K253363 · Auxilium Biotechnologies, Inc. · JXI · Jan 17, 2026 · Neurology
Device Facts
Record ID
K253363
Device Name
NeuroSpan Bridge
Applicant
Auxilium Biotechnologies, Inc.
Product Code
JXI · Neurology
Decision Date
Jan 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5275
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NeuroSpan Bridge is indicated for the repair of peripheral nerve discontinuities where a tensionless repair can be made (i.e., by flexion of the extremity).
Device Story
NeuroSpan Bridge is a biodegradable, micro-channeled tubular implant for peripheral nerve repair; provides protective environment for axonal growth across transected nerve stumps. Device features internal micro-channels (200-300 μm ID) and overhangs for suturing; made of polycaprolactone (PCL). Implanted by surgeons in clinical settings; provides physical bridge between nerve ends. Surgeon places nerve stumps in apposition to internal structure; sutures device via overhangs. Device resorbs over time. Benefits include support for nerve regeneration in peripheral nerve injuries.
Clinical Evidence
No human clinical data. Evidence consists of GLP animal studies: 1) Rat sciatic nerve transection (10 mm gap, n=72) at 3 and 6 months; 2) Rabbit sciatic nerve transection (30 mm gap, n=39) at 5 months. Endpoints included neurobehavioral/functional assessments, retrograde tracing, and histopathology. Results showed similar nerve regeneration, inflammatory response, and functional outcomes between NeuroSpan Bridge and NeuraGen predicate. Bench testing confirmed mechanical resilience (tensile, suture retention, compression) and biodegradation profile.
Technological Characteristics
Tubular implant; material: polycaprolactone (PCL); internal structure: micro-channels (200-300 μm ID). Dimensions: 1.5-3.0 mm ID, 1-3 cm length. Energy source: none (passive). Sterilization: Ethylene oxide (ISO 11135:2014). Biocompatibility per ISO 10993-1. Mechanical testing per ASTM D4169-22 (transportation).
Indications for Use
Indicated for repair of peripheral nerve discontinuities in patients where tensionless repair is achievable via extremity flexion.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 17, 2026
Auxilium Biotechnologies, Inc.
Jacob Koffler
CEO
333 Jackson Plaza
Ann Arbor, Michigan 48103
Re: K253363
Trade/Device Name: NeuroSpan Bridge
Regulation Number: 21 CFR 882.5275
Regulation Name: Nerve Cuff
Regulatory Class: Class II
Product Code: JXI
Dated: September 30, 2025
Received: December 19, 2025
Dear Jacob Koffler:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253363 - Jacob Koffler
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253363 - Jacob Koffler
Page 3
Sincerely,
Yen-chih Lin-S
Digitally signed by Yen-chih Lin-S
Date: 2026.01.17 14:20:50
-05'00"
for Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K253363
Device Name
NeuroSpan Bridge 3 x 30 (PRT001-3030);
NeuroSpan Bridge 3 x 25 (PRT001-3025);
NeuroSpan Bridge 3 x 20 (PRT001-3020);
NeuroSpan Bridge 3 x 15 (PRT001-3015);
NeuroSpan Bridge 3 x 10 (PRT001-3010);
NeuroSpan Bridge 2.5 x 30 (PRT001-2530);
NeuroSpan Bridge 2.5 x 25 (PRT001-2525);
NeuroSpan Bridge 2.5 x 20 (PRT001-2520);
NeuroSpan Bridge 2.5 x 15 (PRT001-2515);
NeuroSpan Bridge 2.5 x 10 (PRT001-2510);
NeuroSpan Bridge 2 x 30 (PRT001-2030);
NeuroSpan Bridge 2 x 25 (PRT001-2025);
NeuroSpan Bridge 2 x 20 (PRT001-2020);
NeuroSpan Bridge 2 x 15 (PRT001-2015);
NeuroSpan Bridge 2 x 10 (PRT001-2010);
NeuroSpan Bridge 1.5 x 30 (PRT001-1530);
NeuroSpan Bridge 1.5 x 25 (PRT001-1525);
NeuroSpan Bridge 1.5 x 20 (PRT001-1520);
NeuroSpan Bridge 1.5 x 15 (PRT001-1515);
NeuroSpan Bridge 1.5 x 10 (PRT001-1510)
Indications for Use (Describe)
The NeuroSpan Bridge is indicated for the repair of peripheral nerve discontinuities where a tensionless repair can be made (i.e., by flexion of the extremity).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
K253363 Page 1 of 8
510(k) SUMMARY
NeuroSpan Bridge
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Auxilium Biotechnologies, Inc.
