K253517 · Reselute, Inc. · HSB · Mar 10, 2026 · Orthopedic
Device Facts
Record ID
K253517
Device Name
Reselute Tibial Nail
Applicant
Reselute, Inc.
Product Code
HSB · Orthopedic
Decision Date
Mar 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reselute Traditional Tibial Intramedullary Nail is intended to stabilize fractures of the proximal and distal tibia and the tibial shaft, open and closed tibial shaft fractures, certain pre and post isthmic fractures and tibial malunions and non-unions.
Device Story
Single-use intramedullary fixation system for tibia; includes tibial nail, cross-locking screws, and endcaps. Implanted by orthopedic surgeons in clinical/OR settings to stabilize bone fragments; prevents movement via locking screws; endcaps prevent proximal bone ingrowth. Benefits patient by providing mechanical stability to facilitate healing of fractures, malunions, and non-unions.
Clinical Evidence
Bench testing only. Mechanical performance validated via static/dynamic bending, static torsion, dynamic screw bending, metallic bone screw testing, and MR compatibility testing per ASTM F1264-16, F543-23, F2052, F2213, F2182, and F2119.
Technological Characteristics
Intramedullary fixation rod; material: titanium alloy (Ti-6Al-4V) per ASTM F136-13; includes cross-locking screws and endcaps; various diameters/lengths; single-use; mechanical performance validated per ASTM F1264-16 and F543-23.
Indications for Use
Indicated for patients requiring stabilization of proximal, distal, or tibial shaft fractures, including open/closed shaft fractures, pre/post isthmic fractures, malunions, and non-unions.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 10, 2026
Reselute, Inc.
Brianna Prindle
Quality and Regulatory Head
701 W Main St.
Suite 410
Durham, North Carolina 27701
Re: K253517
Trade/Device Name: Reselute Tibial Nail
Regulation Number: 21 CFR 888.3020
Regulation Name: Intramedullary fixation rod
Regulatory Class: Class II
Product Code: HSB
Dated: February 18, 2026
Received: February 18, 2026
Dear Brianna Prindle:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253517 - Brianna Prindle
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253517 - Brianna Prindle
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FARZANA SHARMIN -S
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253517 | ? |
| Please provide the device trade name(s). | | ? |
| Reselute Tibial Nail | | |
| Please provide your Indications for Use below. | | ? |
| The Reselute Traditional Tibial Intramedullary Nail is intended to stabilize fractures of the proximal and distal tibia and the tibial shaft, open and closed tibial shaft fractures, certain pre and post isthmic fractures and tibial malunions and non-unions. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253517 - Page 1 of 3
# Resolute
# 510(k) Summary
Date Prepared: March 10, 2025
The purpose of this submission is to seek clearance for a new device. This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
A. **510(k) Sponsor:**
Reselute Medical
701 W Main St.
Suite 410
Durham, NC 27701
B. **Primary Correspondent:**
Brianna Prindle
Quality and Regulatory Consultant
brianna@reselutemedical.com
(786) 521-0501 (direct)
C. **Premarket Notification:**
Trade Name: Reselute Tibial Nail
Regulation Name: Intramedullary Fixation Rod
Common Name: Rod, Fixation, Intramedullary And Accessories
Regulation Number: 21 CFR 888.3020
Product Code: HSB
Classification: II
Review Panel: Orthopedic
D. **Indications for Use:**
The Reselute Tibial Intramedullary Nail is intended to stabilize fractures of the proximal and distal tibia and the tibial shaft, open and closed tibial shaft fractures, certain pre and post isthmic fractures and tibial malunions and non-unions.
E. **Predicate Devices:**
The Reselute Tibial Intramedullary Nail is substantially equivalent to the following devices:
| 510(k) | Trade Name | Manufacturer |
| --- | --- | --- |
| Primary Predicate Device | | |
| K143245 | GPC Intramedullary Nailing Systems | GPC Medical Limited |
| Additional Predicate Device | | |
| K230761 | Trigen Meta Tibial Nail | Smith and Nephew |
Reselute, Inc.
{5}
K253517 - Page 2 of 3
# Resolute
## F. Device Description:
The Resolute Tibial Intramedullary Nail System is a single use intramedullary device for the fixation, correction, or stabilization of the tibia. The Resolute Tibial Intramedullary Nail implant system includes a tibial nail offered at various diameters and lengths to accommodate a range of patient anatomies, cross-locking screws to prevent movement of the nail relative to the bone fragments, and optional endcaps for preventing proximal bone ingrowth. Each implant component is manufactured from titanium alloy (Ti-6Al-4V), which conforms to ASTM F136-13 for medical implants.
## G. Comparison of Characteristics and Intended Use:
The subject Resolute Tibial Intramedullary Nail System is substantially equivalent to the primary predicate, GPC Intramedullary Nailing Systems (K143245) in material, design, technological characteristics, product features, and mechanical performance. The subject Resolute Tibial Intramedullary Nail System, primary predicate, and additional predicate, the Trigen Meta Tibial Nail, are all provided in similar sizing and manufactured from the same Ti-6Al-4V material per ASTM F136-13. The subject Resolute Tibial Intramedullary Nail System and primary predicate are provided with locking screws and endcaps.
The subject Resolute Tibial Intramedullary Nail System is substantially equivalent to the primary predicate, GPC Intramedullary Nailing Systems (K143245) in intended use and indications for use. The subject Resolute Tibial Intramedullary Nail System and primary predicate, the GPC Intramedullary Nailing Systems, are both intended for use for tibial fracture, tibial malunion and tibial non-unions. The subject and predicate device have identical indications for use.
## H. Performance Testing:
The subject Resolute Tibial Intramedullary Nail System was subjected to the following mechanical performance tests to support the assertion of substantial equivalence:
- Static bending per ASTM F1264-16
- Static torsion per ASTM F1264-16
- Dynamic bending per ASTM F1264-16
- Dynamic screw bending per ASTM F1264-16
- Metallic bone screw testing per ASTM F543-23
- MR Testing per ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119
The results of this non-clinical testing demonstrate that the mechanical performance of the subject Resolute Tibial Intramedullary Nail System is sufficient for its intended use and no different questions of safety or efficacy were identified during device testing; therefore, the subject Resolute Tibial Intramedullary Nail System is considered substantially equivalent to the GPC Intramedullary Nailing Systems (K143245).
Resolute, Inc.
{6}
K253517 - Page 3 of 3
# Resolute
I. Conclusions:
The Resolute Tibial Intramedullary Nail System was shown to be substantially equivalent in performance and has similar technological characteristics as the GPC Intramedullary Nailing Systems predicate through comparison in areas including design, intended use, implant materials, product features, mechanical performance, and function. Based on the data submitted, the Resolute Tibial Intramedullary Nail System does not raise any different questions about safety or efficacy. Therefore, the subject Resolute Tibial Intramedullary Nail System is considered substantially equivalent to the GPC Intramedullary Nailing Systems (K143245).
Resolute, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.