Real-world data from a company-held dataset and published investigator-initiated studies were used to support the long-term safety, signal stability (electrode impedance), and clinical effectiveness of the device.
UNEEG real-world dataset; Retrospective analysis of real-world monitoring data
Patients with epilepsy; Sample Size: 5
Not applicable for this study
Electrode impedance stability
Investigator-initiated studies (NCT02946151, NCT05241678, NCT04061707); Investigator-initiated clinical studies; Follow-up/Duration: Mean exposure 38-380 days
Subjects with uncontrolled epilepsy; Sample Size: 34 total (10, 9, 15)
Not applicable for this study
Safety (adverse events), performance (adherence), and signal quality
Indications for Use
The UNEEG EpiSight System is an electroencephalographic (EEG) recording and transmitting device system with a subcutaneous implant. It is indicated for continuous EEG recording in patients aged 18 and older with uncontrolled epilepsy. The UNEEG EpiSight System is a prescription device intended to aid in physician's remote assessment and monitoring of the patient's condition. Remote assessment and monitoring are defined as the availability of EEG data for review by a healthcare provider located at a different location from where the data is being collected. The medical use of the data acquired by the UNEEG EpiSight System is to be performed under the direction and interpretation of a licensed medical professional. The device does not provide diagnostic conclusions regarding the patient's condition.
Device Story
System comprises subcutaneous implant (UNEEG SubQ), wearable recorder, smartphone app, and cloud service. Implant measures subcutaneous EEG via two bipolar channels; housing placed behind ear; lead placed subcutaneously. Wearable recorder attaches to skin via adhesive baseplate; powers implant via inductive link; stores EEG data. Data offloaded via Bluetooth Low Energy to smartphone app; uploaded to UNEEG Cloud/ProConnect for healthcare provider review in 3rd party EEG software. Used in home/outpatient settings; operated by patient. Provides objective, continuous EEG data to aid physician assessment of epilepsy; does not provide diagnostic conclusions. Benefits include less invasive implantation (local anesthesia, no skull recess) compared to predicate, enabling targeted lead placement near seizure initiation sites.
Clinical Evidence
Clinical evidence includes a 12-week prospective study (ULTRA, N=64) and literature-based data. Primary endpoints: seizure detection sensitivity (TPF) and false positive rate (FPR). Results: Mean TPF 0.785 (95% CI: 0.666–0.892); mean FPR 0.014/24h (95% CI: 0.000–0.036). EEG quality demonstrated non-inferiority to scalp EEG. Impedance remained stable (<5 kΩ). Safety profile showed acceptable adverse event rates; one serious device-related infection reported.
Technological Characteristics
Implant: Alumina ceramic/titanium housing, 3 platinum/iridium electrodes, ETFE insulation. Sensing: 2-channel bipolar subcutaneous EEG. Connectivity: Inductive link (implant-to-recorder), Bluetooth Low Energy (recorder-to-app), secure internet (cloud). Sterilization: Ethylene Oxide. MR Conditional (1.5T/3T). Battery: Rechargeable external recorder. Software: Data acquisition, transmission, and cloud-based decryption/conversion to EDF+ format.
Indications for Use
Indicated for continuous EEG recording in patients aged 18 and older with uncontrolled epilepsy. Prescription use only.
Regulatory Classification
Identification
The Minder System is an electroencephalographic (EEG) recording and transmitting device implanted under the scalp. It is a prescription device indicated to acquire, transmit, and store EEGs continuously from patients between 18-75 years of age with drug-resistant epilepsy who are intolerant or not indicated for more conservative monitoring tools. The device consists of sub-scalp implanted electrodes that are connected to a transmitter or storage device. It is intended to aid in a physician's remote assessment and monitoring of the patient's condition by making EEG data available for review by a healthcare provider at a different location. The device is not intended to be implanted in brain tissue.
Special Controls
In combination with the general controls of the FD&C Act, the sub-scalp implanted electroencephalogram system for remote patient monitoring is subject to the following special controls:
(1) In vivo performance testing must demonstrate that the device performs as intended following implantation to continuously acquire, transmit, and store electrical brain activity and evaluate all adverse effects, including tissue injury.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use for the duration of the operational life of the device. The technical specifications of the device's hardware and software must be fully characterized, such as:
(i) Data acquisition testing:
(A) Input dynamic range;
(B) Signal/noise ratio;
(C) Frequency response;
(D) Accuracy and resolution of recorded signal; and
(E) Quality and accessibility of continuous wireless transmission.
(ii) Mechanical testing:
(A) Impact testing; and
(B) Integrity testing following implantation and explantation.
(3) Performance testing must evaluate the compatibility of the device in a magnetic resonance environment.
(4) The patient-contacting components of the device must be demonstrated to be biocompatible.
(5) Performance testing must demonstrate electrical safety, thermal safety, mechanical safety, electromagnetic compatibility (EMC), and wireless coexistence of the device in the intended use environment.
(6) Software verification, validation, and hazard analysis must be performed.
(7) Performance testing must demonstrate the sterility of the device.
(8) Performance data must support the shelf-life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf-life.
