K253513 · Premier Dental Products Company · EBF · Nov 7, 2025 · Dental
Device Facts
Record ID
K253513
Device Name
MultiMatch Flow Chameleon AntiMicrobial Composite
Applicant
Premier Dental Products Company
Product Code
EBF · Dental
Decision Date
Nov 7, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
• Direct restorations in all cavity classes (I-V) • Base / liner under direct restoration • Veneering of discolored anterior teeth • Splinting of mobile teeth • Extended pit and fissure sealing in molars and premolars • Repair of composite/ceramic veneers • Retention of thermoplastic aligners The addition of anti-microbial particles to MultiMatch Flow Chameleon AntiMicrobial Composite reduces demineralization and the total amount of bacteria at the material surface, which is part of the caries-formation process.
Device Story
MultiMatch Flow Chameleon AntiMicrobial Composite is a low-viscosity, light-cured, radiopaque, nano-hybrid resin composite. It incorporates silica-based antimicrobial particles to reduce surface bacteria and demineralization. Used by dental professionals in clinical settings for direct restorations, splinting, sealing, and aligner retention. The material features a 'chameleon effect' allowing a single shade to match all 16 VITA shades. Applied via syringe or unit dose capsules, the composite is layered and light-cured. The antimicrobial property provides a prophylactic benefit by inhibiting caries-formation processes at the material-tooth interface. Clinical decision-making is supported by the material's mechanical strength, esthetic versatility, and documented antimicrobial performance.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 4049:2019, ISO 9917:2007, ISO 29022:2013, and ISO 17304:2013. Tests included flexural/compression strength, water sorption/solubility, cure depth/time, bond strength, and shrinkage. Biocompatibility testing (ISO 10993) confirmed no cytotoxicity, sensitivity, irritation, or genotoxicity. In vitro antimicrobial studies demonstrated significant reduction in total bacteria and demineralization volume compared to predicate, with efficacy against S. mutans, S. oralis, A. naeslundii, V. dispar, F. nucleatum, and C. albicans.
Technological Characteristics
Low-viscosity, light-cured, radiopaque nano-hybrid composite. Composition: UV-curable resins, glass filler, photo-initiator, co-initiator, silica-based antimicrobial particles, pigments. Standards: ISO 4049:2019, ISO 9917:2007, ISO 29022:2013, ISO 17304:2013. Delivery: Syringe and unit dose capsules. Non-sterile.
Indications for Use
Indicated for patients of any age requiring tooth restoration or repair, including direct restorations (Class I-V), base/liner placement, veneering of discolored teeth, splinting of mobile teeth, pit and fissure sealing, repair of composite/ceramic veneers, and retention of thermoplastic aligners.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 7, 2025
Premier Dental Products Company
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Drive,
Suite 510k
Saint Paul, Minnesota 55114
Re: K253513
Trade/Device Name: MultiMatch Flow Chameleon AntiMicrobial Composite
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth shade resin material
Regulatory Class: Class II
Product Code: EBF, DYH, EBC
Dated: November 5, 2025
Received: November 6, 2025
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253513 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253513 - Prithul Bom
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253513 | ? |
| Please provide the device trade name(s). | | ? |
| MultiMatch Flow Chameleon AntiMicrobial Composite | | |
| Please provide your Indications for Use below. | | ? |
| • Direct restorations in all cavity classes (I-V)
• Base / liner under direct restoration
• Veneering of discolored anterior teeth
• Splinting of mobile teeth
• Extended pit and fissure sealing in molars and premolars
• Repair of composite/ceramic veneers
• Retention of thermoplastic aligners
The addition of anti-microbial particles to MultiMatch Flow Chameleon AntiMicrobial Composite reduces demineralization and the total amount of bacteria at the material surface, which is part of the caries-formation process. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
premier
Inspired solutions for daily dentistry
K253513
# 510(k) SUMMARY
Premier Dental's MultiMatch Universal Chameleon Restorative Composite
## 1. Submitter Information
Premier Dental Products Company
1710 Romano Drive
Plymouth Meeting, PA 19462
Contact Person: Jessica Huang, Sr. Director of Quality/Regulatory
Phone: 610-239-6069
Facsimile: 610-239-6171
Date Prepared: November 3, 2025
## 2. Name of Device
Proprietary Name: MultiMatch Flow Chameleon AntiMicrobial Composite
Common Name: Tooth Shade Resin Material
Review Panel: Division of Dental and ENT Devices (DHT1B)
Classification Name: Material, Tooth Shade, Resin
Regulatory Class: 2
Product Code: EBF
Regulation Number: 21 CFR §872.3690
## 3. Predicate Devices
| Product | Applicant | 510(k) No. | Product Code | Predicate |
| --- | --- | --- | --- | --- |
| Infinix™ Flowable Composite | Nobio Ltd. | K201010 | EBF | Primary |
| Flowtain | Reliance Orthodontic Products, Inc. | K083051 | DYH | Reference |
## 4. Device Description
Premier's MultiMatch Flow Chameleon AntiMicrobial Composite is a low viscosity, light-cured, radiopaque, flowable nano-hybrid composite. It is for use in direct Class I-V restorations as a base liner, indirect restorations, and for retention of clear aligners. The restorative has an antimicrobial additive which reduces demineralization and total amount of bacteria at the material surface. The restorative is available in one chameleon shade to fit a variety of tooth-colored shades.
