← Product Code [GEI](/productcode/GEI) · K253484

# LUTRONIC GENIUS (K253484)

_Cynosure Lutronic Technology Corporation · GEI · Jun 29, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K253484

## Device Facts

- **Applicant:** Cynosure Lutronic Technology Corporation
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jun 29, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The LUTRONIC GENIUS is intended for use in dermatologic and general surgical procedures for electrocoagulation/contraction and hemostasis.

## Device Story

Lutronic Genius is an electrosurgical unit delivering bipolar radiofrequency (RF) energy to tissue via disposable handpiece tips containing microneedle arrays (e.g., 7x7 array). The device consists of a main body, handpiece, footswitch, and LCD touch screen. During operation, the clinician holds the handpiece at right angles to the target tissue; microneedles penetrate the epidermis, delivering RF energy to generate an electro-thermal reaction. This reaction causes tissue coagulation and contraction. Used in clinical settings by healthcare professionals for dermatologic and general surgical procedures. The device provides controlled thermal effects, aiding in hemostasis and tissue management. Benefits include precise energy delivery for coagulation and contraction, with adjustable depth and energy settings to suit clinical requirements.

## Clinical Evidence

No clinical data. Bench testing only. Thermal effect study performed on ex vivo tissues (Yucatan pig skin, bovine liver, bovine heart muscle) comparing Lutronic Genius (M49D/P14D tips) to Morpheus8 (Tip 40/Tip 12). Results showed consistent thermal lesion profiles and tissue contraction correlated with energy levels, demonstrating performance equivalence to the predicate.

## Technological Characteristics

Bipolar RF electrosurgical device; 460 kHz frequency; max output 50W. Features disposable microneedle array tips (various configurations). System includes 10.1" LCD touch screen, air cooling, and footswitch/fingerswitch activation. Electrical: 110-120V, 60Hz. Sterilization: EO gas. Biocompatibility per ISO 10993-1, -10, -23. Electrical safety per IEC 60601-1, 60601-2-2. EMC per IEC 60601-1-2. Usability per IEC 60601-1-6.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Lutronic GENIUS Radiofrequency System ([K180945](/device/K180945.md))
- The InMode System with the Morpheus8 Applicators ([K231790](/device/K231790.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** **U.S. FOOD & DRUG**  
ADMINISTRATION

June 29, 2026

Cynosure Lutronic Technology Corporation  
Nicole Getman  
Sr. Regulatory Affairs Specialist  
55 Samwon-Ro  
Deogyang  
Goyang Gyeonggi, 10550  
Republic Of Korea

Re: K253484

Trade/Device Name: Lutronic Genius

Regulation Number: 21 CFR 878.4400

Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories

Regulatory Class: Class II

Product Code: GEI

Dated: May 29, 2026

Received: May 29, 2026

Dear Nicole Getman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food & Drug Administration  
10903 New Hampshire Avenue  
Silver Spring, MD 20993  
www.fda.gov

{1}

K253484 - Nicole Getman

Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K253484 - Nicole Getman

Page 3

Sincerely,

Colin K.
Chen -S

Digitally signed by Colin
K. Chen -S
Date: 2026.06.29
13:31:05 -04'00'

Colin K. Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |
| --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253484  |
|  Please provide the device trade name(s).  |   |
|  LUTRONIC GENIUS  |   |
|  Please provide your Indications for Use below.  |   |
|  The LUTRONIC GENIUS is intended for use in dermatologic and general surgical procedures for electrocoagulation/contraction and hemostasis.  |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |

