The Medical Imaging Calibration Feature (MICF) is a software-only medical device intended to calibrate compatible general purpose off-the-shelf (OTS) desktop displays. The MICF is intended to be used by trained medical practitioners for the purpose of enabling primary diagnostic interpretation and medical image review. The MICF is intended to be used with specific macOS host and Apple display platforms. The MICF is not compatible with head-mounted displays, laptops, tablets, phones, or other portable displays that are not desktop displays. The device is not intended for mammography.
Device Story
MICF is a software-only SaMD that calibrates general-purpose Apple desktop displays for primary diagnostic radiology. It operates on macOS host platforms using an external third-party colorimeter (Calibrite Display Plus HL) to measure display output. The software adjusts display parameters to meet DICOM standards and AAPM TG-18/TG-270 quality control requirements. Used by trained medical practitioners in clinical facilities as part of a QA program; the software ensures the display meets technical tolerances for diagnostic image review. It does not perform image processing itself but enables the display hardware to accurately render medical images, supporting clinical decision-making by ensuring consistent, calibrated visual performance.
Clinical Evidence
No clinical data. Bench testing only. Validation included end-to-end physical testing of calibrated displays per IEC 62563-1, IEC 62563-2, AAPM TG-18, and AAPM TG-270 standards to confirm compliance with primary diagnostic radiology requirements.
Technological Characteristics
Software-only SaMD; runs on macOS host platforms. Calibrates Apple desktop displays (min 27", 5120x2880 resolution, 47-120 Hz). Uses external colorimeter (Calibrite Display Plus HL). Complies with DICOM, IEC 62563-1, IEC 62563-2, AAPM TG-18, and AAPM TG-270. Connectivity via standard macOS host interfaces.
Indications for Use
Indicated for trained medical practitioners to calibrate compatible off-the-shelf Apple desktop displays for primary diagnostic radiology and medical image review. Not for use with mobile devices, laptops, tablets, head-mounted displays, or mammography.
Regulatory Classification
Identification
A radiologic quality assurance instrument is a device intended for medical purposes to measure a physical characteristic associated with another radiologic device.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 1, 2026
Apple, Inc.
Lynda Ikejimba
Princ. Regulatory Affairs Assoc.
1 Apple Park Way
Cupertino, California 95014
Re: K253582
Trade/Device Name: Medical Imaging Calibration Feature (MICF)
Regulation Number: 21 CFR 892.1940
Regulation Name: Radiologic Quality Assurance Instrument
Regulatory Class: Class I
Product Code: SHN
Dated: March 2, 2026
Received: March 2, 2026
Dear Lynda Ikejimba:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253582 - Lynda Ikejimba
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253582 - Lynda Ikejimba
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

For:
Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253582 | ? |
| Please provide the device trade name(s). | | ? |
| Medical Imaging Calibration Feature (MICF) | | |
| Please provide your Indications for Use below. | | ? |
| The Medical Imaging Calibration Feature (MICF) is a software-only medical device intended to calibrate compatible general purpose off-the-shelf (OTS) desktop displays. The MICF is intended to be used by trained medical practitioners for the purpose of enabling primary diagnostic interpretation and medical image review.
The MICF is intended to be used with specific macOS host and Apple display platforms. The MICF is not compatible with head-mounted displays, laptops, tablets, phones, or other portable displays that are not desktop displays. The device is not intended for mammography. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Apple Inc.
MICF Traditional 510(k) Submission
K253582
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
# 1. Submitter
| Applicant | Apple Inc.
One Apple Park Way
Cupertino, CA 95014 |
| --- | --- |
| Submission Correspondent | Lynda Ikejimba, PhD
Principal Regulatory Affairs Associate
Phone: 650-695-9346
Email: Lc_ikejimba@apple.com |
| Secondary Correspondent | Sam Surette
US Regulatory Affairs Manager
Phone: 628-629-4161
Email: ssurette@apple.com |
| Date Prepared | March 27, 2026 |
# 2. Device Name and Classification
Subject Device:
| Name of Device | Medical Imaging Calibration Feature (MICF) |
| --- | --- |
| Classification Name | Radiologic quality assurance instrument |
| Regulatory Class | 21 CFR 892.1940 |
| Product Code | SHN |
| 510(k) Review Pannel | Radiology |
# 3. Predicate Device
Radiologic quality assurance instrument, 21 CFR 892.1940.
