AeroDR TX c02

K253446 · Dk Medical Systems Co., Ltd. · KPR · Apr 10, 2026 · Radiology

Device Facts

Record IDK253446
Device NameAeroDR TX c02
ApplicantDk Medical Systems Co., Ltd.
Product CodeKPR · Radiology
Decision DateApr 10, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2
AttributesPediatric

Indications for Use

AeroDR TX c02 is a stationary X-ray system intended for obtaining radiographic images of various anatomical parts of the human body, both pediatrics and adults, in a clinical environment. AeroDR TX c02 is not intended for mammography, angiography, interventional, or fluoroscopy use.

Device Story

Stationary X-ray system; captures X-ray signals via detector; digitizes signals into DICOM images using Elui imaging software. Operated by licensed radiologic technologists in clinical environments. Supports patient postures including sitting, standing, or lying. Output used by clinicians for diagnostic screening of orthopedic, respiratory, or vertebral disc conditions. System components include X-ray tube, collimator, patient table, detector stand, and ceiling-suspended tube support. Benefits include diagnostic imaging of various anatomical parts.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, and IEC 60601-2-54 verified safety and EMC. Software, cybersecurity, and image quality performance were previously reviewed and cleared under K242119; no new clinical data required.

Technological Characteristics

Stationary X-ray system. Components: E7884X X-ray tube, BL-80 collimator, PT5-G1 K table, WS5-G1 K detector stand, CS5-G1 K ceiling support. Software: Elui imaging software (unchanged from K242119). Standards: IEC 60601-1 (3.2), IEC 60601-1-2 (4.1), IEC 60601-1-3 (2.2), IEC 60601-2-28 (3.0), IEC 60601-2-54 (2.0).

Indications for Use

Indicated for obtaining radiographic images of anatomical parts (head, spine, chest, abdomen) in pediatric and adult patients for diagnostic screening of orthopedic, respiratory, or vertebral disc conditions. Contraindicated for mammography, angiography, interventional, or fluoroscopy.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 10, 2026 Dk Medical Systems Co., Ltd. % Dong Ha Lee RA Consultant KMC, Inc. #509, Daerung Post Tower, 43, Digital-Ro 26-Gil, Guro-Gu, SEOUL, 08389 SOUTH KOREA Re: K253446 Trade/Device Name: AeroDR TX c02 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: March 13, 2026 Received: March 13, 2026 Dear Dong Ha Lee: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253446 - Dong Ha Lee Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253446 - Dong Ha Lee Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by GABRIELA M. RODAL -S for Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K253446 | | | Device Name AeroDR TX c02 | | | Indications for Use (Describe) | | | AeroDR TX c02 is a stationary X-ray system intended for obtaining radiographic images of various anatomical parts of the human body, both pediatrics and adults, in a clinical environment. AeroDR TX c02 is not intended for mammography, angiography, interventional, or fluoroscopy use. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} | 510(k) #: K253446 | 510(k) Summary | Prepared on: 2026-03-13 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | DK MEDICAL SYSTEMS Co., Ltd | | | Applicant Address | 52, Chupalsandan 1-Gil, Paengseong-Eup Pyeongtaek-Si Gyeonggi-do 17998 Korea, South | | | Applicant Contact Telephone | 82-2-535-7231 | | | Applicant Contact | Mr. YoungWoo Kang | | | Applicant Contact Email | ywkang@dk.co.kr | | | Correspondent Name | KMC, Inc | | | Correspondent Address | #509, Daerung Post tower, 43, Digital-ro 26-gil, Guro-gu Seoul 08389 Korea, South | | | Correspondent Contact Telephone | 82-70-8965-5554 | | | Correspondent Contact | Mr. DongHa Lee | | | Correspondent Contact Email | dhlee@kmcerti.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | AeroDR TX c02 | | | Common Name | System, X-Ray, Stationary | | | Classification Name | Stationary x-ray system | | | Regulation Number | 892.1680 | | | Product Code(s) | KPR | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K242119 | INNOVISION-EXII | KPR | | Device Description Summary | | 21 CFR 807.92(a)(4) | | AeroDR TX c02 can receive X-ray signals from X-ray irradiation and digitize them into X-ray images by converting digital images to DICOM image format using Elui imaging software. AeroDR TX c02 is a general radiography X-ray system and not for mammography nor fluoroscopy. In addition, the system must be operated by a user who is trained and licensed to handle a general radiography X-ray system to meet the regulatory requirements of a Radiologic Technologist. Target areas for examinations include the head, spine, chest, and abdomen for diagnostic screening of orthopedic, respiratory, or vertebral discs. The system can capture a patient's postures, such as sitting, standing, or lying. This system can be used for patients of all ages. The x-ray detectors used with the system are unchanged from the predicate (K242119). The device software is unchanged from the predicate and it has been cleared as cybersecure under K242119. | | | {5} | Intended Use/Indications for Use | 21 CFR 807.92(a)(5) | | --- | --- | | AeroDR TX c02 is a stationery X-ray system intended for obtaining radiographic images of various anatomical parts of the human body, both pediatrics and adults, in a clinical environment.AeroDR TX c02 is not intended for mammography, angiography, interventional, or fluoroscopy use. | | | Indications for Use Comparison | 21 CFR 807.92(a)(5) | | Indications for use is the same as the predicate device (existing device) (K242119). | | | Technological Comparison | 21 CFR 807.92(a)(6) | | The predicate device (existing device) is an own manufactured device cleared by FDA (K242119).The subject device (Modified device) is the same as the predicate device (existing device) except for the below.1. Trade Name1) Trade name is changed. (AeroDR TX c02)2. X-ray System Components (Configuration)1) Alternate X-ray tube is changed. (E7884X)2) Alternate Collimator is added. (BL-80)3) Patient table is changed. (PT5-G1 K)4) X-ray detector stand is changed. (WS5-G1 K)5) Ceiling suspended x-ray tube support is changed. (CS5-G1 K)The differences (changes) do not affect to the x-ray image generated from the x-ray system.The differences (changes) between the subject device and predicate device were tested and verified about the safety and effectiveness accordance with FDA recognized standards. The test results show that it did not raise new safety and effectiveness concerns.The subject device is determined to be Substantially Equivalent (SE) to the predicate device. | | | Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) | | Non-Clinical Tests1. EMC and Safety test: EMC and Safety tests were conducted to verify that the subject device (modified device) met all design changes as it is Substantially Equivalent (SE) to the predicate device (existing device). We claim conformance to the following standards.1) IEC 60601-1 (Edition 3.2, 2020-08)2) IEC 60601-1-2 (Edition 4.1, 2020-09)3) IEC 60601-1-3 (Edition 2.2, 2021-01)4) IEC 60601-2-28 (Edition 3.0, 2017-06)5) IEC 60601-2-54 (Edition 2.0, 2022-09)2. Software and Cybersecurity test: The device software and cybersecurity characteristics are unchanged from the predicate and the test results have been reviewed under K242119.3. Performance test: X-ray image quality are unchanged form the predicate and the clinical image evaluation results have been reviewed under K242119.Conclusion: The differences (changes) do not affect to the x-ray image generated from the x-ray system.The differences (changes) between the subject device and predicate device were tested and verified about the safety and effectiveness accordance with FDA recognized standards. The test results show that it did not raise new safety and effectiveness concerns.The subject device is determined to be Substantially Equivalent (SE) to the predicate device. | |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...