← Product Code [DFH](/productcode/DFH) · K253358

# Diazyme Human Kappa (k) Free Light Chain Assay; Diazyme Human Lambda (¿) Free Light Chain Assay (K253358)

_Diazyme Laboratories, Inc. · DFH · Dec 19, 2025 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K253358

## Device Facts

- **Applicant:** Diazyme Laboratories, Inc.
- **Product Code:** [DFH](/productcode/DFH.md)
- **Decision Date:** Dec 19, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5550
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K253358 · Dec 19, 2025 | Diazyme Human Kappa (k) Free Light Chain Assay; Diazyme Human Lambda (¿) Free Light Chain Assay | Diazyme Laboratories, Inc. | Retrospective clinical serum samples from patients with confirmed MGUS and non-MGUS conditions | The sponsor used retrospective clinical samples to demonstrate the performance of the FLC assays in identifying MGUS and distinguishing it from other polyclonal immunostimulation conditions, and to evaluate serial monitoring in stable vs. progressive MGUS patients. | MGUS; Retrospective cohort; Clinical performance; Serum samples |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective cohort study | 270 clinically confirmed MGUS patients and 266 non-MGUS patients with polyclonal immunostimulation; Sample Size: 536 | Not applicable for this study | Positive rate in MGUS patients and negative agreement in non-MGUS patients |
| Retrospective longitudinal cohort study; Follow-up/Duration: 3 to 5 serial draws per patient | 84 subjects with clinically stable MGUS and 2 subjects with progressive clinical status (MGUS to MM); Sample Size: 86 | Not applicable for this study | Stability vs. progression classification based on FLC concentration changes |

## Indications for Use

The Diazyme Human Kappa (κ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS), in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only. The Diazyme Human Lambda (λ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS), in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.

## Device Story

Latex particle enhanced immunoturbidimetric assays for quantitative measurement of Kappa and Lambda FLC in serum. Input: patient serum samples. Operation: FLC in sample binds to anti-FLC antibodies coated on carboxyl polystyrene latex particles, causing agglutination; degree of turbidity measured optically on Roche cobas c 501 analyzer. Output: FLC concentration calculated via 6-point calibration curve interpolation. Used in clinical laboratories by technicians. Results used by physicians alongside other laboratory/clinical findings to aid diagnosis/monitoring of multiple myeloma and evaluation of MGUS. Benefits: provides quantitative data to assist in identifying and monitoring plasma-cell disorders.

## Clinical Evidence

Retrospective clinical studies evaluated MGUS performance. Study 1 (N=270 MGUS, N=266 non-MGUS) showed 53.7% positive rate in MGUS patients and 94.0% negative agreement in non-MGUS patients with polyclonal hypergammaglobulinemia. Study 2 (N=86) evaluated serial monitoring in 84 stable and 2 progressive MGUS subjects; 76/84 stable subjects were correctly identified as stable, and 2/2 progressive subjects were identified as progressive. Bench testing and analytical performance were referenced from previous submissions (K153394, K211648).

## Technological Characteristics

Latex particle enhanced immunoturbidimetric assay. Reagents: Tris buffer, carboxyl polystyrene particles coated with polyclonal rabbit/goat anti-Kappa or rabbit anti-Lambda FLC antibodies. Instrumentation: Roche cobas c 501 analyzer. Quantitative measurement via optical turbidity. Standalone reagent kit; requires external calibrators and controls.

## Regulatory Identification

An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.

## Predicate Devices

- Diazyme Human Kappa (κ) Free Light Chain Assay ([K211648](/device/K211648.md))
- Diazyme Human Lambda (λ) Free Light Chain Assay ([K211648](/device/K211648.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K253358

B Applicant

Diazyme Laboratories, Inc.

C Proprietary and Established Names

Diazyme Human Kappa (κ) Free Light Chain Assay
Diazyme Human Lambda (λ) Free Light Chain Assay

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  DFH, DEH | Class II | 21 CFR 866.5550 - Immunoglobulin (Light Chain Specific) Immunological Test System | IM - Immunology  |

## II Submission/Device Overview:

A Purpose for Submission:

Modification of a previously cleared device: Addition of intended use as an aid in the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS).

B Measurand:

Kappa (κ) Free Light Chain (FLC)
Lambda (λ) Free Light Chain (FLC)

C Type of Test:

Latex particle enhanced immunoturbidimetric, quantitative

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov

{1}

III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

The Diazyme Human Kappa (κ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS), in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.

