K253327 · M.I.Tech Co., Ltd. · ESW · Oct 30, 2025 · General, Plastic Surgery
Device Facts
Record ID
K253327
Device Name
HANAROSTENT Esophagus Upper (CCC)
Applicant
M.I.Tech Co., Ltd.
Product Code
ESW · General, Plastic Surgery
Decision Date
Oct 30, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HANAROSTENT® Esophagus Upper (CCC) is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistula.
Device Story
Self-expanding tubular prosthesis designed to maintain esophageal luminal patency and occlude fistulas. Device consists of Nitinol wire stent, gold radiopaque markers, fully covered silicone membrane, and polymeric repositioning lassos. Delivered via over-the-wire (OTW) delivery system; stent loaded into distal delivery device; deployed by pulling outer sheath. Used in clinical settings by physicians. Stent provides mechanical support to esophageal wall; silicone membrane prevents tissue ingrowth and seals fistulas. Single-use device.
Clinical Evidence
Bench testing only. Tests included foreshortening, expansion force, compression force, guidewire passage, deployment force, deployment accuracy, dimensions, bonding strength, and lasso tensile strength. No animal or clinical performance data submitted.
Technological Characteristics
Self-expanding stent composed of Nitinol wire, gold radiopaque markers, and silicone membrane. Polymeric repositioning lassos and delivery device. Single-use. Mechanical expansion via delivery sheath withdrawal.
Indications for Use
Indicated for patients with esophageal strictures caused by intrinsic or extrinsic malignant tumors and concurrent esophageal fistula requiring luminal patency maintenance.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 30, 2025
M.I. Tech Co., Ltd.
% Monica Montanez
Principal Strategy Consultant
NAMSA
381 Lark Lane
Bailey, Colorado 80421
Re: K253327
Trade/Device Name: HANAROSTENT Esophagus Upper (CCC)
Regulation Number: 21 CFR 878.3610
Regulation Name: Esophageal Prosthesis
Regulatory Class: Class II
Product Code: ESW
Dated: September 29, 2025
Received: September 30, 2025
Dear Monica Montanez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253327 - Monica Montanez
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253327 - Monica Montanez
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253327 | ? |
| Please provide the device trade name(s). | | ? |
| HANAROSTENT Esophagus Upper (CCC) | | |
| Please provide your Indications for Use below. | | ? |
| The HANAROSTENT® Esophagus Upper (CCC) is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistula. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
M.I.Tech
www.mitech.co.kr
HEAD OFFICE
174, HABUK2-GIL, JINWI-MYEON, PYEONGTAEK-SI, GYEONGGI-DO,
17706, REPUBLIC OF KOREA
TEL +82 31 662 5645 FAX +82 31 662 5648
K253327
Page 1 of 3
510(k) Summary
| Preparation Date: | October 20, 2025 | |
| --- | --- | --- |
| Submitter: | M.I.Tech Co., Ltd.
174 Habuk 2-gil, Jinwi-myeon,
Pyeongtaek-si, Gyeonggi-do
17706, Republic of Korea
Phone: 82-31-662-5645
Fax: 82-31-662-5648 | |
| Primary Contact: | Jeongeun Park
Regulatory Affairs Manager
M.I.Tech Co., Ltd.
174 Habuk 2-gil, Jinwi-myeon,
Pyeongtaek-si, Gyeonggi-do
17706, Republic of Korea
Email: jepark@mitech.co.kr
Phone: 82-10-4456-9150
Fax: 82-70-8282-5642 | |
| Subject Devices: | Trade Name: HANAROSTENT® Esophagus Upper (CCC)
Device: Prosthesis, Esophageal
Regulation Description: Esophageal prosthesis
Review Panel: Gastroenterology/Urology
Regulation Number: 21 CFR 878.3610
Device Class: Class II
Product Code: ESW
Regulation Medical Specialty: General & Plastic Surgery | |
| Intended Use / Indications for Use: | The HANAROSTENT® Esophagus Upper (CCC) is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistula. | |
Page 1 of 3
{5}
M.I.Tech
www.mitech.co.kr
HEAD OFFICE
174, HABUK2-GIL, JINWI-MYEON, PYEONGTAEK-SI, GYEONGGI-DO,
17706, REPUBLIC OF KOREA
TEL +82 31 662 5645 FAX +82 31 662 5648
K253327
Page 2 of 3
| Device Description: | This self-expanding tubular prosthesis is designed to maintain patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. It consists of a self-expandable metal stent and an over the wire (OTW) delivery device.
The self-expandable metal stent is made of nickel titanium alloy (Nitinol) wire, radiopaque markers made of gold wire, fully covered silicone membrane, and one repositioning lasso at each end of the stent made of polymeric materials. The delivery device is made of polymeric materials. The stent is loaded into the distal part of the delivery device and expanded in the body by pulling the outer sheath of the delivery device. The HANAROSTENT® Esophagus Upper (CCC) is intended for single use only. |
| --- | --- |
| Predicate Device: | Trade Name: HANAROSTENT® Esophagus (CCC)
Applicant: M.I.Tech Co., Ltd.
510(k) Number: K201160
Clearance Date: July 23, 2021
Device: Prosthesis, Esophageal
Regulation Description: Esophageal prosthesis
Review Panel: Gastroenterology/Urology
Regulation Number: 21 CFR 878.3610
Device Class: Class II
Product Code: ESW
Regulation Medical Specialty: General & Plastic Surgery |
| Comparison to the Predicate: | Both the subject and predicate devices are designed to maintain patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.
The subject device and the predicate device have the same material, manufacturing process, sterilization method, and operation method, but have different stent diameters, type of stent proximal head, diameters and lengths of stent proximal head. |
| Technological Characteristics: | The intended use of the subject device, HANAROSTENT® Esophagus Upper (CCC), is identical to the predicate device, HANAROSTENT® Esophagus (CCC).
The subject device is identical to the delivery device of predicate device in working length, outer diameter, method of placement, and guidewire compatibility.
The subject device has same stent materials as the predicate device and is similar in terms of performance. |
| FDA Guidance Documents: | The following FDA guidance document were consulted in preparing this premarket submission:
• *Guidance for The Content of Premarket Notifications for Esophageal and Tracheal Prostheses*, issued April 28, 1998 |
Page 2 of 3
{6}
M.I.Tech
www.mitech.co.kr
HEAD OFFICE
174, HABUK2-GIL, JINWI-MYEON, PYEONGTAEK-SI, GYEONGGI-DO,
17706, REPUBLIC OF KOREA
TEL +82 31 662 5645 FAX +82 31 662 5648
K253327
Page 3 of 3
| Performance Testing: | Bench testing was performed to confirm the safety and effectiveness of the proposed subject devices as compared to the predicate devices. Performance testing was performed as per the design control system. The following tests were conducted: • Foreshortening • Expansion force • Compression force • Guidewire passage • Deployment force • Deploying accuracy • Dimensions • Bonding strength • Lasso tensile strength No animal and clinical performance data is submitted in this 510(k). |
| --- | --- |
| Substantial Equivalence: | The subject devices are substantially equivalent to the predicate devices when evaluating intended use and technological characteristics. • The subject device has exact same intended use/indications for use as the predicate. • The subject devices and predicate device are substantially equivalent with only minor technological differences. • These differences do not raise new questions of safety and effectiveness. |
| Conclusion: | This comparison demonstrates the subject device are substantially equivalent to the predicate device. The subject devices are as safe and effective as the predicate device and will perform as intended. Therefore, M.I. Tech respectfully requests market clearance for the subject devices. |
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.