(9) Performance testing must validate the reprocessing instructions for the reusable components of the device.
(10) Physician and patient labeling must include the following:
(i) Detailed description of the device technical parameters, explanation of all device outputs, care instructions to maintain the operation of the device, operational life and duration of device implantation, and troubleshooting options;
(ii) A summary of any in vivo performance testing conducted to demonstrate device safety and performance under anticipated use conditions;
(iii) Conditions of use that may impact the accuracy, reliability, or functionality of the device measurement and operation;
(iv) Conditions that may preclude use and implantation of the device, such as skin integrity or thickness;
(v) Magnetic resonance imaging related information;
(vi) Statement that the device is not a standalone diagnostic device and does not replace clinical decision making; and
(vii) A shelf-life.
(11) Physician labeling must include:
(i) Instructions for the preparation and implantation of the device, including recommended site of implantation and method of implantation, and explantation of the device; and
(ii) Validated instructions for reprocessing of any reusable component.
(12) Patient labeling must include:
(i) A patient implant card, including magnetic resonance imaging related information;
(ii) Information about the associated surgical procedure; and
(iii) A discussion of available monitoring alternatives.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 11, 2026
Uneeg Medical A/S
Camilla Wismar
Senior Director, Regulatory & Quality
Borupvang 2
Alleroed, 3450
Denmark
Re: K253607
Trade/Device Name: UNEEG EpiSight System
Regulation Number: 21 CFR 882.1360
Regulation Name: Sub-scalp implanted electroencephalogram system for remote patient monitoring
Regulatory Class: Class II
Product Code: SEM
Dated: May 12, 2026
Received: May 12, 2026
Dear Camilla Wismar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253607 - Camilla Wismar
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253607 - Camilla Wismar
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Patrick Antkowiak -S
for
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K253607
?
Please provide the device trade name(s).
?
UNEEG EpiSight System
Please provide your Indications for Use below.
?
The UNEEG EpiSight System is an electroencephalographic (EEG) recording and transmitting device system with a subcutaneous implant. It is indicated for continuous EEG recording in patients aged 18 and older with uncontrolled epilepsy.
The UNEEG EpiSight System is a prescription device intended to aid in physician's remote assessment and monitoring of the patient's condition. Remote assessment and monitoring are defined as the availability of EEG data for review by a healthcare provider located at a different location from where the data is being collected.
The medical use of the data acquired by the UNEEG EpiSight System is to be performed under the direction and interpretation of a licensed medical professional. The device does not provide diagnostic conclusions regarding the patient's condition.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
UNEEG medical
# 510(k) Summary
## UNEEG EpiSight System, K253607
Contact Details (21 CFR 807.92(a)(1))
**Applicant Name:** UNEEG Medical A/S
**Applicant Address:** Borupvang 2
3450 Alleroed
Denmark
**Applicant Contact Telephone:** +45 4063 8000
**Applicant Contact:** Mrs. Camilla Wismar
**Applicant Contact Email:** cawi@uneeg.com
Device Name (21 CFR 807.92(a)(2))
**Device Trade Name:** UNEEG EpiSight System
**Common Name:** Sub-scalp implanted electroencephalogram system for remote patient monitoring
**Classification Name:** Sub-Scalp Implanted Electroencephalogram System For Remote Patient Monitoring
**Regulation Number:** 882.1360
**Product Code(s):** SEM
Legally Marketed Predicate Devices (21 CFR 807.92(a)(3))
| Predicate # | Predicate Trade Name | Product Code |
| --- | --- | --- |
| DEN240062 | Minder System | SEM |
Device Description Summary (21 CFR 807.92(a)(4))
The UNEEG EpiSight System is an electroencephalographic (EEG) recording and transmitting device system with a subcutaneous implant. The wearable recorder receives the transmitted EEG data from the implant. While the recorder is charging, EEG data will be automatically offloaded from the recorder through a Bluetooth Low Energy connection to the app installed on the patient's mobile phone. The patient's EEG is then transferred to the healthcare facility where it can be analyzed by the healthcare professionals using a 3rd party EEG review system.
Page 1 of 16
{5}
UNEEG medical
The UNEEG EpiSight System provides objective, reliable, long term, and continuous EEG recordings allowing for a more complete understanding of a patient's condition and aiding a physician's clinical decision making regarding how best to treat the patient.
The UNEEG EpiSight System consists of an implantable device (UNEEG SubQ), a wearable recorder (UNEEG EpiSight Recorder), a dedicated app for the patient's smartphone (UNEEG EpiSight App) and cloud data transfer service (UNEEG Cloud and UNEEG ProConnect). The related accessories are a blunt needle (Introducing Needle), a baseplate (UNEEG EpiSight Baseplate) and a charger (UNEEG EpiSight Charger).
The implantable device measures the subcutaneous electroencephalogram (EEG) from two bipolar channels using a common reference. The implant consists of a housing that holds an electronic circuit and a lead with three electrodes which measure the EEG and the impedance. A specialized blunt needle is used for surgical insertion of the implant. The implant is placed subcutaneous pointing in the direction where the electrical activity is intended to be measured. The housing is placed behind the ear, and the lead can be placed in different positions.