The Diazyme Human Lambda (λ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS), in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

The following statements are in the package insert of the Assays:

Warning: Kappa FLC result for a given specimen determined with assays and/or instrument platforms from different manufacturers can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the FLC assay used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of serially monitoring a patient, the assay method used for determining the Kappa FLC levels is changed, additional sequential testing should be carried out. Prior to changing assays, the laboratory should confirm baseline values for patients being serially monitored.

Warning: Lambda FLC results for a given specimen determined with assays and/or instrument platforms from different manufacturers can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the FLC assay used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of serially monitoring a patient, the assay method used for determining the Lambda FLC levels is changed, additional sequential testing should be carried out. Prior to changing assays, the laboratory should confirm baseline values for patients being serially monitored.

Evaluation of monoclonal gammopathy of undetermined significance (MGUS):

- The performance has not been sufficiently studied in Light Chain MGUS patients.

K253358 - Page 2 of 9

{2}

- Patients with renal disease or inflammation may have elevated levels of kappa and lambda free light chains (FLC).
- The performance has not been fully evaluated in all races/ethnicities in the intended use population.

## D Special Instrument Requirements:

Roche cobas c 501 analyzer

## IV Device/System Characteristics:

### A Device Description:

No modification is made to the kit components for the Diazyme Human Kappa (κ) and Lambda (λ) Free Light Chain Assays cleared in K153394.

The Diazyme Human Kappa (κ) Free Light Chain Assay is comprised of the following reagents:
- Reagent R1: Tris buffer solution
- Reagent R2: Carboxyl Polystyrene particles coated with polyclonal rabbit and goat anti-Kappa FLC antibodies
- Calibrators: A five-level set in liquid form (5 x 1 mL)

The Diazyme Human Lambda (λ) Free Light Chain Assay is comprised of the following reagents:
- Reagent R1: Tris buffer solution
- Reagent R2: Carboxyl Polystyrene particles coated with polyclonal rabbit anti-Lambda FLC antibodies
- Calibrators: A five-level set in liquid form (5 x 1 mL)

The controls are required materials but not provided for both assays. For each assay, the control set includes two levels of controls in serum based liquid form.

### B Principle of Operation:

No modification is made to the principle of operation for the Diazyme Human Kappa (κ) Free Light Chain Assay or the Diazyme Human Lambda (λ) Free Light Chain Assay.

The assays are based on a latex enhanced immunoturbidimetric assay. Kappa FLC or Lambda FLC in the sample binds to specific anti-Kappa FLC or anti-Lambda FLC antibody, which is coated on latex particles, and causes agglutination. The degree of the turbidity caused by agglutination can be measured optically and is proportional to the amount of Kappa FLC or Lambda FLC in the sample. The instrument calculates the Kappa FLC or Lambda FLC concentration by interpolation of obtained signal of a 6-point calibration curve prepared from calibrators of known concentrations.

### V Substantial Equivalence Information:

K253358 - Page 3 of 9

{3}

K253358 - Page 4 of 9

A Predicate Device Name(s):

Diazyme Human Kappa (κ) Free Light Chain Assay
Diazyme Human Lambda (λ) Free Light Chain Assay

B Predicate 510(k) Number(s):

K211648

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K253358
(Candidate Device) | K211648
(Predicate Device)  |
| --- | --- | --- |
|  Device Trade Name | Diazyme Human Kappa (κ) Free Light Chain Assay
Diazyme Human Lambda (λ) Free Light Chain Assay | Diazyme Human Kappa (κ) Free Light Chain Assay
Diazyme Human Lambda (λ) Free Light Chain Assay  |
|  General Device Characteristic Similarities  |   |   |
|  Analyte | Kappa FLC, Lambda FLC | Same  |
|  Assay Type | Quantitative | Same  |
|  Test Method | Turbidimetry | Same  |
|  Specimen Type | Serum | Same  |
|  Traceability | Internal Reference preparation | Same  |
|  Units | mg/L | Same  |
|  Detection Antibody | Polyclonal anti-human Kappa/anti-human Lambda FLC Ig fraction coated on polystyrene latex microparticles | Same  |
|  Analytical Measuring Interval | Kappa: 2.9–150 mg/L
Lambda: 3.5–200 mg/L | Same  |
|  Reference Interval | Kappa: 2.4–20.7 mg/L
Lambda: 4.2–27.7 mg/L
Ratio: 0.22–1.74 | Same  |
|  General Device Characteristic Differences  |   |   |
|  Intended Use/Indications for Use | The Diazyme Human Kappa (κ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in | The Diazyme Human Kappa (κ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers.  |

{4}

|  Device & Predicate Device(s): | K253358
(Candidate Device) | K211648
(Predicate Device)  |
| --- | --- | --- |
|   | conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS) in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.

