K253404 · Insera Therapeutics, Inc. · Jun 24, 2026 · CV
Device Facts
Record ID
K253404
Device Name
CLEAR™ Aspiration System
Applicant
Insera Therapeutics, Inc.
Decision Date
Jun 24, 2026
Decision
SESE
Submission Type
Traditional
Attributes
Therapeutic
Indications for Use
The CLEAR Aspiration System is intended for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
Device Story
CLEAR Aspiration System removes thrombus from peripheral vasculature via mechanical aspiration. System components: CLEAR Aspiration Pump, INSERA Canister, Non-sterile intermediate tubing, INSERA Aspiration Tubing, and INSERA Catheter. Pump and canister reside outside sterile field; catheter introduced over-the-wire through guide catheter/sheath to target vessel. Pump provides vacuum suction (0-29 inHg) with selectable modes (Cyclical, Uniform, Saver). Clinician operates system to aspirate occluded vessels. Output is aspirated material collected in canister. Benefits include effective thrombus removal to restore vessel patency.
Clinical Evidence
No human clinical data provided. Evidence consists of non-clinical bench testing (dimensional, suction pressure, flow rate, tensile strength, electrical safety per IEC 60601-1, EMC per IEC 60601-1-2, biocompatibility per ISO 10993, shelf life, and sterilization validation) and an animal study comparing the subject device to the predicate. Animal study showed similar in-vivo safety profiles, with no significant vessel injury or abnormal histology noted.
Technological Characteristics
System includes a digital aspiration pump, collection canister (1,200ml), aspiration tubing, and hydrophilic-coated catheters. Materials are biocompatible interventional grade. Vacuum range 0-29 inHg; flow rate 2.0-2.3 SCFM. Operates on 115/230Vac. Sterilization via Ethylene Oxide (EO). Complies with IEC 60601-1 and IEC 60601-1-2. Software-controlled digital pump interface.
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from peripheral arterial and venous vessels.
Predicate Devices
Penumbra Embolectomy Aspiration System (Indigo® Aspiration System) (K142870)
Reference Devices
Indigo® Aspiration System - Aspiration Catheter 12 and Separator 12 (K192981)
Indigo Aspiration System® with modified 110 Aspiration Tubing (K180939)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 24, 2026
Insera Therapeutics, Inc.
% Lina Kontos
Regulatory Counsel
Hogan Lovells US LLP
555 Thirteenth St. NW,
Washington, DC 20004 USA
Re: K253404
Trade/Device Name: CLEAR™ Aspiration System
Regulation Number: 21 CFR 870.5150
Regulation Name: Embolectomy catheter
Regulatory Class: Class II
Product Code: QEZ
Dated: May 22, 2026
Received: May 22, 2026
Dear Lina Kontos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253404 - Lina Kontos
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253404 - Lina Kontos
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by
KRINA M. PATEL -S
Date: 2026.06.24
16:26:23 -04'00'
For,
Gregory O'Connell
Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253404
Device Name
CLEAR™ Aspiration System
Indications for Use (Describe)
The CLEAR Aspiration System is intended for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-4740 EF
{4}
Traditional 510(k) Notice
CLEAR™ Aspiration System
insera
TECHNOLOGIES
K253404
Page 1 of 6
# 510(k) Summary
## Contact Details
**Applicant Name:** Insera Therapeutics, Inc.
**Applicant Address:** 5546 Serrano Ave, Dallas, TX 75248, USA
**Applicant Contact Telephone:** +1 (646) 270-3531
**Applicant Contact:** Vallabh Janardhan, MD, FAHA
**Applicant Contact Email:** vallabh@inseratherapeutics.com
**Date prepared:** June 24, 2026
## Device Name
**Device Trade Name:** CLEAR™ Aspiration System
**Common Name:** Embolectomy catheter
**Classification Name:** Peripheral Mechanical Thrombectomy With Aspiration
**Regulation Number:** 870.5150
**Product Code(s):** QEZ (Class II)
## Legally Marketed Predicate Devices
The legally marketed predicate device and reference devices of the subject device, CLEAR™ Aspiration System, are identified in **Table A** below.
**Table A - List of Predicate and Reference Devices for the CLEAR™ Aspiration System**
| 510k Number | Name of Device | Name of Manufacturer |
| --- | --- | --- |
| Predicate Device | | |
| K142870 | Predicate for Pump and Entire Aspiration System: Penumbra Embolectomy Aspiration System (Indigo® Aspiration System) | Penumbra, Inc. |
| Reference Devices | | |
| K192981 | Reference for Aspiration Catheter & System: Indigo® Aspiration System - Aspiration Catheter 12 and Separator 12 | Penumbra, Inc. |
| K180939 | Reference for Aspiration Tubing & System: Indigo Aspiration System® with modified 110 Aspiration Tubing | Penumbra, Inc. |
{5}
Traditional 510(k) Notice
CLEAR™ Aspiration System
insera
K253404
Page 2 of 6
## Indications for Use
The CLEAR Aspiration System is intended for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
## Device Description Summary
The CLEAR™ Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The CLEAR™ Aspiration System is comprised of several compatible components, working together to achieve the intended use:
- INSERA™ Canister and Non-sterile (intermediate) Tubing
- CLEAR™ Aspiration Pump
- INSERA™ Aspiration Tubing
- INSERA™ Catheter
The CLEAR Aspiration Pump, the INSERA Canister, plus the Non-sterile (intermediate) Tubing are provided non-sterile and are used outside the sterile field. The canister and the non-sterile (intermediate) tubing come pre-assembled, are single-use and function as a repository for aspirated material. The INSERA Canister is placed into the receptacle of the CLEAR Aspiration Pump while the Non-sterile Tubing is connected to the vacuum inlet port, located on the top of the pump.
