K253484 · Cynosure Lutronic Technology Corporation · GEI · Jun 29, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253484
Device Name
LUTRONIC GENIUS
Applicant
Cynosure Lutronic Technology Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LUTRONIC GENIUS is intended for use in dermatologic and general surgical procedures for electrocoagulation/contraction and hemostasis.
Device Story
Lutronic Genius is an electrosurgical unit delivering bipolar radiofrequency (RF) energy to tissue via disposable handpiece tips containing microneedle arrays (e.g., 7x7 array). The device consists of a main body, handpiece, footswitch, and LCD touch screen. During operation, the clinician holds the handpiece at right angles to the target tissue; microneedles penetrate the epidermis, delivering RF energy to generate an electro-thermal reaction. This reaction causes tissue coagulation and contraction. Used in clinical settings by healthcare professionals for dermatologic and general surgical procedures. The device provides controlled thermal effects, aiding in hemostasis and tissue management. Benefits include precise energy delivery for coagulation and contraction, with adjustable depth and energy settings to suit clinical requirements.
Clinical Evidence
No clinical data. Bench testing only. Thermal effect study performed on ex vivo tissues (Yucatan pig skin, bovine liver, bovine heart muscle) comparing Lutronic Genius (M49D/P14D tips) to Morpheus8 (Tip 40/Tip 12). Results showed consistent thermal lesion profiles and tissue contraction correlated with energy levels, demonstrating performance equivalence to the predicate.
Technological Characteristics
Bipolar RF electrosurgical device; 460 kHz frequency; max output 50W. Features disposable microneedle array tips (various configurations). System includes 10.1" LCD touch screen, air cooling, and footswitch/fingerswitch activation. Electrical: 110-120V, 60Hz. Sterilization: EO gas. Biocompatibility per ISO 10993-1, -10, -23. Electrical safety per IEC 60601-1, 60601-2-2. EMC per IEC 60601-1-2. Usability per IEC 60601-1-6.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electrocoagulation, tissue contraction, and hemostasis. Prescription use only.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
The InMode System with the Morpheus8 Applicators (K231790)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 29, 2026
Cynosure Lutronic Technology Corporation
Nicole Getman
Sr. Regulatory Affairs Specialist
55 Samwon-Ro
Deogyang
Goyang Gyeonggi, 10550
Republic Of Korea
Re: K253484
Trade/Device Name: Lutronic Genius
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: May 29, 2026
Received: May 29, 2026
Dear Nicole Getman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253484 - Nicole Getman
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253484 - Nicole Getman
Page 3
Sincerely,
Colin K.
Chen -S
Digitally signed by Colin
K. Chen -S
Date: 2026.06.29
13:31:05 -04'00'
Colin K. Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | |
| --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253484 |
| Please provide the device trade name(s). | |
| LUTRONIC GENIUS | |
| Please provide your Indications for Use below. | |
| The LUTRONIC GENIUS is intended for use in dermatologic and general surgical procedures for electrocoagulation/contraction and hemostasis. | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
LUTRONIC GENIUS
Page 9 of 46
{4}
K253484
Page 1 of 4
CYNO\(_{\text{SURE}}\) LUTRONIC
# 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a)(1) – Submitter Information | |
| --- | --- |
| Name | Lutronic Corporation |
| Address | Lutronic Center 219, Sowon-Ro Deogyang-Gu, Goyang-Si, KR 410220 |
| Phone number | 888-588-7644 |
| Fax number | N/A |
| Establishment Registration Number | 3004483538 |
| Name of contact Person | Nicole Getman Manager, Regulatory Affairs 3/5 Carlisle Road Westford, MA, 01886 |
| Date prepared | 06/26/2026 |
| 807.92(a)(2) - Name of device | |
| Trade or proprietary name | LUTRONIC GENIUS |
| Common or usual name | Electrosurgical Unit and Accessories |
| Classification name | Electrosurgical, Cutting & Coagulation & Accessories |
| Classification panel | General and Plastic Surgery |
| Regulation | 21 CFR 878.4400 |
| Product Code(s) | GEI |
| 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | |
| | Lutronic GENIUS Radiofrequency System K180945(12/10/2018) Predicate The InMode System with the Morpheus8 Applicators K231790 (07/20/2023) Secondary Predicate |
| 807.92(a)(4) - Device description | |
| | The LUTRONIC GENIUS includes the system main body, a handpiece equipped with disposable handpiece tips, footswitch, and an LCD touch screen control panel. The RF energy is delivered using disposable handpiece tips consisting of a 7x7 bipolar microneedle array. The bipolar microneedle array delivers radiofrequency energy to the target tissue: The microneedles come in light contact with the epidermis of the patient and minimally penetrate the epidermis during the treatment. The handpiece being held at right angles to the target tissue. As the RF energy passes into the skin via the needles, it generates an electro-thermal reaction capable of coagulating the tissue. |
1
{5}
K253484
Page 2 of 4
| 807.92(a)(5) Intended use of the device | |
| --- | --- |
| Indications for use | The LUTRONIC GENIUS is intended for use in dermatologic and general surgical procedures for electrocoagulation/contraction and hemostasis. |
| Company | Lutronic Corporation | Lutronic Corporation | InMode Ltd. |
| --- | --- | --- | --- |