LUTRONIC GENIUS

Page 9 of 46

{4}

K253484

Page 1 of 4

CYNO\(_{\text{SURE}}\) LUTRONIC

# 510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

|  807.92(a)(1) – Submitter Information  |   |
| --- | --- |
|  Name | Lutronic Corporation  |
|  Address | Lutronic Center 219, Sowon-Ro Deogyang-Gu, Goyang-Si, KR 410220  |
|  Phone number | 888-588-7644  |
|  Fax number | N/A  |
|  Establishment Registration Number | 3004483538  |
|  Name of contact Person | Nicole Getman Manager, Regulatory Affairs 3/5 Carlisle Road Westford, MA, 01886  |
|  Date prepared | 06/26/2026  |
|  807.92(a)(2) - Name of device  |   |
|  Trade or proprietary name | LUTRONIC GENIUS  |
|  Common or usual name | Electrosurgical Unit and Accessories  |
|  Classification name | Electrosurgical, Cutting & Coagulation & Accessories  |
|  Classification panel | General and Plastic Surgery  |
|  Regulation | 21 CFR 878.4400  |
|  Product Code(s) | GEI  |
|  807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed  |   |
|   | Lutronic GENIUS Radiofrequency System K180945(12/10/2018) Predicate The InMode System with the Morpheus8 Applicators K231790 (07/20/2023) Secondary Predicate  |
|  807.92(a)(4) - Device description  |   |
|   | The LUTRONIC GENIUS includes the system main body, a handpiece equipped with disposable handpiece tips, footswitch, and an LCD touch screen control panel. The RF energy is delivered using disposable handpiece tips consisting of a 7x7 bipolar microneedle array. The bipolar microneedle array delivers radiofrequency energy to the target tissue: The microneedles come in light contact with the epidermis of the patient and minimally penetrate the epidermis during the treatment. The handpiece being held at right angles to the target tissue. As the RF energy passes into the skin via the needles, it generates an electro-thermal reaction capable of coagulating the tissue.  |

1

{5}

K253484

Page 2 of 4

|  807.92(a)(5) Intended use of the device  |   |
| --- | --- |
|  Indications for use | The LUTRONIC GENIUS is intended for use in dermatologic and general surgical procedures for electrocoagulation/contraction and hemostasis.  |

|  Company | Lutronic Corporation | Lutronic Corporation | InMode Ltd.  |
| --- | --- | --- | --- |
|  Model | LUTRONIC GENIUS | LUTRONIC GENIUS Radiofrequency System | The InMode System with the Morpheus8 Applicators  |
|  K No. | TBD | K180945 | K231790  |
|  Type | Bipolar RF (Radiofrequency) | Bipolar RF (Radiofrequency) | Bipolar RF (Radiofrequency)  |
|  Frequency | 460 kHz | 460 kHz | 50-60 Hz  |
|  Max Output Power [W] | 50W | 50W | 65W  |
|  Treatment Activation | Footswitch or Fingerswitch | Footswitch or Fingerswitch | Footswitch or Fingerswitch  |
|  Rx/OTC | Prescription | Prescription | Prescription  |
|  Treatment Duration (Time) | 10 – 600 ms | 10 -1000 ms | UKN  |
|  Needle Depth | Max 4.0 mm | Max 3.5 mm | Max 4mm Max 7 mm  |
|  No. of Needles | M49D: 49 each (7*7) G49D: 49 each (7*7) P16D: 16 each (4*4) P14D: 14 each (7*2) G25D: 25 each (1st line: 4, 2nd line: 3, Total 7 lines) | G49D: 49 each (7*7) | 12, 24, T-pin, & 40 pin tip head Resurfacing Tip: 24 each Prime Tip: 12 each Morpheus 8 Tip: 24 each Body 3D Tip: 40 each  |
|  Needle Diameter | 200 micrometers | 200 micrometers | UKN  |
|  Accessories | G49D Tip M49D Tip P16D Tip P14D Tip G25D Tip GeniusM Handpiece | G49D Tip M49D Tip GeniusSkin Handpiece Genius M Handpiece GeniusX Handpiece | Resurfacing Tip Prime Tip, Morpheus8 Tip Body 3D Tip Morpheus8 Applicator Morpheus8 Body  |

