Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Glean Urodynamics System Abdominal Sensor (GUS-1000-A)
Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Glean Urodynamics System Abdominal Sensor (GUS-1000-A)
Applicant
Bright Uro, Inc.
Product Code
EXQ · Gastroenterology, Urology
Decision Date
Apr 16, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1620
Device Class
Class 2
Indications for Use
The Glean Urodynamics System is a urodynamics analyzer system that is intended to quantify the pressure and flow characteristics of the lower urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies. The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment.
Device Story
Multi-channel urodynamics system; measures bladder and abdominal pressure to calculate detrusor pressure; performs Uroflow, CMG, UPP, and Micturition Studies. System components: Bladder Sensor, Abdominal Sensor, Insertion Tool, Uroflowmeter. Sensors are single-use, silicone-housed, inserted into bladder and anorectal canal; store data locally; wirelessly transmit data to clinician tablet/app after removal. Clinician uses Glean Mobile App to prepare sensors, log symptoms, and download data; Glean Web App used for data viewing/analysis. Patient uses Glean Mobile App as digital voiding diary. Enables ambulatory urodynamic monitoring in clinical settings. Output assists clinicians in diagnosing incontinence, retention, and micturition disorders.
Clinical Evidence
Bench testing only. Evaluations included biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), software verification and validation, cybersecurity risk management, human factors/usability engineering, packaging/shelf-life testing, and bioburden testing (USP <62>). Battery service life and performance verification tests were also conducted. All pre-determined acceptance criteria were met.
Technological Characteristics
Multi-channel urodynamics system. Components: silicone-housed bladder/abdominal sensors, insertion tool, uroflowmeter. Sensing principle: electronic pressure sensors (0-320 cm H2O range, 20 Hz sampling). Connectivity: wireless data transmission to mobile app. Software: Glean Mobile App (Clinician/Patient) and Glean Web App. Sterilization: not specified (packaging testing per ISO 11607).
Indications for Use
Indicated for adult patients requiring urodynamic testing, including Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP), and Micturition Studies, for the diagnosis of urinary incontinence, abnormal urinary retention, or neurological micturition disorders.
Regulatory Classification
Identification
A urodynamics measurement system is a device used to measure volume and pressure in the urinary bladder when it is filled through a catheter with carbon dioxide or water. The device controls the supply of carbon dioxide or water and may also record the electrical activity of the muscles associated with urination. The device system may include transducers, electronic signal conditioning and display equipment, a catheter withdrawal device to enable a urethral pressure profile to be obtained, and special catheters for urethral profilometry and electrodes for electromyography. This generic type of device includes the cystometric gas (carbon dioxide) device, the cystometric hydrualic device, and the electrical recording cystometer, but excludes any device that uses air to fill the bladder.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 16, 2026
Bright Uro, Inc.
Suranjan Roychowdhury, Ph.D., RAC
Chief Product Development Officer
3 Goddard
Irvine, CA 92618
Re: K253537
Trade/Device Name: Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Glean Urodynamics System Abdominal Sensor (GUS-1000-A)
Regulation Number: 21 CFR§ 876.1620
Regulation Name: Urodynamics Measurement System
Regulatory Class: II
Product Code: EXQ, EXY
Dated: November 18, 2025
Received: November 18, 2025
Dear Suranjan Roychowdhury:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253537 - Suranjan Roychowdhury, Ph.D., RAC
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253537 - Suranjan Roychowdhury, Ph.D., RAC
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Negeen Haghighi -S
for
Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology and Urology Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253537 | ? |
| Please provide the device trade name(s). | | ? |
| Glean Urodynamics System Male Delivery System (GUS-1000-M);
Glean Urodynamics System Female Delivery System (GUS-1000-F);
Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter));
Glean Urodynamics System Abdominal Sensor (GUS-1000-A) | | |
| Please provide your Indications for Use below. | | ? |
| The Glean Urodynamics System is a urodynamics analyzer system that is intended to quantify the pressure
and flow characteristics of the lower urinary tract. The system can be used in adult patients only to perform
standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and
Micturition Studies.
The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal
urinary retention, or neurological cases of micturition disorder. The system is intended to be used as
medical diagnostic equipment. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253537
Page 1 of 5
# 510(k) Summary
## 1. Submitter Information
510 (k) submitter: Bright Uro, Inc.
