BedDot (G1) is a non-contact, bed-mounted vital signs monitor intended to provide continuous measurement of respiratory rate (RR) within a validated operating range of 12–24 breaths/min, and movement in adult patients lying in a supine, resting position in a hospital ward or clinic setting. The operation of the BedDot (G1) has been studied in adults (weight ≥ 50Kg and weight ≤115Kg) during sleep and resting conditions.
Device Story
BedDot (G1) is a non-contact, bed-mounted monitor for adult patients. System uses a vertical geophone sensor module placed under the bed frame to detect micro-vibrations caused by physiological activity. Signals are digitized at 100 Hz and transmitted via Wi-Fi to a backend server. The server processes data to derive respiratory rate, bed occupancy, and body movement. Results are displayed on a secure web-based interface for clinicians. Device is intended for use in hospital wards or clinics under still conditions. It provides trend data for clinical review but does not provide real-time alarms for life-threatening events. Benefits include continuous, non-invasive monitoring of respiratory status and movement without patient contact.
Clinical Evidence
Clinical validation study (n=52 adults) conducted at University of Georgia. Primary endpoint: RR accuracy (12–24 breaths/min) compared to manually scored etCO2 waveforms (Mindray BeneVision N17). Results: MAE 1.77 breaths/min, RMSE 2.80 breaths/min, STD 2.57 breaths/min. Subgroup analyses (sex, age, BMI) showed consistent performance. No adverse events reported.
Technological Characteristics
Non-contact, bed-mounted system. Sensing: Vertical geophone (magnetic element/coil). Energy: +5VDC via AC power module. Connectivity: Wi-Fi to server-based backend. Form factor: Main unit with USB-C and mounting plate. Software: Embedded computing module for signal processing. Non-sterile, reusable, no patient-contacting components.
Indications for Use
Indicated for adult patients (weight 50-115kg) in hospital or clinic settings requiring continuous respiratory rate (12-24 breaths/min) and movement monitoring while in a supine, resting position. Not indicated for active patient monitoring, life-threatening situations, or patients with arrhythmias.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 25, 2026
Intelligent Dots LLC.
% Boyle Wang
General Manager
Abmed Medical Technology Inc.
800 East Wishkah Street, Suite 1021
Aberdeen, Washington 98520
Re: K253274
Trade/Device Name: BedDot (G1)
Regulation Number: 21 CFR 868.2375
Regulation Name: Breathing Frequency Monitor
Regulatory Class: Class II
Product Code: BZQ
Dated: May 18, 2026
Received: May 18, 2026
Dear Boyle Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253274 - Boyle Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253274 - Boyle Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
PRAKHYAT
SINGH -S
Digitally signed by PRAKHYAT
SINGH -S
Date: 2026.06.25 14:55:14 -04'00'
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253274
Device Name
BedDot (G1)
Indications for Use (Describe)
BedDot (G1) is a non-contact, bed-mounted vital signs monitor intended to provide continuous measurement of respiratory rate (RR) within a validated operating range of 12–24 breaths/min, and movement in adult patients lying in a supine, resting position in a hospital ward or clinic setting. The operation of the BedDot (G1) has been studied in adults (weight ≥ 50Kg and weight ≤115Kg) during sleep and resting condition.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-4740 EF
{4}
510(k) Summary
K253274
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
### 1.0 Submitter's Information
Name: Intelligent Dots LLC.
Address: 147 Technology Pkwy, Peachtree Corners Norcross, GA 30092.
Tel: (470) 647-8588
Contact: Dr. WenZhan Song
Email: info@intelligentdots.com
### Designated Submission Correspondent
Contact: Mr. Boyle Wang
Name: Abmed Medical Technology Inc.
Address: 800 East Wishkah Street, Suite 1021 Aberdeen, WA98520
Tel: 303-8000162
Email: info@abmed.us
Date of Preparation: June 25,2026
### 2.0 Device Information
Trade name: BedDot (G1)
Common name: Monitor, breathing frequency
Classification name: Breathing frequency monitor
Model(s): G1
Production code: BZQ
Regulation number: 21 CFR 868.2375
Classification: Class II
Panel: Anesthesiology
### 3.0 Predicate&Reference Device Information
#### Primary Predicate#
Manufacturer: EarlySense Ltd.
Trade name: EarlySense InSight System
510(k) number: K152911
### 4.0 Indication for Use Statement
BedDot (G1) is a non-contact, bed-mounted vital signs monitor intended to provide continuous measurement of respiratory rate (RR) within a validated operating range of 12–24 breaths/min, and movement in adult patients lying in a supine, resting position in a hospital ward or clinic setting. The operation of the BedDot (G1) has been studied in adults (weight ≥ 50Kg and weight ≤115Kg) during sleep and resting conditions.
Page 1 of 9
{5}
510(k) Summary
K253274
### 5.0 Device Description
The BedDot (G1) (hereinafter referred to as the test device or BedDot) is a non-contact bed occupancy and vital signs monitor based on a vibration sensing technology using seismic sensors that detect very small vibrations. When installed under a bed, the BedDot can be used to continuously measure the respiration rate (RR) of adult patients lying in a resting position. The system also provides information on bed occupancy and body movements.
