomical landmarks; system uses infrared optical active/passive tracking to monitor instrument position relative to patient anatomy; used in OR by surgeons; output is real-time visual guidance on monitors for instrument positioning; assists in screw/needle placement, bone resection, and interbody device placement; benefits include increased precision in surgical instrument navigation; Maestro Air Motor and Elite Q Attachments provide powered cutting/drilling capabilities with integrated navigation tracking.
Clinical Evidence
No clinical testing performed. Evidence consists of non-clinical design verification and validation, including software testing (IEC 62304), system accuracy verification (ASTM F2554-22), biocompatibility (ISO 10993-1), sterilization validation (ISO 17665-1, ISO 11137-2), electrical safety/EMC (IEC 60601-1/1-2), and cadaveric lab simulated use testing.
Technological Characteristics
Infrared optical active/passive tracking; DICOM-based image processing; supports CT, MR, PET, 2D imaging; accuracy: 2mm positional, 2° trajectory; components include computer platform, touchscreen monitors, navigated powered instruments (Maestro Air Motor), and trackers. Biocompatibility per ISO 10993-1; electrical safety per IEC 60601-1; accuracy per ASTM F2554-22.
Indications for Use
Indicated for adult and pediatric (adolescent) patients undergoing orthopedic and neurological spine procedures requiring computer-assisted planning and intraoperative stereotaxic guidance for screw/needle placement, bone resection, and lumbar interbody device placement.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 19, 2026
Stryker Leibinger GmbH & Co KG
Andrea Wallen-Gerding
Principal Regulatory Affairs Specialist
Bötzinger Straße 41
Freiburg Im Breisgau, D-79111
Germany
Re: K253623
Trade/Device Name: Spine Guidance 5.4 Software; Spine Guidance 5.4 Upgrade; Spine Guidance 5.4 Update; Spine Guidance 5.4 Software Enable; Spine Guidance 5.4 Sales Demo; Q Drill Tracker; NavLock Adapter; Elite Q Attachments; Elite Cutting Accessories; Maestro Air Motor
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: November 18, 2025
Received: November 18, 2025
Dear Andrea Wallen-Gerding:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253623 - Andrea Wallen-Gerding
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253623 - Andrea Wallen-Gerding
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Page 1 of 1.
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253623 | ? |
| Please provide the device trade name(s). | | ? |
| Spine Guidance 5.4 Software; Spine Guidance 5.4 Upgrade; Spine Guidance 5.4 Update; Spine Guidance
5.4 Software Enable; Spine Guidance 5.4 Sales Demo; Q Drill Tracker; NavLock Adapter; Elite Q
Attachments; Elite Cutting Accessories; Maestro Air Motor. | | |
| Please provide your Indications for Use below. | | ? |
| Spine Guidance 5.4 Software Indications for Use:
The Q Guidance System, when used with the Spine Guidance 5 Software, is intended as a planning and
intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and
pediatric (adolescent) patients.
The system is indicated for orthopedic and neurological spine procedures in which the use of computer
assisted planning and surgery may be appropriate. The system can be used for intraoperative stereotaxic
guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be
identified.
The system assists in the positioning of instruments for surgical procedures on the spine and pelvis,
including:
- Screw and Needle Placement in the spine or pelvis
- Bone resection in the spine
- Interbody device placement in the lumbar spine (adults only) | | |
| Elite Q Attachments and Cutting Accessories:
The Elite Q Attachments and Cutting Accessories are intended to be used with the Stryker Consolidated
Operating Room Equipment (CORE™) 2 Console and the Pi Drive 2 Motor or the Pi Drive 2 Plus Motor.
The Elite Q Attachments, when used with a compatible motor and a stereotactic instrument tracker, are also
indicated as an accessory to the Stryker Spine Guidance Software.
Specific applications include but are not limited to minimally invasive spine (MIS) Spine applications, such
as foraminotomy, facetectomy; orthopedic and neurological spine applications, such as corpectomy, pedicle
subtraction osteotomy; and laminotomy / laminectomy. | | |
| Maestro Air Motor:
The Stryker Consolidated Operating Room Equipment (CORE™) Maestro Air Pneumatic System is a
pneumatically operated surgical instrument system. The pneumatic motor provides power to operate
removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including
craniotomy and spinal surgery; as well as ear, nose and throat (ENT), orthopedic, and general surgical
applications including maxillofacial, craniofacial and sternotomy surgeries.
The Maestro Air Motor, when used with a compatible Elite Q Attachment and a stereotactic instrument
tracker, is also indicated as an accessory to the Stryker Spine Guidance Software. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253623
Stryker Leibinger GmbH & Co. KG.