Address: 3 Goddard
Irvine, CA 92618, United States
Applicant Contact Person: Dr. Suranjan Roychowdhury
Chief Product Development Officer
Phone: 949-216-0873
Email: suranjan@brighturo.com
Preparation date: April 15, 2026
## 2. Device Name
Trade Name of the Device: Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Abdominal Sensor (GUS-1000-A)
Common Name: Urodynamics measurement system
Classification Name: Cystometer, Electrical Recording
Classification Regulation: 21 CFR 876.1620
Device Class: II
Panel: Gastroenterology/Urology
Product Code: EXQ, EXY
## 3. Predicate Devices
| | 510(k)# | Trade Name |
| --- | --- | --- |
| Primary Predicate | K243052 | Glean Urodynamics System |
| Secondary Predicate | 510(k) Exempt | Laborie Goby System |
The predicate devices have not been subject to a design related recall.
No reference devices were used in this submission.
## 4. Device Description
The subject device is a modified Glean Urodynamics System. The specific modification to the subject device consists of adding an Abdominal Sensor component (GUS-1000-A) to the system. The Abdominal Sensor
{5}
K253537
Page 2 of 5
measures abdominal pressure and enables detrusor pressure to be calculated using the Glean Urodynamics System (GUS) to conduct multi-channel urodynamics studies. Addition of the Abdominal Sensor component enables the subject device to perform multi-channel urodynamics studies. The modified Glean Urodynamics System (GUS) is a multi-channel urodynamic system indicated for standard Urodynamic tests such as Uroflow (UF), Cystometrogram (CMG), Urethral Pressure Profile (UPP), and Micturition Studies (MS).
The Glean Urodynamics System obtained market clearance through the original 510(k) Premarket Notification K243052. No changes were made to the intended use, indications for use, or technological characteristics of the predicate device's three physical components, the Bladder Sensor, Insertion Tool, and Uroflowmeter. The Glean Mobile App (Clinician) and Glean Web App software were modified to include use with the Abdominal Sensor. The Glean Mobile App (Patient) was not modified.
The multi-channel GUS consists of four physical component elements: Bladder Sensor, Abdominal Sensor, Insertion Tool, and Uroflowmeter, as well as three software applications: Glean Mobile App (Clinician), Glean Mobile App (Patient), and Glean Web App. The patient may use the Glean Mobile App as a digital voiding diary, logging fluid input, leakage, urgency, and other urological symptoms. The clinician may use the Glean Mobile App to prepare the Sensor for insertion, log symptoms, and download data. The Glean Web App may be used by clinicians to view and analyze data.
The Bladder Sensor is inserted through the urethra into the bladder using the Insertion Tool. Once inserted, the Sensor has a Removal String that hangs out of the urethra to enable removal of the Sensor. The Sensor may stay in the bladder for the entire duration of monitoring while collecting data. The Sensor stores data that may be wirelessly transmitted to the Glean Mobile App (Clinician) once it is removed from the body.
The Abdominal Sensor is manually inserted into the anorectal canal. Once inserted, the Abdominal Sensor has a removal string that extends from the anus to enable removal of the Sensor. The Sensor may stay in the rectum for the entire duration of monitoring while collecting data. The Sensor stores data that may be wirelessly transmitted to the Glean Mobile App used by the clinician once it is removed from the body.
The Uroflowmeter is used to measure voided volume and flow. The Glean Mobile App (Clinician) wirelessly receives data from the Uroflowmeter after the patient has completed a voiding cycle.
## 5. Indications For Use
The Glean Urodynamics System is a urodynamics analyzer system that is intended to quantify the pressure and flow characteristics of the lower-urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies.
The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment.
