Vitalograph VitaloJAK Clinic Model 7100 is indicated for use to aid in the assessment of cough in adults and pediatric patients aged 8 years and older under supervision presenting with suspected Chronic Cough. The Vitalograph VitaloJAK ClinicModel 7100 is a non-invasive battery-operated device intended to acquire, record and store ambulatory cough sounds from patients. The device stores the data on internal memory for later playback, review, and analysis on a windows-based PC.
Device Story
Non-invasive, battery-operated ambulatory cough monitor; records cough sounds via acoustic sensor applied to suprasternal notch. Device stores audio data on internal 32GB eMMC memory for up to 72 hours. Operated by patient in home, clinic, or hospital settings under clinician supervision. Features include event marking and mute button; no integrated display (uses status LED). Data downloaded via USB to Windows-based PC running VitaloJAK Studio software for manual review and analysis by healthcare providers. Facilitates assessment of chronic cough by providing objective recording of cough activity; does not provide automated diagnosis or interpretation.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included ISO 10993 biocompatibility (cytotoxicity, sensitization, irritation), human factor validation, battery life verification, mechanical design verification, and 3rd party audio sensor testing.
Technological Characteristics
Battery-operated (3x AA); acoustic sensor (suprasternal notch); ABS plastic housing; hydrocolloid adhesive. 32GB eMMC memory; USB 2.0/3.0 connectivity. Sampling rate 8 kHz. Standards: EN 60601-1, EN 60601-1-11, EN 60601-1-2. Biocompatibility per ISO 10993-1 (prolonged contact).
Indications for Use
Indicated for adults and pediatric patients aged 8 years and older under supervision presenting with suspected Chronic Cough.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 3, 2025
Vitalograph Ltd.
% Paul Dryden
Consultant
ProMedic Consulting LLC
131 Bay Point Dr NE
St. Petersburg, Florida 33704
Re: K253293
Trade/Device Name: VitaloJAK Clinic (Model 7100)
Regulation Number: 21 CFR 870.2800
Regulation Name: Medical Magnetic Tape Recorder
Regulatory Class: Class II
Product Code: DSH
Dated: November 12, 2025
Received: November 12, 2025
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253293 - Paul Dryden
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K253293 - Paul Dryden
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ethan L. Nyberg -S
Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253293 | |
| Device Name Vitalograph VitaloJAK Clinic (Model 7100) | |
| Indications for Use (Describe) Vitalograph VitaloJAK Clinic Model 7100 is indicated for use to aid in the assessment of cough in adults and pediatric patients aged 8 years and older under supervision presenting with suspected Chronic Cough. The Vitalograph VitaloJAK ClinicModel 7100 is a non-invasive battery-operated device intended to acquire, record and store ambulatory cough sounds from patients. The device stores the data on internal memory for later playback, review, and analysis on a windows-based PC. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K253293
# 510(k) Summary
Page 1 of 6
Date Prepared: 12-Nov-25
Submitter: Vitalograph (Ireland) Ltd
Gort Road Industrial Estate
Ennis Co. Clare V95 HFT4
Ireland
+353 65 6864100
Submitter Contact: James O-Keeffe - CTO
Submission Correspondent: Paul Dryden, ProMedic, LLC
Proprietary or Trade Name: Vitalograph VitaloJAK Clinic Model 7100
Common / Usual Name: Medical Magnetic Tape Recorder
Classification CFR: 21 CFR 870.2800
Product Code: DSH
Predicate: Vitalograph VitaloJAK Model 7100
Common / Usual Name: Medical Magnetic Tape Recorder
Classification CFR: 21 CFR 870.2800
Product Code: DSH
## Device Description:
The Vitalograph VitaloJAK Clinic Model 7100 is a non-invasive battery-operated device intended to acquire, record and store ambulatory cough sounds from patients. The device stores the data on a removable memory card or internal memory for later playback, review, and analysis on a windows-based PC.
The Vitalograph VitaloJAK Clinic Model 7100 is indicated for use in cough recording sessions. It provides the means to sense voice and cough activity from a subject, record this information and store it on an SD memory card. The Vitalograph VitaloJAK Clinic Model 7100 is not life supporting equipment, it does not screen, treat or prevent any condition or disease. The subjects may be of any user population, weight range, health, or condition.
## Device Modifications:
The following is the summary of changes in the device under evaluation (Vitalograph VitaloJAK Clinic):
- Name changed to include “Clinic”. Therefore, the device name is “Model 7100 Vitalograph VitaloJAK Clinic”; to differentiate between the device variants and specify it’s use on advice of clinician in the outpatient clinical set-up.
- Maximum test duration of 24 hours is changed to 96 hours.
