Assisted Reproduction Laser System (Model ILS-400M)
K253333 · Hua Yue Medical Technology Co., Ltd. · MRX · Jun 17, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K253333
Device Name
Assisted Reproduction Laser System (Model ILS-400M)
Applicant
Hua Yue Medical Technology Co., Ltd.
Product Code
MRX · Obstetrics/Gynecology
Decision Date
Jun 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Assisted Reproduction Laser System (Model ILS-400M) is for use in assisted reproduction procedures to ablate or thin the zona pellucida of an embryo to facilitate assisted hatching or recovery of cells for pre-implantation genetic diagnosis (blastomeres). The device can also be used on blastocyst stage embryos for biopsy of trophectoderm cells for pre-implantation diagnosis procedures, and blastocyst collapse prior to vitrification procedures.
Device Story
Assisted Reproduction Laser System (Model ILS-400M) is an accessory for inverted microscopes used in embryology labs. System inputs include user-defined paths via PC software and manual firing triggers (mouse/foot switch). It utilizes a 1480 nm laser (ablation and pilot beams) to perform precise zona pellucida thinning or ablation. Operation modes include single-pulse (manual) and multi-pulse (automated path-following). The device provides visual feedback via integrated camera and software for measurement and image storage. It assists embryologists in embryo biopsy and vitrification preparation by enabling controlled, non-contact tissue removal. Benefits include increased precision and safety during delicate embryo manipulation procedures.
Clinical Evidence
Bench testing only. Performance testing included laser output verification per FDA guidance, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and laser safety (IEC 60825-1). Software and cybersecurity were validated per 2023 FDA guidance documents.
Technological Characteristics
Laser: 1480 nm, 400 mW, 5-2000 μs pulse length. Class 1 laser safety (IEC 60825-1:2014). Components: Laser control unit, movement control unit, mirror module, 20x/40x objectives. Connectivity: PC-based control software. Standards: IEC 60601-1, IEC 60601-1-2, AAMI ES60601-1.
Indications for Use
Indicated for use in assisted reproduction procedures on embryos to ablate or thin the zona pellucida for assisted hatching, recovery of blastomeres for pre-implantation genetic diagnosis, biopsy of trophectoderm cells, and blastocyst collapse prior to vitrification.
Regulatory Classification
Identification
The assisted reproduction laser system is a device that images, targets, and controls the power and pulse duration of a laser beam used to ablate a small tangential hole in, or to thin, the zona pellucida of an embryo for assisted hatching or other assisted reproduction procedures.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Assisted Reproduction Laser Systems.” See § 884.1(e) for the availability of this guidance document.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 17, 2026
Hua Yue Medical Technology Co., Ltd.
Zhenzhi Shi
R&D Director
First Floor, C2 Bldg., #11 Kaiyuan Rd.
Huangpu District
Guangzhou, Guangdong 510530
CHINA
Re: K253333
Trade/Device Name: Assisted Reproduction Laser System (Model ILS-400M)
Regulation Number: 21 CFR 884.6200
Regulation Name: Assisted Reproduction Laser System
Regulatory Class: II
Product Code: MRX
Received: May 15, 2026
Dear Zhenzhi Shi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20903
www.fda.gov
{1}
K253333 - Zhenzhi Shi
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253333 - Zhenzhi Shi
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253333
Device Name
Assisted Reproduction Laser System (Model ILS-400M)
Indications for Use (Describe)
The Assisted Reproduction Laser System (Model ILS-400M) is for use in assisted reproduction procedures to ablate or thin the zona pellucida of an embryo to facilitate assisted hatching or recovery of cells for pre-implantation genetic diagnosis (blastomeres). The device can also be used on blastocyst stage embryos for biopsy of trophectoderm cells for pre-implantation diagnosis procedures, and blastocyst collapse prior to vitrification procedures.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K253333
Page 1 of 4
# 510(k) Summary – K253333
## 1. Submitter Information
Company Name: Hua Yue Medical Technology Co., Ltd.