333 Jackson Plaza
Ann Arbor, MI 48103
Phone: (858) 699-9443
Email: j.koffler@auxiliumbio.com
Contact: Dr. Jacob Koffler, CEO
Tel: Phone: (858) 699-9443
Email: j.koffler@auxiliumbio.com
Date Prepared: January 16, 2026
Name of Device
Trade name: NeuroSpan Bridge
Common name: Cuff, Nerve
Classification name: Nerve cuff (21 CFR 882.5275)
Regulatory class: II
Product code: JXI
Predicate Device
Trade name: NeuraGen Nerve Guide (K011168)
Regulation: 882.5275 - Nerve cuff
Regulatory class: II
Product code: JXI
Device Description
The NeuroSpan Bridge is a biodegradable micro-channeled implant for the repair of peripheral nerve discontinuities. The NeuroSpan Bridge provides a protective environment for peripheral nerve repair after injury, and is designed to be an interface between the two transected nerve stumps in order to provide a bridge for axonal growth across the nerve gap.
The NeuroSpan Bridge is flexible while being strong enough to hold sutures. The NeuroSpan Bridge is made from polycaprolactone (PCL) and features overhangs on
{5}
K253363 Page 2 of 8
both ends for suturing, with marks showing the overhang depth. The overhang allows the surgeon to place the nerve stump with apposition to the internal microchannel structure and suture without damaging it. The NeuroSpan Bridge has multiple channels throughout the device for axon growth. The NeuroSpan Bridge is provided sterile, for single use, and in multiple diameter and length variations.
# Indications for Use
The NeuroSpan Bridge is indicated for the repair of peripheral nerve discontinuities where a tensionless repair can be made (i.e., by flexion of the extremity).
# Comparison of Technological Characteristics with the Predicate Device
The subject NeuroSpan Bridge device has similar technological characteristics to the predicate device. Both are flexible tubular implants made from biodegradable materials and are supplied sterile for single use in a range of sizes. Both devices are designed to support nerve regeneration after injury. Both subject and predicate devices are absorbed by the body over time. Both the subject and predicate devices come in a range of lengths and diameters. Both devices are implanted using the same techniques.
The subject device includes a microtubular structure in its interior lumen. The predicate device has an open lumen. The subject device is made from polycaprolactone. The predicate is made from collagen.
| | NeuroSpan Bridge (Subject device) | Neurogen Nerve Guide (K011168) (Predicate device) | Comparison |
| --- | --- | --- | --- |
| Regulation | 21 CFR 882.5275 | 21 CFR 882.5275 | Same |
| Product Code | JXI | JXI | Same |
| Resorbable | Yes | Yes | Same |
| Suturable | Yes | Yes | Same |
| Material | Poly-ε-caprolactone | Collagen | Equivalent function |
{6}
K253363 Page 3 of 8
| Indications For Use | The NeuroSpan Bridge is indicated for the repair of peripheral nerve discontinuities where a tensionless repair can be made (i.e., by flexion of the extremity). | NeuroGen™ Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity. | Same |
| --- | --- | --- | --- |
| Physical structure | Tubular structure with microchannels | Hollow tubular structure | Internal structure of the devices is different. |
| Mode of Action | Create microenvironment for neuron regeneration | Create microenvironment for neuron regeneration | Same |
| Size | 1.5 – 3.0 mm ID
1-3 cm length
Micro-channels: 200-300 μm ID | 1.5 mm - 7mm ID
2-3 cm length | Equivalent function |
| Biocompatibility | Yes | Yes | Same |
| Preparation for use | Rehydration | Rehydration | Same |
| Reusable | No | No | Same |
| Sterilization Method | Ethylene oxide | Ethylene oxide | Same |
| Packaging | Thermoformed tray with Tyvek lining | Double blister packaging | Equivalent function |
| Shelf life | 2 years | Unknown | Equivalent function |
## Performance Data
The technical characteristics of the subject and predicate devices have been tested using comparative bench tests, which showed very similar results. In addition,
{7}
K253363 Page 4 of 8
performance testing, including biocompatibility testing and animal testing, demonstrates similar safety and performance of the subject device and its predicate. The following performance testing has been conducted:
## Stability testing of the product and packaging
Stability of the NeuroSpan Bridge was demonstrated by the real-time and accelerated aging methodologies for the product and the packaging materials.
