K253598 · Carbon Medical Technologies, Inc. · FBK · Mar 31, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K253598
Device Name
InjecSURE Injection System
Applicant
Carbon Medical Technologies, Inc.
Product Code
FBK · Gastroenterology, Urology
Decision Date
Mar 31, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InjecSURE® Injection System is intended for use as an accessory to currently marketed cystoscopes to allow delivery of injectable materials into urethral tissues of the lower urinary tract during cystoscopic procedures.
Device Story
InjecSURE Injection System is a sterile, single-use accessory for cystoscopes; used in hospitals, clinics, and physician offices. Device consists of a 20 Ga, 5-inch needle (pencil-point or spinal tip) and a disposable sheath. Sheath attaches to a cystoscope to provide a guided pathway for the needle to reach target urethral tissues under direct visualization. Physician operates the system in tandem with a cystoscope to deliver injectable materials. Output is the physical delivery of substances into tissue; assists in clinical procedures for the lower urinary tract. Benefits include precise, guided access to urethral tissues for therapeutic injections.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through non-clinical bench testing, including sterility, packaging, shelf-life, dimensional inspection, simulated use, and functional performance (fragmentation, needle force, compatibility, leakage, and insertion force). Biocompatibility evaluated per ISO 10993-1.
Technological Characteristics
Sterile, single-use injection needle (20 Ga) and sheath. Introducer cannula > 0.036 inch OD. Luer connection interface. Compatible with 2.7 mm cystoscopes. Sterilized via EtO. Materials evaluated per ISO 10993. Mechanical operation; no energy source or software.
Indications for Use
Indicated for female patients requiring delivery of injectable materials into urethral tissues of the lower urinary tract during cystoscopic procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 24, 2026
Carbon Medical Technologies, Inc.
Jared Klein
Manager, Regulatory Affairs
1290 Hammond Rd.
St, Paul, Minnesota 55110
Re: K253598
Trade/Device Name: InjecSURE Injection System
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FBK
Dated: March 5, 2026
Received: March 5, 2026
Dear Jared Klein:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253598 - Jared Klein
Page 2
Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-
{2}
K253598 - Jared Klein
comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARK J. ANTONINO -S
Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
InjecSURE Injection System
Page 10 of 57
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253598 | ? |
| Please provide the device trade name(s). | | ? |
| InjecSURE Injection System | | |
| Please provide your Indications for Use below. | | ? |
| The InjecSURE® Injection System is intended for use as an accessory to currently marketed cystoscopes to allow delivery of injectable materials into urethral tissues of the lower urinary tract during cystoscopic procedures. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253598
Page 1 of 5
# 510(k) Summary – InjecSURE® Injection System
## 1. Submitter’s Name, Address and Date of Submission
Jared Klein
Manager, Regulatory Affairs
Carbon Medical Technologies, Inc.
1290 Hammond Road
Saint Paul, MN 55110
Phone: 651-653-8512
Fax: 651-407-1975
Submitted: March 23, 2026
## 2. Device Name
Trade Name: InjecSURE® Injection System
Common Name: Injection Needle
Classification Panel: Gastroenterology-Urology
Classification Name: Endoscope And Accessories, 21 CFR 876.1500
Product Code: FBK
Device Class: Class II
## 3. Predicate Device(s)
The Carbon Medical Technologies, Inc. InjecSURE Injection System’s legally marketed predicate device is the Repris Bladder Injection System, manufactured by URO-1, Inc.
**Predicate Device**: Repris Bladder Injection System (K180214)
This predicate device has not been subject to a design-related recall.
## 4. Reference Device(s)
The legally marketed devices identified in Table 1 were used as reference devices to support specific aspects of non-clinical performance testing and are not relied upon as predicate devices for purposes of substantial equivalence.
Table 1. Reference Device(s)
| Reference Device | Manufacturer | 510(k) | Product Code |
| --- | --- | --- | --- |
| USA Elite System Accessory Working Element (EAWE) | CIRCON ACMI | K890328 | HIH |
| Injection Needle | Advanced Uroscience | K982890 | FBK |
| LiNA OperaScope™ | LiNA Medical | K193007 | HIH / FAJ |
None of the identified reference devices have been subject to a design-related recall.
K253598 510(k) Summary
Page 1 of 5
{5}
K253598
Page 2 of 5
# 5. Indications for Use
The InjecSURE® Injection System is intended for use as an accessory to currently marketed cystoscopes to allow delivery of injectable materials into urethral tissues of the lower urinary tract during cystoscopic procedures.
# 6. Device Description
The InjecSURE Injection System includes sterile, single-use components designed for one patient, featuring a 20 Ga, 5-inch (127 mm) needle—available with either a pencil-point or spinal tip—and a disposable sheath. The disposable sheath is compatible with 2.7 mm cystoscopes of various lengths.
The sheath attaches to a cystoscope to provide a guided pathway for the needle to reach the desired location.
# 7. Comparison of Technological Characteristics
The InjecSURE Injection System [subject device] is substantially equivalent (SE) to the URO-1 Repris Bladder Injection System [predicate device] as they share the same intended use and fundamental principles of operation.
Both devices function as cystoscopic injection accessories intended to deliver legally marketed injectable materials into tissue structures of the lower urinary tract under direct visualization. Each device consists of an introducer (sheath), injection needle, fluid ports, and luer connection interfaces, and is advanced in tandem with a cystoscope.
The table below compares the InjecSURE Injection System [subject device] to the Repris Bladder Injection System [predicate device] with respect to indications for use, intended use, principles of operation, device components, performance characteristics, sterilization modality, duration of use, and patient-contacting materials.
Although certain technological differences exist, including needle gauge, introducer dimensions, working length, sterilization modality, and specific anatomical site within the lower urinary tract, these differences do not alter the intended use or fundamental principles of operation. The technological differences can be evaluated through performance testing and do not raise new questions of safety or effectiveness.