| Model | LUTRONIC GENIUS | LUTRONIC GENIUS Radiofrequency System | The InMode System with the Morpheus8 Applicators |
| K No. | TBD | K180945 | K231790 |
| Type | Bipolar RF (Radiofrequency) | Bipolar RF (Radiofrequency) | Bipolar RF (Radiofrequency) |
| Frequency | 460 kHz | 460 kHz | 50-60 Hz |
| Max Output Power [W] | 50W | 50W | 65W |
| Treatment Activation | Footswitch or Fingerswitch | Footswitch or Fingerswitch | Footswitch or Fingerswitch |
| Rx/OTC | Prescription | Prescription | Prescription |
| Treatment Duration (Time) | 10 – 600 ms | 10 -1000 ms | UKN |
| Needle Depth | Max 4.0 mm | Max 3.5 mm | Max 4mm Max 7 mm |
| No. of Needles | M49D: 49 each (7*7) G49D: 49 each (7*7) P16D: 16 each (4*4) P14D: 14 each (7*2) G25D: 25 each (1st line: 4, 2nd line: 3, Total 7 lines) | G49D: 49 each (7*7) | 12, 24, T-pin, & 40 pin tip head Resurfacing Tip: 24 each Prime Tip: 12 each Morpheus 8 Tip: 24 each Body 3D Tip: 40 each |
| Needle Diameter | 200 micrometers | 200 micrometers | UKN |
| Accessories | G49D Tip M49D Tip P16D Tip P14D Tip G25D Tip GeniusM Handpiece | G49D Tip M49D Tip GeniusSkin Handpiece Genius M Handpiece GeniusX Handpiece | Resurfacing Tip Prime Tip, Morpheus8 Tip Body 3D Tip Morpheus8 Applicator Morpheus8 Body |
2
{6}
K253484
Page 3 of 4
| User Interface | 10.1" Touch LCD Display | 10.1" Touch LCD Display | UKN |
| --- | --- | --- | --- |
| Sterilization Method | EO gas sterilization | EO gas sterilization | Gamma sterilization |
| Material of Contact Part | Handpiece Tip | Handpiece Tip | Handpiece Tip |
| System Cooling | Air Cooling | Air Cooling | UKN |
| Electrical Rating | Single phase AC110-120V, 60Hz Power consumption: 300VA (Fuse: 250V/6.3A) | Single phase AC110-120V, 60Hz Power consumption: 300VA (Fuse: 250V/6.3A) | 100-240 VAC |
| Cold Air Chiller Device Compatible | Yes | Yes | UKN |
| Weight | 27 kg | 27 kg | 30 kg |
| Dimension | 452 mm (W) x 582 mm (L) x 1824.5 mm (H) | 452 mm (W) x 582 mm (L) x 1540 mm (H) | 46cm W x 46cm D x 100cm H |
807.92(b)(1) NON-CLINICAL TESTS SUBMITTED
# Electrical Safety
IEC 60601-1 :2005+ A 1:2012+ A2:2020 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance, FDA Recognition# 19-49
IEC 60601-2-2:2017 + AMD1: 2023 Medical electrical equipment-Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, FDA Recognition# 6- 389
# EMC
IEC 60601-1-2: 2014+AI :2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests, FDA Recognition# 19-36
IEC TS 60601-4-2:2024 Medical electrical equipment-Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems, FDA Recognition# 19-50
# Usability
IEC 60601-1-6: 201 0+AI :2013+A2:2020 -Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability, FDA Recognition# 5-132
# Biocompatibility
ISO 10993-1:2018 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity, FDA Recognition# 2-258
ISO 10993-10:2021 Biological evaluation of medical devices -Part 10: Tests for skin sensitization, FDA Recognition# 2-296
ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation, FDA Recognition# 2-291
3
{7}
K253484
Page 4 of 4
# **Software**
IEC 62304:2006+AI :2015 Medical Device-Software Life Cycle Processes, FDA Recognition# 13-79
# **Performance**
**Study Title:** Thermal Tests for the Evaluation of GENIUS RF Microneedling Device Vs Morpheus8 As predicate on Three Different Types of Ex Vivo Tissues
The purpose of the study is to assess the thermal effect of the Genius RF microneedling Device, equipped with M49D and P14D tips, using various parameters applied to ex vivo three types of tissues, Yucatan pig skin, Bovine liver, and muscle (derived from bovine heart). The study aims to evaluate the thermal effects of different modes and settings in accordance with the criteria outlined in the FDA Guidance for Industry: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery.
Morpheus8 microneedling device, equipped with Tip 40 (body) and Tip 12 (prime) is used as a predicate device in the proposed study. The study is being conducted for regulatory considerations by the sponsor to provide performance data for the Genius RF, demonstrating equivalence in its safety and treatment effect features.
The treatment of an ex vivo tissue model, bovine skin, liver, and muscle, using the Genius RF device, equipped with M49D and P14D tips compared to the predicate Morpheus8 RF device equipped with Tip 40 (body) and Tip 12 (prime) was performed. Treatment at various penetration depths demonstrated the expected correlation with energy levels and consistent thermal lesion profiles across the range of energy level settings, within the accuracy expected in histology analyses. Morpheus8 served as the predicate for the Genius device. The Genius, equipped with different tip configurations, demonstrated the formation of CMZs and tissue contraction comparable to those formed by Morpheus8. Considering the above, the Test and Predicate Devices demonstrated the expected correlation with energy levels and consistent thermal lesion profiles across the range of energy level settings, within the accuracy expected in histology analyses. It may be concluded that the Test Device is equally safe when compared to the predicate device.
# **807.92(b)(2) CLINICAL TESTS SUBMITTED**
No clinical tests were performed for this submission.
# **807.92(b)(3) Conclusion**
Based on the comparison of the technological characteristics, the specifications for the LUTRONIC GENIUS are the same or a subset of the predicate/secondary predicate device specifications. The indication is the same as the secondary predicate device, The InMode System with the Morpheus8 Applicators.
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.