ACUSON Maple Diagnostic Ultrasound System; ACUSON Maple Select Diagnostic Ultrasound System
K253487 · Siemens Medical Solutions USA, Inc. · IYN · Nov 17, 2025 · Radiology
Device Facts
Record ID
K253487
Device Name
ACUSON Maple Diagnostic Ultrasound System; ACUSON Maple Select Diagnostic Ultrasound System
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Nov 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The ACUSON Maple ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, Musculoskeletal, and Intraoperative applications using different ultrasound transducers for different applications. The system supports the Ultrasound-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult patients with hepatic steatosis. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system. The ACUSON Maple Select ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Vascular, Urology, Echocardiography, and Musculoskeletal applications using different ultrasound transducers for different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
Device Story
Mobile, software-controlled diagnostic ultrasound system; acquires harmonic echo data via various transducers; displays images in B-Mode, M-Mode, Doppler (PW, CW, Color, Tissue), 3D/4D, and elastography modes. Operated by trained healthcare professionals in clinical settings. Includes analysis packages: Arterial Health Package (AHP) for Intima Media Thickness measurement; Virtual Touch UDFF for hepatic steatosis assessment; Virtual Touch pSWE for tissue stiffness. Output displayed on flat panel; used by physicians for clinical diagnosis and patient communication. Benefits include non-invasive anatomical measurement and diagnostic aid for liver and cardiovascular health.
Clinical Evidence
No clinical studies required for system equivalence. UDFF feature performance validated via two studies (N=31 each). Study 1: Pearson correlation 0.97 vs reference device (p=0.459), test-retest ICC 0.97. Study 2: Agreement with MRI-PDFF (mean absolute difference 2.95 percentage points, Bland-Altman limits -4.16 to 10.05), AUROC 0.85 for clinical accuracy. pSWE performance verified via phantom testing (3.0–75.0 kPa range).
Technological Characteristics
Mobile ultrasound system; harmonic imaging; 2D, M-mode, Doppler, 3D/4D modes. Materials/components comply with ISO 10993-1. Connectivity: standard ultrasound interface. Software: includes Auto IMT, Virtual Touch UDFF, Virtual Touch pSWE. Safety standards: IEC 60601-1, IEC 60601-2-37, IEC 62359. Sterilization: per manufacturer instructions for non-critical devices (ISO 17664-2).
Indications for Use
Indicated for adult patients with hepatic steatosis (UDFF tool) and general diagnostic ultrasound imaging (Abdominal, OB/GYN, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, Musculoskeletal, Intraoperative).
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
ACUSON Maple and ACUSON Maple Select Diagnostic Ultrasound System (K232617)
Reference Devices
ACUSON Sequoia Diagnostic Ultrasound System (K251481)
ACUSON Juniper Diagnostic Ultrasound System (K230207)
ACUSON NX3 Elite Diagnostic Ultrasound System (K192835)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Siemens Medical Solutions USA, Inc.
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Drive, Suite 510k
Saint Paul, Minnesota 55114
November 17, 2025
Re: K253487
Trade/Device Name: ACUSON Maple Diagnostic Ultrasound System
ACUSON Maple Select Diagnostic Ultrasound System
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic pulsed doppler imaging system
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, OIJ
Dated: November 12, 2025
Received: November 12, 2025
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253487 - Prithul Bom
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**YANNA S. KANG -S**
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{2}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253487 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| ACUSON Maple Diagnostic Ultrasound System
ACUSON Maple Select Diagnostic Ultrasound System | | |
| Please provide your Indications for Use below. | | ? |
| ACUSON Maple
The ACUSON Maple ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, Musculoskeletal, and Intraoperative applications using different ultrasound transducers for different applications. The system supports the Ultrasound-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult patients with hepatic steatosis. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
ACUSON Maple Select
The ACUSON Maple Select ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Vascular, Urology, Echocardiography, and Musculoskeletal applications using different ultrasound transducers for different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
Operating Modes
2D-mode
– 2D-mode with Harmonic imaging
– 2D-mode with Simultaneous imaging for biplane transducers*
Color flow Doppler
– Color (velocity)
– Power (energy)
Pulsed Wave Doppler
– Pulsed Wave Doppler Tissue Imaging
– High Pulsed Repetition Frequency Pulsed Wave Doppler
Continuous Wave Doppler
– Steerable Continuous Wave Doppler for phased array transducers
– Auxiliary Continuous Wave Doppler for pencil transducers
M-mode | | |
{3}
- M-mode with Harmonic imaging
- Anatomical M-mode
3D/4D Volume Imaging
Combined Modes
2D-mode with Color
2D-mode with Doppler
2D-mode with Color and Doppler
2D-mode with M-mode
2D-mode with M-mode and Color
2D-mode with Elastography
2D-mode with Contrast Agent Imaging
2D-mode with Anatomical M-mode
*Available only on the ACUSON Maple system.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
