K253613 · Paragon 28, Inc. · KTT · Dec 19, 2025 · Orthopedic
Device Facts
Record ID
K253613
Device Name
Monkey Rings External Ring Fixation System
Applicant
Paragon 28, Inc.
Product Code
KTT · Orthopedic
Decision Date
Dec 19, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Monkey Rings™ External Fixation System is intended to be used in adults and pediatric patients that are children aged 2 years to less than 12 years and adolescents aged 12 through 21 years (up to but not including the 22nd birthday). The Monkey Rings External Ring Fixation System is intended to treat the following patient indications in pediatric patients and adults for the treatment and fixation of: • Open and closed fractures • Post-traumatic joint contracture which has resulted in loss of range of motion • Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction • Pseudoarthrosis, infected union, non-union, or malunion of long bones • Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction • Correction of bony or soft tissue deformity (e.g., orthoplastic surgery) • Correction of segmental bony or soft tissue defects • Joint arthrodesis • Management of comminuted intra-articular fractures • Bone transport The Monkey Rings External Ring Fixation System is indicated in adults for: • Osteotomy • Revision procedure where other treatments or devices have been unsuccessful • Bone reconstruction procedures • Fusions and replantations of the foot • Charcot foot reconstruction • Offloading and/or immobilization of ulcers and/or wounds of the foot and ankle • Lisfranc dislocations • Ankle distraction (arthrodiastasis) • Septic fusion
Device Story
Modular ring-based external fixation system; stabilizes bone segments during fracture management, deformity correction, or limb lengthening. Components include rings, connecting rods, struts, bolts, fasteners, clamps, plates, pins, and Kirschner wires. Surgeon-assembled in various configurations for static fixation or gradual adjustment. New Universal Joint Strut allows precise modification of strut length and controlled repositioning of fixation rings; provides angular flexibility for off-axis fixation while maintaining structural stability. Used in clinical settings by surgeons. Output is a mechanical frame construct; provides physical stabilization and controlled distraction/compression of bone segments. Benefits include improved intraoperative adjustability and postoperative frame modification capabilities.
Clinical Evidence
Bench testing only. Performance assessed per ASTM F1541-17, including static axial grip, static torsional grip, static compression of rings, dynamic axial compression, and evaluation of pin bending/torsional strength. Sterilization validation confirmed sterility assurance level and maintenance of material/mechanical integrity.
Indicated for adults and pediatric patients (ages 2-21) for treatment/fixation of open/closed fractures, post-traumatic joint contracture, fractures requiring distraction, pseudoarthrosis, infected/non/mal-union, limb lengthening, deformity correction, segmental defects, joint arthrodesis, comminuted intra-articular fractures, and bone transport. Adults only: osteotomy, revision procedures, bone reconstruction, foot fusions/replantations, Charcot foot, offloading/immobilization of foot/ankle ulcers, Lisfranc dislocations, ankle distraction, and septic fusion.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Paragon 28 External Ring Fixation System (K212895)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 19, 2025
Paragon 28, Inc.
Jacqueline Sloan
Regulatory Affairs Specialist I
14445 Grasslands Dr.
Englewood, Colorado 80112
Re: K253613
Trade/Device Name: Monkey Rings External Ring Fixation System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: KTT
Dated: November 17, 2025
Received: November 18, 2025
Dear Jacqueline Sloan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253613 - Jacqueline Sloan
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253613 - Jacqueline Sloan
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Peter G. Allen
Digitally signed by Peter G. Allen -S
Date: 2025.12.19 14:01:29
-05'00'
for Lixin Liu, Ph.D.