## 5. Indications for Use
^{}[]
{5}
premier
Inspired solutions for daily dentistry
Premier's MultiMatch Flow Chameleon Antimicrobial Composite has the following indications for use:
- Direct restorations in all cavity classes (I-V)
- Base / liner under direct restoration
- Veneering of discolored anterior teeth
- Splinting of mobile teeth
- Extended pit and fissure sealing in molars and premolars
- Repair of composite/ceramic veneers
- Retention of thermoplastic aligners
The addition of anti-microbial particles to MultiMatch Flow Chameleon AntiMicrobial Composite reduces demineralization and the total amount of bacteria at the material surface, which is part of the caries-formation process.
# 6. Summary of Technological Characteristics
Premier's to MultiMatch Flow Chameleon AntiMicrobial Composite is substantially equivalent to the predicate device, Nobio's Infinix Flowable Composite (K201010), in terms of indications for use, composition, physical properties, and technological characteristics. Both the subject device and the predicate device are methacrylate-based light curing resin systems containing filler particles to form a flowable paste which can be applied into a restoration site directly. The composite is designed to have high mechanical strength and durability while maintaining excellent handling and esthetic characteristics.
Premier's MultiMatch Flow formulation differs from the predicate device as it exhibits chameleon effect, which enables MultiMatch Flow to match all 16 VITA shades with only one universal shade. In contrast, the predicated device, Infinix Flowable (K201010) offers 3 shades in A1, A2, and A3.
The subject device, Premier's MultiMatch Flow, and the predicated device, Infinix Flowable (K201010), both contain silica-based antimicrobial particles. Both subject and predicated devices have the indication of reducing demineralization. Premier's MultiMatch Flow also has an indication for reducing the total bacteria amount on material surface.
A table comparing the key features of the subject and predicate devices is provided below.
| | Subject Device: MultiMatch Flow Chameleon AntiMicrobial Composite | Predicated Device: Nobio Infinix Flowable Composite (K201010) | Reference device: Flowtain (K083051) |
| --- | --- | --- | --- |
| Indications for Use | • Direct restorations in all cavity classes (I-V)
• Base / liner under direct restoration
• Veneering of discolored anterior teeth | •Restorations in the posterior region (Class I and II)
•Anterior restorations (Class III, IV) | • For the bonding of lingual retainers and splinting materials, creation of archwire stops and indirect |
^{}[]
{6}
premier
Inspired solutions
for daily dentistry
| | • Splinting of mobile teeth
• Extended pit and fissure sealing in molars and premolars
• Repair of composite/ceramic veneers
• Retention of thermoplastic aligners
The addition of anti-microbial particles to MultiMatch Flow Chameleon AntiMicrobial Composite reduces demineralization and the total amount of bacteria at the material surface, which is part of the caries-formation process. | • Class V restorations (cervical caries, root erosion, wedge-shaped defects)
• Veneering of discolored anterior teeth
• Splinting of mobile teeth
• Extended fissure sealing in molars and premolars
• Repair of composite/ceramic veneers
The addition of the QASi particles to the Infinix Universal Composite reduces demineralization, which is part of the caries-formation process. | bonding of orthodontic brackets.
• Can be used for the retention of thermoplastic aligners.