The implant has a lifetime of 1 year after implantation. The safety of the implant beyond 1 year has not been established. Prior to expiration of this period the implant must be removed.
The wearable recorder supplies power to the implant and receives and stores the measured EEG. This is facilitated through a wireless inductive link whose function requires close transcutaneous alignment between the implant and the wearable recorder. The recorder enables communication and must be fixed to the skin behind the ear with a disposable baseplate. The baseplate situates the recorder onto the skin on top of the implant. The recorder can be dismounted from the baseplate without removing the baseplate from the skin if needed.
The recorder has a button for switching the recorder on/off and self-reporting of seizures, and a light indicator (LED) indicates functions during use. The recorder is wireless and rechargeable; charging is done by placing the recorder in the charger. While the recorder is charging, EEG data will be transferred from the recorder to the app installed on the patient's smartphone through a Bluetooth connection.
The app is the primary user interface in both setup and daily use of the UNEEG EpiSight System. The app guides proper alignment of the recorder when attaching recorder and baseplate on top of the implant. The app allows for self-reporting of seizures, and the patient can see average recording hours and total number of self-reported seizures. The app ensures that data from the recorder are uploaded to the healthcare facility when it is connected to the internet via either mobile data or Wi-Fi.
The recorded EEG data is transferred to UNEEG ProConnect at the healthcare facility through the UNEEG Cloud. Once transferred, the data will be decrypted and converted to a readable EDF+ format which can be analyzed by the healthcare professional using a 3rd party EEG review system.
Page 2 of 16
{6}
UNEEG medical
### **Intended Use/Indication for Use (21 CFR 807.92(a)(5))**
The UNEEG EpiSight System is an electroencephalographic (EEG) recording and transmitting device system with a subcutaneous implant. It is indicated for continuous EEG recording in patients aged 18 and older with uncontrolled epilepsy.
The UNEEG EpiSight System is a prescription device intended to aid in physician's remote assessment and monitoring of the patient's condition. Remote assessment and monitoring are defined as the availability of EEG data for review by a healthcare provider located at a different location from where the data is being collected.
The medical use of the data acquired by the UNEEG EpiSight System is to be performed under the direction and interpretation of a licensed medical professional. The device does not provide diagnostic conclusions regarding the patient's condition.
### **Indication for Use Comparison (21 CFR 807.92(a)(5))**
The indication for use of the subject device, UNEEG EpiSight System, and the predicate device, Minder System, is highly similar. The differences do not affect safety and effectiveness of intended use.
### **Technological Comparison (21 CFR 807.92(a)(6))**
The subject device, UNEEG EpiSight System, has very similar principles of operation and technological characteristics as the predicate device Minder System (DEN240062). The minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are the same between the subject and predicate device:
- Indication for use
- Battery type
- Wireless communication
- Electrode shape
- Sterilization method (implantable device)
- Features of wearable device user interface
- Cloud and data storage
- Features of patient app configuration and communication
The following technological differences exist between subject and predicate device:
- Implantation procedure / lead location
- Number of electrodes
- Electrode array length
- Method of wearing the external device
- MR Conditions
- Battery charging
Page 3 of 16
{7}
The table below compares the principles of operation and operational/technological characteristics of the UNEEG EpiSight System and the predicate device.
| | Subject Device | Primary Predicate Device | Equivalence Discussion |
| --- | --- | --- | --- |
| **Device Name** | UNEEG EpiSight System (UNEEG Medical A/S) | Minder System (Epiminder Pty. Ltd.) DEN240062 | n/a |
| **Regulation and Product Codes** | 21 CFR 882.1360 SEM | 21 CFR 882.1360 SEM | Identical |
| **Indications for Use** | The UNEEG EpiSight System is an electroencephalographic (EEG) recording and transmitting device system with a subcutaneous implant. It is indicated for continuous EEG recording in patients aged 18 and older with uncontrolled epilepsy. The UNEEG EpiSight System is a prescription device intended to aid in physician's remote assessment and monitoring of the patient's condition. Remote assessment and monitoring are defined as the availability of EEG data for review by a healthcare provider located at a different location from where the data is being collected. The medical use of the data acquired by the UNEEG EpiSight System is to be performed under the direction and interpretation of a licensed medical professional. The device does not provide diagnostic conclusions regarding the patient's condition. | The Minder System is an electroencephalographic (EEG) recording and transmitting device implanted under the scalp. It is a prescription device indicated to acquire, transmit, and store EEGs continuously from patients between 18-75 years of age with drug-resistant epilepsy who are intolerant or not indicated for more conservative monitoring tools. The Minder System is intended to aid in a physician's remote assessment and monitoring of the indicated patient's condition. Remote patient assessment and monitoring for this use is defined as the patient's EEG data is available for review by a healthcare provider located at a different location from the patient and where the data is being collected. The medical use of the data acquired by the Minder System is to be performed under the direction and interpretation of a licensed medical professional. The Minder System does not provide any diagnostic conclusions about the patient's condition. | Highly similar. The differences in wording choice have the same meaning and do not affect safety and effectiveness of the device. |