Address: First Floor, C2 building, No.11 Kaiyuan Road, Huangpu District, Guangzhou City, Guangdong Province, 510530, P.R. China
Phone: +86-020-34821111
Contact Person (Title): Shi Zhenzhi (R&D Director)
E-mail: register@huayueco.com
## 2. Date of Preparation: June 16, 2026
## 3. Subject Device Information
Trade/Proprietary Name: Assisted Reproduction Laser System (Model ILS-400M)
Common Name: Assisted Reproduction Laser System
Regulation Number: 21CFR 884.6200
Regulation Name: Assisted Reproduction Laser System
Product Code: MRX
Review Panel: Obstetrics/Gynecology
Regulatory Class: II
## 4. Predicate Device Information
510(k) Number: K141434
Trade/Proprietary Name: Saturn 5 Laser System
The predicate device has not been subject to a design-related recall.
## 5. Device Description
The Assisted Reproduction Laser System (Model ILS-400M) is composed of a laser control unit (including Laser Control Module, Optical Fiber and the Power Adapter), movement control unit (including Movement Control Module, Power Adaptor), Mirror Module, Laser Objective (40x and 20x) and Assisted Reproduction Laser System Control Software. The system has been designed to be fitted to commercially available inverted microscopes and is also used with a PC and camera to provide an optical system and means of firing the laser. A Foot Switch for firing the laser is available as an optional extra device component.
This Assisted Reproduction Laser System (Model ILS-400M) delivers two lasers: the ablation laser and the pilot laser. Both laser beams are launched from the end of a fiber optic and collimated by an achromatic collimator. The Assisted Reproduction Laser System Control software controls the movement of the laser, fires the laser and allows the user to take measurements and generate and store images.
The Assisted Reproduction Laser System (Model ILS-400M) includes a single-pulse mode and multi-pulse mode. The multi-pulse mode enables users to ablate a series of holes along a set of pre-determined paths. The preset lines can be either straight or curved. After drawing the shooting path,
1 / 4
{5}
K253333
Page 2 of 4
users can adjust the laser drilling spacing by adjusting the drilling density. The multi-pulse mode is as safe as the single-pulse mode. Users can choose the usage mode according to their application scenarios.
### 6. Indications for Use
The Assisted Reproduction Laser System (Model ILS-400M) is for use in assisted reproduction procedures to ablate or thin the zona pellucida of an embryo to facilitate assisted hatching or recovery of cells for pre-implantation genetic diagnosis (blastomeres). The device can also be used on blastocyst stage embryos for biopsy of trophectoderm cells for pre-implantation diagnosis procedures, and blastocyst collapse prior to vitrification procedures.
### 7. Substantial Equivalence Discussion
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Parameter | Subject device Assisted Reproduction Laser System (Model ILS-400M) K253333 | Predict device Saturn 5 Laser System K141434 | Comparison |
| --- | --- | --- | --- |
| Indications for Use | The Assisted Reproduction Laser System (Model ILS-400M) is for use in assisted reproduction procedures to ablate or thin the zona pellucida of an embryo to facilitate assisted hatching or recovery of cells for pre-implantation genetic diagnosis (blastomeres). The device can also be used on blastocyst stage embryos for biopsy of trophectoderm cells for pre-implantation diagnosis procedures, and blastocyst collapse prior to vitrification procedures. | For use in assisted reproduction procedures to ablate or thin the zona pellucida of an oocyte or embryo to facilitate assisted hatching or recovery of cells for pre-implantation genetic diagnosis (blastomeres). The device can also be used on blastocyst stage embryos for biopsy of trophectoderm cells for pre-implantation diagnosis procedures, and blastocyst collapse prior to vitrification procedures. | The indications for use for the subject and predicate devices differ, as the subject device is to be used to ablate or thin the zona pellucida of an embryo, while the predicate is to be used for the same uses for oocytes or embryos. This difference does not raise new questions of safety or effectiveness. |