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253613 | ? |
| Please provide the device trade name(s). | | ? |
| Monkey Rings External Ring Fixation System | | |
| Please provide your Indications for Use below. | | ? |
| The Monkey Rings™ External Fixation System is intended to be used in adults and pediatric patients that
are children aged 2 years to less than 12 years and adolescents aged 12 through 21 years (up to but not
including the 22nd birthday). | | |
| The Monkey Rings External Ring Fixation System is intended to treat the following patient indications in
pediatric patients and adults for the treatment and fixation of: | | |
| • Open and closed fractures | | |
| • Post-traumatic joint contracture which has resulted in loss of range of motion | | |
| • Fractures and disease which generally may result in joint contractures or loss of range of motion and
fractures requiring distraction | | |
| • Pseudoarthrosis, infected union, non-union, or malunion of long bones | | |
| • Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction | | |
| • Correction of bony or soft tissue deformity (e.g., orthoplastic surgery) | | |
| • Correction of segmental bony or soft tissue defects | | |
| • Joint arthrodesis | | |
| • Management of comminuted intra-articular fractures | | |
| • Bone transport | | |
| The Monkey Rings External Ring Fixation System is indicated in adults for: | | |
| • Osteotomy | | |
| • Revision procedure where other treatments or devices have been unsuccessful | | |
| • Bone reconstruction procedures | | |
| • Fusions and replantations of the foot | | |
| • Charcot foot reconstruction | | |
| • Offloading and/or immobilization of ulcers and/or wounds of the foot and ankle | | |
| • Lisfranc dislocations | | |
| • Ankle distraction (arthrodiastasis) | | |
| • Septic fusion | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253613 Page 1 of 6
510(K) SUMMARY
Manufacturer: Paragon 28, Inc.
14445 Grasslands Dr.
Englewood, CO 80112
Contact: Jacqueline Sloan
Regulatory Affairs Specialist I Paragon 28, Inc.
14445 Grasslands Dr.
Englewood, CO 80112
Phone: 720-617-6843
Jackie.Sloan@zimmerbiomet.com
Date Prepared: November 16, 2025
Device Trade Name: Monkey Rings External Ring Fixation System
Device Class and Common Name: Class II, appliance, fixation, nail/blade/plate combination, multiple component
Classification: 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances and accessories.
Product Code: KTT
Indications for Use: The Monkey Rings™ External Fixation System is intended to be used in adults and pediatric patients that are children aged 2 years to less than 12 years and adolescents aged 12 through 21 years (up to but not including the 22nd birthday).
The Monkey Rings™ External Fixation System is indicated in adults and pediatric patients or the treatment and fixation of:
- Open and closed fractures
- Post-traumatic joint contracture which has resulted in loss of range of motion
{5}
K253613 Page 2 of 6
- Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
- Pseudoarthrosis, infected union, non-union, or malunion of long bones
- Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction
- Correction of bony or soft tissue deformity (e.g., orthoplastic surgery)
- Correction of segmental bony or soft tissue defects
- Joint arthrodesis
- Management of comminuted intra-articular fractures
- Bone transport
The Monkey Rings External Ring Fixation System is indicated in adults for:
- Osteotomy
- Revision procedure where other treatments or devices have been unsuccessful
- Bone reconstruction procedures
- Fusions and replantations of the foot
- Charcot foot reconstruction
- Offloading and/or immobilization of ulcers and/or wounds of the foot and ankle
- Lisfranc dislocations
- Ankle distraction (arthrodiastasis)
- Septic fusion
{6}
K253613 Page 3 of 6
## Device Description:
The Monkey Rings External Ring Fixation System is a modular, ring-based external fixation system intended for the treatment and fixation of a variety of orthopedic conditions in pediatric and adult patients. The system is designed to stabilize bone segments during fracture management, deformity correction, or limb lengthening procedures. It consists of multiple components, including rings, connecting rods, struts, bolts, fasteners, clamps, plates, pins, and Kirschner wires (K-wires), which can be assembled in various configurations to provide either static fixation or gradual adjustment of bone position. The modular design allows surgeons to tailor the construct to meet specific anatomical and clinical needs.
The subject components included in this submission are the Universal Joint Struts and updated sterilization parameters for the system. The Universal Joint Strut is a newly introduced component designed to allow precise modification of strut length and controlled repositioning of the fixation rings over time. Its universal joint mechanism provides enhanced angular flexibility, enabling improved intraoperative adjustability and postoperative frame modification while maintaining structural stability and fixation integrity. All other components of the Monkey Rings External Ring Fixation System, including rings, connecting rods, and associated fixation hardware, were previously cleared and remain unchanged in design and function.