Section 6 – 510(k) Summary
## 6.0 Submitter Information
### 6.1 This Premarket Notification is submitted by:
Stryker Leibinger GmbH & Co. KG
Bötzinger Straße 41
79111 Freiburg, Germany
### 6.2 Contact Information
Primary Contact: Andrea N. Wallen-Gerding
Principal Regulatory Affairs Specialist
Telephone: (269) 491-9234
Email: andrea.wallengerding@stryker.com
Date Prepared: September 9, 2025
Alternate Contact: Mariah Knight
Principal Regulatory Affairs Specialist
Telephone: 971-294-0565
Email: mariah.knight@stryker.com
### 6.3 Device Name
Table 6-1: Device Name
| Subject (Modified) Device Information | |
| --- | --- |
| Trade/ Proprietary Name | Spine Guidance 5.4 Software; Spine Guidance 5.4 Upgrade; Spine Guidance 5.4 Update; Spine Guidance 5.4 Software Enable; Spine Guidance 5.4 Sales Demo; Q Drill Tracker; NavLock Adapter; Elite Q Attachments; Elite Cutting Accessories; Maestro Air Motor |
| Common Name | Stereotaxic Instruments |
| Classification | Class II |
| Classification Product Code | OLO |
| Classification Name | Orthopedic Stereotaxic Instrument |
| Classification Regulation | 21 CFR 882.4560 |
| Review Panel | Orthopedic |
Traditional 510(k)
{5}
Stryker Leibinger GmbH & Co. KG.
Section 6 – 510(k) Summary
## 6.4 Predicate Devices
The following are the legally marketed predicate devices for the subject device included in this Traditional 510(k):
Table 6-2: Predicate Device List
| Subject Device | Predicate Device | Predicate Device 510(k) | Predicate Device Product Code | Manufacturer |
| --- | --- | --- | --- | --- |
| Spine Guidance 5.4 Software | Spine Guidance 5.3 Software | K252871 | OLO | Stryker Leibinger GmbH & Co. KG |
| Maestro Air Motor | Maestro Air Motor | K150801 | HBB | Stryker Instruments |
| Elite Q Attachments | Elite Q Attachments | K252871 | OLO | Stryker Instruments |
| Elite Cutting Accessories | Elite Cutting Accessories | K252871 | OLO | Stryker Instruments |
## 6.5 Device Description
### 6.5.1 Spine Guidance 5.4 Software
The Spine Guidance 5.4 System is an image guided stereotaxic, planning, and intraoperative guidance system intended to enable open or percutaneous computer-assisted surgery. It assists the surgeon in precisely positioning manual and powered instruments and locating patient anatomy during spinal surgery.
The system is comprised of the Spine Guidance 5.4 Software, the Q Guidance System (computer platform), Mako Spine System, navigated accessories / instruments (e.g., powered drills, pointers), and various system components (e.g., calibration devices, navigation adapters, patient / instrument trackers, etc.). The system provides intraoperative guidance to the surgeon using passive and active wireless optical tracking technologies.
The computer platform consists of a computer, camera, big touchscreen monitor, and a small touchscreen monitor. The Spine Guidance 5.4 Software functionality is described in terms of its capabilities that feature planning, registration, and navigation of medical devices.
The Spine Guidance 5.4 Software includes all the existing features from the Spine Guidance 5.3 Software described in K252871. Updates to the software include the integration of the Q Drill Tracker as a new system component, ability to navigate the Maestro Air Motor, integration of additional burs, and now allows the NavLock Adapter to be used with two additional trackers.
### 6.5.2 Maestro Air Motor
The Maestro Air Motor is intended to be used with a compatible attachment and cutting accessory for cutting, drilling, reaming, decorticating, shaping, and smoothing of bone and bone cement in a variety of surgical procedures. The motor is usable in placement or cutting of screws, metals, wires, pins, and other fixation devices. The motor can also be navigated when performing computer assisted surgery using Spine Guidance 5.4 Software.
Traditional 510(k)
K253623, Page 2 of 7.
{6}
Stryker Leibinger GmbH & Co. KG.
Section 6 – 510(k) Summary
## 6.5.3 Elite Q Attachments
The Elite Q Attachments serve as the interface between a compatible motor and a distally attached cutting accessory. The Elite Q Attachments provide a location for mounting a stereotactic instrument tracker and support navigation of a high-speed drill as well as the system’s instrument tip proximity detection feature.