- Addition of VitaloJAK Studio software application. The VitaloJAK Studio App provides device configuration to perform recording and download the results onto a workstation.
- Microphone and LCD/ display have been removed.
- Redesign of user interface with the four buttons on the front removed and changed to one button on the front of the device for marking events and to mute the device. The second button on the side of the device is to start recording. The on/off button is located inside the battery door. Housing dimensions are modified.
- Memory capacity is increased to 32GB eMMC. The predicate device had a Compact Flash memory card.
{5}
K253293
510(k) Summary
Page 2 of 6
# Comparison Table
Table of Comparison of Proposed to Predicate Device
| Product | Modified Device Vitalograph VitaloJAK Clinic Model 7100 | Predicate Device Vitalograph VitaloJAK Model 7100 (Cough Monitor) (K110525) | Substantially equivalent to predicate? |
| --- | --- | --- | --- |
| Model and Device Name | Vitalograph VitaloJAK Clinic Model 7100 | Vitalograph VitaloJAK Model 7100 | Yes, same name and model number except “Clinic” is added to the name of the changed device as it is used on prescription by clinician in the out-patient department. No impact on product safety and performance. All VitaloJAK variants are used under supervision. |
| Indications for Use | Vitalograph VitaloJAK Clinic Model 7100 is indicated for use to aid in the assessment of cough in adults and pediatric patients aged 8 years and older under supervision presenting with suspected Chronic Cough. The Vitalograph VitaloJAK Clinic Model 7100 is a non-invasive battery-operated device intended to acquire, record and store ambulatory cough sounds from patients. The device stores the data on internal memory for later playback, review, and analysis on a windows-based PC. | The VitaloJAK is indicated for use in cough recording sessions. It provides the means to sense voice and cough activity from a subject, record this information and store it on an SD memory card. The subject device is not interpretative and does not deliver a diagnosis. | Yes, both are indicated for use for similar indication i.e. cough recording, however the wording is updated in the changed device to differentiate it from Intended use statement and to include specific condition of Chronic cough. Although both the devices are indicated for chronic cough, but it is only now specified in the changed device. |
| Patient Population | Adults and pediatric patients aged 8 years and older under supervision | Adults and pediatric patients aged 8 years and older under supervision | Similar, though the population was not specified in the predicate, but test data was provided. |
| Environment of Use | Hospital, clinic or home setting | Hospital, clinic or home setting | Similar |
{6}
K253293
# 510(k) Summary
Page 3 of 6
| Product | Modified Device
Vitalograph VitaloJAK Clinic Model 7100 | Predicate Device
Vitalograph VitaloJAK Model 7100 (Cough Monitor) (K110525) | Substantially equivalent to predicate? |
| --- | --- | --- | --- |
| Principle of Operation | Recording the audio inputs from the Cough Sensor | Recording the audio inputs from the Cough Sensor and Microphone | Yes, the cough sensor is identical. Microphone is removed in the modified device but it does not impact the safety and performance of the device. |
| Essential Performance Test Limits | Recordings are downloaded to a PC for subsequent analysis via USB or flashcard. A valid sound recording must be present on the cough channel to perform audio analysis. Minimum requirement for a valid audio file for EN60601 testing is no flat line data, indicating no audio was recorded. | Recordings are downloaded to a PC for subsequent analysis by trained Audio Analysts via compact flashcard. A valid sound recording must be present on one of the channels to perform audio analysis. Minimum requirement for a valid audio file for EN60601 testing is no flat line data, indicating no audio was recorded | Yes, Similar, added USB downloaded |
| Acoustic sensor applied on suprasternal notch | Yes | Yes | Yes, Identical |
| Non-invasive | Yes | Yes | Yes, Identical |
| Mute | Yes – Center button on front of the device to mute and to register events during the recording session | No | No muting option in predicate but this feature does not affect safety or effectiveness compared to the predicate |
| Display | No LCD/ display. LED to monitor various operational states | Monochrome LCD display. Bi-colour status LED to indicate device state | There is a change in the design of the device. LED function is identical however, LCD/Display is removed to minimise operations to make it user friendly. No impact on product safety and performance. |
| Maximum Test Duration | 72 hours on internal memory | 24 hours | Yes, similar (now can record four consecutive sessions of 24 hours for up to 72 hours) |
{7}
K253293
Page 4 of 6
| Product | Modified Device
Vitalograph VitaloJAK Clinic Model 7100 | Predicate Device
Vitalograph VitaloJAK Model 7100 (Cough Monitor) (K110525) | Substantially equivalent to predicate? |
| --- | --- | --- | --- |
| False Positive Cough Rate | None (coughs are manually identified) | None (coughs are manually identified) | Similar |
| Power Supply | 3 x AA Batteries.