## Primary Predicate:
Paragon 28 External Ring Fixation System (K212895)
## Reference Device:
Taylor Spatial Frame (K201253)
{7}
K253613 Page 4 of 6
## Technological Comparison:
The subject device, Monkey Rings External Ring Fixation System, and the predicate device, the Paragon 28 External Ring Fixation System, share similar technological characteristics. Both systems consist of circular rings connected by adjustable struts, allowing precise multi-planar adjustments. Both use stainless steel or titanium pins and tensioned wires to secure the rings to bone, with optional hinges or joints to permit controlled joint movement. The struts on both devices are manually adjustable and include lockout mechanisms to maintain corrections and prevent unintended movement.
The primary difference between the subject device and the predicate device is the addition of a universal joint strut, which enables off-axis circular ring fixation for deformity correction. The universal joint strut can be adjusted to change its length, allowing the ring placement to be gradually modified over time. It also incorporates a lockout feature to control when lengthening or shortening occurs. Using six universal joint struts around a ring construct allows independent movement while maintaining overall frame stability, whereas the predicate device uses ball joint struts with locking ends to stabilize the frame.
The differences between design modifications amid the universal joint strut and the previous ball joint struts are minor and do not affect the intended use, performance, or safety profile. Therefore, they do not raise different questions of safety or effectiveness.
{8}
K253613 Page 5 of 6
Table 1: Technological Comparison
| Feature / Characteristic | Subject Device: Monkey Rings External Fixation System | Predicate Device: Paragon 28 External Fixation System | Comparison / Notes |
| --- | --- | --- | --- |
| Device Type | Circular external ring fixation system | Circular external ring fixation system | Same device type; intended for multi-planar deformity correction |
| Rings & Struts | Circular rings connected by adjustable struts; includes universal joint struts | Circular rings connected by adjustable struts; uses ball joint struts | Both allow precise multi-planar adjustment; subject device adds universal joint strut for off-axis fixation |
| Adjustable Mechanism | Manual struts with lockout mechanisms; universal joint allows controlled lengthening/shortening | Manual struts with lockout mechanisms; ball joint struts | Universal joint provides independent ring movement while maintaining frame stability; overall locking function remains comparable |
| Pins & Wires | Stainless steel or titanium pins; tensioned wires | Stainless steel or titanium pins; tensioned wires | Material and fixation method identical |
| Optional Hinges / Joints | Available for controlled joint movement | Available for controlled joint movement | Same functionality |
| Intended Use | Stabilization of fractures, osteotomies, and deformity corrections | Stabilization of fractures, osteotomies, and deformity corrections | No change in intended use |
| Safety / Performance Impact | Addition of universal joint strut; minor design changes | Original design | Differences do not affect safety, performance, or raise different questions of effectiveness |
| Lockout Feature | Yes, on all struts including universal joint | Yes, on all struts | Functionally similar; subject device offers added off-axis capability |
{9}
K253613 Page 6 of 6
## Performance Testing:
ASTM F1541-17 – Static Axial Grip of Connector (A2), Static Torsional Grip of Connectors and Fixators (A2 and A3), Static Compression of Rings (A3), Dynamic Axial Compression of Construct (A7), Evaluation of Bending and Torsional Strength of Pins (A5), Evaluation of Static Compression of Subassembly (A6).
The performance of these configurations was assessed in accordance with ASTM F1541-17 under established design control procedures. In addition, a comprehensive sterilization assessment, documented in TR-23080701, was completed to confirm that the selected sterilization process achieves the required sterility assurance level and that the device maintains its material, mechanical, and functional integrity following sterilization. These activities collectively demonstrate that the evaluation methods are appropriate and sufficient to support the safety and performance of the subject device.
## Conclusions:
The Monkey Rings External Ring Fixation System has the same intended use and similar technological characteristics as the predicate device, the Paragon 28 External Ring Fixation System (K212895). Differences in design do not raise different questions of safety or effectiveness. The subject device is therefore considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.