# 4. Device Description
The Medical Imaging Calibration Feature (MICF) is a software-only device intended to calibrate a compatible general purpose off-the-shelf (OTS) Apple display to enable use of the display in primary diagnostic radiology. Specifically, the MICF is a Software as a Medical Device (SaMD), intended to provide adjustments to the display in accordance with the Digital Imaging and Communications in Medicine (DICOM) standard for medical imaging. As part of a clinical facility's quality assurance (QA) program, the MICF fulfills calibration and visual quality control (QC) testing of the radiologist's display. The MICF is intended to be used by trained medical practitioners authorized by their facility's QA program to perform display calibration.
One Apple Park Way, Cupertino, CA 95014
{5}
Apple Inc.
MICF Traditional 510(k) Submission
# 5. Indications for Use
The Medical Imaging Calibration Feature (MICF) is a software-only medical device intended to calibrate compatible general purpose off-the-shelf (OTS) desktop displays. The MICF is intended to be used by trained medical practitioners for the purpose of enabling primary diagnostic interpretation and medical image review.
The MICF is intended to be used with specific macOS host and Apple display platforms. The MICF is not compatible with head-mounted displays, laptops, tablets, phones, or other portable displays that are not desktop displays. The device is not intended for mammography.
# 6. Comparison with the Predicate Device
The subject and predicate device, as described by the identification (21 CFR 892.1940), are both instruments used for radiologic QA. Technologically, the subject device is a software-only device intended to calibrate a compatible general purpose OTS desktop display via a connection with a compatible macOS general purpose computer and external third-party colorimeter. The predicate device also provides display calibration and has historically provided radiologic calibration and QC testing to other medical displays. The MICF can be used as part of a medical facility's QA program, consistent with the intended use of the predicate device. Since both the subject and predicate device (per 21 CFR 892.1940) are intended to enable use of a display for primary radiology, they have the same intended use.
While the MICF adjusts a general purpose OTS Apple display and the predicate provides calibration for a medical display, these technological differences do not raise different questions of safety or effectiveness, given that software devices that provide calibration are well characterized and have a known history of low risk. Furthermore, the differences in technological characteristics are addressed through acceptable scientific methods, from which the data demonstrate substantial equivalence.
The subject device has been appropriately verified and validated through non-clinical testing to ensure that the device is substantially equivalent to the predicate. A complete comparison of the subject and predicate devices can be found in Table 1 below.
One Apple Park Way, Cupertino, CA 95014
{6}
Apple Inc.
MICF Traditional 510(k) Submission
Table 1. MICF Comparison with the Predicate Device
| Item | Subject Device
Medical Imaging Calibration Feature
(MICF) | Predicate Device
21 CFR 892.1940 |
| --- | --- | --- |
| Device Name | Medical Imaging Calibration Feature | N/A |
| Manufacturer | Apple, Inc. | N/A |
| Regulation Number | 21 CFR 892.1940 | 21 CFR 892.1940 |
| Product Code | SHN | N/A |
| Regulation Name | Radiologic quality assurance instrument | Radiologic quality assurance instrument |
| Regulation Subset | Display calibration software | Display calibration software |
| Device Classification | Class I | Class I |
| OTC/Prescription | Rx Only | Rx Only |
| Intended Use | Quality assurance device intended to calibrate a display to performance standards to enable use in primary diagnostic radiology. | Quality assurance device intended to calibrate a display to performance standards to enable use in primary diagnostic radiology. |
| Indications for Use | The Medical Imaging Calibration Feature (MICF) is a software-only medical device intended to calibrate compatible general purpose off-the-shelf (OTS) desktop displays. The MICF is intended to be used by trained medical practitioners for the purpose of enabling primary diagnostic interpretation and medical image review.
The MICF is intended to be used with specific macOS host and Apple display platforms. The MICF is not compatible with head-mounted displays, laptops, tablets, phones, or other portable displays that are not desktop displays. The device is not intended for mammography. | N/A |
| Principle of Operation | Software-only medical device. Used with compatible OTS display, general purpose host computer, and external colorimeter. | Hardware and/or software medical device(s). Used with medical device displays, general purpose host computer, and external or internal colorimeter. |
| Use Environment | Anywhere primary diagnostic radiology is performed. | Anywhere primary diagnostic radiology is performed. |
One Apple Park Way, Cupertino, CA 95014
{7}
Apple Inc.
