K253536 · Epineuron Technologies, Inc. · ETN · Feb 27, 2026 · Ear, Nose, Throat
Device Facts
Record ID
K253536
Device Name
Evala Nerve Stimulator (EPNR002)
Applicant
Epineuron Technologies, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Feb 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Evala is a single-use device intended to provide electrical stimulation of exposed motor nerves or muscle tissue to locate and identify nerves and to test nerve and muscle excitability.
Device Story
Handheld, sterile, single-use device; used intraoperatively by surgeons to locate/identify nerves and test nerve/muscle excitability. Device delivers charge-balanced, biphasic electrical stimulus via bipolar probe to nerve tissue; surgeon monitors resulting muscle responses. Internal battery-powered; embedded firmware controls stimulation parameters. User interface includes visual LED indicators and haptic (vibratory) feedback on device body to confirm stimulation delivery and status. Bipolar probe configuration allows localized stimulation field; adjustable output current levels accommodate various nerve sizes. Benefits include precise nerve identification and integrity testing during surgery.
Clinical Evidence
Bench testing only. Evaluated per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10 for electrical/thermal/mechanical safety. Biocompatibility verified per ISO 10993 series. Usability/human factors tested per IEC 62366-1. Sterility, packaging, and shelf-life validated per ISO 14937, ISO 11737, ISO 11607-1, and ASTM standards.
Technological Characteristics
Handheld, battery-powered, sterile, single-use stimulator. Bipolar probe configuration. Delivers biphasic rectangular waveform (no net DC). Embedded firmware control. Biocompatibility per ISO 10993; packaging per ISO 11607-1, ASTM D4332-22, ASTM D4169-23e1; sterility per ISO 14937, ISO 11737; shelf life per ASTM F1980-21. Electrical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10.
Indications for Use
Indicated for prescription use by surgeons to provide electrical stimulation of exposed motor nerves or muscle tissue to locate and identify nerves and to test nerve and muscle excitability.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 27, 2026
Epineuron Technologies, Inc.
Mike Willand
Chief Technology Officer, Co-Founder
1875 Buckhorn Gate
Unit 602
Mississauga, ON L4W5P1
Canada
Re: K253536
Trade/Device Name: Evala Nerve Stimulator (EPNR002)
Regulation Number: 21 CFR 874.1820
Regulation Name: Surgical Nerve Stimulator/Locator
Regulatory Class: Class II
Product Code: ETN
Dated: January 29, 2026
Received: January 29, 2026
Dear Mike Willand:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253536 - Mike Willand
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253536 - Mike Willand
Page 3
Sincerely,
**SHUCHEN PENG -S**
Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253536 | ? |
| Please provide the device trade name(s). | | ? |
| Evala Nerve Stimulator (EPNR002) | | |
| Please provide your Indications for Use below. | | ? |
| Evala is a single-use device intended to provide electrical stimulation of exposed motor nerves or muscle tissue to locate and identify nerves and to test nerve and muscle excitability. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K253536 | 510(k) Summary | Prepared on: 2026-02-04 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Epineuron Technologies Inc. | |
| Applicant Address | 1875 Buckhorn Gate Unit 602 Mississauga ON L4W5P1 Canada | |
| Applicant Contact Telephone | +1-905-206-0466 | |
| Applicant Contact | Mike Willand | |
| Applicant Contact Email | mwilland@epineurontech.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Evala Nerve Stimulator (EPNR002) | |
| Common Name | Surgical nerve stimulator/locator | |
| Classification Name | Stimulator, Nerve | |
| Regulation Number | 874.1820 | |
| Product Code(s) | ETN | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K150005 | Checkpoint Head & Neck | ETN |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Evala Nerve Stimulator is a small handheld device used by a surgeon to deliver electrical stimulation intraoperatively to test nerve integrity and muscle excitability. This is a sterile disposable device designed to be simple to use with one-handed control.
The device employs a bipolar stimulating probe to deliver a charge-balanced, biphasic electrical stimulus to nerve tissue, allowing surgeons to monitor the resulting muscle responses. The device contains embedded firmware that controls the function of the stimulator. The firmware cannot be modified by the user. The stimulator is powered by an internal battery which is not user replaceable. The surgeon is notified of stimulation delivery and device status via the visual indicators at the front of the device and vibratory (haptic) feedback felt on the body of the device. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| Evala is a single-use device intended to provide electrical stimulation of exposed motor nerves or muscle tissue to locate and identify nerves and to test nerve and muscle excitability. | | |
| Indications for Use Comparison | | 21 CFR 807.92(a)(5) |
| The predicate device (Checkpoint Head & Neck) has the same indications for use as the Evala Nerve Stimulator. | | |
| Technological Comparison | | 21 CFR 807.92(a)(6) |
| The subject and predicate device are similar based on the following technological elements: | | |
{5}
hand-held, sterile, single-use, disposable, integral stimulus probe, battery powered, intended for prescription use, regulated current, electrode material, software/firmware/microprocessor control, LED based visual indicator to notify the user that selected stimulus is being delivered to the probe, and biphasic rectangular waveform with no net DC current.
The differences between the two devices include the bipolar probe configuration, output current, frequency, and haptic motor. The bipolar probe facilitates a more localized and precise stimulation field compared to a monopolar probe. The subject device offers additional output current levels which allow users to further adjust the stimulation field to interface nerves of various sizes. Although the output frequency of the subject device is 4Hz higher, functionally there is no difference physiologically when using the subject device compared to the predicate device. The haptic motor is used to provide additional feedback to the user to communicate stimulation delivery and device status.
The underlying mechanism of operation of the subject device remains the same as the predicate device. The differences do not raise new questions regarding safety, or effectiveness.
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The Evala Nerve Stimulator was evaluated with recognized consensus standards for EMC, electrical, thermal & mechanical safety including IEC 60601-1, IEC 60601-1-2, and applicable subparts of IEC 60601-2-10.
Testing was conducted to verify device requirements & performance (internal verification & validation testing including software verification & validation in accordance with IEC 62304), biocompatibility (ISO 10993-1, 10993-5, 10993-7, 10993-11, 10993-12, and 10993-23), usability and human factors (IEC 62366-1), labeling (ISO 15223-1 & IEC 60601-1), packaging (ISO 11607-1, ASTM D4332-22 & ASTM D4169-23e1), sterile barrier (ASTM F2096 & ASTM F88), sterility (ISO 14937, ISO 11737), and shelf life (ASTM F1980-21).
The above testing was conducted to demonstrate substantial equivalence between the predicate and the subject device.
Final Assessment of Substantial Equivalence
The Evala Nerve Stimulator and Checkpoint Head & Neck stimulator are identical in their intended use, principles of operation, fundamental technology, and general operation. The differences in the user interface and probe configuration do not raise new questions regarding safety and effectiveness. Therefore, the Evala Nerve Stimulator is substantially equivalent to the Checkpoint Head & Neck.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.