“FLNT Base and FLNT Temp” are used for the fabrication removable or temporary dental structures, such as crowns and bridges using milling technology using CAD/CAM. Indications for Use: - Temporary anterior and posterior crowns; - Temporary anterior and posterior bridges; - Removable structures for dentures
Device Story
FLNT Base and FLNT Temp are PMMA-based polymer materials supplied as blocks, discs, or rods; used by dental professionals to fabricate temporary dental prostheses (crowns, bridges, dentures) via CAD/CAM milling technology. Device transforms raw polymer material into patient-specific dental structures through subtractive manufacturing. Healthcare providers design restorations using CAD software; milled output is placed in the patient's mouth as a temporary or removable structure. Benefits include restoration of dental function and aesthetics during treatment. Device is used in dental clinics or laboratories.
Clinical Evidence
No clinical data. Evidence consists of bench testing including flexural strength, water sorption, water solubility, and residual monomer content, compared against ISO 20795 and ISO 10477 standards. Biocompatibility evaluated per ISO 10993 series.
Technological Characteristics
Materials: Polymethyl methacrylate (PMMA) and colorants. Form factor: Blocks, discs, rods. Processing: Polymerization and CAD/CAM milling. Standards: ISO 20795 (FLNT Base), ISO 10477 (FLNT Temp), ISO 10993 (biocompatibility). Non-sterile.
Indications for Use
Indicated for patients requiring temporary anterior/posterior crowns, temporary anterior/posterior bridges, or removable denture structures.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 2, 2026
Liaoning Upcera Co., Ltd
% Jinfeng Ning
Manager
Manton Business and Technology Services
37 Winding Ridge
Oakland, New Jersey 07436
Re: K253380
Trade/Device Name: "FLNT Base" and "FLNT Temp"
Regulation Number: 21 CFR 872.3770
Regulation Name: Temporary Crown And Bridge Resin
Regulatory Class: Class II
Product Code: EBG, EBI
Dated: January 28, 2026
Received: January 28, 2026
Dear Jinfeng Ning:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253380 - Jinfeng Ning
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253380 - Jinfeng Ning
Page 3
Sincerely,

Bobak
Shirmohammadi -
S
For Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253380 | ? |
| Please provide the device trade name(s). | | ? |
| “FLNT Base” and “FLNT Temp” | | |
| Please provide your Indications for Use below. | | ? |
| “FLNT Base” and “FLNT Temp” are used for the fabrication removable or temporary dental structures, such as crowns and bridges using milling technology using CAD/CAM.
Indications for Use:
- Temporary anterior and posterior crowns;
- Temporary anterior and posterior bridges;
- Removable structures for dentures | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary
Page 1 of 5
# K253380
# 510(k) Summary:
This summary of 510k safety and effectiveness information is being submitted
In accordance with the requirements of 21CFR 807.92
# 1 Submitter & Foreign Manufacture Identification
Liaoning Upcera Co., Ltd
No.122 Xianghuai Road, Economic Development Zone, Benxi, Liaoning, China
Tel: (086)-24-45565006
Submitter's FDA Registration Number: 3010582952
www.upcera-dental.com
# 2 Contact Person

Primary Contact
Jinfeng Ning
Manton Business and Technology Services
37 Winding Ridge, Oakland, NJ 07436
Tel: 608-217-9358
Email: cyshen@aol.com
Secondary Contact
Charles Shen
Manton Business and Technology Services
37 Winding Ridge, Oakland, NJ 07436
Tel: 608-217-9358
Email: cyshen@aol.com
# 3 Date of Summary: November 23, 2025
# 4 Device Name:
Proprietary Name: FLNT Base
Common Name: Removable Denture
Classification Name: Resin, Denture, Relining, Repairing, Rebasing
Device Classification: II
Regulation Number: 21 CFR 872.3760
Panel: General Dental
Product Code: EBI
Proprietary Name: FLNT Temp
Common Name: Temporary Crown and Bridge Resin
Classification Name: Crown and Bridge, Temporary, Resin
{5}
Device Classification: II
Regulation Number: 21 CFR 872.3770
Panel: General Dental
Product Code: EBG
## 5 Predicate Device Information:
(1) K223706, “PMMA Block”, manufactured by “Huliang (Shanghai) Bio-Tech Co., Ltd.”
## 6 Device Description:
“FLNT Base” is a homogeneous polymer material composed of high-quality polymethyl methacrylate (PMMA) and cross-linking agents to improve the network structure through a unique polymerization molding technology. This material is further fabricated into a kind of temporary prosthesis, such as removable dentures. The FLNT Base is composed of PMMA and colorant. Restorations are designed and manufactured by a dental professional using CAD technology.
