K253611 · Mcnicoll Surgical, Inc. · OUR · Jan 29, 2026 · Orthopedic
Device Facts
Record ID
K253611
Device Name
Falco Fusion System
Applicant
Mcnicoll Surgical, Inc.
Product Code
OUR · Orthopedic
Decision Date
Jan 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Falco Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroilittis.
Device Story
The Falco Fusion System is a surgical implant set used to stabilize or fuse the sacroiliac (SI) joint. The system consists of cannulated, fully threaded, fenestrated titanium alloy screws (11mm and 13mm diameters; 35mm-75mm lengths), disposable Kirschner wires, and reusable surgical instruments. Used in a clinical setting by surgeons, the device is implanted across the SI joint (typically 2 to 4 screws) to provide mechanical stabilization. The screws feature a dual-pitch, self-drilling, and self-tapping design, eliminating the need for pilot hole drilling. By stabilizing the joint, the system aims to reduce patient pain and improve mobility. The device is delivered non-sterile and requires steam sterilization by the user before the procedure.
Clinical Evidence
Bench testing only. Mechanical performance evaluated via axial pull-out strength, driving torque, torsional properties, and static/dynamic cantilever bending per ASTM F543 and ASTM F2193. Biocompatibility assessed per ISO 10993-1. Reprocessing validation for cleaning and steam sterilization performed per ISO 19227 and ANSI/AAMI ST98:2022.
Technological Characteristics
Cannulated, fully threaded, fenestrated screws. Materials: Ti-6Al-4V ELI (ASTM F136) for implants; 316-LVM stainless steel (ASTM F138) for Kirschner wires; medical-grade stainless steel (ASTM F899) and silicone (USP Class VI) for instruments. Dual-pitch, self-drilling, self-tapping design. Delivered non-sterile; steam sterilized by user. Passivation per ASTM F86 and ASTM A967.
Indications for Use
Indicated for patients requiring sacroiliac joint fusion due to sacroiliac joint disruptions or degenerative sacroiliitis.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Life Spine SIMPACT Sacroiliac Joint Fixation System (K201538)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 29, 2026
McNicoll Surgical, Inc.
Etienne Robichaud
Chief Executive Officer
702-1750 Ave. De Vitre
Quebec City, QC G1J1Z6
Canada
Re: K253611
Trade/Device Name: Falco Fusion System
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: OUR
Dated: November 10, 2025
Received: November 18, 2025
Dear Etienne Robichaud:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253611 - Etienne Robichaud
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253611 - Etienne Robichaud
Page 3
Sincerely,
MAZIAR SHAH-MOHAMMADI -S
For: Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
K253611 - Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253611 | ? |
| Please provide the device trade name(s). | | ? |
| Falco Fusion System | | |
| Please provide your Indications for Use below. | | ? |
| The Falco Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroilittis. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253611 - Page 1 of 3
510(k) Summary
Falco Fusion System – Sacroiliac Joint Fusion System
Submitted By:
McNicoll Surgical Inc.
702-1750 Ave de Vitre
Quebec City, QC
G1J 1Z6
Canada
Telephone: +1 (833) 488-8688
510(k) Contact:
Etienne Robichaud
CEO
EtienneRobichaud@mcnicollsurgical.com
Date Prepared: November 10th, 2025
Trade Name: Falco Fusion System
Common Name: Sacroiliac Joint Fusion System
Classification: OUR, CFR 888.3040, Class II
Primary Predicate: Outlet Sacroiliac Joint Fusion System (K181881)
Additional Predicates: Blue Topaz Sacroiliac Screw System (K213590)
Life Spine SIMPACT Sacroiliac Joint Fixation System (K201538)
Device Description:
The Falco Fusion System is designed to stabilize or fuse the sacroiliac (SI) joint, helping to reduce pain and improve mobility in patients suffering from SI joint disruptions or degenerative sacroiliitis. The system includes titanium alloy screws (Ti-6Al-4V ELI, ASTM F-136) in diameters of 11mm and 13mm, with lengths ranging from 35mm to 75mm. Additionally, the system comes with disposable Kirschner wires (316-LVM, ASTM F138) and reusable surgical (Stainless Steel, ASTM F899) instruments, providing a comprehensive solution for sacroiliac joint fusion procedures.
Indications for Use:
The Falco Fusion System is intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.
Technological Characteristics:
Falco Fusion implants are cannulated, fully threaded screw with fenestration. It features dual-pitch, self-drilling, and self-tapping screws for efficient bone purchase without the need for pilot hole drilling. Typically, 2 to 4 screws are implanted across the sacroiliac joint to achieve stabilization and fusion.
The subject device has similar technological characteristics to the predicate device, including indications for use, design, materials, and principles of operation. Differences, if any, do not raise new questions of safety or effectiveness and have been evaluated through bench testing.
{5}
K253611 - Page 2 of 3
## Material:
The implants are made from Ti-6Al-4V ELI titanium alloy (ASTM F136) and are intended for long-term implantation. Kirschner wires are made from 316-LVM stainless steel (ASTM F138) and are single-use. Instruments are manufactured from medical-grade stainless steel (ASTM F899) and silicone (USP Class VI for handles) and are intended for reuse following steam sterilization.
The system is delivered non-sterile, and all implants and instruments are organized in a reusable instrument tray with screw caddies. Screws and Kirshner wires are replaced after each use.
## Performance Data:
To support substantial equivalence, the Falco Fusion System underwent a series of non-clinical performance tests. Mechanical testing included:
- Axial pull-out strength (ASTM F543)
- Driving torque (ASTM F543)
- Torsional properties (ASTM F543)
- Static and dynamic cantilever bending (ASTM F2193)
All testing met predetermined acceptance criteria and demonstrated that the subject device performs comparably to the predicate. Although testing was performed per ASTM F543 and F2193, the device design and testing approach were also reviewed against the recommendations in ASTM F3574-22, and the system is compliant with applicable elements of that standard.
The system and associated Class II accessories/instruments undergo validated passivation and final cleaning processes. Passivation of the titanium alloy implants is performed in accordance with ASTM F86, while passivation of the stainless steel Class II instruments (e.g., guide wires) is performed in accordance with ASTM A967. Final cleaning validation for the subject device system is performed per ISO 19227. These processes support the biocompatibility assessment conducted in accordance with ISO 10993-1.
The system is delivered non-sterile. Validated steam sterilization instructions are provided in the labeling for end-user processing. Reusable instruments were validated for cleaning and sterilization in accordance with ANSI/AAMI ST98:2022.
## Substantial Equivalence:
The Falco Fusion System has the same intended use and similar technological characteristics as the predicate device. Performance testing, including mechanical evaluation and reprocessing validation, demonstrates the device is as safe and effective as the predicate.
Any differences do not raise new questions of safety or effectiveness. The device is therefore substantially equivalent to the predicate.
{6}
K253611 - Page 3 of 3
## Predicate Devices:
| 510(k) # | Device Name | Manufacturer | Product Code | Type |
| --- | --- | --- | --- | --- |
| K181881 | Outlet Sacroiliac Joint Fusion System | SIJ Surgical | OUR | Primary |
| K201538 | Life Spine SIMPACT Sacroiliac Joint Fixation System | Life Spine Inc. | OUR, HWC | Additional |
| K213590 | Blue Topaz Sacroiliac Screw System | Osseus Fusion Systems, LLC | OUR | Additional |
## Conclusion:
Based on the indications for use, technological characteristics, and non-clinical performance testing, the Falco Fusion System is substantially equivalent to the predicate device and does not raise new questions of safety or effectiveness
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.