Placement of the INSERA Catheter is achieved through over-the-wire introduction through a guide catheter or vascular sheath, followed by the advancement of the catheter, over a guidewire, to the target vessel. Once the desired location has been reached, the guidewire is removed, and the procedure can commence. The INSERA Catheter is used with the CLEAR Aspiration Pump to aspirate thrombus from an occluded vessel. The CLEAR™ Aspiration System supports several presets suction levels: Cyclical, Uniform, and Saver.
## Predicate Comparison
The CLEAR™ Aspiration System is substantially equivalent to the claimed predicate device; Penumbra Embolectomy Aspiration System (Indigo® Aspiration System) (K142870) based on comparisons of the intended use and technological characteristics.
Table B - Comparison Table of Subject Device to Predicate Device
| Attributes | CLEAR™ Aspiration System (Subject Device) | Indigo® Aspiration System (K142870) (Predicate Device) |
| --- | --- | --- |
| Regulation # | 21 CFR Part 870.5150 | 21 CFR Part 870.5150 |
| Regulation Name | Embolectomy Catheter | Embolectomy Catheter |
| Regulatory Class | II | II |
{6}
Traditional 510(k) Notice
CLEAR™ Aspiration System
![insera logo]() insera
INTERNATIONAL
K253404
Page 3 of 6
| Product Code | QEZ | DXE |
| --- | --- | --- |
| **Indications for Use** | The CLEAR Aspiration System is intended for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems. | The Penumbra Embolectomy Aspiration System (INDIGO™ Aspiration System) is intended for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems. Not for use in the coronaries or the neurovasculature. |
| **Aspiration Pump (Non-Sterile)** | | |
| **Control Panel/ Display** | Digital | Analog |
| **Vacuum Range** | 0 -29 inHg (98.2 kPa) | 0 -29 inHg (98.2 kPa) |
| **Flow Rate** | Flow Rate 2.0 – 2.3 SCFM (0-60.9 LPM) | Flow Rate 0-0.8 SCFM (0-23 LPM) |
| **Temperature** | Operating: 18°C - 24°C (65°F - 75°F) | Operating: 18°C - 24°C (65°F - 75°F) |
| **Relative Humidity (RH)** | Operating: 10% - 85% (<85% RH) | Operating: <75% RH |
| **Pressure** | Operating: 81kPa at Sea Level – 1828.8 meters (6000 feet); Range 70kPa – 105kPa | Operating: 81kPa at Sea Level – 1828.8 meters (6000 feet); Range 70kPa – 105kPa |
| **Voltage** | 115Vac/230Vac | 100-115Vac (Model PMX110) 230Vac (Model PMX220) |
| **Frequency** | 50 Hz/60 Hz | 50 Hz/60 Hz |
| **Duty Cycle** | Non-continuous 75% (90 mins on, 30 mins off) | Non-continuous 97.8% (45 mins on, 1 min off) |
| **IEC 60601-1 Compliance** | Yes | Yes |
| **IEC 60601-1-2 Compliance** | Yes | Yes |
| **Collection Canister and Intermediate Tubing (Non-Sterile)** | | |
| **Components (Materials)** | 1,200ml Setup: INSERA Collection Canister (1,200ml) & (Non-sterile Tubing) Intermediate Tubing | 1,000ml Setup: Indigo™ Collection Canister (1,000ml) & (Non-sterile Tubing) Intermediate Tubing Filter |
| **Volume** | 1,200ml | 1,000ml |
| **Use** | Single use, disposable | Single use, disposable |
| **Aspiration Tubing (Sterile)** | | |
| **Materials** | Biocompatible, commonly utilized for interventional devices. | Biocompatible, commonly utilized for interventional devices. |
| **Dimensions** | Usable Length = 100.0' Inner Diameter = | Usable Length = 112.0' Inner Diameter = |
{7}
Traditional 510(k) Notice
CLEAR™ Aspiration System
insera
K253404
Page 4 of 6
| | 0.110"Outer Diameter = 0.188"Distal length = N/A (Single Construction) | 0.110"Outer Diameter = 0.188"Distal length = 7.0" |
| --- | --- | --- |
| Sterilization | Ethylene Oxide (EO) | Ethylene Oxide (EO) |
| Use | Single use, disposable | Single use, disposable |
| Aspiration Catheter (Sterile) | | |
| Materials | Biocompatible, commonly utilized for interventional devices. | Biocompatible, commonly utilized for interventional devices. |
| Coating | Hydrophilic | Hydrophilic |
| Dimensions | SAME plus larger diameters | Appropriately sized for the target vessel |
| Sterilization | Ethylene Oxide (EO) | Ethylene Oxide (EO) |
### Performance Data
The following non-clinical performance bench tests were conducted for the CLEAR™ Aspiration System:
#### - Design verification testing
##### o Catheter
■ dimensional verification
■ suction pressure
■ flow rate
■ catheter/guidewire compatibility
■ coating integrity and particulate evaluation
■ lumen burst pressure
■ tensile strength
■ elongation