2

{6}

K253484

Page 3 of 4

|  User Interface | 10.1" Touch LCD Display | 10.1" Touch LCD Display | UKN  |
| --- | --- | --- | --- |
|  Sterilization Method | EO gas sterilization | EO gas sterilization | Gamma sterilization  |
|  Material of Contact Part | Handpiece Tip | Handpiece Tip | Handpiece Tip  |
|  System Cooling | Air Cooling | Air Cooling | UKN  |
|  Electrical Rating | Single phase AC110-120V, 60Hz Power consumption: 300VA (Fuse: 250V/6.3A) | Single phase AC110-120V, 60Hz Power consumption: 300VA (Fuse: 250V/6.3A) | 100-240 VAC  |
|  Cold Air Chiller Device Compatible | Yes | Yes | UKN  |
|  Weight | 27 kg | 27 kg | 30 kg  |
|  Dimension | 452 mm (W) x 582 mm (L) x 1824.5 mm (H) | 452 mm (W) x 582 mm (L) x 1540 mm (H) | 46cm W x 46cm D x 100cm H  |

807.92(b)(1) NON-CLINICAL TESTS SUBMITTED

# Electrical Safety

IEC 60601-1 :2005+ A 1:2012+ A2:2020 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance, FDA Recognition# 19-49

IEC 60601-2-2:2017 + AMD1: 2023 Medical electrical equipment-Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, FDA Recognition# 6- 389

# EMC

IEC 60601-1-2: 2014+AI :2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests, FDA Recognition# 19-36

IEC TS 60601-4-2:2024 Medical electrical equipment-Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems, FDA Recognition# 19-50

# Usability

IEC 60601-1-6: 201 0+AI :2013+A2:2020 -Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability, FDA Recognition# 5-132

# Biocompatibility

ISO 10993-1:2018 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity, FDA Recognition# 2-258

ISO 10993-10:2021 Biological evaluation of medical devices -Part 10: Tests for skin sensitization, FDA Recognition# 2-296

ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation, FDA Recognition# 2-291

3

{7}

K253484

Page 4 of 4

# **Software**

IEC 62304:2006+AI :2015 Medical Device-Software Life Cycle Processes, FDA Recognition# 13-79

# **Performance**

**Study Title:** Thermal Tests for the Evaluation of GENIUS RF Microneedling Device Vs Morpheus8 As predicate on Three Different Types of Ex Vivo Tissues

The purpose of the study is to assess the thermal effect of the Genius RF microneedling Device, equipped with M49D and P14D tips, using various parameters applied to ex vivo three types of tissues, Yucatan pig skin, Bovine liver, and muscle (derived from bovine heart). The study aims to evaluate the thermal effects of different modes and settings in accordance with the criteria outlined in the FDA Guidance for Industry: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery.

Morpheus8 microneedling device, equipped with Tip 40 (body) and Tip 12 (prime) is used as a predicate device in the proposed study. The study is being conducted for regulatory considerations by the sponsor to provide performance data for the Genius RF, demonstrating equivalence in its safety and treatment effect features.

The treatment of an ex vivo tissue model, bovine skin, liver, and muscle, using the Genius RF device, equipped with M49D and P14D tips compared to the predicate Morpheus8 RF device equipped with Tip 40 (body) and Tip 12 (prime) was performed. Treatment at various penetration depths demonstrated the expected correlation with energy levels and consistent thermal lesion profiles across the range of energy level settings, within the accuracy expected in histology analyses. Morpheus8 served as the predicate for the Genius device. The Genius, equipped with different tip configurations, demonstrated the formation of CMZs and tissue contraction comparable to those formed by Morpheus8. Considering the above, the Test and Predicate Devices demonstrated the expected correlation with energy levels and consistent thermal lesion profiles across the range of energy level settings, within the accuracy expected in histology analyses. It may be concluded that the Test Device is equally safe when compared to the predicate device.

# **807.92(b)(2) CLINICAL TESTS SUBMITTED**

No clinical tests were performed for this submission.

# **807.92(b)(3) Conclusion**

Based on the comparison of the technological characteristics, the specifications for the LUTRONIC GENIUS are the same or a subset of the predicate/secondary predicate device specifications. The indication is the same as the secondary predicate device, The InMode System with the Morpheus8 Applicators.

4

---

**Source:** [https://fda-staging.innolitics.com/device/K253484](https://fda-staging.innolitics.com/device/K253484)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com