USB powered during configuration/ download 5V/500mA | 4 x AA Alkaline Batteries. | No USB in predicate |
| Dimensions | 84mm x 92mm x 28mm | 137 mm x 76 mm x 45.5 mm | Similar device dimensions. |
| Weight | 250g – device only | 225 g – device only | Similar |
| Operating Temperature | 5 – 40°C | 0–50°C | Similar |
| Operating Humidity | 15%–90%RH | 10%–95%RH | Similar |
| Atmospheric pressure | 700 – 1060hPa | 700 – 1060hPa | Similar |
| IP rating | IP22 when in pouch - protected against touch by fingers >12.5mm and water spray <15 degrees from vertical. | IP22 when in pouch - protected against touch by fingers >12.5mm and water spray <15 degrees from vertical. | Similar |
| Service Life | No claim | Same | Similar |
| Shelf-life of sensor | 1 year | 1 year | Similar |
| Memory and Communications | Micro-SD Card: min 8GB, Internal Memory: 32GB, Write speed: 20 MB/sec.
USB 2.0/3.0 for data download and device configuration. Custom USB cable is required, supplied with device. | Compact Flash Card.
Size: 4GB,
Write speed: 20 MB/sec
No USB cable | Technology upgrade with increase storage capacity in changed device as compared to predicate. Device is as safe and efficient as predicate |
| Sampling Rate | 8 kHz | 8 kHz | Similar |
{8}
K253293
Page 5 of 6
| Product | Modified Device
Vitalograph VitaloJAK Clinic Model 7100 | Predicate Device
Vitalograph VitaloJAK Model 7100
(Cough Monitor) (K110525) | Substantially equivalent to predicate? |
| --- | --- | --- | --- |
| Minimum PC System Requirements | Processor Speed: 2GHz or greater
RAM: 2GB (Min), 4GB (Recommended)
Disk Space: 1GB or greater.
Operating System: Windows 7 or above
Monitor: 1280 x 800 pixel
Other: Net framework 4.5.1, USB Port | Doesn’t required to be connected to a PC, therefore, no minimum requirements | Predicate device does not connect to PC. This does not raise new risks compared to the predicate. |
| Safety Standards | EN 60601-1:2006+A1:2013+A2:2021 | EN ISO 60601:2006 {IEC 60601 - 1:2005} | Similar |
| Home Use Standard | EN 60601-1-11:2015+A1:2021 | EN 60601-1: 2006 | Similar |
| EMC Standards | EN 60601-1-2:2015+A1:2021 | EN60601-1-2:2001 | Similar |
| Materials in Patient contact | Laminated hydrocolloid medical grade adhesive
Device housing – ABS Plastic
Surface contact, Intact skin, Prolonged duration ISO 10993-1 testing | Laminated hydrocolloid medical grade adhesive
Device housing – ABS Plastic
Surface contact, Intact skin, Limited duration ISO 10993-1 testing | Identical materials. Testing performed is the same for limited or prolonged duration of contact |
{9}
K253293
510(k) Summary
Page 6 of 6
# Discussion of Substantial Equivalence Between Subject and Predicate
## Indications for Use –
The indications for use for the subject and predicate device are similar.
**Discussion** – The change in the length of possible recording time from 24 to 72 hours and inclusion of the population detail which were not presented in the predicate, but testing provided.
## Technology and construction –
The technology of the subject and predicate device is identical.
**Discussion** – There is no difference in the technology and construction between the subject and predicate device with exception of removal of the microphone which is no longer needed.
## Environment of Use –
The environment of use is identical.
**Discussion** – The environments of use are identical to the predicate.
## Patient Population –
The patient population of the subject device and predicate is the same.
**Discussion** – The population is not specified in the predicate device.
## Non-clinical Testing
Performance testing demonstrated that the subject device met its acceptance criteria. Testing included:
- ISO 10993-5 – Cytotoxicity
- ISO 10993-10 – Sensitization
- ISO 10993-10 – Irritation
Though the duration of use of the patient contacting materials has changed from limited to prolonged according to ISO 10993-1, the required tests are the same.
## Bench testing
- Post Aging Performance Testing
- Drop Testing
- Transportation Testing
- Testing to the Recommended Operating Conditions
- Durability Testing on Labelling
- Human Factor Validation Testing
- Battery Life Verification
- Mechanical Design Verification
- Real Time Age Testing
- Accelerated Age Testing
- RTC Battery Life Testing
- Design Verification 3rd Party Audio Testing
- Audio Sensor Testing
- Software Testing
- Electronic Design Verification
## Substantial Equivalence Conclusion
Through testing, the subject device has demonstrated it is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.