## Transportation testing
The transportation testing was conducted to demonstrate the ability of the NeuroSpan Bridge and its packaging to withstand the distribution environment. The testing was conducted according to ASTM D4169-22.
## Pyrogenicity testing
NeuroSpan Bridge is labeled as non-pyrogenic. Endotoxin testing was conducted to demonstrate that the product is non-pyrogenic.
## Biocompatibility testing
Biocompatibility testing was conducted according to FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."
| Test | Test Summary | Conclusions |
| --- | --- | --- |
| Cytotoxicity | The device was evaluated for potential cytotoxic effects following ISO 10993-5 guidelines. | The cytotoxicity test results do not raise any safety concerns. |
| Sensitization | The device was evaluated for the potential to cause sensitization in guinea pigs based on ISO 10993-10. The test animals showed no evidence of inducting sensitization. | Non-sensitizing |
| Irritation | The device was evaluated for the potential to cause irritation following intracutaneous injection in rabbits based on ISO 10993-23. The extracts of the test article show no evidence of irritation. | Non-irritant |
{8}
K253363 Page 5 of 8
| Test | Test Summary | Conclusions |
| --- | --- | --- |
| Pyrogenicity | The device was evaluated for the potential to induce a pyrogenic response following intravenous injection in rabbits based on ISO 10993-11. The total rise of temperature was within acceptable range. | Non-pyrogenic |
| Genotoxicity - Bacterial Reverse Mutation | The device was evaluated for the potential to induce reverse mutations in bacteria per ISO 10993-3. The result indicated that the extracts of the device were not genotoxic. | Non-genotoxic |
| Genotoxicity - Chromosomal Abnormality | The device was evaluated for the potential of clastogenicity per ISO 10993-3. It showed that this device is clastogenicity negative in cultured mammalian cells. | Non-genotoxic |
| Implantation Effects | The device was evaluated in a series of animal studies. | No adverse tissue responses |
| Acute systemic toxicity | The device was evaluated for acute systemic toxicity in mice based on ISO 10993-11. No change was observed in general conditions or in necropsy. | No acute systemic toxicity |
| Subacute/ Subchronic and Chronic toxicity | The device was evaluated for systemic effects following implant. No abnormality occurred in general conditions or surgical sites after implantation. Manufacturing residual testing was conducted | No systemic toxicity
No added risk for manufacturing residuals |
{9}
K253363 Page 6 of 8
| Test | Test Summary | Conclusions |
| --- | --- | --- |
| Hemocompatibility - Hemolysis testing | The device was evaluated for hemolytic potential when in contact with blood based on ISO 10993-4 and ASTM F756 standard. The test article extracts demonstrated that the test article was non-hemolytic. | The test article is non-hemolytic |
| Carcinogenicity | Endpoint was addressed with a toxicological risk assessment approach. | Justified by risk assessment. |
| Neurotoxicity | Neurotoxicity was evaluated in a series of animal studies. | No neurotoxic effects |
## Sterility Testing
Sterilization of the NeuroSpan Bridge is based on ISO 11135:2014 Annex E Single Batch Release Process.
