Diazyme Human Kappa (k) Free Light Chain Assay; Diazyme Human Lambda (¿) Free Light Chain Assay
Device Facts
| Record ID | K253358 |
|---|---|
| Device Name | Diazyme Human Kappa (k) Free Light Chain Assay; Diazyme Human Lambda (¿) Free Light Chain Assay |
| Applicant | Diazyme Laboratories, Inc. |
| Product Code | DFH · Immunology |
| Decision Date | Dec 19, 2025 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5550 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K253358 · Dec 19, 2025 | Diazyme Human Kappa (k) Free Light Chain Assay; Diazyme Human Lambda (¿) Free Light Chain Assay | Diazyme Laboratories, Inc. | Retrospective clinical serum samples from patients with confirmed MGUS and non-MGUS conditions | The sponsor used retrospective clinical samples to demonstrate the performance of the FLC assays in identifying MGUS and distinguishing it from other polyclonal immunostimulation conditions, and to evaluate serial monitoring in stable vs. progressive MGUS patients. | MGUS; Retrospective cohort; Clinical performance; Serum samples |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Retrospective cohort study | 270 clinically confirmed MGUS patients and 266 non-MGUS patients with polyclonal immunostimulation; Sample Size: 536 | Not applicable for this study | Positive rate in MGUS patients and negative agreement in non-MGUS patients |
| Retrospective longitudinal cohort study; Follow-up/Duration: 3 to 5 serial draws per patient | 84 subjects with clinically stable MGUS and 2 subjects with progressive clinical status (MGUS to MM); Sample Size: 86 | Not applicable for this study | Stability vs. progression classification based on FLC concentration changes |
Indications for Use
The Diazyme Human Kappa (κ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS), in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only. The Diazyme Human Lambda (λ) Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on validated analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis and monitoring of multiple myeloma and aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS), in conjunction with other laboratory and clinical findings. For in-vitro diagnostic use only.
Device Story
Latex particle enhanced immunoturbidimetric assays for quantitative measurement of Kappa and Lambda FLC in serum. Input: patient serum samples. Operation: FLC in sample binds to anti-FLC antibodies coated on carboxyl polystyrene latex particles, causing agglutination; degree of turbidity measured optically on Roche cobas c 501 analyzer. Output: FLC concentration calculated via 6-point calibration curve interpolation. Used in clinical laboratories by technicians. Results used by physicians alongside other laboratory/clinical findings to aid diagnosis/monitoring of multiple myeloma and evaluation of MGUS. Benefits: provides quantitative data to assist in identifying and monitoring plasma-cell disorders.
Clinical Evidence
Retrospective clinical studies evaluated MGUS performance. Study 1 (N=270 MGUS, N=266 non-MGUS) showed 53.7% positive rate in MGUS patients and 94.0% negative agreement in non-MGUS patients with polyclonal hypergammaglobulinemia. Study 2 (N=86) evaluated serial monitoring in 84 stable and 2 progressive MGUS subjects; 76/84 stable subjects were correctly identified as stable, and 2/2 progressive subjects were identified as progressive. Bench testing and analytical performance were referenced from previous submissions (K153394, K211648).
Technological Characteristics
Latex particle enhanced immunoturbidimetric assay. Reagents: Tris buffer, carboxyl polystyrene particles coated with polyclonal rabbit/goat anti-Kappa or rabbit anti-Lambda FLC antibodies. Instrumentation: Roche cobas c 501 analyzer. Quantitative measurement via optical turbidity. Standalone reagent kit; requires external calibrators and controls.
Indications for Use
Indicated for use in patients requiring diagnosis and monitoring of multiple myeloma and evaluation of monoclonal gammopathy of undetermined significance (MGUS). Used in conjunction with other laboratory and clinical findings. For prescription use only.
Regulatory Classification
Identification
An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.
Predicate Devices
- Diazyme Human Kappa (κ) Free Light Chain Assay (K211648)
- Diazyme Human Lambda (λ) Free Light Chain Assay (K211648)