■ corrosion resistance
##### o Aspiration tubing
■ Dimensional/Visual Inspection
■ Suction Connector /Canister Lid (Luer/Port) Compatibility
■ Rotating Luer/RHV (Luer/Port) Compatibility
■ Aspiration Tubing Lumen Ovalization under Vacuum
■ Aspiration Tubing Joint Leak under Vacuum
##### o Pump
■ Flow Control (Digital) Switch Function
■ Aspiration System Compatibility with Aspiration Tubing
■ Suction Connector / Tubing Joint Tensile Strength
■ Rotating Male Luer / Tubing Joint Tensile Strength
- Durability test to determine the number of hours of continuous operation of the CLEAR™ Aspiration System
- Usability test to evaluate the usability, kink resistance, trackability, and performance attributes of the CLEAR™ Aspiration System compared to predicate device; Indigo®
{8}
Traditional 510(k) Notice
CLEAR™ Aspiration System
insera
K253404
Page 5 of 6
Aspiration System.
- Electromagnetic Compatibility (EMC) test in compliance with IEC 60601-1-2
- Electrical Safety test in compliance with IEC 60601-1
- ISO 10079-1 compliance assessment
- Packaging Validation test
- Software Validation and Cybersecurity
- Sterilization Validation in accordance with ISO 11138
- Shelf Life Tests: Package integrity tests in accordance with ISTA 3A and shelf life validation in accordance with ASTM F1980.
- Biocompatibility: Biocompatibility studies were selected in accordance with ISO 10993-1 as summarized in the table below:
Table C – List of biocompatibility tests
| Test | Relevant Standards | Test Result |
| --- | --- | --- |
| **INSERA Aspiration Tubing** | | |
| MEM Elution | ISO 10993-5 ISO 10993-12 | Pass |
| Maximization Test for Delayed-Type Hypersensitivity | ISO 10993-10 ISO 10993-12 | Pass |
| Intracutaneous Reactivity Test | ISO 10993-23 ISO 10993-12 | Pass |
| In-vivo Thrombogenicity | ISO 10993-4 | Pass |
| **INSERA Catheter** | | |
| MEM Elution | ISO 10993-5 ISO 10993-12 | Pass |
| Maximization Test for Delayed-Type Hypersensitivity | ISO 10993-10 ISO 10993-12 | Pass |
| Intracutaneous Reactivity Test | ISO 10993-23 ISO 10993-12 | Pass |
| Acute System Toxicity Test | ISO 10993-11 ISO 10993-12 | Pass |
| Rabbit Pyrogen-Material Mediated | ISO 10993-11 ISO 10993-12 | Pass |
| Hemolysis ASTM | ASTM F756-17 ISO 10993-12 | Pass |
| Complement Activation Assay C3a and CS5-9 | ISO 10093-12 ISO 10993-1 ISO 10993-4 | Pass |
{9}
Traditional 510(k) Notice
CLEAR™ Aspiration System
insera
TECHNOLOGICALS
K253404
Page 6 of 6
## Animal Study
An animal study was conducted to evaluate the safe use of the CLEAR™ Aspiration System including the CLEAR™ Aspiration Pump, INSERA™ Catheters, INSERA™ Aspiration Tubing, and INSERA™ Canister and Non-sterile Tubing compared to the predicate device namely Penumbra's Indigo® Aspiration System and showed that the in-vivo safety profile was similar between the subject and predicate devices at both the acute and chronic time points. The studies concluded that:
- No vessel injury was noted on the final angiograms following the vessel response procedure.
- No abnormal gross or significant histology findings were noted in the test vessel segments. If any, these findings did not affect vascular integrity and were less than or similar to the control (predicate) device.
- The use of the devices resulted in no significant vascular response in these experimental conditions.
## Conclusion
Insera Therapeutics, Inc. believes that the subject device: CLEAR™ Aspiration System, is substantially equivalent to the claimed primary predicate device: Penumbra's Indigo® Aspiration System (K142870), based on comparisons of the intended use and technological characteristics. While some technological characteristic differences exist between the subject device and the primary predicate device, the differences do not raise different questions of safety or effectiveness. Reference devices support the use of the catheter, pump, and aspiration system together. Further, non-clinical performance data confirm that these differences do not raise any additional risks. Additionally, the performance data demonstrates that the subject device is substantially equivalent to the legally marketed primary predicate device.
The CLEAR™ Aspiration System therefore meets the Federal Food, Drug and Cosmetic Act criteria for 510(k) clearance of this device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.