For each sterilization batch, the testing conducted will include product sterility, method suitability (bacteriostasis and fungistasis), and ethylene oxide residuals. For final product release all sterility tests (comparative resistance, product sterility, bacteriostasis and fungi, and ethylene oxide residuals) will need to pass their respective assessment criteria.
A batch sterilization has been performed. This report is in Steris Batch Sterilization Validation Report and Appendices. There were no deviations reported in this testing.
Sterility validation testing included assessments of EO residuals and bacterial endotoxins.
## Bench Testing
Bench testing was conducted to assess the mechanical resilience of the NeuroSpan Bridge device in tension and compression, as well as its ability to retain sutures under load and the rate of biodegradation. Results were compared against the performance of the predicate device, NeuroGen, under the same conditions.
| Test | Test Method Summary | Results |
| --- | --- | --- |
| Tensile strength | The device is cut into a piece, and then pull both ends of the test piece and measure the strength when it breaks. | Pass |
{10}
K253363 Page 7 of 8
| Suture retention strength | After passing a suture with a needle though the wetted specimen, the device is sutured to blocks. The specimen is pulled, and the strength at which the device breaks is measured. | Pass |
| --- | --- | --- |
| Compression strength | A device is placed on a load cell and subjected to transverse compression cycles up to 50%. Devices were considered passing if channels remained open following compression. | Pass |
| Degradation | Devices chronically implanted in rat and rabbit animal models were assessed for material degradation after explantation. | Degradation profile characterized. |
# Animal Testing
Two GLP animal studies evaluated the effect of NeuroSpan Bridge on nerve regeneration. Study 1 was a comparison of the NeuroSpan Bridge, Neuragen Nerve Guide and an injury-only cohort using the Sprague Dawley rat sciatic nerve transection (10 mm defect) model at 3- and 6-month endpoints. A total of 72 animals were used across the two timepoints and three cohorts. Study endpoints included neurobehavioral assessments, functional assessments (e.g., cranial nerves, motor, sensory and coordination), retrograde nerve tracing, clinical pathology, gross pathology and histopathology for the assessment of safety, functional performance, nerve regeneration, inflammatory responses, and biodegradation. The results of functional testing were similar between the test (NeuroSpan Bridge) and control group (Neuragen Nerve Guide) at both timepoints, with some improvement in scores but there was no consistent trend noted for improvement in test (NeuroSpan Bridge) or control (Neuragen Nerve Guide) groups over the injury only group.
The gross and histological responses at the implantation sites were highly similar among all animals in both the test and control animals at the 3-month and 6-month time points. Histological observations of the test article demonstrated inflammation consistent with normal, expected bioresorption and healing from the surgical procedures with mean inflammation scores decreasing from 3 to 6 months.
Study 2 compared the subject device and predicate in a New Zealand white rabbit sciatic nerve transection (30 mm gap) model. A total of 39 rabbits were evenly distributed across three cohorts (test, control and injury-only) at a 5-month timepoint. Study endpoints included neurobehavioral assessments, functions assessments (e.g., posture, gait, toe spread, circumference of the limb under the knee), retrograde nerve tracing, clinical pathology, gross pathology, histopathology and bio-resorption for the assessment of safety and functional performance. The response to implant and functional outcomes were similar between the test and control groups. The gross and histological response at the test article implantation sites were highly similar among all animals and were similar between the test and control animals. Both test and control articles presented minimal to mild, common, and expected changes at the device/tissue interface related to
{11}
K253363 Page 8 of 8
encapsulation of the test and control articles and healing from the surgical procedures. A similar level of nerve regeneration as assessed by muscle weight and histological evidence was observed in test (NeuroSpan Bridge) and control (Neuragen Nerve Guide) groups.
## Conclusions
Performance testing demonstrated that the subject device performs similarly as the predicate device. Thus, the NeuroSpan Bridge is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.