| **Implant Site** | Subcutaneous, implant housing placed behind the ear with lead location adjustable based on seizure location. | Subgaleal space, placed ear to ear posterior to the vertex, with one pair of recording contacts over each hemisphere. The telemetry unit is secured in a postauricular bone well. | Different. Reduced risk of surgical complications due to less invasive implantation procedure. Implant body location is similar but lead location is different from |
Page 4 of 16
{8}
**UNEEG**medical
| | Subject Device | Primary Predicate Device | Equivalence Discussion |
| --- | --- | --- | --- |
| | | | predicate. Differences allow for more targeted lead placement in the UNEEG EpiSight System allowing for placement closer to seizure initiation site. Additionally, the UNEEG SubQ implant does not require a skull recess to be created for placement of the lead and can be implanted under local anesthesia rather than general. |
| **System Characteristics** | | | |
| **MR Conditions** | MR Conditional (only implant) | Not publicly available | Non-clinical testing confirms the subject device conditional MR labeling. |
| **Battery type** | Rechargeable (external recorder) | Rechargeable (external wearable) | Identical |
| **Wireless Communications** | Implant to external recorder: inductive link Recorder to phone app: Bluetooth low energy | Implant to external wearable: inductive link Recorder to phone app: Bluetooth low energy | Identical |
| **Implantable device** | | | |
| **Number of electrodes** | 3 (2 channels) | 4 (2 bipolar pairs) | Different The differences do not affect the safety and effectiveness of the subject device. Non-clinical and clinical testing confirms equivalent device performance to the predicate including signal quality and ability to record relevant EEG patterns. |
| **Electrode shape** | Cylindrical | Cylindrical | Identical |
| **Individual Electrode surface area** | 1.10 mm diameter | 0.6 mm diameter [cochlear] | Different The differences do not affect the safety and effectiveness of the subject device. Non-clinical and clinical testing confirms equivalent device performance to the |
Page 5 of 16
{9}
**UNEEG**medical
| | Subject Device | Primary Predicate Device | Equivalence Discussion |
| --- | --- | --- | --- |
| | | | predicate including signal quality and ability to record relevant EEG patterns. |
| **Electrode array length** | 103 mm long electrode lead | 253 mm-long electrode lead | Different The differences do not affect the safety and effectiveness of the subject device. Non-clinical and clinical testing confirms equivalent device performance to the predicate including signal quality and ability to record relevant EEG patterns. |
| **Individual electrode length** | 10 mm | Not publicly available | Any differences do not affect the safety and effectiveness of the subject device. Non-clinical and clinical testing confirms equivalent device performance to the predicate including signal quality and ability to record relevant EEG patterns. |
| **Electrode spacing** | 20 mm | Not publicly available | Any differences do not affect the safety and effectiveness of the subject device. Non-clinical and clinical testing confirms equivalent device performance to the predicate including signal quality and ability to record relevant EEG patterns. |
| **Electrode material** | Platinum/Iridium Alloy | Platinum [cochlear] | Different The differences do not affect the safety and effectiveness of the subject device. Non-clinical and clinical testing confirms equivalent device performance to the predicate including signal quality and ability to record relevant EEG |
Page 6 of 16
{10}
**UNEEG**medical
| | Subject Device | Primary Predicate Device | Equivalence Discussion |
| --- | --- | --- | --- |
| | | | patterns. |
| **Lead body insulation material** | ETFE (signal wire insulation) | Silicone elastomer [cochlear] | Different Non-clinical and clinical testing confirms equivalent device performance to the predicate including signal quality and ability to record relevant EEG patterns. |
| **Implant housing material** | Alumina Ceramic/Titanium | Titanium [cochlear] | Different Non-clinical and clinical testing confirms equivalent device performance to the predicate including signal quality and ability to record relevant EEG patterns. |
| **Wire features** | Multi-strand | Not publicly available | Any differences do not affect the safety and effectiveness of the subject device. Non-clinical and clinical testing confirms equivalent device performance to the predicate including signal quality and ability to record relevant EEG patterns. |
| **Biocompatibility** | Yes, ISO 10993-1 | Yes | Identical |
| **Sterilization method** | Ethylene Oxide | Ethylene Oxide [cochlear] | Identical |
| **External Device** | | | |
| **User interface** | Integrated controls and visual indicators that allow the user to turn the device on/off, log seizures and monitor device status. | Integrated controls that allow the user to turn the device on/off and monitor device status. | Highly similar |
| **Method of wearing the external device** | The recorder is mounted in a baseplate containing an adhesive that can adhere to the skin. It is placed on the user's scalp behind the ear (aligning with the implant). | The Minder Wearable is placed over the ear and includes a wired connection to a receiver which magnetically connects to the implant. | Different. Non-clinical and clinical testing confirms equivalent device performance and safety to the predicate including signal quality |
Page 7 of 16