MICF Traditional 510(k) Submission
| Item | Subject Device
Medical Imaging Calibration Feature
(MICF) | Predicate Device
21 CFR 892.1940 |
| --- | --- | --- |
| Intended Users | Qualified healthcare practitioners who have been authorized by their facility to use display calibration software. | Qualified healthcare practitioners who have been authorized by their facility to use display calibration software. |
| Display Performance | Tested in conformance with DICOM, IEC 62563-1, and IEC 62563-2 standards for the purpose of enabling medical image review. Met acceptance criteria for primary radiology displays. | Tested in conformance with DICOM, IEC 62563-1, and IEC 62563-2 standards for the purpose of enabling medical image review. Met acceptance criteria for primary radiology displays. |
| Display active screen size (diagonal) | Minimum 68 cm/ 27" | N/A |
| Display resolution | 5120 x 2880 with no scaling | N/A |
| Display frame rate | 47 - 120 Hz | N/A |
| Luminance (maximum, minimum, recommended) | Recommended: 600 nits (for TG-18) or 350 nits (for TG-270) | N/A |
| Color imaging | Yes | N/A |
| Gray imaging | Yes | Yes |
| DICOM calibration tool(s) | Internal calibration software
External compatible colorimeter:
Calibrite Display Plus HL Model Number CCDIS3PL | Internal calibration software,
Internal/external colorimeter |
| Quality control procedures | AAPM TG-18 Compliant
AAPM TG-270 Compliant | AAPM TG-18 Compliant
AAPM TG-270 Compliant |
The subject device and predicate devices shares the same intended use and represent the same generic device type as as established by 21 CFR 892.1940. While there are differences of technological characteristics, these differences do not raise different questions of safety and effectiveness. Data from acceptable methods, including results from software verification, hardware verification, and end-to-end physical bench validation testing substantiate the intended use of the subject device and demonstrate that the subject device is substantially equivalent to the predicate device. Thus, the MICF is substantially equivalent to the predicate device, radiologic quality assurance instrument (21 CFR 892.1940).
## 7. Summary of Non-Clinical Testing
Apple conducted non-clinical testing on the MICF with passing results, supporting a determination of substantial equivalence. Non-clinical testing included the following:
One Apple Park Way, Cupertino, CA 95014
{8}
Software and Hardware Verification
As part of the software development life cycle, extensive software and hardware verification testing was conducted to asses the safe and effective use of the MICF with its supporting components. Software verification assessed if the software design was implemented and integrated per the integration plan (i.e. design outputs met design input requirements). Software verification assessed all code paths of the calibration algorithm, test and validation functions, and calibration reporting. Hardware verification assessed the interoperability of various macOS host platforms to support use of the MICF. Testing was conducted in accordance with with IEC 82304-1 Health Software - General requirements for product safety and IEC 62304 Medical Device Software - Software life cycle process Edition 1.1 2015. Pre-calibration testing was also conducted on the Apple display following recommendations in FDA's guidance document titled Content of Premarket Submissions for Device Software Functions. Results of verification testing demonstrated that the compatible Apple platforms supported use of the MICF, and that before MICF calibration the Apple display met technical requirements expected of displays used primary radiology.
### Bench Validation Testing
The bench validation testing demonstrated that the MICF provides sufficient adjustments to the display in accordance with requirements for medical imaging use. Validation consisted of end-to-end physical testing of the calibrated display in accordance with IEC 62563-1 Medical electrical equipment -- Medical image display systems -- Part 1: Evaluation methods, IEC 62563-2 Medical electrical equipment - Medical image display systems - Part 2: Acceptance and constancy tests for medical image displays, and AAPM TG-18 and TG-270 technical practice parameters. The results validated that the medical calibration adjustments provided by MICF meet the technological tolerance requirements specified in IEC 62563-2 and that the calibrated display meets the technical requirements necessary for use in primary diagnostic radiology. Furthermore, the results demonstrated that the MICF performs as intended when used in conjunction with an external colorimeter and that the general purpose computing platforms (i.e. Apple display) do not adversely affect the ability of the MICF to perform its calibration functions.
### General Purpose Computing Platform Assessment
The MICF is a software-only device, available on the compatible macOS host and Apple display general purpose computing platforms. For this reason, medical device hardware testing is not applicable. However, as a multiple function device product, the potential impacts of the general purpose computing platforms on MICF functionality were assessed as part of the device's risk management, verification, and validation activities, and determined to be acceptable.
## 8. Summary of Clinical Testing
Clinical testing data was not submitted.
## 9. Conclusion
The Medical Imaging Calibration Feature is substantially equivalent to the predicate device as both devices share the same intended use, and where there are differences in technological characteristics the differences do not raise different questions of safety and effectiveness. Furthermore, the scientific methods used for evaluating the different technological characteristics of the subject device are acceptable. Results of the testing support a decision of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.