{4}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
# 510(k) Summary
# K253487
Date: Nov 14, 2025
1. Sponsor: Siemens Medical Solutions USA, Inc.
Ultrasound Division
22010 South East 51st Street, Issaquah,
Washington 98029
Primary Contact: JungIn Choi
Regulatory Affairs Specialist
+82 1046958694
Secondary Contact: Shilpa Rapaka
Regulatory Affairs Manager
+1 512-913-1053
2. Device Name: ACUSON Maple Diagnostic Ultrasound System
ACUSON Maple Select Diagnostic Ultrasound System
Common Name: Diagnostic Ultrasound System
Classification: Regulatory Class: II
Classification Panel: Radiology
| Classification Name | 21 CFR Section | Product Code |
| --- | --- | --- |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX |
| Biopsy Needle Guide Kit | 892.1560 | OIJ |
Manufacturing Site: Siemens Healthcare s.r.o.
Panattoni Park Kosice Airport ul. Andreja
Kvasa 5 040 17, Kosice-Barca, Slovakia
Page 1 of 9
{5}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
# 3. Legally Marketed Predicate and Reference Device
The ACUSON Maple Diagnostic Ultrasound System and ACUSON Maple Select Diagnostic Ultrasound System are multi-purpose diagnostic ultrasound systems with proprietary software and optional accessories and are substantially equivalent to the company's own Ultrasound devices.
- Predicate device:
ACUSON Maple and ACUSON Maple Select Diagnostic Ultrasound System (K232617)
- Reference devices:
ACUSON Sequoia Diagnostic Ultrasound System (K251481)
ACUSON Juniper Diagnostic Ultrasound System (K230207)
ACUSON NX3 Elite Diagnostic Ultrasound System (K192835)
# 4. Device Description
The ACUSON Maple Diagnostic Ultrasound System and ACUSON Maple Select Diagnostic Ultrasound System are multi-purpose, mobile, software-controlled, diagnostic ultrasound systems with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Their function is to acquire harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M mode, Doppler Tissue Image mode, Amplitude Doppler Mode, combination modes, Harmonic Imaging and 3D Imaging modes, or Harmonic Imaging and 4D imaging modes on a flat panel display for diagnostic ultrasound imaging.
# 5. Indications for Use
## ACUSON Maple
The ACUSON Maple ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, Musculoskeletal, and Intraoperative applications using different ultrasound transducers for different applications.
Page 2 of 9
{6}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
The system supports the Ultrasound-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult patients with hepatic steatosis.
The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
## ACUSON Maple Select
The ACUSON Maple Select ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Vascular, Urology, Echocardiography, and Musculoskeletal applications using different ultrasound transducers for different applications.