| System configuration | •Laser control unit •Movement control unit •Mirror Module •Laser Objective(40X,20X) •Assisted Reproduction Laser System Control Software | • Laser control unit • 40X objective • Microscope adapters • Mirror module • Motor module (Active version only) •Laser application software | The subject and predicate devices have similar components, but differences exist (e.g., Objective magnification) |
2 / 4
{6}
K253333
Page 3 of 4
| | - • Camera (optional) - • Foot switch (optional) | - • Camera (optional) - • Foot switch (optional) | However, these differences do not raise different questions of safety and effectiveness. |
| --- | --- | --- | --- |
| Operating Modes | Active Mode – The direction and path of the laser beam is computer-controlled; embryo is stationary. | Fixed Mode – The direction of the laser beam is fixed; user manually guides embryo. Active Mode – The direction and path of the laser beam is computer-controlled; embryo is stationary. | The subject device only has active mode capability. However, the differences do not raise different questions of safety and effectiveness. |
| Laser Modes | Single-Pulse mode: Allows a single ablation of the target (Active Mode) Multi-Pulse mode: Allows to ablate a series of holes for biopsy procedures (Active Mode) | Single Pulse: Allows a single ablation of the target (Fixed Mode and Active Mode) Biopsy Mode: Allows to ablate a series of holes for biopsy procedures (Active Mode only) | Same |
| Laser Specifications | $$\lambda = 1480$$ nm Power 400 mW Pulse length 5-2000 $$\mu$$s | $$\lambda = 1480$$ nm Power 400 mW Pulse length 5-2000 $$\mu$$s | Same |
| Laser Classification | Class 1 | Class 1 | Same |
| Microscope Compatibility | Compatible with various models of inverted microscopes. | Compatible with various models of inverted microscopes. | Same |
| Laser Firing Mechanism | Mouse; foot switch | Mouse; foot pedal | Similar |
| Hole Size Indicator | Yes | Yes | Same |
The indications for use of the subject and predicate device are similar, with the subject device specifying use for only embryos. Since the subject and predicate device intended use populations overlap (i.e., both are used for ablating or thinning the zona pellucida of an embryo), no new questions of safety or effectiveness are raised.
3 / 4
{7}
K253333
Page 4 of 4
The subject and predicate devices have similar technological features, including laser specification, firing mechanism, laser modes, and microscope compatibility. However, as shown in the table above, there are technological differences between the subject and predicate device, including differences in available operating modes and system configurations. The different technological characteristics of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness and were sufficiently evaluated in the non-clinical testing outlined below.
## 8. Summary of non-clinical Performance Testing
The following performance tests were conducted on the Assisted Reproduction Laser System (Model ILS-400M):
### Laser Performance Testing
Laser Performance Testing in accordance with the 2004 FDA guidance document “Assisted Reproduction Laser Systems – Class II Special Controls Guidance Document for Industry and FDA Staff.
### Electrical Safety and Electromagnetic Compatibility (EMC)
- Electrical safety
- IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020
- AAMI ES60601-1:2005/AMD1:2012, AAMI ES60601-1:2005/AMD2:2021
- Electromagnetic compatibility (EMC)
- IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020
- ANSI AAMI IEC60601-1-2:2014+A1:2021
- IEC TS 60601-4-2: 2024
- Laser Safety Testing
Both the pilot laser and ablation laser are classified as Class I in accordance with IEC 60825-1:2014.
### Software and Cybersecurity
Software verification and validation testing that met the requirements of the FDA Guidance Document, “Content of Premarket Submissions for Device Software Functions” issued on June 2023.
Cybersecurity was evaluated per the 2023 FDA guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”.
## 9. Conclusion
The results of the performance testing described above demonstrate that Assisted Reproduction Laser System (Model ILS-400M) is as safe and effective as the predicate device and supports a determination of substantial equivalence.
4 / 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.