The Single-use Balloon Dilatation Catheter is intended to endoscopically dilate strictures of the alimentary tract (involving biliary tract). The device is supplied sterile and intended for single use only.
Device Story
Single-use balloon dilatation catheter; used for endoscopic dilation of alimentary/biliary tract strictures. Device consists of catheter assembly, double-lumen tube, balloon, tip, and two-way valve; rapid exchange series includes torque device and rapid-exchange guidewire port. Operated by physician via flexible endoscope; requires external guidewire and balloon dilatation pressure pump (ISO 80369-7:2021 compliant connector). Balloon inflated to specific diameters (6-20mm) and lengths (30-80mm) to mechanically dilate strictures. Benefits patient by restoring patency to narrowed tracts. Used in clinical/endoscopy settings.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including biocompatibility (ISO 10993), sterility (ISO 11135), balloon appearance, dimensional verification, leakage, tensile force, radiopacity, compliance, burst pressure, fatigue, relief time, and simulated use testing.
Technological Characteristics
Materials: Pebax. Energy: Hydraulic (via external pressure pump). Dimensions: 2.3mm OD; 1800-2400mm length. Sterilization: Ethylene Oxide. Connectivity: ISO 80369-7:2021 compliant injection port. Configuration: Standard or rapid exchange (with torque device and guidewire port).
Indications for Use
Indicated for endoscopic dilation of alimentary tract strictures, including the biliary tract, in patients requiring such intervention.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 24, 2026
Hangzhou AGS MedTech Co., Ltd.
Kai Zhang
Official Correspondent
Building 2, 5 and 7, No.389 Xingzhong Road
Linping District
Hangzhou, Zhejiang 311103
China
Re: K253393
Trade/Device Name: Single-use Balloon Dilatation Catheter
Regulation Number: 21 CFR 876.5010
Regulation Name: Biliary Catheter And Accessories
Regulatory Class: Class II
Product Code: FGE, KNQ
Dated: May 21, 2026
Received: May 21, 2026
Dear Kai Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253393 - Kai Zhang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253393 - Kai Zhang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee, Ph.D.,
MBA
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
Submission Number (if known)
K253393
Device Name
Single-use Balloon Dilatation Catheter
Indications for Use (Describe)
The Single-use Balloon Dilatation Catheter is intended to endoscopically dilate strictures of the alimentary tract (involving biliary tract). The device is supplied sterile and intended for single use only.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
**AGS**
K253393^{}[] Page 1 of 4
# **510(k) Summary**
We submit this 510(k) Summary as per 21 CFR 807.92, it meets the content and format regulatory requirements.
# **CH2.03.1 Submitter**
| Submitted by/Owner: | Hangzhou AGS MedTech Co., Ltd. Building 2, 5 and 7, No.389 Xingzhong Road, Linping District, 311103 Hangzhou, Zhejiang, PEOPLE'S REPUBLIC OF CHINA |
| --- | --- |
| Establishment Registration Number: | 3010288205 |
| Registration Status: | Active |
| Contact Person: | Kai Zhang Phone: +86-0571-87671223 Fax: +86-0571-87671230 Email: zhangkai@bioags.com |
| Date Prepared: | September 23, 2025 |
# **CH2.03.2 Proposed Device**
| Trade Name: | Single-use Balloon Dilatation Catheter |
| --- | --- |
| Device Name: | Single-use Balloon Dilatation Catheter |
| Common Name: | Single-use Balloon Dilatation Catheter |
| Regulation class: | Class II |
| Regulation Number: | 876.5010 |
| Regulation Description: | Biliary catheter and accessories |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | FGE, KNQ |
| Product Code Name: | FGE: Stents, drains and dilators for the biliary ducts KNQ: Dilator, esophageal FDS: Gastroscope and accessories, flexible/rigid FDF: Colonoscope and accessories, flexible/rigid |
# **CH2.03.3 Predicate Device**
| Trade Name: | Reliant(TM) Multistage Dilatation Balloon Catheter |
| --- | --- |
| Device Name: | Reliant(TM) Multistage Dilatation Balloon Catheter |
| Common Name: | Reliant(TM) Multistage Dilatation Balloon Catheter |
| 510(k) Number: | K180418 |
| Regulation class: | Class II |
| Regulation Number: | 876.5010 |
| Regulation Description: | Biliary catheter and accessories |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | FGE, KNQ |
CH2.03-2
{5}
AGS
K253393
Page 2 of 4
| Product Code Name: | FGE: Stents, drains and dilators for the biliary ducts KNQ: Dilator, esophageal |
| --- | --- |
#### **CH2.03.4 Device Description**
The Single-use Balloon Dilatation Catheter is available in two types: standard and rapid exchange series. For the standard series, the Single-use Balloon Dilatation Catheter consists of a catheter assembly, double-lumen tube, balloon, tip, and two-way valve. For the rapid exchange series, it additionally includes a torque device, and there is a small hole on the catheter which is about 200mm far away from the distal end, it is used to insert the guide wire rapidly. The balloon can be inflated to three distinct sizes, 6-7-8mm, 8-9-10mm, 10-11-12mm, 12-13.5-15mm, 15-16.5-18mm, 18-19-20mm diameters; the balloon length can be 30mm, 55mm and 80mm. The effective length of the catheter is 1800mm, 1950mm, 2100mm and 2400mm with an outer diameter of 2.3mm. The device is provided sterile via Ethylene Oxide sterilization and is intended for single use only.