{11}
| | Subject Device | Primary Predicate Device | Equivalence Discussion |
| --- | --- | --- | --- |
| | | | and ability to record relevant EEG patterns. |
| **Size of external device** | Length: 35 mm Width: 34 mm Thickness: 7 mm | Not publicly available | Any differences do not affect the safety and effectiveness of the subject device. Non-clinical testing confirms equivalent device performance and safety to the predicate. |
| **Externally contacting materials** | Adhesive on baseplate: Acrylate/polyethylene | Not publicly available | Any differences do not affect the safety and effectiveness of the subject device. Non-clinical testing confirms equivalent device performance and safety to the predicate. |
| **Battery charging** | Wireless charging, the recorder is fully charged within 8 hours. | Batteries are removed from the external device and charged through a wired connection. [Cochlear] | Different. Non-clinical testing confirms equivalent device performance and safety to the predicate. |
| **Cloud and Data Storage** | | | |
| **Configuration** | Software for data transferring and storage. | Software for data transferring and storage. | Identical |
| **Purpose** | Allow healthcare professionals to access patient data for viewing in a 3^{rd} party system. | Allow healthcare professionals to access and download patient data for viewing. | Highly similar with identical functionality to review EEG data on a 3^{rd} party EEG review system. |
| **Communication** | Secure internet connection. | Secure internet connection | Identical |
| **Patient App** | | | |
| **Configuration** | Software app installed on the patient's mobile phone that collects and transmits EEG data from the UNEEG EpiSight Recorder via Bluetooth and uploads it to UNEEG Cloud for secure transition storage before accessing in a hospital facility. | Software app installed on a mobile device collects and transmits EEG data from the Minder Wearable via Bluetooth and uploads it to Minder Cloud for secure storage and processing. It can also record entries for a patient's seizure diary. | Both systems allow for self-registration of seizures. Differences do not affect safety or effectiveness. |
| **Communication** | Blue tooth low energy | Blue tooth low energy | Identical |
Page 8 of 16
{12}
UNEEG medical
# **Benefit/Risk Summary**
The subject device presents a different benefit-risk profile primarily due to its shorter labeled duration of use of up to 1 year compared to the predicate device that has up to 3 years duration of use. The shorter duration of implantation may reduce the cumulative benefit of long-term monitoring over time because continuous EEG data only is available in a shorter period. However, this potential reduction in duration-related benefit is balanced by a less invasive implantation procedure for the subject device. Specifically, the subject device does not require creation of a skull recess, bone removal, or tunneling, and it can be implanted under local anesthesia rather than general anesthesia. These differences reduce procedural burden and are expected to lower the risks associated with surgical implantation compared with the predicate device. In addition, the subject device allows targeted lead placement closer to the suspected seizure initiation site, while non-clinical and clinical testing support that the device provides clinically meaningful EEG recording performance and an acceptable safety profile for the duration of use period. Therefore, when used as intended for up to 1 year, the benefits of the subject device outweigh its risks and support substantial equivalence to the predicate device.
# **Non-Clinical and/or Clinical Tests Summary & Conclusions (21 CFR 807.92(b))**
# **Performance Bench Testing**
The following bench testing was conducted to ensure appropriate safety and performance of the UNEEG EpiSight System.
Implant bench testing:
- Measurement and transmission of data and impedance
- Electrode Dielectric Strength
- Implant surface temperature
- Lifetime of use
- Recording performance per IEC 80601-2-26
- Flexural Stresses, Mechanical shocks and Tensile forces
- Immunity to diagnostic ultrasound
- Pressure changes and Vibration
- Hermetic seal
Wearable bench testing:
- Environmental conditions
- Free fall, vibration, shock
- Surface Temperature
- Battery Safety including charging protection
System bench testing:
- Data Transfer from the implant to the recorder
- Power of the implant through inductive link
Page 9 of 16
{13}
UNEEG medical
### **Biocompatibility Testing**
Biocompatibility testing was performed for the UNEEG EpiSight System in accordance with ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA Guidance issued on September 8, 2023: Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'.
The implantable device, UNEEG SubQ, is categorized as an implant medical device with long-term (>30 days) direct contact with tissue. The risks associated with the implant device resulting in adverse tissue reaction or systemic toxicity effects have been evaluated through chemical characterization and subsequent toxicological risk assessment on the final finished device. The toxicological risk assessment concluded that the extractables from the test article, UNEEG SubQ, are unlikely to pose a toxicological safety concern.
Based on the toxicological risk assessment, the implant was evaluated for the following biocompatibility endpoints: cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), irritation/intracutaneous reactivity (ISO 10993-23), acute systemic toxicity (ISO 10993-11), material-mediated pyrogenicity (ISO 10993-11), subacute/subchronic systemic toxicity (ISO 10993-11), genotoxicity (ISO 10993-3), implantation (ISO 10993-6), chronic toxicity (ISO 10993-11) and carcinogenicity (ISO 10993-3). Evaluation of these biocompatibility endpoints were found to be adequate, with assessments supporting implantation for up to 1 year.