{6}
K253537
Page 3 of 5
# 6. Comparison of Technological Characteristics with Predicate Device
| Technological/ Performance Characteristics | Subject Device | Primary Predicate Device | Secondary Predicate Device |
| --- | --- | --- | --- |
| | Glean Urodynamic System with Abdominal Sensor | Glean Urodynamic System | Laborie Goby System (Exempt) |
| Intended Use / Indications for Use | The Glean Urodynamics System is a urodynamic analyzer system that is intended to quantify the pressure and flow characteristics of the lower-urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies. The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment. | The Glean Urodynamics System is a urodynamic analyzer system that is intended to quantify the pressure and flow characteristics of the lower-urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies. The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment. | The Goby™ system is a Urodynamic Analyzer system that is intended to quantify the pressure, flow, and EMG characteristics of the lower urinary tract. Using the available transducers, the system can be used for performing standard Urodynamic tests such as Uroflow, CMG, UPP, and Micturition Studies. The major application of Urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of Micturition disorder. The device is intended to be used as medical diagnostic equipment. |
| Principles of Operation | An electronic pressure sensor is used to measure pressure in the bladder and in the abdomen. This sensor is connected to an electrical recording device that stores measurements from the sensor. Once monitoring is complete, the sensor is removed from the patient's body, measurements are transferred from the sensor to a tablet, and the sensor discarded. | An electronic pressure sensor is used to measure pressure in the bladder. This sensor is connected to an electrical recording device that stores measurements from the sensor. Once monitoring is complete, the sensor is removed from the patient's body, measurements are transferred from the sensor to a tablet, and the sensor discarded. | A catheter is used to measure pressure in the bladder and in the abdomen. The catheter is connected to a sensor outside the patient. The sensor is connected to an electrical recording device that stores measurements from the sensor. Once monitoring is complete, the catheter is removed from the patient's body and discarded. |
| Anatomical location of use | The silicone housing with sensor and electronics are placed in the patient's bladder and anorectal cavity. | The silicone housing with sensor and electronics are placed in the patient's bladder. | The catheter tip is positioned in the bladder and anorectal cavity. The proximal portion of the catheter is taped to the patient's body. |
| Mode of Operation | Continuous | Continuous | Continuous |
| Pressure (Range) | 0 to 320 cm H2O | 0 to 320 cm H2O | 0 to 320 cm H2O (PLM001) -50 to 200 cm H2O (COM762) |
| Pressure Sampling Rate | 20 Hz (minimum rate) | 20 Hz | 20 Hz |
| Sensor/Catheter diameter | Less than 5.35 mm or 16 Fr (bladder) 17.5 mm maximum (abdominal) | Less than 5.35 mm or 16 Fr (bladder) | 7 Fr (bladder) 7 Fr (abdominal) |
| Sensor/ Catheter Material | Silicone | Silicone | TPU-PVD |
| Environment of Use | Clinic | Clinic | Clinic |
| Enables ambulatory urodynamic | Yes | Yes | Yes |
{7}
K253537
Page 4 of 5
| Technological/ Performance Characteristics | Subject Device | Primary Predicate Device | Secondary Predicate Device |
| --- | --- | --- | --- |
| monitoring | Glean Urodynamic System with Abdominal Sensor | Glean Urodynamic System | Laborie Goby System (Exempt) |
| Single Use Sensor/ Catheter | Yes | Yes | Yes |
| Method of Removal | String | String | Catheter |
| Integrated EMG | Yes | Yes | Yes |
| Integrated Uroflowmetry | Yes | Yes | Yes |
As evidenced by the above table, both the subject and the predicate devices have the same intended use, but the subject and predicate devices have different technological characteristics. Additionally, several types of performance testing were conducted on the subject device to evaluate its safety and effectiveness. Overall, it was established that the differences in technological characteristics between the subject and the predicate devices do not raise different questions of safety or effectiveness.
## 7. Non-Clinical Testing
Below is a list of the evaluations and tests that were performed and successfully completed for the subject device per the specified guidance and standards:
- Biocompatibility evaluation according to ISO 10993-1:2018 - Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process and FDA Guidance “Use of International Standard ISO 10993-1” (2016).
- Electrical Safety testing according to IEC 60601-1: 2020 - Medical electrical equipment – Basic safety and essential performance
- Electromagnetic Compatibility testing according to IEC 60601-1-2: 2020 - General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests
- Software Verification and Validation Testing according to FDA’s Guidance “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices”
- Cybersecurity risk management activities according to FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued September 27, 2023
- Human Factors Engineering/Usability Engineering (HFE/UE) processes were used as part of the design and development process for the subject device according to FDA Guidance, “Applying Human Factors and Usability Engineering to Medical Devices”
- Packaging testing according to ISO 11607-1:2019/Amd. 1:2023- Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including Amendment 1 (2023)] and Shelf Life testing
- Bioburden (for abdominal sensor only): USP <62> Microbiological Examination of Nonsterile Products: Tests for Specified Microorganisms
Additionally, battery service life and performance verification test data were submitted to establish performance and durability of the subject device.
{8}
K253537
Page 5 of 5
All pre-determined acceptance criteria were met.
## 8. Conclusions
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.