#### **CH2.03.5 Indication for use statement**
The Single-use Balloon Dilatation Catheter is intended to endoscopically dilate strictures of the alimentary tract (involving biliary tract). The device is supplied sterile and intended for single use only.
#### **CH2.03.6 A description of the accessories**
For Single-use Balloon Dilatation Catheter, no other devices and no other products that are not devices provided with it as accessories.
In clinical use:
1. The Single-use Balloon Dilatation Catheter must be used in conjunction with a flexible endoscope. The product is inserted through the flexible endoscope and advanced through its channel to the patient's lesion site, extending from the distal end of the endoscope channel.
2. The Single-use Balloon Dilatation Catheter must be used together with a guidewire. The guidewire is inserted through the guidewire lumen interface of the product and advanced until it extends from the distal end of the product.
3. The Single-use Balloon Dilatation Catheter must be used in combination with a balloon dilatation pressure pump. The injection port connector of this product complies with ISO 80369-7:2021 standards, enabling connection and fluid delivery with a compatible balloon dilatation pressure pump to inflate the balloon.
#### **CH2.03.7 Comparison of Technology Characteristics**
Our proposed device Single-use Balloon Dilatation Catheter is substantially equivalent to the predicate devices. The differences between the Single-use Balloon Dilatation Catheter and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below:
Table CH2.03.7 Comparison of technical characteristics
CH2.03-3
{6}
K253393^{}[] Page 3 of 4
| Item | | Proposed device Hangzhou AGS MedTech Co., Ltd. | Predicate device (K180418) Micro-Tech (Nanjing) Co., Ltd | Comparison |
| --- | --- | --- | --- | --- |
| Technical | Configuration | For the standard series, the Single-use Balloon Dilatation Catheter consists of Component, double-lumen tube, balloon, tip, and two-way valve. For the rapid exchange series, it additionally includes a torque device in Component, and there is a small hole on the catheter which is about 200mm far away from the distal end, it is used to insert the guide wire rapidly. | Tip, balloon, marker band, handle junction, and guidewire. | Similar. The predicate device includes an additional pre-loaded guidewire, while the subject device features an additional rapid exchange port. Apart from these differences, the main structures of the two products are identical. |
| | Effective length of catheter(mm) | 1800, 1950, 2100, 2400 | 1800, 2300 | Similar. The working length range of our product is 1800-2400mm, which is substantially consistent with the predicate device's range of 1800-2300mm. |
| Biological | Main Patient contact Material | Pebax | Pebax | Similar. Except for the pre-loaded guidewire, the proposed device and the predicate device share the same primary materials that contact the human body. |
### CH2.03.8 Performance Data
The Single-use Balloon Dilatation Catheter meets all design specifications and medical device standards for biocompatibility (ISO 10993) and sterility (ISO 11135). All necessary verification and validation tests have been performed on the proposed device to assure substantial equivalence to the predicate device. The tests include: Balloon appearance, Dimensional Verification, Leakage Testing, Peak Tensile Force, Radiopacity, Balloon Compliance (Diameter vs. Pressure), Balloon Rated Burst Pressure, Balloon Fatigue, Relief Time of Balloon, Guide Wire Compatibility, Endoscope Compatibility, and Simulated Use.
The non-clinical performance meets the design specification and shows substantial equivalence to the predicate device.
CH2.03-4
{7}
AGS
K253393^{}[] Page 4 of 4^{}[] ---
# **CH2.03.09 Clinical Test**
No Clinical test is included in this submission.
# **CH2.03.10 Conclusion**
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS MedTech Co., Ltd has demonstrated that proposed device Single-use Balloon Dilatation Catheter is substantially equivalent to the predicate devices.
CH2.03-5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.