“FLNT Temp” is a homogeneous polymer material composed of high-quality polymethyl methacrylate (PMMA) and cross-linking agents to improve the network structure through a unique polymerization molding technology. It is indicated for manufacturing temporary crowns and bridges. Restorations are designed and manufactured by a dental professional using CAD technology.
“FLNT Base and FLNT Temp” are supplied in different shapes, such as blocks, discs, rods. It is also supplied in different colors, and single or multi-layer aesthetic effect.
## 7 Indications for Use:
“FLNT Base and FLNT Temp” are used for the fabrication removable or temporary dental structures, such as crowns and bridges using milling technology using CAD/CAM.
Indications for Use:
- Temporary anterior and posterior crowns;
- Temporary anterior and posterior bridges;
- Removable structures for dentures
## 8 Technological Comparison with Predicate Device
The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
510(k) Summary
{6}
Table 5.1: Comparison of Intended Use, Design, Material, and Processing
| Description | Subject Device | Predicate Device (K223706) |
| --- | --- | --- |
| Indication for Use | The devices are used for the fabrication removable or temporary dental structures, such as crowns and bridges using milling technology using CAD/CAM. Indications for Use: - Temporary anterior and posterior crowns; - Temporary anterior and posterior bridges; - Removable structures for dentures | PMMA BLOCK is used for the fabrication removable or temporary dental structures, such as crowns and bridges using milling technology using CAD/CAM. Indications for Use: - Temporary anterior and posterior crowns; - Temporary anterior and posterior bridges; - Removable structures for dentures; - Removable structures for therapeutic restorations (night guards, bite splints or occlusal splints). |
| Basic Design | Blocks, disc, and rod | Blocks, disc, and rod |
| Materials | PMMA and colorants | PMMA and colorants |
| Processing | Polymerization | Polymerization |
| Dimension | Various | Various |
| Single Use | Yes | Yes |
| Shade | 10 shades for “FLNT Base” 25 shades for “FLNT Temp” | Various shades |
| Aesthetic Effect | Single and multilayer aesthetic effect. | Information not available |
| Sterile | Non-sterile | Non-sterile |
Our device is essentially identical to the predicate device in terms of indications for use, design, material, and processing between our device and the predicate devices. The minor differences do not raise any safety and effectiveness concerns.
# 9 Summary of Biocompatibility
The subject device is substantially equivalent to the predicate devices that have been legally marketed for years and with no clinical adverse events. The formulation of new device does not contain any non-conventional chemicals compared to the legally marketed predicate device.
Biocompatibility tests were performed following the current version of ISO 10993 standards. The following biocompatibility end points were evaluated:
510(k) Summary
{7}
Cytotoxicity (ISO 10993-5)
- Irritation & Sensitization (ISO 10993-10)
- Acute Systematic Toxicity (ISO 10993-11)
- Subacute Subchronic Systematic Toxicity (ISO 10993-11)
- Material Mediated Pyrogenicity (ISO 10993-11)
- Implantation (ISO 10993-6)
- Chronic Systematic Toxicity (ISO 10993-11)
Genotoxicity (ISO 10993-3)
Carcinogenicity (ISO 10993-3)
## 10 Summary of Non-Clinical Testing:
The following table shows similarities and differences of the key performance between our device and the predicate devices.
Table 5.2: Comparison of Performance Testing
| Description | Subject Device | Predicate Device (K223706) |
| --- | --- | --- |
| Flexural Strength | FLNT Base > 65 PMa
FLNT Temp > 50 PMa | ISO 20795 > 65 PMa
ISO 10477 > 50 PMa |
| Water Sorption | FLNT Base ≤ 32 μg/mm³
FLNT Temp ≤ 40 μg/mm³ | ISO 20795 ≤ 32 μg/mm³
ISO 10477 ≤ 40 μg/mm³ |
| Water Solubility | FLNT Base ≤ 1.6 μg/mm³
FLNT Temp ≤ 7.5 μg/mm³ | ISO 10795 ≤ 1.6 μg/mm³
ISO 10477 ≤ 7.5 μg/mm³ |
| Residue Monomer | FLNT Temp < 2.2% | ISO 20795 < 2.2% |
Tests were conducted following applicable procedures outlined in the FDA recognized consensus standard of ISO 20795 (for FLNT Base) and ISO 10477 (for FLNT Temp), and results met all relevant requirements in the test standards, and are comparable to the predicate device in performance.
## 11 Summary of Clinical Study:
Clinical Study is not performed for this device.
510(k) Summary
{8}
510(k) Summary
Page 5 of 5
## 12 Substantial Equivalence Conclusion
It has been shown in this 510(k) submission that “FLNT Base and FLNT Temp” and their predicate devices have similar indications for use, similar composition, and biocompatibility, similar manufacturing process, and similar performance.
The difference between the “FLNT Base and FLNT Temp” and their predicate device do not raise any question regarding its equivalence.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, the subject device is respectively substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.