Therefore, the InjecSURE Injection System is substantially equivalent to the cited predicate device. The Substantial Equivalence Comparison is presented in Table 2.
K253598 510(k) Summary
Page 2 of 5
{6}
K253598
Page 3 of 5
Table 2. Substantial Equivalence Comparison.
| Characteristic | Subject Device | Predicate Device |
| --- | --- | --- |
| Device Name | InjecSURE Injection System | Repris Bladder Injection System |
| Manufacturer: | Carbon Medical Technologies, Inc. | URO-1, Inc. |
| 510(k) Number: | K253598 | K180214 |
| Indications for Use: | As an accessory to currently marketed cystoscopes to allow delivery of injectable materials into urethral tissues of the lower urinary tract during cystoscopic procedures. | Injection of drugs to address abnormal physiology in the lower urinary tract of adults. |
| Intended Use: | To deliver legally marketed injectable materials into tissue structures during cystoscopic procedures. It is provided sterile for single use. | To deliver a variety of legally marketed drugs into tissue structures during cystoscopic procedures. It is provided sterile for single use. |
| Route of Advancement: | Advanced to the lower urinary tract in tandem with a cystoscope. | Advanced to the lower urinary tract in tandem with a cystoscope |
| Target Populations: | Female | Female |
| Location of Injection: | Lower urinary tract tissue (Urethral tissues) | Lower urinary tract tissue (Bladder Wall) |
| Site of Use: | Hospitals, clinics, and physician offices | Hospitals, clinics, and physician offices |
| Device Features | | |
| Components: | Introducer Injection Needle Fluid Ports | Introducer Injection Needle Fluid Ports Syringe |
| Size of Needle: | 20 Ga. | 23 ga. |
| Size of Introducer Cannula: | > 0.036 inch | 0.064 inch (1.6 mm) |
| Connection Type(s) | Luer connector | Luer connector |
| Length of Needle Assembly: | 12.7 cm | 32.4 cm |
| Introducer working length | 11.0 – 13.0 cm | 8.06 inch (20 cm) |
| Introducer size (Outer Diameter) | < 8.4 mm (6.1 - 8.4 mm) | 7 mm |
| Introducer Compatibility | Various currently marketed cystoscopes | Various currently marketed cystoscopes |
| Performance | | |
| Duration of Use | ≤ 24 hours | ≤ 24 hours |
| Sterilization | Provided sterile. Sterilized by EtO | Provided sterile. Sterilized by gamma radiation (gamma). |
| Shelf-Life | 60 Months | Not Publicly Available |
K253598 510(k) Summary
Page 3 of 5
{7}
K253598
Page 4 of 5
| Characteristic | Subject Device | Predicate Device |
| --- | --- | --- |
| Device Name | InjecSURE Injection System | Repris Bladder Injection System |
| Frequency of Use | Single patient use. | Single patient use. |
| Patient Contact Materials | Evaluated in accordance with ISO 10993 for the nature and duration of contact. | Evaluated in accordance with ISO 10993 for the nature and duration of contact. |
The InjecSURE Injection System is designed for use with currently marketed cystoscopes and has the same intended use and fundamental principles of operation as the predicate device. Differences in needle gauge, introducer dimensions, working length, sterilization modality, and specific anatomical site within the lower urinary tract do not constitute a new intended use and do not introduce new mechanisms of action. Performance testing demonstrates that these technological differences do not raise new questions of safety or effectiveness.
The reference devices were used solely to support certain performance test methodologies and acceptance criteria. They are not relied upon as predicate devices and are therefore not included in the substantial equivalence comparison table.
## 8. Performance Data
Non-clinical performance data:
The following testing has been performed to demonstrate substantial equivalence to the predicate device. The InjecSURE Injection System met all the requirements for non-clinical performance testing, confirming that the design output meets the design inputs and specifications for the device.
- Sterility Testing
- ISO 11135:2014, ISO 11737-1:2018, ISO 11737-2:2019
- Packaging Testing
- ISO 11607-1:2019, ISO 11607-2:2019, ASTM D4169-22 and D4332-22
- Shelf-life Testing
- ASTM F1980-21, ASTM F88/F88M-15, ASTM F1929-23
- Dimensional Inspection and Testing
- ISO 8600-4:2023
- Simulated Use Testing
- ISO 62366-1:2015
- Functional Performance Testing, including:
- Fragmentation testing (ISO 7864:2016, ISO 8871-5:2016)
- Needle penetration force
K253598 510(k) Summary
Page 4 of 5
{8}
K253598
Page 5 of 5
- Needle/Hub separation (pull) force
- Cystoscope compatibility
- Irrigation flow rate
- Leakage integrity
- Visual field clearance
- Urethral insertion force
The InjecSURE Injection System met all pre-determined acceptance criteria for non-clinical performance testing and demonstrated that the device performs as intended. The results support determination that the technological differences do not raise new questions of safety or effectiveness.
**Biocompatibility:**
Biocompatibility studies were performed on relevant patient contacting materials in accordance with ISO 10993-1 and 2023 FDA guidance document “Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process”.
**Clinical performance data:**
No clinical study is deemed necessary since the substantial equivalence has been sufficiently demonstrated by non-clinical studies.
## 9. Conclusion
The InjecSURE Injection System [subject device] is substantially equivalent (SE) to the URO-1 Repris Bladder Injection System [predicate device] as they share the same intended use and fundamental principles of operation. Based on technical comparison and non-clinical performance testing described above, the technological differences do not raise new questions of safety and effectiveness.
Therefore, the InjecSURE Injection System is substantially equivalent to the referenced predicate device.
K253598 510(k) Summary
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.