The Introducing needle is categorized as an external communicating medical device with limited (<24 hours) direct contact with tissue. The device was evaluated for the following biocompatibility endpoints: cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), irritation/intracutaneous reactivity (ISO 10993-23), acute systemic toxicity (ISO 10993-11), and material-mediated pyrogenicity (ISO 10993-11). Evaluation of these biocompatibility endpoints were found to be adequate.
The UNEEG EpiSight Recorder and UNEEG EpiSight Baseplate are categorized as surface medical devices with long-term (>30 days) direct contact with intact skin. The devices were evaluated for the following biocompatibility endpoints: cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation/intracutaneous reactivity (ISO 10993-23). Evaluation of these biocompatibility endpoints were found to be adequate.
### **Magnetic Resonance Imaging (MRI)**
The implant is MR conditional. MRI scan can be performed safely on a person with the implant only under very specific conditions (1.5T and 3T). All external devices used in combination with the implant, such as the UNEEG EpiSight Recorder, are MR Unsafe.
---Page 10 of 16
{14}
UNEEG medical
# **Clinical Performance Data Summary (ULTRA)**
# Study Design:
The clinical study was a 12-week open-label, prospective study with a paired comparative design for pivotal evaluation of the safety and effectiveness of the 24/7 EEG SubQ in subjects with epilepsy involving the temporal lobe (NCT04526418). The study enrolled subjects in the United States at 5 sites and outside the United States (OUS) in Europe at 3 sites. The study included subjects with uncontrolled epilepsy and evidence supporting temporal lobe involvement. Subjects were implanted and followed during a 12-week period, for two sites up to a 52-week period of outpatient EEG recording, including a required Epilepsy Monitoring Unit (EMU) admission for concurrent video-EEG. Data was collected through device recordings, clinical examinations, seizure logs, and structured questionnaires.
# Objectives:
Primary Objective – Effectiveness:
- Electrographic seizure recording sensitivity (TPF) of an ipsilaterally implanted device of at least 0.8: proportion of identified ipsilateral seizures as compared to ipsilateral video-EEG during admission to the Epilepsy Monitoring Unit
- False positive rate of at most 0.2: number of falsely identified ipsilateral seizures (false positives) per 24 hours as compared to ipsilateral video-EEG data.
Secondary Objectives – Safety:
- Nature and frequency of adverse events (AE) during the study and their severity, outcomes, and relationship to the device.
Secondary Objectives – Effectiveness:
- Median number of hours of subject device use per day during the study period of outpatient recording.
- Quality of EEG recorded by 24/7 EEG SubQ system as compared to video-EEG recordings in similar position during simulated home use and change in subcutaneous EEG quality at study completion.
- Change in electrode impedance over the study period of outpatient recording.
# Enrollment results:
Subjects enrolled in the study were equally distributed between the sites in US and Europe. The study enrolled 64 subjects, 32 subjects in U.S. and 32 subjects OUS, 62 implanted, 9 discontinued, 60 completed the EMU, 55 completed 12 weeks follow up and 25 completed the 52 weeks follow up.
The study population reflects a clinically diverse epilepsy population, see Table 1, with a broad range of seizure burdens, ages of onset, and time spent alone during day and night, see Table 2.
Page 11 of 16
{15}
**Table 1: Baseline demographics of study population.**
| | Overall (N=64) |
| --- | --- |
| **Sex** | |
| Female | 36 (56.3%) |
| Male | 28 (43.8%) |
| **Age (years)** | |
| Mean (SD) | 38.7 (13.2) |
| Median [Min, Max] | 37.0 [18.0, 69.0] |
| **BMI** | |
| Mean (SD) | 29.1 (7.03) |
| Median [Min, Max] | 27.5 [19.2, 61.7] |
| **Race** | |
| White | 46 (71.9%) |
| Black | 5 (7.8%) |
| Asian or Pacific Islander & White | 1 (1.6%) |
| Not applicable | 1 (1.6%) |
| Missing | 11 (17.2%) |
| **Ethnicity (Hispanic/Latino or not)** | |
| Hispanic/Latino | 2 (3.1%) |
| Not Hispanic/Latino | 51 (79.7%) |
| Not applicable | 11 (17.2%) |
**Table 2: Subject Epilepsy History**
| | Overall (N=64) |
| --- | --- |
| **Age at onset of epilepsy** | |
| Mean (SD) | 21.3 (15.0) |
| Median [Min, Max] | 17.0 [0, 66.0] |
| **Years since epilepsy onset** | |
| Mean (SD) | 17.0 (13.5) |
| Median [Min, Max] | 14.0 [0, 51.0] |
### Results – Safety:
Safety analysis was performed on all 62 subjects implanted with the UNEEG SubQ device. A total of 84 AEs were reported throughout the study period. Most of the AEs and SAEs observed were reported to be unrelated to the device or study procedure, see Table 3. There was one serious, device-related adverse event (AEs) reported during this study caused by an infection at the implant site following the surgical procedure. Two other unrelated SAEs were reported. The remaining AEs were classified as mild (63 AEs) and moderate (18 AEs), and the most frequently reported event was “headache” (12 AEs), followed by “discomfort” (6 AEs), and “implant pain” (6 AEs).