The Diazyme Human Lambda (λ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS) in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only. | The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.

The Diazyme Human Lambda (λ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.  |

VI Standards/Guidance Documents Referenced:

Not applicable

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Refer to K153394

2. Linearity:

Refer to K211648

K253358 - Page 5 of 9

{5}

3. Analytical Specificity/Interference:
Refer to K153394

4. Assay Reportable Range:
See K153394, K211648

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Refer to K153394

6. Detection Limit:
Refer to K211648

7. Assay Cut-Off:
Refer to K153394 and K184338

## B Comparison Studies:

1. Method Comparison with Predicate Device:
Not applicable

2. Matrix Comparison:
Not applicable

## C Clinical Studies:

Clinical performance of the Diazyme Human Kappa (κ) Free Light Chain Assay and the Diazyme Human Lambda (λ) Free Light Chain Assay as an aid in the diagnosis and monitoring of multiple myeloma was previously demonstrated (Refer to K153394 and K220001).

The performance of the Diazyme Human Kappa (κ) Free Light Chain Assay and the Diazyme Lambda (λ) Free Light Chain Assay as an aid in the evaluation of MGUS was investigated from the following studies:

### Study 1:

A study was performed by testing a total of 270 serum samples from clinically confirmed MGUS patients (with IFE positivity) and a total of 266 samples from clinically defined non-MGUS patients with polyclonal immunostimulation (confirmed with negative SPEP/SIFE). Clinical diagnostic criteria and classification for MGUS and related plasma-cell disorders were, as practiced clinically, fulfilled, but were not limited to the criteria outlined by the 'International Myeloma Working Group (IMWG)' consensus. The result of the device was compared to the clinical diagnosis for each sample.

K253358 - Page 6 of 9

{6}

The cohort of 270 MGUS samples included 205 non-IgM MGUS, 40 IgM MGUS and 25 light chain (LC) MGUS. All samples were tested for FLC kappa and lambda levels with the candidate devices, the Diazyme Human Kappa (κ) Free Light Chain Assay and the Diazyme Human Lambda (λ) Free Light Chain Assay on the Roche cobas c 501 analyzer. FLC κ/λ ratios were calculated for each sample. The test results for MGUS positive or negative were based on the following criteria:

- For Non-LC MGUS, the test was considered as "MGUS Positive" or "Abnormal" when abnormal FLC κ/λ ratio (outside reference interval, i.e., &lt; 0.22 or &gt; 1.74) with IFE positivity was determined.
- For LC MGUS, the test was considered as "MGUS Positive" when abnormal FLC κ/λ ratio (outside reference interval, i.e., &lt; 0.22 or &gt; 1.74) with elevated FLC kappa (&gt;20.7 mg/L) or elevated FLC lambda FLC (&gt;27.7 mg/L) was determined.

# Results:

The results showed that 145 out of 270 MGUS cases were tested positive by the Diazyme Human Kappa (κ) Free Light Chain Assay and the Diazyme Human Lambda (λ) Free Light Chain Assay with a positive rate of 53.7% (95% CI: 47.6% - 59.8%). The distribution of the cohort and the positivity rate for each clinical type of MGUS are summarized in the following table:

|  MGUS Type | N | n (n/N%) of MGUS Positive by Diazyme FLC Assays  |
| --- | --- | --- |
|  Non-IgM MGUS (all) | 205 | 92 (44.9%)  |
|  IgG κ | 85 | 48 (56.5%)  |
|  IgG λ | 80 | 26 (32.5%)  |
|  IgA κ | 25 | 13 (52.0%)  |
|  IgA λ | 15 | 5 (33.3%)  |
|  IgD κ | 0 | 0 (0%)  |
|  IgD λ | 0 | 0 (0%)  |
|  |   |   |
|  IgM MGUS (all) | 40 | 28 (70.0%)  |
|  IgM κ | 30 | 24 (80.0%)  |
|  IgM λ | 10 | 4 (40.0%)  |
|  |   |   |
|  LC-MGUS (all) | 25 | 25 (100%)  |
|  κ | 15 | 15 (100%)  |
|  λ | 10 | 10 (100%)  |
|  |   |   |
|  TOTAL | 270 | 145 (53.7%)  |