## 6.5.4 Elite Cutting Accessories
The Elite Cutting Accessories are used for cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement in a variety of surgical procedures. The cutting accessories can also be used in the placement or cutting of screws, metal, wires, pins, and other fixation devices. The cutting accessories support navigation of a high-speed drill as well as the system’s instrument tip proximity detection feature.
## 6.6 Indications for Use
### 6.6.1 Spine Guidance 5.4 Software
The Q Guidance System, when used with the Spine Guidance 5 Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients.
The system is indicated for orthopedic and neurological spine procedures in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative stereotaxic guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified.
The system assists in the positioning of instruments for surgical procedures on the spine and pelvis, including:
- Screw and Needle Placement in the spine or pelvis
- Bone resection in the spine
- Interbody device placement in the lumbar spine (adults only)
### 6.6.2 Maestro Air Motor
The Stryker Consolidated Operating Room Equipment (CORE™) Maestro Air Pneumatic System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as ear, nose and throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries.
The Maestro Air Motor, when used with a compatible Elite Q Attachment and a stereotactic instrument tracker, is also indicated as an accessory to the Stryker Spine Guidance Software.
### 6.6.3 Elite Q Attachments and Cutting Accessories
The Elite Q Attachments and Cutting Accessories are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE™) 2 Console or CORE Maestro Air Pneumatic System and a compatible motor.
The Elite Q Attachments, when used with a compatible motor and a stereotactic instrument tracker, are also indicated as an accessory to the Stryker Spine Guidance Software.
Traditional 510(k)
K253623, Page 3 of 7.
{7}
Stryker Leibinger GmbH & Co. KG.
Section 6 – 510(k) Summary
Specific applications include but are not limited to minimally invasive spine (MIS) Spine applications, such as foraminotomy, facetectomy; orthopedic and neurological spine applications, such as corpectomy, pedicle subtraction osteotomy; and laminotomy / laminectomy.
## 6.7 Comparison of Technological Characteristics
A comparison of the technological characteristics of the subject devices included in the scope of this Traditional 510(k) is included in the tables below.
## 6.7.1 Technological Comparison between the Spine Guidance 5.4 Software and the Spine Guidance 5.3 Software
The technological comparison between the subject device (Spine Guidance 5.4 Software) and the predicate device (Spine Guidance 5.3 Software) is included in Table 6-3 below. The Spine Guidance 5.3 software application received 510(k) clearance per 510(k) number K252871.
Table 6-3: Technological Comparison between Spine Guidance 5.3 Software (Subject Device) and the Predicate Device
| Item | Subject Device: Spine Guidance 5.4 Software | Predicate Device: Spine Guidance 5.3 Software (K252871) |
| --- | --- | --- |
| Indications for Use | The Q Guidance System, when used with the Spine Guidance 5 Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients. | The Q Guidance System, when used with the Spine Guidance 5 Software, is intended as a planning and intraoperative guidance system to enable open or percutaneous computer assisted surgery in adult and pediatric (adolescent) patients. |
| | The system is indicated for orthopedic and neurological spine procedures in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative stereotaxic guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified. | The system is indicated for orthopedic and neurological spine procedures in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative stereotaxic guidance where a reference to a rigid anatomical structure such as the skull, pelvis or spine can be identified. |
| | The system assists in the positioning of instruments for surgical procedures on the spine and pelvis, including:
• Screw and Needle Placement in the spine or pelvis
• Bone resection in the spine
• Interbody device placement in the lumbar spine (adults only) | The system assists in the positioning of instruments for surgical procedures on the spine and pelvis, including:
• Screw and Needle Placement in the spine or pelvis
• Bone resection in the spine
• Interbody device placement in the lumbar spine (adults only) |
| Main System Components | • Q Guidance System
• Mako Spine System
• Various reusable and single use instruments for use with the SG5.4 Software | • Q Guidance System
• Mako Spine System
• Various reusable and single use instruments for use with the SG5.3 Software |
Traditional 510(k)
K253623, Page 4 of 7.
{8}
Stryker Leibinger GmbH & Co. KG.