The BedDot is intended for use by clinicians or other trained medical personnel under still conditions only. It is not indicated for active patient monitoring, does not provide alarms for timely response in life-threatening situations, and is not intended for use in patients with arrhythmias.
The BedDot system consists of the main unit and compatible accessories, including a power adapter, USB-C cable, and adhesive metal plate for mounting. The main unit incorporates a vertical geophone sensor module, an analog-to-digital converter (ADC) board, and an embedded computing module. The geophone contains a magnetic element suspended within a coil assembly; micro-vibrations induced by physiological activity cause relative motion between the magnet and coils, generating an electrical signal. The ADC digitizes the signal at 100 Hz for subsequent processing by the embedded computing module.
The following parameters are estimated:
- Respiratory Rate;
- Bed Occupancy;
- Body Movements.
The BedDot operates in a direct device-to-server configuration. Each BedDot device integrates micro-vibration sensing technology and includes a built-in Wi-Fi module for secure data transmission to a designated backend application hosted on a server. The backend application processes incoming sensor data to derive respiratory rate (RR), bed occupancy status, and body movement parameters. Processed information is presented through a secure web-based user interface that allows authorized users to review current parameter status and historical trend data.
The device is non-sterile, reusable, and contains no patient-contacting components.
### 6.0 Technological Characteristic Comparison Table
{6}
510(k) Summary
| Parameter | BedDot (G1) (K253274) | EarlySense InSight System (K152911) | Substantial Equivalence |
| --- | --- | --- | --- |
| Product Code | BZQ | BZQ | Same |
| Regulation Number | 868.2375 | 868.2375 | Same |
| Classification Name | Breathing Frequency Monitor | Breathing Frequency Monitor | Same |
| Manufacturer | Intelligent Dots, Inc. | EarlySense Ltd. | -- |
| Intended Use | BedDot (G1) is a non-contact, bed-mounted vital signs monitor intended to provide continuous measurement of respiratory rate (RR) within a validated operating range of 12–24 breaths/min, and movement in adult patients lying in a supine, resting position in a hospital ward or clinic setting. The operation of the BedDot (G1) has been studied in adults (weight ≥ 50Kg and weight ≤115Kg) during sleep and resting conditions. | The EarlySense InSight system is intended for continuous measurement of respiration rate, heart rate and movement, in an automatic contact less manner, in hospital or clinic setting. The system is indicated for use in children, adolescents and adults. The operation of the EarlySense has been studied in children (weight ≥ 10 Kg) and adults (weight <111 Kg) during sleep and resting condition. | Similar, general intended use with respiratory rate (RR)-only functionality and narrower adult-only patient population. |
| Sensing Unit | Seismic Sensor (under mattress) | Piezoelectric sensor (under mattress/chair) | Similar, both technologies are passive, sensing methods that detect mechanical vibrations generated by patient physiological activity without direct patient contact and therefore does not raise different questions of safety or effectiveness |
| Signals acquired | Vibration signal generated by chest movement (respiration), and body movements. | Piezoelectric signal generated by variations of pressure applied on the device due to body motion, chest movement (respiration), and vibrations created by the heart. | Similar, both devices acquire vibrations associated with respiration and body movement and process these signals to derive the |
Page 4 of 9
{7}
510(k) Summary
| | | | same physiological information. |
| --- | --- | --- | --- |
| SpO2 Module | Not supported | Not supported | Same |
| Bedside Unit | Processing unit; no display | Processing unit; no display | Same |
| Energy Source | +5VDC Micro C connector supplied by AC power module | AC power | Different Does not raise new questions of safety or effectiveness and is addressed through electrical safety and power reliability testing |
| Clinical Environment | Hospital or clinic | Hospital or clinic | Same |
| Patient Population | Adults | Children, adolescents, adults | Different, Narrower intended population |
| Bedside Visual/Audio | No bedside clinical alarm; Device status indicators only | LEDs and buzzer | Different, the absence of alarm functionality in the subject device represents a narrower intended functionality and does not adversely affect the device's ability to perform its intended monitoring functions. |
| User Interface Display at Bedside | None | None | Same |
| Network Communication | LAN / Wi-Fi | LAN / Wi-Fi | Same |
| Monitoring Parameters | RR, Movement, Bed Occupancy | HR, RR, Movement, Bed/Chair Occupancy | Similar, the differences represent a narrower monitoring scope for the subject device and do not alter the safety and |
Page 5 of 9
{8}
510(k) Summary
| | | | effectiveness of the device. |
| --- | --- | --- | --- |
| Visual Status Indication | RR(Range: 12-24 bpm) Motion, Bed Occupancy | HR(Range: 30 - 170bpm RR(6-45 bpm) Motion, Bed/Chair Occupancy | Similar, difference represents a narrower set of displayed physiological parameters and does not affect the device's ability to perform its intended respiratory monitoring, occupancy detection, and movement monitoring functions. |
| Accuracy | MAE of 1.77 breaths/min, and RMSE of 2.80 breaths/min based on clinical study results | Not publicly Available | The subject device's RR accuracy is supported by clinical validation testing against a manually reviewed etCO_{2} reference methodology. |
| Displayed Information | Displayed remotely via central station | Displayed remotely via central station | Same |
Page 6 of 9
{9}
510(k) Summary
# **Substantial Equivalence Summary:**
The subject device, BedDot (G1), and the predicate device, EarlySense InSight System (K152911), have substantially equivalent intended use in that both are non-contact monitoring systems intended to provide physiological information, including respiratory rate (RR), movement, and occupancy-related status, for review by healthcare professionals in clinical care environments. Both devices utilize contact-free sensing technologies positioned beneath the patient support surface to detect low-amplitude mechanical signals associated with physiological activity. The predicate device uses a piezoelectric sensing element embedded in a bed sensor panel, while the subject device uses a geophone-based vibration sensing element installed under the bed frame. Despite this difference in sensing hardware, both devices employ the same fundamental measurement approach: acquisition of mechanical micro-motion signals generated by cardiac activity, respiration, and body movement, followed by signal processing to derive clinically relevant physiological parameters.