The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
## Operating Modes
- 2D-mode
- 2D-mode with Harmonic imaging
- 2D-mode with Simultaneous imaging for biplane transducers*
- Color flow Doppler
- Color (velocity)
- Power (energy)
Page 3 of 9
{7}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
Pulsed Wave Doppler
- Pulsed Wave Doppler Tissue Imaging
- High Pulsed Repetition Frequency Pulsed Wave Doppler
Continuous Wave Doppler
- Steerable Continuous Wave Doppler for phased array transducers
- Auxiliary Continuous Wave Doppler for pencil transducers
M-mode
- M-mode with Harmonic imaging
- Anatomical M-mode
3D/4D Volume Imaging
Combined Modes
2D-mode with Color
2D-mode with Doppler
2D-mode with Color and Doppler
2D-mode with M-mode
2D-mode with M-mode and Color
2D-mode with Elastography
2D-mode with Contrast Agent Imaging
2D-mode with Anatomical M-mode
*Available only on the ACUSON Maple system
## 6. Substantial Equivalence and Comparison of Technological Comparison to Predicate and Reference Devices
The purpose of this Traditional 510(k) is to introduce the following changes compared to the predicate and reference device:
- Software features:
- The addition of Auto IMT feature, which is substantially equivalent to the Auto IMT feature previously cleared on ACUSON Sequoia (K251481).
- The addition of Virtual Touch UDFF (Ultrasound-Derived Fat Fraction) feature which is substantially equivalent to the Virtual Touch UDFF feature previously cleared on the ACUSON Sequoia (K251481).
Page 4 of 9
{8}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
- The addition of Virtual Touch pSWE (Point Shear Wave Elastography) feature which is substantially equivalent to the Virtual Touch pSWE feature previously cleared on the ACUSON Juniper (K230207).
- New transducers and needle guides:
- The addition of 10BCC3 transducer, which is substantially equivalent to the BP10-3 transducer previously cleared on the ACUSON NX3 Elite (K192835).
- The addition of 6C1, 18H5, 9MC3, 10V4, 5P1, 11L4 transducers which are substantially equivalent to the previously cleared transducers on the ACUSON Juniper (K230207).
- The addition of needle guides for 10BCC3 transducer which is substantially equivalent to needle guides for BP10-3 transducer previously cleared on the ACUSON NX3 Elite (K192835).
- The addition of needle guides for 6C1, 9MC3, 11L4 transducers which are substantially equivalent to the needle guides for 6C1, 9MC3, 11L4 transducers previously cleared on the ACUSON Juniper (K230207).
The intended use, indications for use, use environment, technological characteristics, acoustic output, software features, hardware, compatible transducers, safety, and effectiveness of the subject devices are substantially equivalent to the predicate device and/or reference device.
# 7. Nonclinical Data
The ACUSON Maple and Maple Select Diagnostic Ultrasound Systems comply with the following standards and guidance documents:
- IEC 62359: Edition 2.1 2017-09, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields / Combines IEC 62359 (2010-10) and AMD 1 (2017-09)
- ANSI/AAMI ES60601-1:2005/A2:2021: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005/A1(2012)/A2(2020))
- IEC 60601-1:2005/A1(2012)/A2(2020), Medical electric equipment – Part 1: General requirements for basic safety and essential performance / This document and its separate amendments continue to be valid together with the consolidated version
- IEC 60601-1-2 Edition 4.1 2020-09, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-18: Edition 3.0 2009-08, Medical electrical equipment - Part 2-18: Particular
Page 5 of 9
{9}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
requirements for the basic safety and essential performance of endoscopic equipment
- IEC 60601-2-37 Edition 3.0 2024, Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- ISO 10993-1 Fifth edition 2018-08, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- IEC 60601-1-6:2010+A1:2013+A2:2020 Medical Electrical Equipment Part 1-6, General Requirements for Basic Safety and Essential Performance- Collateral standard: Usability
- ANSI AAMI ISO 14971: Medical devices- Applications of risk management to medical devices, 2019
- IEC 62304: Medical Device Software – Software life cycle process, 2006 + A 2015
- ISO 13485:2016 Medical devices – Quality management systems- Requirements for regulatory purposes
- ISO 17664-2: 2021-02 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices.
- FDA Ultrasound Guidance document, "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers," issued in February 2023
Non-Clinical verification testing has been performed addressing system level requirements according to system and design specifications, and risk control measures. Design Control activities to assure the safety and effectiveness of the subject devices include but are not limited to the following:
- Requirements Review
- Risk Analysis and Risk Management
- Product Specifications
- Design Reviews
- Safety Testing
- Verification and Validation
## 8. Clinical Data
## Summary of Clinical Tests
Since the ACUSON Juniper and Maple Diagnostic Ultrasound Systems use the same technology and principles as existing predicate devices, clinical studies were not required to support substantial equivalence.
