K253285 · Elidah, Inc. · QAJ · Jan 23, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K253285
Device Name
Elitone for Men
Applicant
Elidah, Inc.
Product Code
QAJ · Gastroenterology, Urology
Decision Date
Jan 23, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5330
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K253285 · Jan 23, 2026
Elitone for Men
Elidah, Inc.
Systematic literature review of clinical studies
A systematic literature review was conducted to establish a historical control cohort of 215 subjects to compare against the prospective study cohort, accounting for natural recovery of continence following prostatectomy.
Historical control; Literature review; Prostatectomy; Urinary incontinence
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Historical Control Cohort (via Systematic Literature Review); Systematic literature review / Historical control cohort; Follow-up/Duration: Assessments within the first three months post-surgery
Men post-prostate surgery receiving no active treatment or self-guided pelvic floor muscle exercises; Sample Size: 215
Not applicable for this study
24-hour pad weight reduction
Indications for Use
Elitone for Men is a non-implanted muscle stimulator designed to aid early continence recovery in men immediately following prostate surgery.
Device Story
Elitone for Men is a non-implanted electrical muscle stimulator for home use by male patients following prostate surgery. The device consists of a small control unit and a disposable perineal electrode component. The patient applies the electrode to the perineal skin and uses the control unit to deliver therapeutic electrical stimulation, which induces pelvic floor muscle contraction and relaxation. The patient manually adjusts stimulation intensity. Regular use over several weeks is intended to accelerate continence recovery post-prostatectomy. The device provides a therapeutic benefit by augmenting pelvic floor muscle function beyond natural recovery.
Clinical Evidence
Prospective case series (n=16) evaluated Elitone for Men in men post-prostate surgery (mean age 65.3). Subjects used the device 4-5 times/week for 12 weeks. Primary endpoint: reduction in 24-hour pad weight compared to a historical control cohort (n=215) from a systematic literature review. Results: Mean pad weight reduction was 61% at 6 weeks and 92% at 12 weeks for the device group, compared to 45% and 68% respectively for the control group. No adverse events reported.
Technological Characteristics
Transcutaneous electrical muscle stimulator; QAJ product code. Consists of a control unit and disposable perineal electrode. Shares identical components, materials, and performance parameters with the predicate Elitone, except for electrode shape. Operates via electrical stimulation to pelvic floor muscles. Non-implanted, external application.
Indications for Use
Indicated for men immediately following prostate surgery to aid early continence recovery.
Regulatory Classification
Identification
A transcutaneous electrical continence device consists of cutaneous electrodes that are used to apply external stimulation to reduce urinary incontinence.
Special Controls
In combination with the general controls of the FD&C Act, the transcutaneous electrical continence device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must characterize the electrical stimulation, including the following: Waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(2) The patient-contacting materials must be demonstrated to be biocompatible.
(3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) Instructions for use, including specific instructions regarding the proper placement of electrodes;
(ii) A summary of electrical stimulation parameters; and
(iii) Cleaning instructions and reuse information.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 23, 2026
Elidah, Inc.