Page 12 of 16
{16}
UNEEG medical
**Table 3: Summary of adverse events**
| Classification | Number of events (n = 84) | |
| --- | --- | --- |
| | Relationship to device | Relationship to procedure |
| Not related | 47 | 44 |
| Unlikely | 12 | 7 |
| Possible | 12 | 2 |
| Probable | 6 | 12 |
| Causal relationship | 7 | 19 |
Results – Effectiveness:
- Usage (adherence): The median total usage time for day and night was 13.9 hours and 8.6 hours, respectively, corresponding to adherence rates of daytime: 92.6% and nighttime: 95%. This is very high and supports clinically relevant usage patterns. No relevant difference was observed across age, sex, geography, race, or ethnicity.
- EEG quality assessment: The EEG recordings from the subcutaneous implant demonstrated non-inferiority compared with concurrent scalp EEG recordings across all predefined task groups simulating home use, based on a model-based analysis of pooled task-group scores and the protocol-defined non-inferiority margin (lower bound of the 95% CI > -0.5).
- Electrode impedance: The mean impedance across all subjects was 2.33 kΩ, and thus below the clinically recommended 5 kΩ. The population level impedance increase was -4.30 Ω/year with the upper bound of the 95% confidence interval for the slope being 59.27 Ω/year. This was below the predefined threshold of 540 Ω per year. This points to a robust electrical connection between the electrodes and the tissue throughout the study which is a proxy for good signal quality.
- Seizure detection: At the subject level, perfect ipsilateral seizure detection sensitivity (TPF = 1) was observed for more than half of the patients. The mean TPF was 0.785 (95% CI: 0.666–0.892), see Table 4. For the false positive rate (FPR), no false positive events were observed for all but two subjects. The observed mean FPR was 0.014 false positives per 24 hours (95% CI: 0.000–0.036). This indicates a low risk of false seizure detections.
**Table 4: Bootstrap analysis of true positive fraction (TPF) and false positive rate (FPR).**
| | FAS Population (N=39) | | |
| --- | --- | --- | --- |
| | Mean | 95% CI^{1} | P-value^{1,2,3} |
| True Positive Fraction (TPF) | 0.785 | 0.666, 0.892 | 0.593 |
| False Positive Rate (FPR) | 0.014 | 0.000, 0.036 | <0.001 |
| ^{1}Bootstrap 95% CI and p-value based of of sampling on subject with replacement 10,000 times. | | | |
| ^{2}Null hypothesis that the TPF is less than or equal to 0.8. | | | |
| ^{3}Null hypothesis that the FPR is greater than or equal to 0.2. | | | |
**Additional Clinical Performance Data from other studies:**
Findings from other clinical studies published in the literature support the safety and effectiveness of the UNEEG EpiSight System for the intended patient population. Key enrollment, performance, and safety outcomes from the published investigator-initiated studies are summarized in Table 5.
Page 13 of 16
{17}
**Table 5: Clinical evidence for investigator-initiated studies**
| | Study Title | Subcutaneous EEG in epilepsy – proof of concept and clinical applications | Evaluation of sqEEG in subjects with uncontrolled epilepsy with genetic generalised epilepsy syndromes | Subcutaneous EEG: forecasting of epileptic seizures through investigation of long-term dynamics of seizure occurrences, stress, sleep and other factors. |
| --- | --- | --- | --- | --- |
| Study Information | No. | NCT02946151 | NCT05241678 | NCT04061707 |
| | Country | Denmark | Ireland | United Kingdom |
| **Enrollment** | | | | |
| Subjects (numbers) | Enrolled | 10 | 9 | 15 |
| | Completed | 9 | 9 | 10 |
| | Discontinued | 1 | 0 | 5 |
| **Performance** | | | | |
| Exposure (days) | N | 9 | 9 | 11 |
| | Mean ± SD | 99 ± 30 | 61 ± 40 | 380 ± 147 |
| | Median (range) | 98 (39-141) | 54 (15-154) | 433 (45-538) |
| Adherence rate (0-1) | N | 9 | 9 | 11 |
| | Mean ± SD | 0.75 ± 0.15 | 0.74 ± 0.30 | 0.70 ± 0.28 |
| | Median (range) | 0.77 (0.45-0.91) | 0.82 (0.07-0.98) | 0.78 (0.17-0.99) |
| **Safety** | | | | |
| Adverse device events, classified by seriousness. | Serious | 0 | 0 | 0 |
| | Not serious | 13 | 1 | 14 |
| Device deficiencies | Count | 0 | 0 | 15 |
### Results – Safety
Across all studies used to support substantial equivalence, a total of two serious adverse device effects (SADEs), 130 adverse device effects (ADEs), and 243 device deficiencies (DDs) were reported, see Table 6. Among the 243 DDs, 108 were relevant to the subject device, the remaining 135 DDs were reported for previous prototype devices or software.