The 266 samples from patients with confirmed non-MGUS (negative SPE/IFE) polyclonal stimulation (polyclonal hypergammaglobulinemia) were comprised of: connective tissue disease (59); renal disease (46); infection (40); CVD (21); neurological diseases (20); dermatological disease (15); cancers (14); autoimmune diseases (13); B Lymphoma (8);

K253358 - Page 7 of 9

{7}

anemia (6); liver disease (5); diabetes (4) and other miscellaneous diseases/conditions (15) including: polyarteritis nodosa, vascular disease, gout, immunodeficiency, venous insufficiency, thrombocytosis, leukocytosis, thrombotic microangiopathy, lymphadenopathy, COPD, pulmonary fibrosis, interstitial lung disease, lung nodules.

Among 266 non-MGUS samples, 250 of these were determined as negative by the Diazyme Kappa and Lambda FLC assays, indicating a negative agreement of 94.0% (95% CI: 90.4%-96.5%) in this sample cohort. The 16 false positive samples were from patients with the following disorders: cancers (4); renal diseases (3); connective (2); anemia (2); infection/inflammation (2); autoimmune diseases (1); B lymphoma (1); and other diseases/conditions (1).

## Study 2:

Another study was performed by testing 348 serum samples from 84 subjects with clinically stable MGUS and two subjects with progressive clinical status converting from MGUS to multiple myeloma (MM). Among 84 MGUS stable subjects, 69 patients were diagnosed with non-IgM MGUS (29 IgG K, 22 IgG L, 11 IgA K, and seven IgA L), and 11 patients with IgM MGUS (eight IgM K and three IgM L), and four patients with LC MGUS (one LC-K, three LC-L). The two progressive subjects were one IgG L and one LC-L.

At least three and up to five serial draws were collected from each patient and tested with the Diazyme Human Kappa (κ) Free Light Chain Assay and the Diazyme Human Lambda (λ) Free Light Chain Assay on the Roche cobas c 501 analyzer. FLC κ/λ ratio was calculated for each blood draw sample. Since the MGUS has been identified as a precursor to malignant diseases (multiple myeloma or amyloidosis), but not as a disease which requires treatment, no defined criteria in how to interpret consecutive FLC results are available. Therefore, for device evaluation purposes only, the criteria for stable MGUS and progressive MGUS based on test results in this study are defined as follows:

- Stable MGUS defined as &lt; 25% increase in the concentration of the involved FLC in two consecutive assessments. This analysis included MGUS patients with and without abnormal FLC κ/λ ratio.
- Progressive MGUS defined as the FLC κ/λ ratio outside of the reference interval of 0.22 - 1.74, and an increase of ≥ 25% in the concentration of the involved light chain at or preceding the diagnosis of MM, for two consecutive assessments.

## Results:

Seventy-six (76) out of 84 clinically stable MGUS subjects were determined as stable by the test and two out of two clinically progressive subjects were determined as progressive by these tests.

## Limitations

- The performance has not been fully evaluated on all race/ethnicity in the intended use population.
- The study included time points of blood draws not indicative of clinical practice. The algorithm using only FLC has not been validated for progression of disease. Furthermore, a small sample size (two patients) with the subtypes Lambda IgG and LC-L was used to study

K253358 - Page 8 of 9

{8}

progression in Study 2. Risk mitigation strategies include that this test is not a stand-alone test for the evaluation of patients with MGUS and the test is to be utilized in conjunction with serum protein electrophoresis and immunofixation blood tests.

## D Clinical Cut-Off:

Not applicable

## E Expected Values/Reference Range:

Refer to K153394 and K181438

Kappa FLC reference range: 2.4 – 20.7 mg/L

Lambda FLC reference range: 4.2 – 27.7 mg/L

Kappa/Lambda Ratio reference Range: 0.22 – 1.74

## VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K253358 - Page 9 of 9

---

**Source:** [https://fda-staging.innolitics.com/device/K253358](https://fda-staging.innolitics.com/device/K253358)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