Gloria Kolb
CEO
31 Pecks Ln
Suite 11
Newtown, Connecticut 06470
Re: K253285
Trade/Device Name: Elitone for Men
Regulation Number: 21 CFR 876.5330
Regulation Name: Transcutaneous electrical continence device
Regulatory Class: II
Product Code: QAJ
Received: December 23, 2025
Dear Gloria Kolb:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253285 - Gloria Kolb
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253285 - Gloria Kolb
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSICA K. NGUYEN -S
Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253285 | ? |
| Please provide the device trade name(s). | | ? |
| Elitone for Men | | |
| Please provide your Indications for Use below. | | ? |
| Elitone for Men is a non-implanted muscle stimulator designed to aid early continence recovery in men immediately following prostate surgery. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253285, Page 1 of 2
| 510(k) #: K253285 | 510(k) Summary | Prepared on: 2026-01-22 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Elidah, Inc. | |
| Applicant Address | 31 Pecks Ln Suite 11, Newtown CT 06470 United States | |
| Applicant Contact Telephone | 978-435-4324 | |
| Applicant Contact | Mrs. Gloria Kolb | |
| Applicant Contact Email | gloria@elidah.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Elitone for Men | |
| Common Name | Transcutaneous Electrical Continence Device | |
| Classification Name | Cutaneous Electrode Stimulator For Urinary Incontinence | |
| Regulation Number | 876.5330 | |
| Product Code(s) | QAJ | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K183585 | Elitone | QAJ |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Elitone for Men device is an electrical muscle stimulator designed to speed continence recovery in men following prostate surgery. Stimulation is delivered to the pelvic floor muscles and surrounding structures via a disposable electrode component configured to be applied to the skin in the perineal region. A small control unit component generates a therapeutic stimulation and allows the patient to increment or decrement the stimulation intensity. The applied electrical stimulation causes muscle contraction and relaxation in a treatment regimen that, with regular use over multiple weeks, is intended to speed continence recovery. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| Elitone for Men is a non-implanted muscle stimulator designed to aid early continence recovery in men immediately following prostate surgery. | | |
| Indications for Use Comparison | | 21 CFR 807.92(a)(5) |
| The indication for use is specific to use by males following prostate surgery, whereas the predicate is specific to females. This is a necessary distinction because the patient-contacting component is anatomically shaped to fit the male anatomy. The intended use is unchanged, as both the subject and predicate devices deliver electrical muscle stimulation through a perineal-applied electrode to treat urinary incontinence. | | |
| Technological Comparison | | 21 CFR 807.92(a)(6) |
| The subject device has the same technological characteristics as the predicate. Further, the Elitone for Men and Elitone devices are identical (i.e. shared components) except for the electrode component, which have profiles corresponding the male and female anatomy respectively. The materials, construction, and performance parameters of the electrodes are identical. | | |
{5}
K253285, Page 2 of 2
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The FDA guidance document titled "Guidance Document for Powered Muscle Stimulator 510(k)s" was used in determining what types of bench tests were applicable for the controller component. Specifically, the table titled "Output Specifications" served as the template for characterizing performance. The subject Elitone for Men device shares components with the predicate Elitone device. Accordingly, no new bench testing was performed on the controller component. Performance of the electrode component, including for example the maximum current density, was characterized and found to be substantially equivalent to the predicate female electrode.
Clinical data were submitted to support the determination of substantial equivalence. The study was designed as a prospective case series. Sixteen subjects completed the treatment and were included in the analysis. Subjects initiated treatment with the Elitone for Men device approximately 1 month after prostate surgery (mean 35.5 ± 13.3 days post-surgery). The study population had a mean age of 65.3 ± 5.3 years and mean BMI of 27.5 ± 4.9. Subjects were treated 4-5 times per week for 12 weeks. Subjects recorded baseline incontinence symptoms prior to initiating treatment and maintained treatment/symptom logs throughout the study. Effectiveness was assessed by reduction in 24-hour pad weights.
A control group was needed given the significant natural recovery of continence that occurs following prostatectomy, particularly during the first-year post-surgery. A systematic literature review was conducted to establish a historical control cohort. Studies included in the analysis contained: quantitative continence outcomes reported using a 24-hour pad weight test; multiple continence assessments within the first three months following radical prostatectomy; and inclusion of a control cohort consisting of either no active treatment or self-guided pelvic floor muscle exercises. Studies were excluded if they did not include multiple early post-operative assessments or included physician-guided or supervised pelvic floor muscle exercise programs. The control cohort was comprised of a total of 215 subjects. The literature review population was comparable to the study population in terms of post-operative timing, with assessments beginning in the early post-prostatectomy period when natural recovery is most pronounced.
For the Elitone for Men study cohort, mean 24-hour pad weight was 608 ± 598 grams at baseline (n=15), decreased to 238 ± 315 grams at 6 weeks, and further decreased to 47 ± 65 grams at 12 weeks. Men treated with the Elitone for Men device had a mean reduction in 24-hour pad weight at 6 weeks of 61% compared to a 45% reduction in pad weight for the control cohort. At 12 weeks men treated with the device had a mean reduction in 24-hour pad weight of 92% compared to 68% for men in the control group. While the use of a historical control has inherent limitations including study heterogeneity, different patient populations, and varying surgical techniques across the referenced control studies, it provides evidence that the device may offer benefits beyond natural recovery.
No adverse events were reported during the study. Given the low-risk nature of this non-implanted, externally applied device, the observational clinical evidence provides acceptable support for safety and effectiveness in aiding early continence recovery in men immediately following prostate surgery.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.