• Flowtain LV intended to be used for indirect bonding of orthodontic brackets and bonding to enamel surfaces. |
| --- | --- | --- | --- |
| Classification Name | Material, Tooth Shade, Resin | Material, Tooth Shade, Resin | Adhesive, bracket and tooth conditioner, resin |
| Class | 2 | 2 | 2 |
| Product Code | EBF | EBF | DYH |
| Target Population | Any age patient requiring restoration or repair of teeth. | Any age patient requiring restoration or repair of teeth. | Any age patient requiring restoration or repair of teeth. |
| Method of Use | Bonding agent is applied and the composite is layered on and light-cured for each layer. | Bonding agent is applied and the composite is layered on and light-cured for each layer. | Bonding agent is applied and the composite is layered on and light-cured for each layer. |
| Curing Mechanism | Light Cure | Light Cure | Light Cure |
| Composition | UV Curable resins, Glass Filler, Photo-initiator, Co-initiator, Silica-based antimicrobial particles and Pigments | UV Curable resins, Glass Filler, Photo-initiator, Co-initiator, Silica-based antimicrobial particles and Pigments | UV Curable resins, Photo-initiator, Glass filler |
^{}[]
{7}
premier
Inspired solutions for daily dentistry
| Technological Characteristics | Performance of the device meets the requirements of ISO 4049:2019, ISO 9917:2007, ISO 29022:2013, and ISO 17304:2013 | Performance of the device meets the requirements of ISO 4049:2019, ISO 9917:2007, ISO 29022:2013, ISO 17304:2013 | Performance of the device meets the requirements of ISO 29022:2013 |
| --- | --- | --- | --- |
| Sterility | Non-sterile | Non-sterile | Non-sterile |
| Shade | Single shade to match all 16 Vita Shades | 3 shades (A1, A2, A3) | - |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible |
| Antimicrobial Property | Reduces demineralization, Reduces total bacterial amount at material surface | Reduces demineralization | - |
| Delivery System | Syringe and Unit Dose Capsules | Syringe and Unit Dose Capsules | Syringe |
## 7. Non-Clinical Testing
### a. Performance
Performance tests for Premier's MultiMatch Flow Chameleon AntiMicrobial Composite were conducted following on the FDA's recognized consensus standards for dental composites. Bench tests, including flexural strength, compression strength, water sorption, water solubility, cure time, cure depth, shear bond strength, polymerization conversion and shrinkage, radiopacity, and color, were performed on Premier's MultiMatch Flow device against the predicate device Nobio Infinix Flowable Composite (K201010). The subject device passed all evaluations, demonstrated physical characteristics equal or better than the predicated device (K201010). The testing results confirmed Premier's MultiMatch Flow meets its intended physical and performance requirements.
### b. Biocompatibility
To demonstrate the overall biological safety of Premier's MultiMatch Flow Chameleon AntiMicrobial Composite, cytotoxicity, sensitivity, irritation, acute systemic toxicity, and genotoxicity tests were performed following ISO 10993 standards. MultiMatch Flow showed no observed cytotoxicity, no skin sensitivity, no mucosa irritation, no systemic toxic reaction, and no genotoxicity. Based on these results, MultiMatch Flow is biocompatible.
### c. Antimicrobial Testing
In Vitro Bacterial Biomarkers and In Vitro Demineralization tests were performed to support the indications of reducing demineralization and the total amount of bacteria at the material surface, which is part of the caries-formation process. The in vitro studies showed that Premier's MultiMatch Flow significantly reduced the total amount of bacteria at the material
^{}[]
{8}
premier
Inspired solutions for daily dentistry
surface and reduced the demineralization volume versus the predicated Nobio Infinix Flowable Composite (K201010).
Note- In vitro multispecies biofilm studies have shown MultiMatch™ Flow Chameleon Antimicrobial Composite's ability to act on the following organisms¹.
- Streptococcus mutans (GP)
- Streptococcus oralis (GP)
- Actinomyces naeslundii (GP)
- Veillonella dispar (GN)
- Fusobacterium nucleatum (GN)
- Candida albicans (Fungus)
1. Data on file
8. Conclusions
Premier's MultiMatch Flow Chameleon AntiMicrobial Composite is as safe and effective as the predicate device. Premiers MultiMatch Flow has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. In addition, the minor technological differences between Premier's Multimatch Flow and its predicate devices raise no new issues of safety or effectiveness. Non-Clinical Performance data demonstrate that Premier's Multimatch Flow is as safe and effective as predicated device of Infinix Flowable (K201010). Thus, Premier's Multimatch Flow is substantially equivalent.
^{}[]
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.