{10}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
# UDFF Study Summary
Ultrasound-Derived Fat Fraction (UDFF) is a software tool that measures tissue in a selected region of interest and reports an index for assessing fat content in the liver. UDFF can be useful as an aid to a physician managing adult patients with hepatic steatosis.
The 6C1 transducer on the Condor (software platform) is compatible with the UDFF option. The UDFF feature in the Condor (functionality, algorithms for UDFF measurements and calculations, implementation, etc.) is identical (remains unchanged) when compared to the UDFF feature in the reference device.
Two different studies were conducted to evaluate the performance of the UDFF feature on the Condor. The demographic distribution of the study population in the two different studies are presented in the following table:
| Study 1 | Study 2 |
| --- | --- |
| N= 31 subjects | N= 31 subjects |
| Ethnicity/Country | Ethnicity/Country |
| USA = 15 | USA = 31 |
| Korea = 16 | |
| | Sex |
| Sex | Male = 11 |
| Male = 24 | Female = 20 |
| Female = 7 | |
| | Age (years) |
| Age (years) | Mean ±SD = 52.2 ± 10.1 |
| Mean ±SD = 41.4 ± 8.5 | Range = 31 – 75 |
| Range = 24 – 63 | |
| | BMI (kg/m2) |
| BMI (kg/m2) | Mean ± SD = 31.9 ± 8.2 |
| Mean ± SD = 27.5 ± 7.1 | Range = 20.5 – 56 |
| Range = 16.7 – 55.2 | |
{11}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
The study details and results are presented in the following table:
| Study No. | Study Objective | Evaluation Criteria | Results |
| --- | --- | --- | --- |
| 1 | To verify the correlation between the performance of the UDFF feature on Condor - ACUSON Juniper and Maple (Maple is the subject device) and the UDFF feature on ACUSON Sequoia (reference device) | Significance level p-Value < 0.05 in a two-tailed test | Pearson Correlation coefficient between UDFF on Condor and ACUSON Sequoia was 0.97, and the difference is not statistically significant (p=0.459 and t-statistic = -0.750) |
| | To verify the clinical reliability of the UDFF feature on the Condor | Test-retest ICC greater than 0.75 | ICC = 0.97. The reliability of the UDFF feature on Condor is considered excellent |
| 2 | To verify that UDFF of Condor shows good agreement with MRI-PDFF | Absolute mean difference less than 5 percentage points and Bland-Altman limits of agreement within ±15 percentage points | Condor UDFF and MRI’s PDFF had Bland-Altman limits of agreement between -4.16 to 10.05 percentage points with a mean absolute difference of 2.95 percentage points |
| | To verify the clinical accuracy of the Condor UDFF feature | AUROC greater than 0.80 | Performance of UDFF on Condor showed strong performance with AUROC 0.85 |
# pSWE Performance Testing- Phantom Testing
Verification testing for the Liver Elastography optimization was performed using calibrated elasticity phantoms certified by the phantom manufacturer. The table below lists the phantom stiffness range, pSWE modes, and the supported transducers. The verification results met the acceptance criteria for accuracy and variability.
| Exam type | Phantom Stiffness range | Mode | 6C1 |
| --- | --- | --- | --- |
| Abdomen | 3.0 – 75.0kPa | pSWE | ☑ |
Page 8 of 9
{12}
SIEMENS Healthineers
Siemens Medical Solutions USA, Inc.
Ultrasound Division
Traditional 510(k)
# 9. Summary
The subject and predicate devices are substantially equivalent in terms of intended use, indications for use, use environment, technological characteristics, acoustic output, software features, hardware, compatible transducers, safety, and effectiveness. For testing, all predetermined acceptance criteria were met. Results of these tests show that the proposed subject devices continue to meet their intended use, and the modifications made do not raise new or different questions of safety and effectiveness in support of a substantially equivalence determination.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.