# **7.0 Non-clinical Testing Summary**
The following performance data have been conducted to verify that the BedDot meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the targeted device complies with the following standards:
# **Electrical & EMC Safety and Wireless:**
The electrical safety and EMC safety testing was performed to, and passed, the following standards:
- ➤ IEC 60601-1:2005/AMD2:2020, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- ➤ IEC 60601-1-2:2020, Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral standard: Electromagnetic disturbances - Requirements and tests
- ➤ IEC TS 60601-4-2:2024, Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- ➤ IEC 62133-2:2021, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- ➤ USEMCSC C63.27-2021, American National Standard for Evaluation of Wireless Coexistence
- ➤ Electromagnetic Compatibility (EMC) of Medical Devices Guidance Document (June 2022)
- ➤ NSI C63.27-2021, American National Standard for Evaluation of Wireless Coexistence
- ➤ AAMI TIR69:2017 (R2020) Risk Management of Radio-Frequency Wireless Coexistence for Medical Devices and Systems
- ➤ Radio Frequency Wireless Technology in Medical Devices – Guidance for Industry and Food and Drug Administration Staff (August 2020)
# **Risk Analysis**
Risk Analysis was performed on the subject device as recommended by ISO 14971:2019, Medical devices - Application of risk management to medical devices and risk controls were implemented to mitigate all identified hazards through suitable modifications to each components' functionality, labeling and packaging, IFU and User Manual. All foreseeable cybersecurity risks associated with the subject device have been identified in addition to finding timely errors through penetration testing.
Page 7 of 9
{10}
510(k) Summary---
### Summary of Bench Testing
Bench testing included continuous operation, repeated use, drift and durability stability, long-term stability under accelerated aging, environmental interference, temperature variation, motion detection sensitivity, occupancy detection, and drop/mechanical robustness was conducted and the results show that the subject device complies with applicable performance criteria.
### Software Verification and Validation Testing
Software documentation including verification & validation was provided in accordance with FDA Guidance “Content of Premarket Submissions for Device Software Functions”. The Software Validation is in compliance with FDA Guidance (June 2023).
### Cybersecurity
Cybersecurity risk assessment and controls of the Device was carried out per the FDA guidance, “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions Guidance for Industry and Food and Drug Administration Staff (February 2026)”
### Shelf-life & Reliability
Shelf-life & Reliability and testing were conducted and verified on the BedDot.
### Cleaning
Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance (March 2015)
### Biocompatibility Testing
As BedDot (G1) is a contactless device, this testing is not applicable for the subject device.
### 8.0 Clinical Test Summary
A clinical validation study was conducted to evaluate the accuracy of the BedDot (G1) system for measuring respiration rate (RR) in adult subjects under the device’s intended use conditions (resting/supine, still conditions in a healthcare facility setting). A total of 52 adult subjects from the intended population were enrolled and analyzed at the Clinical and Translational Research Unit, University of Georgia, USA.
For respiration rate validation, BedDot RR outputs were evaluated against clinician-manually scored respiratory rates derived from the raw etCO₂ capnography waveform using the FDA-cleared Mindray BeneVision N17 (K202405). Two independent blinded clinicians manually counted respiratory cycles from the raw waveform, with adjudication by a third blinded clinician when needed. The primary RR validation was restricted to the intended operating range of 12–24 breaths/min, consistent with the device’s intended use population. BedDot achieved an RR MAE of 1.77 breaths/min, and RMSE of 2.80 breaths/min, and STD of 2.57 breaths/min.
Demographic subgroup analyses (sex, age, BMI) and time-range analyses demonstrated consistent performance within the evaluated intended-use population. No device-related adverse events, serious adverse events, or device stability issues were observed during the study.
The clinical test results support clinically acceptable accuracy of BedDot for RR within the intended-use environment.
Page 8 of 9
{11}
510(k) Summary---
## 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the subject device is substantially equivalent to the legally marketed predicate device.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.