Page 14 of 16
{18}
UNEEG medical
**Table 6: Reported adverse device effects and device deficiencies (number of events) for the device system, including implant and recorder.** Studies marked with * are ongoing, but data included here is peer-reviewed published.
| Study | Study population | SADEs | ADEs | | | | | DDs (related/unrelated) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Headache | Implant site^{a} | Infection^{b} | Other^{c} | Unclassified or unknown | |
| ULTRA | Epilepsy | 1† | 5 | 15 | 4 | 0 | 0 | 14 / 69 |
| ULTS | Healthy | 0 | 17 | 10 | 5 | 2 | 2 | 6 / 31 |
| Baseplate | Healthy | 0 | 0 | 0 | 0 | 22 | 0 | 53 / 0 |
| P2H | Healthy | 0 | 5 | 6 | 0 | 0 | 0 | 1 / 20 |
| P2D | Type 1 diabetes | 0 | 0 | 10 | 0 | 4 | 0 | 0 / 15 |
| MTLE | Epilepsy | 0 | 2 | 8 | 0 | 2 | 1 | 0 |
| SUBER | Epilepsy | 1 | 4 | 1 | 0 | 0 | 0 | 34 / 0 |
| PREDYct* | Epilepsy | 0 | 0 | 0 | 0 | 3 | 0 | 0 |
| EMIRE* | Epilepsy | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Epi-SubQ | Epilepsy | 0 | 0 | 1 | 0 | 0 | 0 | 0 |
| Filadelfia* | Epilepsy | 0 | 0 | 0 | 1 | 0 | 0 | 0 |
| **Total** | **-** | **2** | **33** | **51** | **10** | **33** | **3** | **108 / 135** |
$^{a}$ Transient pain/soreness, discomfort, bruising, swelling, increased or decreased sensitivity near the implant, protrusion, replacement.
$^{b}$ Related to implant or surgical procedure.
$^{c}$ Skin irritation, itching at operation site, hypoglycemia.
† Infection at implant site following implantation.
## Results – Effectiveness
- Usage (adherence): For the entire population with epilepsy, a total of 157 subjects, the median exposure time was 144 days with a mean ± SD = 216 ± 190 days. Across all epilepsy subjects with adherence information available (n=154), the observed median adherence was 73%, ≈17.5h per 24h, with a mean ± SD of 65±29%.
- EEG quality assessment: A clinical interpretability analysis for the ULTRA study demonstrated non-inferiority when comparing acceptable (score ≥ 3) versus unacceptable mean review scores for sqEEG and scalp-EEG. 33 of 34 paired assessments (97.1%) were classified as acceptable for both modalities, with one discordant pair. The paired difference in proportions (sqEEG minus scalp EEG) was -2.9% (95% CI -14.9% to 7.5%), meeting the predefined non inferiority margin of -15%. In the EpiSubQ study (N=8), Visual scoring (1-5) showed clear, interpretable EEG in 86% of scalp and 85% of sqEEG tasks, with comparable epileptiform morphology. In the P2H study (N=13), PSD comparison (-5 to 5 scale) showed sqEEG quality on par or better than scalp EEG, particularly during artifact-inducing tasks. Finally, in the P2D (N=8) and SUBER studies: assessments of predefined task demonstrated satisfactory EEG quality with clear physiological signals and artifacts across all subjects.
- Electrode impedance: In a meta-analysis of 3 different studies (P2D, MTLE and SUBER) the daily impedance check was analyzed for 27 subjects recording sqEEG for more than 1 year. Electrode impedances remained consistently low (<5 kΩ) across all subjects, with only a minor decreasing trend over time (-0.018 kΩ/month). These findings are further supported by the 3-months studies EpiSubQ (N=9) and P2D (N=8) where all measurements consistently remained low (<5 kOhm). No progressive changes in spectral characteristics or frequency-band power were observed in the reported studies, see Table 7, indicating stable signal quality over claimed implantation durations.
Page 15 of 16
{19}
[LOGO]
UNEEG medical
Table 7: Electrode impedance
| Study | Subjects (N) | Mean electrode impedance (kΩ) | Change in impedance Ω/time period |
| --- | --- | --- | --- |
| ULTRA | 60 | 2.33 | 4.30 per year (95% CI upper bound 59.27 per year) |
| UNEEG real-world dataset | 5 | <5 | -21.49 per year (95% CI -28.35 to -14.62 /year) |
| Meta-analysis (P2H, MTLE and SUBER) | 27 | <5 | -18 per month |
| EpiSubQ | 9 | <5 | - |
| P2D | 8 | <5 | - |
## Conclusion
Performance evaluation across the population demonstrated that the UNEEG EpiSight System enables continuous, ultra long-term EEG monitoring with signal quality comparable to scalp EEG, remaining stable over long-term use. It enables reliable, objective seizure monitoring across various seizure types and locations, often identifies more events than patient diaries, and supports clinical decision-making. Clinical data confirm the device is well-tolerated, unobtrusive, and does not interfere with daily activities.
Safety analysis showed an acceptable profile, with no unanticipated serious adverse device effects. Most adverse events were reported to be non-serious.
In conclusion, results of nonclinical and clinical tests have demonstrated that the UNEEG EpiSight System is safe, effective and performs equivalent to the predicate device for long-term EEG monitoring.
Page 16 of 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.