Section 6 – 510(k) Summary
| Item | Subject Device: Spine Guidance 5.4 Software | Predicate Device: Spine Guidance 5.3 Software (K252871) |
| --- | --- | --- |
| **Modes of Operation** | • Mako Robotic Arm pre-surgery check
• Patient preparation
• System set-up
• Image import
• Planning
• Patient registration
• Mako Robotic Arm setup
• Navigation including trajectory guidance for screw placement | • Mako Robotic Arm pre-surgery check
• Patient preparation
• System set-up
• Image import
• Planning
• Patient registration
• Mako Robotic Arm setup
• Navigation including trajectory guidance for screw placement |
| **Operating Principle** | • The software is installed on the computer that is part of the platform
• Images are imported in DICOM format
• The software displays the images and planned items with navigational information on a monitor | • The software is installed on the computer that is part of the platform
• Images are imported in DICOM format
• The software displays the images and planned items with navigational information on a monitor |
| **Planning Navigation Features** | • Screws (including Screw Placement Suggestion)
• Measurements
• Segmentations (Manual and automatic)
• Local correlation (merge regions)
• Planes
• Smart Zones | • Screws (including Screw Placement Suggestion)
• Measurements
• Segmentations (Manual and automatic)
• Local correlation (merge regions)
• Planes
• Smart Zones |
| **Navigation implant Features** | • Screw Placement
• Interbody/Cage Placement | • Screw Placement
• Interbody/Cage Placement |
| **Localizing and Tracking Technology** | Infrared Optical Active and Passive Tracking | Infrared Optical Active and Passive Tracking |
| **System Accuracy** | The system has a mean accuracy of 2 mm for positional displacement and 2° for trajectory angle displacement. Accuracy values apply to tracking in the workspace. | The system has a mean accuracy of 2 mm for positional displacement and 2° for trajectory angle displacement. Accuracy values apply to tracking in the workspace. |
| **Supported Imaging Modalities** | • Computed Tomography (CT)
• Magnetic Resonance (MR)
• Position emission tomography (PET)
• 2D DICOM images (SCOUT only) | • Computed Tomography (CT)
• Magnetic Resonance (MR)
• Position emission tomography (PET)
• 2D DICOM images (SCOUT only) |
| **Intended Use Environment** | Operating Room | Operating Room |
Traditional 510(k)
K253623, Page 5 of 7.
{9}
Stryker Leibinger GmbH & Co. KG.
Section 6 – 510(k) Summary
## 6.7.2 Technological Comparison between the Maestro Air Motor and its Predicate Device
There have been no changes to the intended use, design, materials, or fundamental scientific technology between the subject and predicate devices. The only change is an update to the indications for use reflecting its use as an accessory with Spine Guidance 5.4 Software.
## 6.7.3 Technological Comparison between the subject and Predicate Elite Q Attachments
There have been no changes to the intended use, design, materials, or fundamental scientific technology between the subject and predicate devices. The only change is an update to their indications reflecting use with the CORE Maestro Air Pneumatic System.
## 6.7.4 Technological Comparison between the subject and Predicate Elite Cutting Accessories
There have been no changes to the intended use, design, materials, or fundamental scientific technology between the subject and predicate devices. The only change is an update to their indications reflecting use with the CORE Maestro Air Pneumatic System.
## 6.8 Summary of Non-Clinical Testing
The function and performance of the subject devices (Spine Guidance 5.4 Software, Maestro Air Motor, Elite Q Attachments, and Elite Cutting Accessories) have been evaluated through non-clinical design verification and validation testing. The results of the evaluation tests demonstrate that the subject devices successfully meet the requirements of their intended use, and the subject devices demonstrate substantial equivalence to the cited predicate devices.
Additional testing was performed on the subject devices to ensure they met their design requirements. A summary of the testing and the results are included in Table 6-3 below.
- Software testing as required by IEC 62304 and FDA Guidance on General Principles of Software Validation, January 11, 2002.
- System accuracy verification per ASTM F2554-22.
- The subject devices were validated with intended users in cadaver labs with simulated use tests to ensure the user needs and intended use requirements were met. All requirements were met and no new issues of safety or effectiveness were raised.
- Biocompatibility verification according to ISO 10993-1:2018 and the FDA Guidance Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, September 2023.
- Sterilization validation of the reusable devices was conducted according to ISO 17665-1 and the FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (March 2015).
- Sterilization validation of the single-use devices was conducted according to per ISO 11137-2 and the FDA Guidance Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile (January 2024).
- Electrical Safety and Electromagnetic Compatibility verification to IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 and IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020.
## 6.9 Summary of Clinical Testing
No clinical testing was required to support this submission.
Traditional 510(k)
K253623, Page 6 of 7.
{10}
Stryker Leibinger GmbH & Co. KG.
Section 6 – 510(k) Summary
## 6.10 Conclusion
The subject devices, Spine Guidance 5.4 Software, Maestro Air Motor, Elite Q Attachments, and Elite Cutting Accessories perform as intended and are substantially equivalent to their respective predicate device with regard to intended use, indications for use, design, principles of operation, technology, materials, and performance. No new issues of safety or effectiveness have been raised. The performance testing supports a determination of substantial equivalence.
Traditional 510(k)
K253623, Page 7 of 7.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.